CATHETER, INTRAVASCULAR OCCLUDING, TEMPORARY, MODEL SOL 3510; WIRE, GUIDE, CATHETER, MODEL QSW1000

K994141 · Medtronic Ave, Inc. · MJN · Jul 21, 2000 · Cardiovascular

Device Facts

Record IDK994141
Device NameCATHETER, INTRAVASCULAR OCCLUDING, TEMPORARY, MODEL SOL 3510; WIRE, GUIDE, CATHETER, MODEL QSW1000
ApplicantMedtronic Ave, Inc.
Product CodeMJN · Cardiovascular
Decision DateJul 21, 2000
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.4450
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Medtronic AVE Solstice™ Solstice Balloon Temporary Occlusion System is intended to be used for temporary vascular occlusion in selectively stopping or controlling blood flow in the peripheral, cerebral, and visceral vasculature.

Device Story

Solstice Temporary Occlusion Balloon System consists of a single-lumen balloon catheter and QS 10 guidewire. Used in peripheral, cerebral, and visceral vasculature to temporarily occlude blood flow. Operation involves advancing QS 10 guidewire through catheter to occlude central lumen, enabling balloon inflation with fluid. Used by clinicians in interventional settings. Provides temporary flow control to assist in vascular procedures. Benefits include localized flow management during interventions.

Clinical Evidence

No clinical data. Safety and effectiveness established through in-vitro bench testing (mechanical/reliability) and in-vivo animal studies. Testing followed guidance for PTCA/Atherectomy catheters and stents.

Technological Characteristics

Single-lumen balloon catheter; QS 10 guidewire. Sterilization: EtO for catheter, e-beam for guidewire. Packaging: Tyvek bags. No specific material standards or software/algorithm components described.

Indications for Use

Indicated for patients requiring temporary vascular occlusion to selectively stop or control blood flow in peripheral, cerebral, and visceral vasculature.

Regulatory Classification

Identification

A vascular clamp is a surgical instrument used to occlude a blood vessel temporarily.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # Summary of Safety and Effectiveness K994141 #### Trade Name: Medtronic AVE Solstice ™ Temporary Occlusion Balloon System. ### Manufacturer: Medtronic AVE, Inc. 3576 Unocal Street Santa Rosa, California, 95403 Contact: Sonny Yamasaki Establishment Registration Number: 2953200 ## Classification Name: Catheter, Intravascular Occluding, Temporary (21 CFR 870.4450) Wire, Guide, Catheter (21 CRF 870.1330) ### Device Classification: Solstice: Class II (21 CFR 870.4450) Panel: Interventional Cardiovascular DCRND QS 10: Class II (21 CFR 870.1330) Panel: Interventional Cardiovascular DCRND ## Intended Use and Product Description: The Medtronic AVE Solstice Temporary Occlusion Balloon Catheter and QS 10 Guidewire System (here after referred to as Solstice and/or QS 10) is designed for use in blood vessels where temporary occlusion is desired. Solstice and QS 10 are intended to be used together for temporary vascular occlusion in selectively stopping or controlling blood flow in the peripheral, cerebral, and visceral vasculature. The Solstice is a single lumen balloon catheter that uses the QS 10 guidewire to occlude the central lumen and allow inflation of the balloon. This is accomplished by advancing any part of the QS 10 through the catheter. The single lumen then becomes occluded allowing fluid to fill the balloon. ### Sterilization, Packaging, and Pyrogenicity: The Solstice Temporary Occlusion Balloon System is packaged in individual dispensing hoops which are sealed inside their own labeled Tyvek bags, The Solstice System is created by placing a Solstice sterile pouch, a QS 10 Guidewire sterilized pouch and the IFUs (one for Solstice and one for QS 10) in a carton with a Solstice System Label. QS 10 Guidewire packaging has also been validated for being sold individually. Solstice Catheter is sterilized using EtO and the QS 10 is e-beam irradiation sterilized. 20 {1}------------------------------------------------ ## Substantial Equivalence: The Medtronic AVE Solstice Temporary Occlusion Balloon System is substantially rno Wouth to the MTI, Equinox™ Occlusion Balloon System (K990487) which includes a 0.10" guide wire (SilverSpeed: K982543) Establishment of equivalence is based on similarities of intended use, design, and physical characteristics as evaluated by physical bench testing, biocompatibility, and animal studies. # Summary of Safety and Effectiveness: Safety and effectiveness were evaluated through biocompatibility testing, reliability testing, mechanical testing, and animal studies. Testing was conducted on final, testing, mountined to atheters and QS 10 Guidewires with guidance in part from "PTCA Catheters Atherectorny Catheters Lasers Intravascular Stents," May 1995. The tests were used to assess the mechanical properties of the Solstice Catheter and QS 10 Guidewire. Based on in-vitro and in vivo testing results, Medtronic AVE believes the Solstice and QS 10 are safe and effective for their intended use. {2}------------------------------------------------ Public Health Service Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular emblem with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES" written around the top half of the circle. Inside the circle is a stylized image of a bird-like figure with three curved lines representing its wings or body. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JUL 2 1 2000 Sonny Yamasaki, Ph.D. Medtronic AVE, Inc. 3576 Unocal Street Santa Rosa, CA 95403 Re: K994141 Medtronic AVE Solstice™ Temporary Occlusion Balloon System Regulatory Class: II (two) Product Code: 74 MJN Dated: April 28, 2000 Received: May 1, 2000 Dear Dr. Yamasaki: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to The general controls the general controls provisions of the Act. provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {3}------------------------------------------------ Page 2 - Sonny Yamasaki, Ph.D. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html". Sincerely yours, Brim E. Harvey James E. Dillard III Director Division of Cardiovascular and Respiratory Devices Office of Device Evaluation enter for Devices and Radiological Health Enclosure {4}------------------------------------------------ # INDICATION FOR USE STATEMENT Company Name: Medtronic AVE Device Name: Indication for Use: Solstice™ Balloon Temporary Occlusion Catheter Balloon System The Medtronic AVE Solstice™ Solstice Balloon Temporary Occlusion System is intended to be used for temporary vascular occlusion in selectively stopping or controlling blood flow in the peripheral, cerebral, and visceral vasculature. Bin E. Hansen K994141 ﺮ ﺍﻟﻤ
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...