K994078 · United Orthopedic Corp. · LPH · Dec 4, 2000 · Orthopedic
Device Facts
Record ID
K994078
Device Name
U1 HIP SYSTEM
Applicant
United Orthopedic Corp.
Product Code
LPH · Orthopedic
Decision Date
Dec 4, 2000
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3358
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The U1 hip system is indicated in hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients with the following conditions: 1. Non-inflammatory degenerative joint disease such as osteoathritis, avascular necrosis, ankylosis, protrusio acetabuli, and painful hip dysplasia; 2. Inflammatory degenerative joint disease such as rheumatoid arthritis; 3. Correction of function deformity; 4. Revision procedures where other treatments or devices have failed; and 5. Treatment of nonunion, femoral neck, and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.
Device Story
The U1 Hip System is a modular total hip joint replacement prosthesis. Components include a tri-wedge femoral stem (cast Co-Cr-Mo alloy, ASTM F75) with porous-coated proximal surface and Morse taper; a 26mm femoral head (wrought Co-Cr-Mo alloy, ASTM F1537); a hemispherical acetabular shell (cast Co-Cr-Mo alloy) with porous coating and screw holes; and a UHMWPE liner. The device is implanted by orthopedic surgeons in a clinical/hospital setting to replace the hip joint. It functions as a mechanical load-bearing implant to restore joint function and reduce pain. Clinical benefit is derived from the restoration of hip biomechanics and stability.
Clinical Evidence
Bench testing only. Finite element analysis was performed to evaluate stem performance (relative motion, load transfer). Porous coating characterization included static tensile and shear strength testing (results >20MPa), volume porosity (30-70%), pore size (100-1000 microns), and coating thickness (500-1500 microns).
Technological Characteristics
Modular hip prosthesis. Materials: Cast Co-Cr-Mo alloy (ASTM F75) for femoral stem and acetabular shell; wrought Co-Cr-Mo alloy (ASTM F1537) for femoral head; UHMWPE (ISO 5834/1) for liner. Features: Tri-wedge femoral stem, Morse taper, beaded porous coating, hemispherical acetabular shell with screw holes. Sterilization method not specified.
Indications for Use
Indicated for skeletally mature patients requiring hip arthroplasty for pain relief or improved function due to non-inflammatory degenerative joint disease (osteoarthritis, avascular necrosis, ankylosis, protrusio acetabuli, painful hip dysplasia), inflammatory degenerative joint disease (rheumatoid arthritis), functional deformity, revision procedures, or proximal femur fractures (nonunion, femoral neck, trochanteric) with head involvement.
Regulatory Classification
Identification
A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.
Predicate Devices
DePuy AML® Hip Prosthesis
Submission Summary (Full Text)
{0}------------------------------------------------
# X994078
## DEC = 4 2000
#### ਹੈ U1 HIP SYSTEM
Summary
### 510(k) Summary of Safety and Effectiveness
| Company: | United Orthopedic Corporation |
|----------------|--------------------------------------------------------|
| Address: | No 57, Park Ave. 2, Science Park, Hsinchu, 300, Taiwan |
| Phone Number: | 886-3-5773351 |
| Fax Number: | 886-3-5777156 |
| Date Prepared: | November 16, 1999. |
| Device Name: | U1 Hip System |
| Common Name: | Hip joint prostheses or replacement |
Hip joint, Semi-Constrained, Metal/Polymer, Porous Classification Name: Uncemented Prosthesis per 21CFR 888.3358. DePuy AML® Hip Prosthesis Predicate Device:
#### Device Description:
The U1 Hip System is designated as a total hip joint replacement. It is a modular type of product system. The followings are the major design features.
Femoral Stem: The component has seven sizes of options, tri- wedge design, straight stem, 1340 neck-stem angle, a Morse type taper to receive modular heads, porous coated surface on proximal portion and satin finished on distal. There are w/ and w/o calcar collar designs. This device is produced from cast Co-Cr-Mo alloy (ASTM F75).
Femoral Head; The head diameter is 26mm. There are four adjustments of options. Std. +3, +6, +9, respectively. It is fabricated from wrought Co-Cr-Mo alloy bars (ASTM F1537) by machining.
Acetabular Component: This device has ten sizes of options, hemispherical design, porous coated surface on metallic shell, clustered bone screw holes, spherical screw holes for variable screw locking angle, PMMA cement plug, easy snap-in and take out mechanism, 16 options for angle adjustment, minimum 4.33mm thickness of UHMWPE liner. The metallic shell is produced from cast Co-Cr-Mo alloy (ASTM F75). The plastic liner is machined from extruded UHMWPE bars (ISO 5834/1)
Bone Screw: This device is a 6.5 mm cancellous bone screw. It has six lengths of options (15 mm ~ 40 mm), and has a shallow screw head.
#### Intended Use:
The U1 hip system is indicated in hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients with the following conditions: Non-inflammatory degenerative joint disease such as osteoathritis, avascular necrosis, 1.
{1}------------------------------------------------
#### 9 U1 HIP SYSTEM
ankylosis, protrusio acetabuli, and painful hip dysplasia;
- 2. Inflammatory degenerative joint disease such as rheumatoid arthritis;
- 3. Correction of function deformity:
- 4. Revision procedures where other treatments or devices have failed; and
- 5. Treatment of nonunion, femoral neck, and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.
### Basis for Substantial Equivalence:
Features comparable to predicate devices, DePuy AML® Hip prosthesis, include Co-Cr-Mo alloy substrate, straight type stem, beaded porous coated surface on the proximal portion of stem and acetabular cup, and Morse taper locked with heads.
### Test Results:
The Ul stem was analyzed using finite element analysis. The results for the U1 stem (relative motion, load transfer, etc.) are somewhat less sensitive to changes in friction condition than they are for the more extensively coated AML. The analysis establishes the essential equivalence of the design concepts for the U1 and AML femoral components.
The static tensile and shear strength of the beaded porous coating are more than 20MPa. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1000 microns and a porous coating thickness between 500 and 1,500 microns.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three lines forming its body and head. The eagle is encircled by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DEC - 4. 2000
Mr. Mellen Liu Regulatory Affairs Coordinator United Orthopedic Corporation 57, Park Avenue 2, Science Park Hsinchu 300, Taiwan, R.O.C.
Re: K994078/S1 Trade Name: U1 Hip System Regulatory Class: II Product Code: LPH Dated: September 8, 2000 Received: September 11, 2000
Dear Mr. Liu:
We have reviewed your Section 510(k) notification of intent to market the device referenced we have and we have determined the device is substantially equivalent (for the indications for use above and we have actering as marketed in interstate commerce prior to May 28, 1976, the stated in the enorours) to teater with the enterest that have been reclassified in enacificin and of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, controls pro horons or use, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations (1 remarket ripps of ally ware found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to mspotions, the Food and Diam may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
{3}------------------------------------------------
Page 2 - Mr. Mellen Liu
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and 11 you desire specific arrivitro diagnostic devices), please contact the Office of Compliance at additionally 607.10 for m. Pro aluguestions on the promotion and advertising of your device, (201) 594-4639. Traditionally, and (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). "Other general information on your responsibilities under the Act may be obtained from the Division of Small monmation on your responsive at its toll-free number (800) 638-2041 or at (301) 443-6597, or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Narls N Milburn
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
、 4 . . . . 1000
Enclosure
{4}------------------------------------------------
510 (k) Number (if known):____________________________________________________________________________________________________________________________________________________
Device Name: _U1 Hip System
Indications for Use:
This device is indicated in hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients with the following conditions:
- · Non-inflammatory degenerative joint disease such as osteoathritis, avascular necrosis, ankylosis, protrusio acetabuli, and painful hip dysplasia;
- · Inflammatory degenerative joint disease such as rheumatoid arthritis;
- . Correction of function deformity;
- Revision procedures where other treatments or devices have failed; and .
- · Treatment of nonunion, femoral neck, and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.
# PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation
Prescription Use __ (per 21 CFR 801.109) OR
Over-The Counter Use
for Mich N. Milliken
(Division Sign-Off)
Division of General Restorative Devices
510(k) Number K994078
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.