← Product Code [LZA](/productcode/LZA) · K994064

# BLUE NITRILE EXAMINATION GLOVES - POWDER FREE (K994064)

_Glove Source, Inc. · LZA · Jan 28, 2000 · General Hospital · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K994064

## Device Facts

- **Applicant:** Glove Source, Inc.
- **Product Code:** [LZA](/productcode/LZA.md)
- **Decision Date:** Jan 28, 2000
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.6250
- **Device Class:** Class 1
- **Review Panel:** General Hospital

## Indications for Use

A patient examination glove is a disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiners.

## Device Story

Nitrile patient examination gloves; disposable; worn on hands/fingers by healthcare providers during examinations; barrier protection to prevent cross-contamination between patient and examiner. Device is non-sterile, powder-free, and available in blue or white. Manufactured in various sizes (XS-XL) with specified dimensions and physical properties (tensile strength, elongation). Used in clinical settings. Benefit: provides protective barrier for infection control.

## Clinical Evidence

No clinical data. Bench testing only. Device performance verified against ASTM D 3578-95 standards, including tensile strength, ultimate elongation, and watertight integrity (AQL 2.5).

## Technological Characteristics

Nitrile rubber material. Powder-free. Dimensions: 240mm length; 0.10mm minimum thickness (cuff, palm, finger). Physical properties: 14 MPa (min) tensile strength; 700% (min) ultimate elongation before aging. Meets ASTM D 3578-95 standards. Non-sterile.

## Regulatory Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

## Predicate Devices

- Nitrile Examination Gloves (Class I, 80 LZA, meeting ASTM D 3578-95)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

JAN 2 8 2000

## RLIAN GLOVINDO

JLN. SUTOMO UJUNG 96 MEDAN 20235 - INDONESIA Tel : 62-61-616944 Tel : 62-4-92777 Fax :62-61-8459180

Page Numbers 1 of 2

### "510 (K)" SUMMARY

K994064

| (1) Name of applicant | Mr. Tony Jap                                                          |
|-----------------------|-----------------------------------------------------------------------|
| Address               | PT. Berlian Glovindo<br>Jl. Sutomo Ujung No. 96<br>Medan<br>Indonesia |
| Phone No.             | 62-61-616944 62-4-671077 671088                                       |
| Fax No.               | 62-61-8459170                                                         |

The contact persons within the firm as well as in U.S.A are given below

| The contact persons within the firm as well as in U.S.A are given below |                        |
|-------------------------------------------------------------------------|------------------------|
| Contact person in firm                                                  | : Mr. Tony Jap         |
|                                                                         | Fax No : 62-61-8459170 |

| Contact person in U.S.A | : Emmy Tjoeng         |
|-------------------------|-----------------------|
|                         | Fax No.: 626-913-1498 |

- (2) Device details Trade Name
: Private lebel-Nirile Examination Gloves Powder Free Blue and White Color : Patient Examination Gloves

- Classification Name : Nitrile 80 LZA Product Code
- (3) Equivalent device legally : Class | Nitrile Examination Gloves 80 LZA marketed Powder Free meeting ASTM D 3578-95
- (4) Intended use : patient examination glove is a disposable device A intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between and patient examiners.

{1}------------------------------------------------

# RLIAN GLOVIND

ILN. SUTOMO UJUNG 96 MEDAN 20235 - INDONESIA

Tel : 62-61-616944 Tel : 62-4-92777 Fax : 62-61-8459180

Page Numbers 2 of 2

- Dimensions a. S XS M XL Sizes ﺎ 240 mm 240 mm 240 mm 240 mm 240 mm Length Width 80 < mm 80±10 mm 95±10 mm 111±10 mm >111mm Thickness 1. Cuff (min) 0.10 mm 0.10 mm 0.10 mm 0.10 mm 0.10 mm 2. Palm (min) 0.10 mm 0.10 mm 0.10 mm 0.10 mm 0.10 mm 3. Finger Tip (min) 0.10 mm 0.10 mm 0.10 mm 0.10 mm 0.10 mm b. Physical Properties Before aging After aging at 100ºC 22 hrs. Tensile Strength 14 Mpa (min.) 14 Mpa (min.) Ultimate Elongation 700 % (min.) 500 % (min.) Performance Requirement C. Characteristic Related Defects Inspection AQL Level Visible defects Stains, Lumps, S-4 2.5 Holes etc. Watertight Holes S-4 2.5 Dimensions S-2 Width Length 4 & Thickness Physical Properties ട-2 Before and 4 after ageing d. Weight of residual powder in medium size gloves : 0.5 ± 0.2 mg e. Bio - Compatibility g. Test Resulth as per ASTM D 3578 -95 (6) Performance data is the same as mentioned immediately above. (7) Clinical date is not needed for gloves or for most devices cleared by the 510 (K) process. (8) Non-dinical data Gloves meet or exceed the ASTM D 3578 Standard.
#### (5) Technological characteristic of the gloves,

Meets FDA pin hole requirement.

Meets labeling daim.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing right, with flowing lines above and below them. The profiles are arranged in a way that suggests movement or progress. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the image.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

CIAN 2 8 2000

Emmy Tjoeng Mr. Official Correspondent for PT. Berlian Glaovindo Jln. Sutomo Ujung 96 Medan 20235 - Indonesia

Re : K994064 Trade Name: Nitrile Examination Gloves - Powder Free Requlatory Class: I Product Code: LZA November 30, 1999 Dated: Received: December 1, 1999

Dear Dr. Tjoeng:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Druq, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in requlatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any

{3}------------------------------------------------

Page 2 - Mr. Tjoeng

obligation you might have under sections 531 through 542 of obligation for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your marice and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in regaradiagnostic devices), please contact the Office of Compliance at (301) 594-4692. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general premation on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at at Its coll free namber (65)/www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

Timothy A. Ulatowski

Directbr Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

JLN. SUTOMO UJUNG 96 MEDAN 20235 - INDONESIA Tel : 62-61-616944 Tel :62-4-92777 Fax :62-61-8459180

(K994064)

ANNEXURE II

#### INDICATION FOR USE

| Applicant          | :  Mr. Tony Jap                                  |
|--------------------|--------------------------------------------------|
| Device Name        | : Nitrile Patient Examination Gloves Powder Free |
| Indication for use | :                                                |

A patient examination glove is a disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiners.

(PLEASE DO NOT WRITE BELOW THIS LINE. CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Qum S. Lin

(Division Sign Off) Division of Dental, Infection Control, and General Hospital Devic 510(K) Number K GU

Prescription Use (Per 21 CFR 801.109)

OR Over-The-Counter Use

---

**Source:** [https://fda-staging.innolitics.com/device/K994064](https://fda-staging.innolitics.com/device/K994064)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).