Use of the LoCon-T Distal Radial Plate System is intended to be used for fixation of unstable distal radius fractures in which closed reduction is not suitable: - Joint destruction and/or subluxation visible on x-ray; . - Failed fracture fixation with or without bone graft; . - Osteotomy and repair of distal radius malunion with or without bone . graft; - Displaced or non-displaced fracture which may or may not involve . angulation or fragmentation of bone; - Dorsal plates are indicated for use with comminuted articular fractures, . shearing fractures of the articular surface, severely comminuted extraarticular fractures, and fractures in which reduction has been lost following fixation with percutaneous pins with or without an external fixator; or - T-plates are indicated for use with volar articular shearing fractures
Device Story
LoCon-T Distal Radial Plate System; metallic bone fixation system for distal radius fractures. Components: dorsal plates, volar T-plate, dorsal plate extender, cortical/cancellous screws, buttress pins. Used by orthopedic surgeons in clinical/surgical settings to secure bone fragments. Provides internal fixation for unstable fractures where closed reduction fails. Benefits patient by stabilizing bone, facilitating healing, and restoring joint alignment.
Clinical Evidence
Bench testing only. Mechanical test data demonstrated that the material and subject device design meet the strength requirements of the predicate device.
Technological Characteristics
Materials: Stainless Steel (ASTM F 139 for plates/extender, ASTM F 138 for screws, ASTM F 1314 for pins). Components: dorsal plates (two sizes, left/right), volar T-plate, extender, 4.0mm cancellous screws (partially/fully threaded), cortical screws, 1.8mm diameter pins. Low-profile design.
Indications for Use
Indicated for patients with unstable distal radius fractures where closed reduction is unsuitable, including joint destruction/subluxation, failed prior fixation, malunion requiring osteotomy, and displaced/non-displaced fractures with or without fragmentation. Dorsal plates specifically for comminuted articular, shearing articular, severely comminuted extra-articular, and post-percutaneous pin fixation loss fractures. T-plates specifically for volar articular shearing fractures.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
Zimmer Forte Distal Radial Plate System
Submission Summary (Full Text)
{0}------------------------------------------------
# FFB 1 4 2000
Image /page/0/Picture/2 description: The image shows the logo for Wright Medical Technology, Inc. The logo features a stylized "W" on the left, followed by the company name in bold, uppercase letters. Below the name is the company's address: 5677 Airline Road, Arlington, TN 38002, and their phone number: 901-867-9971.
#### 510K Summary
Company: Wright Medical Technology, Inc. 5677 Airline Road Arlington, TN 38002 Date: November 30, 1999 Trade Name: LoCon-TTM Distal Radial Plate System Common Name: Metallic Bone Fixation System Predicate Device: Zimmer Forte Distal Radial Plate System
#### Description/Intended Use:
The LoCon-T Distal Radial Plate System includes four side specific dorsal plate, one volar T-plate, a dorsal plate extender, cortical screws, cancellous screws, and buttress pins. The plates and plate extender are manufactured from Stainless Steel (ASTM F 139). The screws are manufactured from Stainless Steel (ASTM F 138). The pins are manufactured from Stainless Steel (ASTM F 1314). The plates, pins, extender, and screws are designed to be lower in profile and secure bone fragments. The dorsal plates are available in two sizes for both right and left configurations. The cancellous screws are available in partially and fully threaded 4.0mm diameter designs ranging in length from 10.0 to 40.0mm in 2.0mm increments. The cortical screws are available in 3 diameters. The pins are 1.8mm in diameter and range in length from 12mm to 30mm.
### Intended Use:
Use of the LoCon-T Distal Radial Plate System is intended to be used for fixation of unstable distal radius fractures in which closed reduction is not suitable:
- Joint destruction and/or subluxation visible on x-ray; .
- Failed fracture fixation with or without bone graft; .
- Osteotomy and repair of distal radius malunion with or without bone . graft;
- Displaced or non-displaced fracture which may or may not involve . angulation or fragmentation of bone;
- Dorsal plates are indicated for use with comminuted articular fractures, . shearing fractures of the articular surface, severely comminuted extraarticular fractures, and fractures in which reduction has been lost following fixation with percutaneous pins with or without an external fixator; or
- T-plates are indicated for use with volar articular shearing fractures
#### Testing Summary:
The LoCon-T Distal Radial Plate System was declared substantially equivalent to the predicate devices. Mechanical test data demonstrated that the material and subject device design meets the strength requirements of the predicate device.
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of a caduceus, a symbol often associated with medicine and healthcare.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
FEB 1 4 2000
Ms. Lynne Witkowski Regulatory Affairs Associate Wright Medical Technology, Inc. 5677 Airline Road Arlington, Tennessee 38002
Re: K994061 Trade Name: LoCon-T™ Distal Radial Plate System Regulatory Class: II Product Code: HRS Dated: November 30, 1999 Received: December 1, 1999
Dear Ms. Witkowski:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{2}------------------------------------------------
# Page 2 - Ms. Lynne Witkowski
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Neil RP Ogden
James E. Dillard III
Acting Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
| 510(k) Number (if known) | K994061 |
|--------------------------|------------------------------------|
| Device Name | LoCon-T Distal Radial Plate System |
| Indications for Use | |
**Intended Use**
Use of the LoCon-T Distal Radial Plate System is intended for...
ystem is intended to be used for fixation of unstable distal radius fractures in which closed reduction is not suitable:
Indications for Use Statement
- Joint destruction and/or subluxation visible on x-ray; .
- Failed fracture fixation with or without bone graft; ●
- Osteotomy and repair of distal radius malunion with or without . bone graft;
- Displaced or non-displaced fracture which may or may not involve . angulation or fragmentation of bone;
- Dorsal plates are indicated for use with comminuted articular . fractures, shearing fractures of the articular surface, severely comminuted extra-articular fractures, and fractures in which reduction has been lost following fixation with percutaneous pins with or without an external fixator; or
- T-plates are indicated for use with volar articular shearing . fractures
## PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED
| Concurrence of CDRH, Office of Device Evaluation (ODE) | |
|--------------------------------------------------------|----------------------|
| (Division Sign-Off) | NRO for J2D |
| Division of General Restorative Devices | |
| 510(k) Number | K994061 |
| Prescription Use (per 21 CFR 801.109) | X |
| | OR |
| | Over-The Counter Use |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.