← Product Code [JEY](/productcode/JEY) · K994060

# THE BIOPLATE RESORBABLE BONE PLATING SYSTEM FOR CRANIOMAXILLOFACIAL SURGERY, BIOLACTATE (K994060)

_Bioplate, Inc. · JEY · Jan 24, 2001 · Dental · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K994060

## Device Facts

- **Applicant:** Bioplate, Inc.
- **Product Code:** [JEY](/productcode/JEY.md)
- **Decision Date:** Jan 24, 2001
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.4760
- **Device Class:** Class 2
- **Review Panel:** Dental
- **Attributes:** Therapeutic

## Indications for Use

The Bioplate Resorbable Bone Plating System is intended for use in conjunction with the Bioplate Rigid Fixation Bone Plating System for craniomaxillofacial indications in which resorbable fixation is desired. Plates and screws are used to align and stabilize fractures of bony tissue while normal tissue healing occurs. The resorbable plates and screws are intended for minimally load bearing fixation for the following indications: - Craniofacial fractures - Reconstructive procedures of the craniomaxillofacial skeleton - Cranial bone fixation - Brow lift procedures

## Device Story

Bioplate Resorbable Bone Plating System consists of plates, mesh, and screws for craniomaxillofacial skeletal fixation. Device provides alignment and stabilization of bony tissue during healing. Manufactured from bioresorbable Poly(L-lactide) copolymer; material resorbs via hydrolysis into metabolizable lactic acids. Intended for use by surgeons in clinical settings for minimally load-bearing applications. System used in conjunction with Bioplate Rigid Fixation Bone Plating System. Provided sterile via gamma radiation; single-use only.

## Clinical Evidence

No clinical data provided; substantial equivalence based on design, material, and intended use comparisons to legally marketed predicate devices.

## Technological Characteristics

Materials: Bioresorbable Poly(L-lactide) copolymer. Form factor: Plates, mesh, and screws. Energy source: None (mechanical fixation). Sterilization: Gamma radiation. Connectivity: None.

## Regulatory Identification

A bone plate is a metal device intended to stabilize fractured bone structures in the oral cavity. The bone segments are attached to the plate with screws to prevent movement of the segments.

## Predicate Devices

- LactoSorb® Trauma Plating System ([K955729](/device/K955729.md))
- Synthes (U.S.A.) Resorbable Fixation System ([K974554](/device/K974554.md))
- Howmedica Leibinger Resorbable Fixation System ([K982531](/device/K982531.md))
- Bioplate® Rigid fixation Bone Plating System for Craniomaxillofacial Surgery ([K980983](/device/K980983.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

## JAN 2 4 2001

K994060

### BIOMEDICS, INC. 6911 Melrose Avenue Los Angeles, California 90038 Tel: (323) 549-9500, FAX: (323) 935-0110

### 510(k) SUMMARY

The Bioplate® Resorbable Bone Plating System for Craniomaxillofacial Surgery consists of a variety of plate, mesh, and screw configurations intended for use in the treatment of fractures and reconstructive procedures of the craniomaxillofacial skeleton in which bioresorbable fixation is desired. Plates, mesh and screws are used to align and stabilize fractures of bony tissue while normal tissue healing occurs.

The plates, mesh, and screws are manufactured of a bioresorbable Poly(L-lactide) copolymer which resorbs by hydrolysis into lactic acids that are metabolized by the body. The material is substantially equivalent to that currently being used in fixation devices and is, in some cases, the exact same copolymer material.

Bioplate® resorbable plates, mesh, and screws are provided sterile by gamma radiation and must not be resterilized.

The substantial equivalence of this device is demonstrated by its similarity in intended use, design, dimensions, and materials to the LactoSorb® Trauma Plating System (K955729), which is manufactured from 82:18 poly(L-lactide-co-glycolide),, the Synthes (U.S.A.) Resorbable Fixation System (K974554), which is manufactured from 70:30 poly(L/DL-lactide), and the Howmedica Leibinger Resorbable Fixation System (K982531), manufactured of material described in the 510(k) as "substantially equivalent to that used in the LactoSorb Device".

The device is also substantially equivalent in its intended use and design to the Bioplate® Rigid fixation Bone Plating System for Craniomaxillofacial Surgery (K980983), which is manufactured of non-resorbable titanium.

{1}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus, a symbol often associated with medicine and healthcare. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular pattern around the caduceus.

#### Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

# JAN 2 4 2001

Mr. Eric V. Hohenstein Engineering & Product Development Manager Bioplate, Incorporated 6911 Melrose Avenue Los Angeles, California 90038

K994060 Re :

The Bioplate® Resorbable Bone Plating Trade Name: System for Craniomaxillofacial Surgery, Biolactate™ Regulatory Class: II Product Code: JEY Dated: October 24, 2000 Received: October 26, 2000

Dear Mr. Hohenstein:

We have reviewed your Section 510 (k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory

{2}------------------------------------------------

Page 2 – Mr. Hohenstein

In addition, FDA may publish further announcements action. accron. In addition in the Federal Register. Please note: this response to your premarket notification submission does ents response of the might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA deberibed in your and equivalence of your device to a legally marketed predicate device results in a classification for your markets problems, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4692. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address

"http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Timothy A. Ulatowski

Tim thy A. Ulatowski Director

Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

### 510(k) Number (if known): K994060

### DEVICE NAME:

The Bioplate Resorbable Bone Plating System for Craniomaxillofacial Surgery

### INDICATIONS FOR USE:

The Bioplate Resorbable Bone Plating System is intended for use in conjunction with the Bioplate Rigid Fixation Bone Plating System for craniomaxillofacial indications in which resorbable fixation is desired. Plates and screws are used to align and stabilize fractures of bony tissue while normal tissue healing occurs.

The resorbable plates and screws are intended for minimally load bearing fixation for the following indications:

- . Craniofacial fractures
- Reconstructive procedures of the craniomaxillofacial skeleton .
- Cranial bone fixation .
- . Brow lift procedures

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

CONCURRANCE OF CDRH, OFFICE OF DEVICE EVALUATION (ODE)

1 Prescription Use . (Per 21 CFR 801.109

ार

Over-The-Counter Use (Optimal Format 1-2-96)

Teall W. Shiffer
for MSR

Division Sign-Off

510(k) Number K994060

---

**Source:** [https://fda-staging.innolitics.com/device/K994060](https://fda-staging.innolitics.com/device/K994060)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com