← Product Code [JTN](/productcode/JTN) · K994046

# HARDYDISK CEFACLOR 30MCG (K994046)

_Hardy Diagnostics · JTN · Feb 1, 2000 · Microbiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K994046

## Device Facts

- **Applicant:** Hardy Diagnostics
- **Product Code:** [JTN](/productcode/JTN.md)
- **Decision Date:** Feb 1, 2000
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.1620
- **Device Class:** Class 2
- **Review Panel:** Microbiology

## Indications for Use

HardyDisk™ Antimicrobial Sensitivity Disks are used for semi-quantitative in vitro susceptibility testing by the agar diffusion test procedure (Kirby-Bauer) of rapidly growing and certain fastidious bacterial pathogens. Standardized methods for agar diffusion testing have been described for Enterobacteriaceae, Staphylococcus spp., Pseudomonas spp., Acinetobacter spp., Listeria monocytogenes, Enterococcus spp., other streptococci and, by modified procedures, Haemophilus influenzae, Neisseria gonorrhoeae and Streptococcus pneumoniae. HardyDisk™ Cefaclor is indicated for in vitro activity against the Enterobacteriacieae, methicillin sensitive Staphylococcus spp., non-beta lactamase producing Haemophilus spp. and Moraxella catarrhalis.

## Device Story

HardyDisk™ Cefaclor 30mcg is an antimicrobial susceptibility test disk used in clinical microbiology laboratories. It utilizes the Kirby-Bauer agar diffusion method to determine bacterial sensitivity to the antibiotic Cefaclor. The disk is placed on agar plates inoculated with patient bacterial isolates; the antibiotic diffuses into the medium, creating a zone of inhibition proportional to the organism's susceptibility. Laboratory technicians measure these zones to categorize bacteria as susceptible, intermediate, or resistant. This information assists clinicians in selecting appropriate antibiotic therapy for patients with bacterial infections. The device provides a standardized, semi-quantitative assessment of antimicrobial activity.

## Clinical Evidence

No clinical data. Performance is established through standardized in vitro agar diffusion testing (Kirby-Bauer method) comparing zone of inhibition results to established interpretive criteria for specific bacterial pathogens.

## Technological Characteristics

Antimicrobial sensitivity disk containing 30mcg of Cefaclor. Operates via passive diffusion in agar medium. Standardized for Kirby-Bauer disk diffusion susceptibility testing. Non-automated, manual laboratory diagnostic component.

## Regulatory Identification

An antimicrobial susceptibility test disc is a device that consists of antimicrobic-impregnated paper discs used to measure by a disc-agar diffusion technique or a disc-broth elution technique the in vitro susceptibility of most clinically important bacterial pathogens to antimicrobial agents. In the disc-agar diffusion technique, bacterial susceptibility is ascertained by directly measuring the magnitude of a zone of bacterial inhibition around the disc on an agar surface. The disc-broth elution technique is associated with an automated rapid susceptibility test system and employs a fluid medium in which susceptibility is ascertained by photometrically measuring changes in bacterial growth resulting when antimicrobial material is eluted from the disc into the fluid medium. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a black and white circular logo. The logo features the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - SSA" arranged around the circumference of the circle. Inside the circle is a stylized graphic of an abstract human figure, possibly representing a family or group of people.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

FEB - 1 2000

Ms. Melissa M. Traylor, RAC Director of Technical Services/Regulatory Affairs Hardy Diagnostics 1430 West McCoy Lane Santa Maria, California 93455

Re: K994046 Trade Name: HardyDisk™ Cefaclor 30mcg Regulatory Class: II Product Code: JTN Dated: November 17, 1999 Received: November 29, 1999

Dear Ms. Traylor:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{1}------------------------------------------------

Page 2

ﺎ ﻓﻲ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Steven Butman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for Hardy Diagnostics. The logo consists of a circular graphic on the left and the text "HARDY DIAGNOSTICS" on the right. The word "HARDY" is on top of the word "DIAGNOSTICS".

anta Maria (

1430 West McCov Lane ta Maria CA 93459 el 18051 346-2766

> alt I ake City Iltat 54 W Cottage Ave. Sandy, UT 84070 Tel: 18001 995-4273 Fax: 1801 | 562-3214

Phoenix Arizona Tel: 18001 995-8456 Fax: 16021464-9878

510K Number K994046

## Indications for Use Statement-HardyDisk™ Cefaclor 30mcg

HardyDisk™ Antimicrobial Sensitivity Disks are used for semi-quantitative in vitro susceptibility testing by the agar diffusion test procedure (Kirby-Bauer) of rapidly growing and certain fastidious bacterial pathogens. Standardized methods for agar diffusion testing have been described for Enterobacteriaceae, Staphylococcus spp., Pseudomonas spp., Acinetobacter spp., Listeria monocytogenes, Enterococcus spp., other streptococci and, by modified procedures, Haemophilus influenzae, Neisseria gonorrhoeae and Streptococcus pneumoniae.

HardyDisk™ Cefaclor is indicated for in vitro activity against the Enterobacteriacieae, methicillin sensitive Staphylococcus spp., non-beta lactamase producing Haemophilus spp. and Moraxella catarrhalis.

<End>

Concurrence of CDRH-ODE

Woody Dubois

510(k) Numbe

Prescription Use (per 21 CFR 801.109)

OR

Over-the Counter-Use (Optional format 1-2-96)

---

**Source:** [https://fda-staging.innolitics.com/device/K994046](https://fda-staging.innolitics.com/device/K994046)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).