← Product Code [JIR](/productcode/JIR) · K994035

# IMMUNODIP URINARY ALBUMIN SCREEN, MODEL 790-01, 790-15 (K994035)

_Diagnostic Chemicals , Ltd. · JIR · Feb 1, 2000 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K994035

## Device Facts

- **Applicant:** Diagnostic Chemicals , Ltd.
- **Product Code:** [JIR](/productcode/JIR.md)
- **Decision Date:** Feb 1, 2000
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.1645
- **Device Class:** Class 1
- **Review Panel:** Clinical Chemistry

## Indications for Use

The ImmunoDip™ Urinary Albumin Screen tests for the presence of elevated levels of albumin in urine. Elevated urmary albumin is also known as microalbuminuria. Elevated albumin is an early sign of possible kidney damage. Detection of elevated urinary albumin can aid in the early detection and monitoring of the course of incipient nephropathy in diabetics and hypertensive patients. For IN VITRO diagnostic use.

## Device Story

ImmunoDip™ Urinary Albumin Screen is an immunochromatographic test strip in a plastic housing; used for semi-quantitative detection of albumin in urine. Input: urine sample. Principle: capillary action carries urine and blue-colored latex beads (conjugated with monoclonal mouse anti-human albumin antibodies) up the strip. Albumin binds to beads; at low levels, beads bind to a lower band (human albumin); at levels >18 mg/L, beads bind to an upper band (goat anti-mouse antibodies), creating a darker color than the lower band. Output: visual comparison of two blue bands to determine if albumin is negative (≤18 mg/L) or positive (>18 mg/L). Used in physician office laboratories (POL) or clinical settings; operated by trained laboratory personnel or POL users. Results aid clinicians in early detection and monitoring of kidney damage in diabetic and hypertensive patients.

## Clinical Evidence

Clinical performance evaluated via precision studies and comparative testing. Internal 20-day precision study showed 100% agreement. Physician Office Laboratory (POL) studies (n=6 proficiency samples) showed 91% agreement with expected results, comparable to 95% agreement by trained lab personnel. Clinical comparison against two predicate methods (DCLare™ and quantitative assays) demonstrated 95% sensitivity, 95-97% specificity, and 95-97% efficiency.

## Technological Characteristics

Immunochromatographic test strip; monoclonal mouse anti-human albumin antibodies bound to colored latex beads; human albumin fixed at lower band; goat anti-mouse antibodies fixed at upper band; plastic housing; capillary action-based migration; semi-quantitative visual readout; no external energy source.

## Regulatory Identification

A urinary protein or albumin (nonquantitative) test system is a device intended to identify proteins or albumin in urine. Identification of urinary protein or albumin (nonquantitative) is used in the diagnosis and treatment of disease conditions such as renal or heart diseases or thyroid disorders, which are characterized by proteinuria or albuminuria.

## Predicate Devices

- DCLare™ ImmunoDip™ stick for Microalbuminuria ([K972337](/device/K972337.md))
- Microalbumin Assay/Array® Analyzer ([K922273](/device/K922273.md))
- N-Assay TIA Microalbumin ([K934146](/device/K934146.md))
- BeSure Plus OneStep Ovulation Prediction test ([K983113](/device/K983113.md))
- One Step Ovulation Predictor ([K991386](/device/K991386.md))
- MiniClinic Ovulation Predictor ([K960233](/device/K960233.md))
- Clearplan Easy Ovulation Test ([K981271](/device/K981271.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

510(K) Notification

Diagnostics Chemicals Limited

Product Cat. No. 790-01, 790-15

Image /page/0/Picture/4 description: The image shows a logo with two stylized letters. The letters appear to be "d" and "d", with the first "d" having a rounded rectangular shape and a vertical line extending upwards. The second "d" is similar but has a curved element at the top, creating a hook-like shape. The logo is black against a white background.

# 20.0 510(k) SMDA SUMMARY

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

| The assigned 510(k) number is: | K994035                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |             |                                   |              |                           |
|--------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|-----------------------------------|--------------|---------------------------|
| Prepared:                      | November 22, 1999                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |             |                                   |              |                           |
| Summiteer:                     | Diagnostic Chemicals Limited                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |             |                                   |              |                           |
| Address:                       | West Royalty Industrial Park<br>Charlottetown<br>P.E.I., C0A 1B0<br>Canada<br>(902) 566-1396                                                                                                                                                                                                                                                                                                                                                                                                                                                      |             |                                   |              |                           |
| Contact:                       | Karen Callbeck                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |             |                                   |              |                           |
| Device:                        | <table><tr><td>Trade Name:</td><td>ImmunoDip™ Urinary Albumin Screen</td></tr><tr><td>Common Name:</td><td>Test for microalbuminuria</td></tr></table>                                                                                                                                                                                                                                                                                                                                                                                            | Trade Name: | ImmunoDip™ Urinary Albumin Screen | Common Name: | Test for microalbuminuria |
| Trade Name:                    | ImmunoDip™ Urinary Albumin Screen                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |             |                                   |              |                           |
| Common Name:                   | Test for microalbuminuria                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |             |                                   |              |                           |
| Classification:                | Division of Clinical Laboratory Devices<br>Panel- Clinical Chemistry<br>Classification Code- 75 JIR (Urinary Protein or Albumin)                                                                                                                                                                                                                                                                                                                                                                                                                  |             |                                   |              |                           |
| Predicate Devices:             | DCLare™ ImmunoDip™ stick for Microalbuminuria (Diagnostic Chemicals<br>Limited- K972337),<br>Microalbumin Assay/Array® Analyzer (Beckman Instruments Inc.-<br>K922273), N-Assay TIA Microalbumin (Crestat Diagnostics/Kamiya<br>Biomedical Company- K934146),<br>BeSure Plus OneStep Ovulation Prediction test (Syntron Bioresearch, Inc,<br>K983113),<br>One Step Ovulation Predictor (Selfcare, Inc.-K991386)<br>MiniClinic Ovulation Predictor (Vanguard Biomedical- K960233), and<br>Clearplan Easy Ovulation Test (Whitehall Labs- K981271). |             |                                   |              |                           |

# Device Description:

ImmunoDip™ Urinary Albumin Screen is an immunochromatographic test strip which is encased in a plastic housing. The ImmunoDip™ Urinary Albumin Screen is placed into a urine sample for at least three minutes and is then removed and read. Results are determined by visually comparing the relative color intensity of two blue bands to obtain a semi-quantitative result of Negative ( 18 mg/L) or Positive (>18 mg/L).

{1}------------------------------------------------

510(K) Notification

Diagnostics Chemicals Limited

Product Cat. No. 790-01. 790-15

Image /page/1/Picture/4 description: The image shows a stylized, bold, and black graphic of the letters 'ddj'. The letters are interconnected, with the 'd' on the left and the 'j' on the right sharing a common vertical stroke. The overall design is simple and modern, with a focus on the interplay of positive and negative space.

## Intended Use:

The ImmunoDip™ Urinary Albumin Screen tests for the presence of elevated levels of albumin in urine. Elevated urmary albumin is also known as microalbuminuria. Elevated albumin is an early sign of possible kidney damage. Detection of elevated urinary albumin can aid in the early detection and monitoring of the course of incipient nephropathy in diabetics and hypertensive patients. For IN VITRO diagnostic use.

# Technological Characteristics:

ImmunoDip™ Urmary Albumin Screen is an immunochromatographic test strip containing monoclonal mouse antibodies against human serum albumin bound to colored latex beads. Human albumin is fixed in a band at the bottom half of the testing region. Goat anti-mouse antibodies are fixed in a band at the top half of the testing region. The dipstick is encased in an open-ended plastic housing.

When the dipstick is placed into a urine sample cup, the urine sample migrates up the test stip, Albumin present in the urine binds with blue colored latex beads present in the strip. Both beads and albumin are carried up the device by capillary action. At low levels of albumin, the great majority of blue beads are bound at the lower band containing human albumin. At higher levels of albumin, many of the beads pass through the lower band and are bound at the upper band. Levels of albumin above the decision level value of 18 mg/L will produce color on the upper band which is darket than the lower band. By observing the appearance of the two lines, the user can semi-quantitatively determine the urine microalbumin concentration as either Negative (s18 mg/L), or Positive(>18 mg/L),

# Assessment of Performance:

The performance characteristics on the ImmunoDip™ Urinary Albumin Screen were evaluated in precision studies to determine the within-day, between -day, within-run and between-run precision. Physician Office Laboratory (POL) studies were carried out in which the performance of trained laboratory technicians was compared against that of POL users. Clinical studies which compared results obtained with the Urinary Albumin Screen against three existing uring albumm assay methods were also carried out.

In an internal 20 day precision study utilizing two levels of albumin in urine, 100% agreement was obtained between two operators and two additional observers in all instances of within-day, betweenday, within-run and between-run testing.

In the POL studies, professional and POL users participated in the evaluations of six proficiency samples over a clinically relevant range of concentrations (4.5 mg/L, 9.0 mg/L, 15 mg/L, 22 mg/L, 36 mg/L and 72 mg/L). The POL results showed an overall agreement with expected results of 91% (range of 89% 94%). This was not significantly different form the results obtained with trained laboratory personnel who reported overall agreement of 95% (range 93%).

{2}------------------------------------------------

510(K) Notification

Diagnostic Chemicals Limited

Product Cat. No. 790-01. 790-15

The majority of deviations from expected results were observed at the two sampling levels nearest the 18 mg/L level that separate the positive and negative results. Excluding the 15 and 22 mg/L samples, both groups exceeded 98% accuracy. No significant differences were observed withinrun (15 replicates), between-runs (5 different days), or between sites (3 POL sites). Results demonstrated that lay users can obtain results equivalent to those obtained by trained laboratory workers.

The performance of ImmunoDip™ Urinary Albumin Screen on clinical urine samples was compared against that of two predicate methods; the sem-quantitative DCLare™ ImmunoDip™ Stick for Microalbuminuria (K972337) and full quantitative urinary albumin testing using the Beckman Array (K922273) and the Kamiya (Crestat) Microalbumin Assay (K934146). Specificity ranged from 95-97%. Sensitivity was 95% against both methodologies. Efficiency ranged from 95-97%. These results indicate that the ImmunoDip™ Urinary Albumin Screen gives performance which is substantially equivalent to that of existing methods for urinary albumin measurement.

# CONCLUSION:

The ImmunoDip™ Urinary Albumin Screen provides a convenient method for screening for microalbuminuria. Studies indicate that use of the device by POL users provides results which are comparable to the results obtained by trained laboratory personnel. The ImmunoDip Urinary Albumin Screen provides results which are comparable to other methods currently used in chical laboratories and physicians office laboratories ..

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the top half of the circle. Inside the circle is a stylized symbol resembling three overlapping lines that curve upwards, often interpreted as representing people. The logo is presented in black and white.

#### FEB 1 2000

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Ms. Karen Callbeck, R.T., B.Sc. Regulatory Affairs Coordinator, Diagnostic Division Diagnostics Chemicals Limited 16 McCarville Street Charlottetown. PE, CIE 2A6 CANADA

Re: K994035

> Trade Name: ImmunoDip Urinary Albumin Screen Regulatory Class: I reserved Product Code: JIR Dated: November 29, 1999 Received: November 29, 1999

Dear Ms. Callbeck:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{4}------------------------------------------------

Page 2

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Steven Butman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

# INDICATIONS FOR USE

### 510(k) Number (if known): K994035

### Device Name: ImmunoDip Urinary Albumin Screen

#### Indications for Use:

There are several kidney disease conditions that can produce high levels of albumin in urine (1). Determining albumin in the wine at the low lowels measured by this test is helpful for early detection and trestment of petients it risk for reast (Listany) disease. Low but eleveled urinary albuminuria, refers to a level of the human protein albumin in nime shout 18 mg . Levels above 18 mg L are not normally found in healthy individuals. These low bet eigificant levels are not detectable with older dipstick assuys. The ImmoDig™ Uniony Albumin Screen classifies samples as postive or negative based on their being above or below a level of 18 mg/L.

Conditions in which elevated levels of albumin in urine may be present include: Type 1 and Type 2 diabetes (2-8); bypertension (9, 10); and real disease found in pregancy (11). There are other best as well. Distans is the largest single cause. One study found 45% of the insulin-timental disbetics develop secious kidney disease (3). Testing for elevelod levels of albumin in wine helps to identify those diabetics who are prone to kidney disease. Scientifics studies indicate that proper control of blood sugar) levels and blood pressure help slow or prevent kidney damage (1,9).

# References:

- Viberti, G. C., Hill, R. D., Jazett, R. J., Argyzopoulos, A., Mahmud, U. & Keen, H. Loncer 1982; i: 1430-1432. ﻠ Microalbuminuria as a predictor of clinical nephropathy in insulin-dependent diabetes mellibus.
- Viberti, G. C., Pickup J. C., Jazett, R. J., Keen, H. N Engl J Med 1979; 300: 638-41. Effect of control of blood 2 glucose on urinary excretion of albumin and fi-microglobulin in insulin-dependent diabetes.
- Mogensen, C. E. & Christensen, C. K. N Engl J Med 1984; 311: 89-93. Predicting diabetic nephropathy in insulin-3. dependent patients.
- Mogasen, C. E. N. Engl J Med 1984; 310: 356-360. Microalbuminuta proceinuria and early 4. mortality in maturity-onset diabetes.
- Deckert, T., Feldt-Rasmussen, B., Borch-Johnsen, K., Jensen, T. & Kofoel-Ezevaldsen, A. Diobetologio 1989; 32. રું. 219-226. Albuminuria reflects widespread vascular damage. The Steno hypothesis.
- Waller, K. V., Ward, K. M., Mahan, J. D. & Wissmatt, D. K. Clin Chem 1989; 35: 755-765. Curcat concepts in ર. proteinuria.
- Mattock, M. B., Keen, H., Viberti, G. C., El-Gohsti, M. R., Murrells, T. J., Scott, G. S., Wing, J. R. & Jackson, P. 7. G. Diaberalogia 1988; 31: 82-87. Coronary heart disease and urinary excention rate in Type 2 (non-insulindependent) diabetic patients.
- Alzaid, A. A., Diobetes Care 1996; 19: 79-89. MicrosIbuminuta in patients with NDDM: An Overview. 8.
- Mathiesen, E. R., Rom, B., Jensen, T., Storm, B. & Decker, T. Diabeter 1990; 39: 245-249. Relationship between 9. blood pressure and urinary albumin excretion in development of microslbuminuris.

# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE)

### Concurrence of CDRH, Office of Device Evaluation (ODE)

Patricia Bernhart for M.J. Cooper
(Division Sign-Off)

Division of Clinical Laboratory Devices 510(k) Number_K 994035

**Prescription Use**
(Per 21 CFR 801.109)

OR

Over the Counter Use

---

**Source:** [https://fda-staging.innolitics.com/device/K994035](https://fda-staging.innolitics.com/device/K994035)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
