← Product Code [JDW](/productcode/JDW) · K993910

# TAC' PIN (K993910)

_New Deal, S.A. · JDW · Feb 1, 2000 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K993910

## Device Facts

- **Applicant:** New Deal, S.A.
- **Product Code:** [JDW](/productcode/JDW.md)
- **Decision Date:** Feb 1, 2000
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3040
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The TAC'pin® is intended to be implanted for fixation of bone fractures or for bone reconstructions.

## Device Story

TAC'pin is a self-tapping, threaded metallic bone fixation wire. The device is attached to a standard surgical power tool, inserted into the bone, and cut at the bone surface by the surgeon. It is used in clinical settings for the fixation of small bone fragments in the foot. By providing compression across bone fragments, the device facilitates stabilization during osteotomy or arthrodesis procedures, aiding in bone healing and reconstruction.

## Clinical Evidence

No clinical data provided; substantial equivalence is based on technological characteristics and intended use comparisons.

## Technological Characteristics

Self-tapping threaded wire; titanium alloy construction; available in 15, 20, and 25 mm thread lengths; designed for use with standard surgical power tools.

## Regulatory Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- Threaded fixation pin (Sgarlato laboratories, [K982931](/device/K982931.md))
- Kirschner wires and Steinmann pins (Syntec-Taichung Medical Instruments, [K983121](/device/K983121.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo and contact information for a company called Newdeal. The logo is in a stylized font, and the contact information includes the company's address, phone number, fax number, and email address. The address is 31 Rue de la Convention, Parc D'Activités Garigliano, 38200 Vienne, France. The phone and fax numbers are (33) 04 74 78 15 15 and (33) 04 74 78 15 16, respectively, and the email address is NEWDEALFR@AOL.COM.

#### 3. SUMMARY OF SAFETY AND EFFECTIVENESS

### SPONSOR IDENTIFICATION A.

NewDeal SA Parc d'Activitiés Garigliano Rue de la Convention 38 200 VIENNE FRANCE

Tél. : (33) 4 74 78 15 15 Fax : (33) 4 74 78 15 16

#### B. ESTABLISHMENT REGISTRATION NUMBER: 9615741

### OFFICIAL CONTACT PERSON C.

Norman F. Estrin, Ph. D., RAC President Estrin Consulting Group, Inc. 9109 Copenhaver Drive Potomac , MD 20854

Tel. : (301) 519-1098 Fax : (301)519-1389

## D. DATE OF PREPARATION OF THIS SUMMARY: November 15, 1999

- E. PROPRIETARY (TRADE) NAME: TAC'pin®
- F. COMMON NAME: Bone fixation wire self tapping threaded wire

1 2000

{1}------------------------------------------------

K9939/6
pg 2082

- CLASSIFICATION NAME AND REFERENCE ઉં, Smooth or threaded metallic hone fixation fastener (21 CFR, Section 888.3040)
- PROPOSED REGULATORY CLASS: Class II BL
- DEVICE PRODUCT CODE: 87JDW I.
- 87 OR PANEL CODE: J.
- DESCRIPTION OF DEVICE: The TAC'pin® is a self tapping wire. One part K is fixed on a standard surgical power tool equipment and when the threaded wire is introduced and positioned as required, the surgeon has to cut the wire at the bone surface. The lengths of threads available are the following : 15 - 20 - 25 mm.
- The TAC'pin® is intended to be implanted for fixation of INTENDED USE: L. bone fractures or for bone reconstructions.
- INDICATIONS FOR USE: The TAC'pin® is indicated for synthesis of small M. bones fragments, in the foot only. Examples include: -Akin type osteotomy -first MP Arthrodesis -phalangeal arthrodesis -small bones osteosynthesis requiring compression.
- N. The TAC'pin® is substantially equivalent to the PREDICATE DEVICE Threaded fixation pin manufactured by Sgarlato laboratories (K982931), and the Kirschner wires and Steinmann pins manufactured by Syntec-Taichung Medical Instruments (K983121).

#### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS 0.

Both the TAC'pin®, the Sgarlato threaded fixation pin, and the Kirschner wires and Steinmann pins from Syntec-Taichung Medical are intended to be implanted for fixation of small bone fragments. Both the TAC'pin® and the Sgarlato threaded fixation pin are made from the same titanium alloy, whereas the Kirschner wires and Steinmann pins from Syntec-Taichung Medical are made from stainless steel. This difference has no significant consequences on the safety or effectiveness of the product. The TAC pin®, the Sgarlato Threaded fixation pin, and the Kirschner wires and Steinmann pins manufactured by Syntec-Taichung Medical Instruments are all threaded pins with comparable ranges of sizes and diameters.

000010

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus or a representation of human figures.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

# FEB - 1 2000

NewDeal SA C/O Norman F. Estrin, Ph.D., RAC President Estrin Consulting Group, Inc. 9109 Copenhaver Drive Potomac, Maryland 20854

Re: K993910

Trade Name: TAC' pin® Regulatory Class: II Product Code: JDW Dated: November 15, 1999 Received: November 17, 1999

Dear Dr. Estrin:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed

{3}------------------------------------------------

Page 2 - Norman F. Estrin, Ph.D., RAC

predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597, or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

Russell Sage

Ser James E. Dillard III Acting Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# k993910

Page 1 of 1

510(k) Number (if known):

TAC' pin ® Device Name:

Indications for Use:

The TAC'pin® is indicated for synthesis of small bones fragments, in the foot only. Examples include:

> -Akin type osteotomy -first MP Arthrodesis -phalangeal arthrodesis -small bones osteosynthesis requiring compression.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

× Prescription Use (Per 21 CFR 801.109)

OR

Over-the-Counter Use

(Optional Format 1-2-96)

ell Payne Surr

(Division Sign-Off) Division of General Restorative Devices 510(k) Number

000008

---

**Source:** [https://fda-staging.innolitics.com/device/K993910](https://fda-staging.innolitics.com/device/K993910)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
