K993893 · Medx, Inc. · KPS · Feb 10, 2000 · Radiology
Device Facts
Record ID
K993893
Device Name
MEDX INTECAM INTERFACE
Applicant
Medx, Inc.
Product Code
KPS · Radiology
Decision Date
Feb 10, 2000
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1200
Device Class
Class 2
Indications for Use
The intended uses of the MEDX InteCam™ Interface accessory together with the NuQuest™ Nuclear Medicine Computer are the acquisition, processing, display and analysis of planar and SPECT nuclear images of all organ systems using approved pharmaceuticals.
Device Story
The MEDX InteCam™ Interface and NuQuest™ Nuclear Medicine Computer acquire, process, display, and analyze nuclear medicine data. The system receives x, y coordinate signals and energy data from gamma camera scintillation detectors following radioisotope injection. The computer replaces traditional CRT/film imaging, enabling immediate viewing, non-linearity correction, and quantitative analysis. It performs tomographic reconstruction from multiple detector positions to generate 3D representations or 2D slices. Operators (technologists/physicians) use the system to observe isotope uptake/clearance, assess tissue viability, and identify abnormal activity (e.g., cancerous tissue). Output is displayed on a monitor or transferred to film. The device facilitates clinical decision-making by providing high-contrast functional imaging of organ systems.
Clinical Evidence
No clinical data provided; substantial equivalence is based on technological characteristics and functional comparison to the predicate device.
Technological Characteristics
System consists of the InteCam™ Interface and NuQuest™ Nuclear Medicine Computer. Operates by processing electrical signals from photomultiplier tubes connected to scintillation crystals. Supports planar and SPECT imaging. Digital computer architecture enables tomographic reconstruction, quantitative count analysis, and temporal observation of radioisotope kinetics.
Indications for Use
Indicated for acquisition, processing, display, and analysis of planar and SPECT nuclear medicine images of all organ systems in patients undergoing diagnostic procedures with approved radiopharmaceuticals.
Regulatory Classification
Identification
An emission computed tomography system is a device intended to detect the location and distribution of gamma ray- and positron-emitting radionuclides in the body and produce cross-sectional images through computer reconstruction of the data. This generic type of device may include signal analysis and display equipment, patient and equipment supports, radionuclide anatomical markers, component parts, and accessories.
{0}------------------------------------------------
K993893
Page 1 of 4
## 510(k) SUMMARY
- Submitter Information: 1.
- 1.1 Submitter MEDX Incorporated 3456 N. Ridge Ave., #100 Arlington Heights, IL 60004 Phone: (847) 463-2020 FAX (847) 463-2019
- 1.2 Manufacturing Facility MEDX Incorporated 3456 N. Ridge Ave., #100 Arlington Heights, IL 60004
- 1.3 Contact: Floyd Rowan
- Date: November XX, 1999 1.4
- 2. Device Name
- Tomographic Gamma Camera Systems 2.1
- Camera, Scintillation (Gamma) 2.2 Classification Name: 901YX Classification Number: Camera Tomographic Nuclear 2.3 Classification Name: 90JWM Classification Number: MEDX InteCam™ Interface 2.4 Trade/Proprietary Name: G.E. Starcam 4000 2.5 Predicate Device: (DC K904174)
- 2. Device Description
- 3.1 Function
The NuQuest™ Nuclear Medicine Computer together with the MEDX InteCam™ Interface is used to acquire data from nuclear medicine gamma cameras. The data consists of a representation of
{1}------------------------------------------------
K993893
the x coordinate and y coordinate of the position and optionally a signal representing the energy of a gamma ray emitted from the decay of radioisotopes injected into a patient. The position of the decay is calculated (a ray from the event to the detector) by the gamma camera. The positions of a large number of decay events are used to form an image of the location of the radioactive material.
This image was originally produced by using photographic film exposed by a Cathode Ray Tube (CRT). A flash of light was produced on the CRT in the proper x,y location of the event. A large number of events produced an image on the film.
The substitution of a computer for the film allowed immediate viewing of the image, a more accurate rendering of the image since the non-linearities present in film were removed, and the capability for quantitative analysis of the data.
This image acquired can be displayed on a computer monitor or transferred to photographic film for review. The collection of data at multiple detector positions allows three dimensional information to be obtained by tomographic means. The computer is then used to reconstruct slices, or two dimensional views, of the data viewed from any angle or orientation.
The computer can also be used to obtain quantitative information such as the number of counts in a particular area of the image, or the rate of change of counts in an area of the image with respect to time.
Unlike most other diagnostic imaging devices (MR, CT, ultrasound). nuclear medicine devices emit no radiation, but rely on decav events from radioisotopes injected into a patient by a licensed technologist or physician.
## 3.2 Scientific Concepts
Diagnostic Nuclear Medicine began in early 1950's with the availability of short half-life radioisotopes. Isotopes such as I131 were injected into the patient and were selectively taken up by organ systems such as the thyroid. Measurement of the resulting
{2}------------------------------------------------
K993893
Page 3 of 4
radioactivity in the organ provided information on both the size of the organ and the relative amount of isotope taken up.
Nuclear Medicine cameras work on a principle similar to television cameras. A collimator (lens) ""focuses" gamma rays on a scintillation crystal. The scintillation crystal converts the gamma rays into light. Photomultiplier tubes are then used to convert the light into an electrical signal proportional to the energy of the detected gamma ray. Early instruments used a single hole lead collimator and detector that was moved in a raster pattern foming a 2-D image of the organ of interest. In the late 1950's methods were developed for directly obtaining a 2-D image by using a large crystal with multiple photomultiplier tubes and electronically calculating the position and energy of the gamma event. Two dimensional projections collected at many positions can be mathematically combined to vield a three dimensional representation of the data. This principle of tomographic reconstruction was discovered early in this century, but it was not until the advent of high speed digital computer that the technique could be successfully applied in diagnostic imaging first to CT then to Nuclear Medicine and MRI.
- 3.3 Nuclear Medicine is currently of great interest because of its high contras, and relatively low cost per study. The ability to attach isotopes to substances that are selectively taken up by specific tissue types can provide very high contrast between the tissue and organ of interest and surrounding tissue. This has tended to compensate for the relatively poor spatial resolution of Nuclear Medicine compared to other modalities such as MRI.
In addition the uptake and clearing of the radioisotopes can be observed temporally, providing an indication of the biological activity of the tissue. This is important when attempting to determine tissue viability, or finding areas of abnormal activity such as cancerous tissue.
- 4.0 Device Intended Use:
- 4.1 The intended uses of the MEDX InteCam™ Interface and the NuQuest Nuclear Medicine Computer are the acquisition. processing, display and analysis of planar and SPECT nuclear
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the number "K993893" in a handwritten style. Below the number, the text "Page 4 of 4" is present, indicating that this is the last page of a document. The text is in a simple, sans-serif font.
medicine images of all organ systems using approved radiopharmaceuticals.
- 5.0 Device Technological Characteristics:
- The characteristics of the MEDX InteCam™ Interface when used 5.1 with the NuQuest™ Nuclear Medicine Computer compare substantially with the predicate device, in both materials used, technology applied, and functional methodology. Differences of note do not affect safety and effectiveness of the device, intended use, or application methods. The device operates in a manner substantially equivalent to other cleared devices in this category, and performs as well as the predicate device.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES" around the perimeter. Inside the circle is a stylized image of a human figure with three wavy lines extending from the head. The logo is black and white.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
FEB 1 0 2000
Floyd R. Rowan Executive Vice President Medx, Incorporated 3456 N. Ridge Ave. #100 Arlington Heights, IL 60004 Re:
K993893 MEDX InteCam™ Interface Dated: November 15, 1999 Received: November 16, 1999 Regulatory class: II 21 CFR 892.1200/Procode: 90 KPS
Dear Mr. Rowan:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been redassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/odth/dsma/dsmamain.html".
Sincerely yours,
CAPT Daniel G. Schultz, M.D. Acting Director, Division of Reproductive, Abdominal, Ear, Nose and Throat. and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Page_1 of_1
510(k) Number (if known):_____________________________________________________________________________________________________________________________________________________
Device Name:_MEDX InteCam™ Interface
Indications For Use:
ं
The intended uses of the MEDX InteCam™ Interface accessory together with the NuQuest™ Nuclear Medicine Computer are the acquisition, processing, display and analysis of planar and SPECT nuclear images of all organ systems using approved pharmaceuticals.
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use<br>(Per 21 CFR 801.109) | |
|------------------------------------------|--|
|------------------------------------------|--|
(Optional Format 3-10-98)
| (Division Sign-Off)<br>Division of Reproductive, Abdominal, ENT,<br>and Radiological Devices | |
|----------------------------------------------------------------------------------------------|---------|
| 510(k) Number | K993893 |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.