GENX CULTURE DISH

K993881 · <Genx> Intl., Inc. · MQK · Feb 14, 2000 · Obstetrics/Gynecology

Device Facts

Record IDK993881
Device NameGENX CULTURE DISH
Applicant<Genx> Intl., Inc.
Product CodeMQK · Obstetrics/Gynecology
Decision DateFeb 14, 2000
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 884.6160
Device ClassClass 2

Indications for Use

The <genX> Culture Dish is intended to be used when culturing tissues and cells. It may be used for the use with sperm and for the culturing of embryos.

Device Story

The <genX> Culture Dish is a 60mm round laboratory vessel with 1.5-inch sidewalls, designed for assisted reproduction. It features internal picket-fenced wall structures that create defined areas to confine cells or embryos while allowing the free flow of fluids, nutrients, and metabolic byproducts. The device is used in clinical laboratory settings by embryologists or laboratory technicians to support the culture of tissues, cells, sperm, and embryos. It serves as a passive containment environment during the incubation process. The device is provided sterile and is intended for single-use applications.

Clinical Evidence

No clinical data; bench testing only. Sterilization validation performed in accordance with ANSI/AAMI/ISO 11137-1994 standards to a Sterility Assurance Level (SAL) of 10^-6.

Technological Characteristics

60mm round culture dish with 1.5-inch sidewalls; features internal picket-fenced wall partitions for cell confinement. Sterilization via Gamma Irradiation (2.5 Mrad). Complies with ANSI/AAMI/ISO 11137-1994 for sterilization validation.

Indications for Use

Indicated for culturing tissues and cells, including holding sperm and culturing embryos in assisted reproductive procedures.

Regulatory Classification

Identification

Assisted reproduction labware consists of laboratory equipment or supplies intended to prepare, store, manipulate, or transfer human gametes or embryos for in vitro fertilization (IVF), gamete intrafallopian transfer (GIFT), or other assisted reproduction procedures. These include syringes, IVF tissue culture dishes, IVF tissue culture plates, pipette tips, dishes, plates, and other vessels that come into physical contact with gametes, embryos or tissue culture media.

Special Controls

*Classification.* Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, and clinical testing). The device, when it is a dish or plate intended for general assisted reproduction technology procedures, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.

Submission Summary (Full Text)

{0}------------------------------------------------ ## FEB 1 4 2000 Page 1 of 2 <genX> international, Inc. 510(K) Premarket Notification ## 510(K) Summary K993881 - DEVICE NAME a) Proprietary Name: <genX> Culture dish Classification Name: Assisted Reproduction Labware b)Submitted bv: <genX> international, inc. 170 Fort Path Rd. Madison, CT 06443 ESTABLISHMENT REGISTRATION No.: 9003605 Telephone number: 203-245-4901 Fax number: 203-245-4994 E-mail: genXintl@aol.com Contact individual: Michael D. Cecchi President - CLASSIFICATION: Class II c) Assisted Reproductive Labware Product: <genX> Culture dish Procode: 85 MQK CFR#: 884.6160 - d) PERFORMANCE STANDARDS Performance Standards under Section 514 of the ACT have not been developed for this device. However, Special controls have been identified by the FDA to provide reasonable assurance of safety and effectiveness of the device in assisted reproductive procedures. - e) PROPOSED LABELS, LABELING, AND ADVERTISING The proposed labeling and instruction material is included in this package in the appropriate section. - f) DESCRIPTION AND INTENDED USE The <genX> Culture Dish is intended to be used when culturing tissues and cells. It may be used for the use with sperm and for the culturing of embryos. The catalog number is GDEC-100 {1}------------------------------------------------ K993881 Page 2 of 2 The basic design is that of a traditional culture dish. Approx 60mm round with 1½ inch sidewalls. This dish does have defined areas for the cell to be confined within. These areas are squares of picket-fenced walls, which will allow fluids, nutrients and byproducts to freely flow among the culturing cell, embryos. ## SUBSTANTIALLY EQUIVALENT ದಿ) The <genX> Culture Dish is substantially equivalent to several products currently marketed in the US. See the appropriate section for the comparisons. (Section 2) - 1. Same intended use - 2. Same design - 3. Same materials configured the same for use. - 4. Perform similar procedures, by method. - 5. Same performance criteria - 6. Same sterility methods and testing - 7. Same intended population ## h) STERILIZATION PROCEDURES AND FACILITIES All dishes manufactured, will be Gamma Irradiated by and according to the guidelines in place. Dosage: 2.5 Mrad Sterility Assurance Level (SAL): of (10 to the -6 ) After sterilization the contractor will issue a "Certificate of Sterilization". Validation assurance is in accordance with ANSI / AAMI / ISO 11137-1994 Standards. See Section 11.0 {2}------------------------------------------------ Image /page/2/Picture/2 description: The image shows the address of the Food and Drug Administration. The address is 9200 Corporate Boulevard, Rockville MD 20850. The text is black and the background is white. FEB 1 4 2000 Mr. Michael D. Cecchi President <genX> international, Inc. 170 Fort Path Road Madison, CT 06443 Re: K993881 <genX> Tissue Culture Dish Dated: November 15, 1999 Received: November 16, 1999 Regulatory Class: II 21 CFR §884.6110/Procode: 85 MQK Dear Mr. Cecchi: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html". Sincerely yours, CAPT Daniel G. Schultz, M.D. Acting Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ <genX> international, Inc. 510(K) Premarket Notification INDICATIONS FOR USE 49 2881 510 (k) Number (if known) Device Names: <genX> Tissue Culture Dish Indication for Use: "To culture tissues and cells" May be used to hold sperm and culture embryos (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation ( ODE ) Prescription Use X Over the Counter Use *_*_ (Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological Devices 510(k) Number. or
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...