← Product Code [KGO](/productcode/KGO) · K993857

# THE ORIGINAL PERRY, STYLE 42 WHITE LATEX POWDERED SURGICAL GLOVES WITH PROTEIN CONTENT LABELING CLAIM (60 MICROGRAMS OR (K993857)

_Ansell Perry · KGO · Jan 31, 2000 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K993857

## Device Facts

- **Applicant:** Ansell Perry
- **Product Code:** [KGO](/productcode/KGO.md)
- **Decision Date:** Jan 31, 2000
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4460
- **Device Class:** Class 1
- **Review Panel:** General, Plastic Surgery

## Indications for Use

The Original Perry Style 42 Surgeon's Glove intended use is to be worn by operating room personnel to protect a surgical wound from contamination.

## Device Story

The Original Perry Style 42 Powdered Surgeon's Gloves are sterile, disposable latex gloves. They are designed for use by operating room personnel to provide a barrier against contamination of surgical wounds. The gloves are manufactured to meet ASTM D 3577-99 standards for physical properties, dimensions, and freedom from holes. A specific feature of this device is a protein label claim, indicating a total water-extractable protein content of 60 micrograms or less per gram, verified via ASTM D 5712-95. The device is used in clinical surgical settings to maintain sterile fields and protect both the patient and the clinician.

## Clinical Evidence

No clinical data provided. Substantial equivalence is based on bench testing, including physical property testing, freedom from holes testing per ASTM D 5151-92, and protein content analysis per ASTM D 5712-95, alongside biocompatibility testing (primary skin irritation and guinea pig sensitization).

## Technological Characteristics

Material: Latex. Standards: ASTM D 3577-99 (Type 1), ASTM D 5151-92 (freedom from holes), ASTM D 5712-95 (protein content). Protein content: ≤ 60 micrograms/gram. Biocompatibility: Passes primary skin irritation and guinea pig sensitization tests. Form factor: Powdered surgical glove.

## Regulatory Identification

A non-powdered surgeon's glove is a device intended to be worn on the hands of operating room personnel to protect a surgical wound from contamination. A non-powdered surgeon's glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Jan 3 1 2000

K993857

. .

Attachment 4

## 510(k) Summary

Summary

Attached

Signature

James R. Chatterton
Typed Name

November 12, 1999
Date

{1}------------------------------------------------

K99 3857

The Original Perry Style 42 Powdered Surgeon's Gloves (Protein Label Claim) Ansell Perry 1875 Harsh Avenue SE Massillon, Ohio 44646 Telephone: 330-833-2811 330-833-6213 Fax:

- [1] Summary
- [2] Ansell Perry Inc. 1875 Harsh Avenue SE Massillon, Ohio 44646

| Contact:   | James R. Chatterton |
|------------|---------------------|
| Telephone: | 330-833-2811        |
| Fax:       | 330-833-6213        |

November 12, 1999

- [3] Trade Name: The Original Perry Style 42 Powdered Surgeon's Gloves (Protein Label Claim) Common Name: Surgical Gloves Classification Name: Surgeon's Glove
- [4] The Original Perry Style 42 Powdered Surgeon's Gloves (Protein Label Claim) meet all of the requirements of ASTM D 3577-99, Type 1.
- [5] The Original Perry Style 42 Powdered Surgeon's Gloves (Protein Label Claim) meet all the current specifications for ASTM D 3577-99 Rubber Surgical Gloves.
- The Original Perry Style 42 Powdered Surgeon's Gloves (Protein Label Claim) are sterile disposable [e] devices intended to be worn by operating room personnel to protect a surgical wound from contamination.
- The Original Perry Style 42 Powdered Surgeon's Gloves (Protein Label Claim) are summarized with [7] the following technological characteristics compared to ASTM or equivalent standards.

| Characteristics     | Standard                                                                                                                                                                                                         |
|---------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Dimensions          | Meets ASTM D 3577-99                                                                                                                                                                                             |
| Physical Properties | Meets ASTM D 3577-99, Type 1                                                                                                                                                                                     |
| Freedom from holes  | Meets ASTM D 3577-99<br>Meets ASTM D 5151-92                                                                                                                                                                     |
| Protein Label Claim | This latex glove contains 60 micrograms or<br>less of total water extractable protein per<br>gram.<br>Meets ASTM D 5712-95 Standard Test<br>Method for Analysis of Protein in Natural<br>Rubber and Its Products |
| Biocompatability    |                                                                                                                                                                                                                  |

Primary Skin Irritation in Rabbits Guinea Pig Sensitization

Passes Passes

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows a handwritten string of characters. The string appears to be "KG93857". The characters are written in a dark ink on a white background. The handwriting is somewhat stylized, with some characters having unique shapes.

The Original Perry Style 42 Powdered Surgeon's Gloves (Protein Label Claim) Ansell Perry 1875 Harsh Avenue SE Massillon, Ohio 44646 Telephone: 330-833-2811 330-833-6213 Fax:

- The performance test data of the non clinical tests are the same as mentioned immediately above. [8]
- Clinical data is not needed for medical gloves or for most devices cleared by the 510(k) process. വി
- It is concluded that The Original Perry Style 42 Powdered Surgeon's Gloves (Protein Label Claim) [10] are as safe, as effective, and perform as well as the glove performance standards referenced above and therefore meet:

ASTM listed standards, FDA hole requirements, and labeling claims for the product.

- [11] This summary will include any other information reasonably deemed necessary by the FDA.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, with three stylized human profiles incorporated into the design.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JAN 31 2000

Mr. James R. Chatterton Vice President Regulatory Ansell Perry Ansell Healthcare Products, Incorporated 1875 Harsh Avenue S.E. Massillon, Ohio 44646

Re : K993857 Trade Name: The Original Perry, Style 42 White Latex Powdered Surqical Gloves With Protein Content Labeling Claim (60 micrograms or less) Regulatory Class: I Product Code: KGO Dated: November 12, 1999 November 15, 1999 Received:

Dear Mr. Chatterton:

We have reviewed your Section 510 (k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual reqistration, listing of devices, good manufacturing practice, labeling, and prohibitions aqainst misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in requlatory action. In addition, FDA may publish further announcements concerning your device in

{4}------------------------------------------------

Page 2 - Mr. Chatterton

the Federal Register. Please note: this response to your premarket notification submission does not affect any premarker nocurright have under sections 531 through 542 of obligation you magne the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as fire ited in your 510 k) premarket notification. The FDA debtired in form of substantial equivalence of your device to a legally marketed predicate device results in a classification for your marketed production permits your device to proceed to the market.

If you desire specific advice for your device on our labeling II you debire bpoort 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4692. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be Information on Journal Manufacturers Assistance obeation free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

Timothy A. Ulatowski

Diredtor Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

#### Attachment 2

### Indications for Use Statement

| 510(k) Number (if known) | K993857<br>None, preamendments device                                                                                                                 |
|--------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name              | The Original Perry Style 42<br>White latex powdered surgical gloves with Protein Content Labeling Claim<br>(60 micrograms or Less)                    |
| Indications for Use      | The Original Perry Style 42 Surgeon's Glove intended use is to be worn by<br>operating room personnel to protect a surgical wound from contamination. |

# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED

### Concurrence of CDRH Office of Device Evaluation (ODE)

Prescription Use Per 21 CFR 801.109

: 中国 上海

出している。

OR

Over-The-Counter Use \$\longrightarrow\$

Qiu S. Lin

(Division Sign-Off) Division of Dental, Infectio and General Hospita 510(k) Number _

10

---

**Source:** [https://fda-staging.innolitics.com/device/K993857](https://fda-staging.innolitics.com/device/K993857)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).