The Humid-Heat™ is intended for use as an active respiratory humidifier to heat and humidify dry breathing gases supplied from a ventilator via an endotracheal tube, a tracheostomy tube, or a face mask to adult patients under intensive care or anesthesia.
Device Story
Humid-Heat™ is an active respiratory humidifier for adult patients in intensive care or anesthesia. It receives dry breathing gases from a ventilator; pumps external water onto a wick within an HME; evaporates water via an electrical heater to provide physiologically optimized gases (37°C, 100% relative humidity). System components include supply unit, heater, temperature probe, HME, and water feed set. Operated via a four-button keypad (minute volume setting, start/stop, power). Device eliminates condensation in circuits, removes need for water traps or heated wires, and reduces water consumption. By preventing delivery of cool, dry gases, it mitigates thickened mucus and decreased cilia activity, supporting long-term ventilation.
Clinical Evidence
Bench testing only. Testing included electrical safety (IEC 601-1), performance, and environmental testing. No clinical data provided.
Technological Characteristics
Active respiratory humidifier. Components: supply unit, heater, temperature probe, HME, water feed set. Principle: external water pumped onto wick, evaporated by electrical heater. Output: 37°C ± 2°C; 44 mg H2O/L (active), 30 mg H2O/L (passive). Compatible with volume and pressure-controlled ventilators. Keypad interface. Electrical safety per IEC 601-1.
Indications for Use
Indicated for adult patients under intensive care or anesthesia requiring mechanical ventilation via endotracheal tube, tracheostomy tube, or face mask to receive heated and humidified breathing gases.
Regulatory Classification
Identification
A respiratory gas humidifier is a device that is intended to add moisture to, and sometimes to warm, the breathing gases for administration to a patient. Cascade, gas, heated, and prefilled humidifiers are included in this generic type of device.
{0}------------------------------------------------
2
# SECTION 2. SUMMARY AND CERTIFICATION
## A. 510(K) SUMMARY
## Summary of Safety and Effectiveness
In accordance with 21 CFR 807.92, the following information constitutes the Hudson RCI AB summary for the Humid-Heat™
| SUBMITTER'S NAME: | Hudson RCI AB (Formerly Louis Gibeck AB) |
|---------------------|-----------------------------------------------|
| ADDRESS: | Box 711<br>SE-194 27 Upplands Väsby<br>Sweden |
| CONTACT PERSON: | Constance Bundy |
| TELEPHONE NUMBER: | 612-574-1976 |
| FAX NUMBER: | 612-571-2437 |
| DATE OF SUBMISSION: | November 1, 1999 |
#### Identification of device 1.
Proprietary Name: Humid-Heat™ Common Name: Respiratory gas humidifier Classification Status: Class II per regulations 868.5450 Product Codes: 73 BTT
#### Equivalent devices 2.
Hudson RCI AB believes the Humid-Heat™ is substantially equivalent to Fisher & Paykel Dual Servo Controlled Heated Respiratory Humidifiers, cleared for marketing under 510(k) K913368.
#### Description of the Device ર્જ
Humid-Heat™ is an active respiratory humidifier that heats and humidifies dry breathing gases supplied from a ventilator via an endotracheal, tracheostomy tube or a face mask to adult patients under intensive care or anesthesia. Patients receive physiologically optimized breathing gases (37°C, 100% relative humidity) instead of the dry and cool breathing gases they would otherwise receive if connected directly to a ventilator. They thus avoid the thickened mucus, decreased cilia activity and secretions trapped in the lower airways that are often caused by exposure to cool, dry gases. By heating and humidifying the supplied breathing gases, Humid-Heat™ overcomes these problems and provides optimal conditions for long-term ventilation.
Heating and humidification take place close to the patient, thereby eliminating the problem of condensation in ventilator circuits. Humid-Heat™ does not require any water traps or heated wires and water consumption is reduced. Humid-Heat™ is suitable for use with volume and pressure-controlled ventilators.
6
{1}------------------------------------------------
The Humid-Heat™ consists of five main components: the supply unit, the heater, the temperature probe, the HME and the water feed set. It is controlled with a keypad consisting of four buttons; up/down buttons to set the minute volume of the ventilator, start button, stop button, and the power switch.
#### Intended use 4.
The Humid-Heat™ is intended for use as an active respiratory humidifier to heat and humidify dry breathing gases supplied from a ventilator via an endotracheal tube, a tracheostomy tube, or a face mask to adult patients under intensive care or anesthesia.
#### Technological characteristics, comparison to predicate device. ਪੰਜ
Like the predicate devices, the Humid-Heat™ is intended to add moisture to, and to warn, the breathing gases for administration to a patient.
| Characteristic | Predicate device: Fisher &<br>Paykel MR850, K983112 | Humid-Heat ™ |
|------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | To warm and add humidity to<br>gases delivered to patients<br>requiring mechanical<br>ventilation. | Same |
| Airway Temperature<br>output | 30-39° C marked<br>29-40° C actual | 37° C± 2° C |
| Moisture output | At least 30 mg H2O/L if<br>temperature of gas delivered<br>to patient is set to 37° C for a<br>given continuous flow. | 44 mg H2O/L active VT 800 ml<br>30 mg H2O/L passive VT 800 ml |
| Technical background | The warmth and moisture are<br>supplied by passing the<br>breathing gas over heated<br>water while the temperature<br>of the gas flowing through the<br>breathing circuit is<br>maintained by a heated wire. | The supply unit pumps external<br>water onto a wick in the HME.<br>The water on the wick is<br>evaporated by an electrical<br>heater. |
| Ventilator systems | Mechanical ventilation or<br>positive pressure breathing<br>assistance | Volume and pressure-controlled<br>ventilators |
| Passive humidification | Maximal 20 mg H2O/L | 30 mg H2O/L passive VT 800 ml |
Comparison table
#### 6. Discussion of performance testing.
An extensive collection of tests have been conducted and successfully completed, including electrical safety (IEC 601-1), performance and environmental testing.
7
{2}------------------------------------------------
#### 7. Conclusion
1.00
It is the conclusion of Hudson RCI AB that the Humid-Heat™ is substantially equivalent to devices already on the market (cleared by the 510(k) process) and presents no new concerns about safety and effectiveness.
1999 - 1999
:
:
.
8
: :
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image is a black and white seal for the Department of Health & Human Services - USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an image of an eagle with its wings spread.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
SEP 1 3 2000
Hundson RCI AB c/o Ms. Constance G. Bundy 6470 Riverview Terrace Fridley, MN 55432
K993773 Re : Hudson RCI AB Humid-Heat™ Regulatory Class: II (two) Product Code: 73 BTT August 17, 2000 Dated: August 18, 2000 Received:
Dear Ms. Bundy:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is che device foreioned (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and In actoration with anoou therefore, market the device, subject to the general controls provisions of the Act. The general controls enc general of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special II yours or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such Failure to comply with the GMP regulation may result in assumptions. regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this conse to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}------------------------------------------------
### Page 2 - Ms. Constance G. Bundy
This letter will allow you to begin marketing your device as described inis lecter will arrow you co icention. The FDA finding of substantial in your 510 (x) promainsice to a legally marketed predicate device equivalence of your device of a region and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling II you destre bpooriis and additionally 809.10 for in vitro reguration (=vices), please contact the Office of Compliance at Graghosere devices // plonally, for questions on the promotion and (Jor) 53 1010. Hauter, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, (901) 37 1000 by reference to premarket notification" (21 CFR 807.97). firstranding by rereation on your responsibilities under the Act may ocher general information of Small Manufacturers Assistance at its be obtained from the 22138-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
James B. Dillard III
Director
Directbr Division of Cardiovascular and Respiratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Page 1 of 1
## B. INDICATIONS FOR USE
510(k) Number K993773
Device Name: Humid-Heat™
## Indications for Use:
The Humid-Heat™ is an active respiratory humidifier that heats and humidifies dry rile fruinid-riour - 18 an actro respirator via an endotracheal tube, a tracheostomy tube, or a face mask to adult patients under intensive care or anesthesia.
(Please do not write below this line - continue on another page if needed)
| Concurrence of CDRH, Office of Device Evaluation (ODE) | |
|--------------------------------------------------------|--|
|--------------------------------------------------------|--|
| (Division Sign-Off) | |
|-------------------------------------------------------------------|---------|
| Division of Cardiovascular, Respiratory, and Neurological Devices | |
| 510(k) Number | K993773 |
| Prescription Use | <div style="text-align:center;">✓</div> |
|----------------------|-----------------------------------------|
| (Per 21 CFR 801.109) | |
| OR | |
| Over the Counter Use | |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.