K993758 · Medical Compression Systems (Dbn) , Ltd. · JOW · May 19, 2000 · Cardiovascular
Device Facts
Record ID
K993758
Device Name
PROAIR 3000 COMPRESSION SYSTEM
Applicant
Medical Compression Systems (Dbn) , Ltd.
Product Code
JOW · Cardiovascular
Decision Date
May 19, 2000
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.5800
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The ProAir 3000 is a prescriptive device that induces controlled compression of the calf, the foot or combined compression of both. The ProAir 3000 is intended for use by patients and medical professionals in treating many conditions, such as: - prevention of deep vein thrombosis (DVT) . - enhancement of blood circulation - reduction of post-operative pain and swelling . - reduction of wound-healing time - stasis dermatitis . - treatment and assist healing of cutaneous ulceration . - venous stasis ulcers . - leg ulcers - chronic venous insufficiencies . - reduction of edema
Device Story
ProAir 3000 is a prescriptive, portable pneumatic compression system for lower limbs. Components include a control unit, calf/foot cuffs, connecting tubes, and electrical transformer. Device applies sequential compression via solenoid-controlled inflation cycles; operates in calf, foot, or combined modes. Used by patients and clinicians in clinical or home settings. Software-based protection monitors pressure via sensors; mechanical valve provides over-pressure relief. Output is controlled pneumatic compression to limbs. Benefits include improved circulation, DVT prevention, and accelerated wound healing.
Clinical Evidence
Comparative study performed to demonstrate safety and performance equivalence to predicate devices. No specific clinical trial data or statistical metrics (e.g., p-values, CIs) provided in the summary.
Technological Characteristics
Pneumatic compression system; portable control unit with solenoid valves and pressure sensors. Complies with IEC 60601-1, IEC 60601-1-2, EN-1441, and Mil-Std-810E. Includes mechanical over-pressure relief valve. Software-based pressure monitoring.
Indications for Use
Indicated for patients requiring controlled compression of the calf and/or foot for DVT prevention, circulation enhancement, pain/swelling reduction, wound healing, and treatment of venous stasis/ulcers, stasis dermatitis, and edema. Contraindicated in patients with gangrene, recent skin grafts, severe arteriosclerosis/ischemic vascular disease, congestive cardiac failure, massive/pulmonary edema, existing DVT, acute thrombophlebitis, acute infections, or pulmonary embolism.
Regulatory Classification
Identification
A compressible limb sleeve is a device that is used to prevent pooling of blood in a limb by inflating periodically a sleeve around the limb.
Predicate Devices
Kendall Model 6325 SCD Sequel Compression System (K942664)
Talley TM300 Sequential Multicom Compression System (K915092)
{0}------------------------------------------------
K993758
# MAY 1 9 2000
## 510(K) SUMMARY
# ProAir 3000 Compression System
510(k) Number K_______________________________________________________________________________________________________________________________________________________________
# Applicant's Name:
Medical Compression Systems (DBN) Ltd. 9 Harugei Malhut st. Tel-Aviv 69714, Israel Tel.: 972-3-647-1615 Fax: 972-3-647-0293
### Contact Person:
Shoshana Friedman, RAC Push-med Ltd. 117 Ahuzah St. Ra'anana 43373, Israel Tel: 972-9- 7718130 Fax: 972-9-7718131
# Date Prepared:
October, 1999
### Trade Name:
ProAir 3000 Compression System
# Classification Name:
Compressible Limb Sleeve
# Classification:
The FDA has classified compressible limb sleeves as class II devices (product code JOW, Regulation No. 870.5800) and they are reviewed by the Cardiovascular Panel.
{1}------------------------------------------------
# Predicate Device:
- · Kendall Model 6325 SCD Sequel Compression System (Kendall Co.), cleared under K942664
- · Talley TM300 Sequential Multicom Compression System (Progressive Medical Technology, Inc.) cleared under K915092
- · Jobst Extremity Pump System 7500 (Jobst Institute, Inc.) cleared under K921562
- · Cowboy XV [PlexiPulse®] (KCI New Technologies, Inc.) cleared under K981311
## Performance Standards:
No performance standards have been established for such device under Section 514 of the Federal Food, Drug, and Cosmetic Act. However, the ProAir 3000 complies with the following voluntary standards: IEC 60601-1 (and amendments), IEC 60601-1-2. EN-1441, and applicable parts of Mil-Std-810E.
## Indications:
The ProAir 3000 is a prescriptive device that induces controlled compression of the calf, the foot or combined compression of both.
The ProAir 3000 is intended for use by patients and medical professionals in treating many conditions, such as:
- prevention of deep vein thrombosis (DVT) .
- . enhancement of blood circulation
- reduction of post-operative pain and swelling .
- . reduction of wound-healing time
- stasis dermatitis .
- treatment and assist healing of cutaneous ulceration .
- venous stasis ulcers .
- . leg ulcers
- chronic venous insufficiencies .
- . reduction of edema
## Contraindications:
The ProAir 3000 compression system should not be used in the following cases: Gangrene, recent skin graft, severe arteriosclerosis or other ischemic vascular disease, congestive cardiac failure, massive edema, pulmonary edema, existing DVT, acute thrombophlebitis, acute infections, and during episodes of pulmonary embolism.
{2}------------------------------------------------
#### Device Description:
The ProAir 3000 is a prescriptive, pneumatic compression device designed to apply sequential compression to the lower limb. The device is composed of four components:
- Portable pneumatic control unit ●
- Pair of cuffs (calf and/or foot-ware) ●
- Pneumatic connecting tubes. .
- . Electrical transformer
During the compression cycle the solenoid valves are activated to insert air into one cuff and then other valves are activated for the other cuff. After inflating the cells in one cuff, the solenoid valves are deactivated and, simultaneously, the second cuff is inflated. Following a complete cycle for both legs there is a relaxation period. A software-based protection limits the pressure in all cuff cells using pressure sensors. In addition, a mechanical valve has been included in the air system for releasing air in case of over pressure in one of the cells. The ProAir 3000 has three operation modes: calf cuff mode, foot cuff mode and combined mode.
## Substantial Equivalence:
Based on a series of safety and performance testing including a comparative study, Medical Compression Systems (DBN) Ltd. believes that the ProAir 3000 is substantially equivalent to its predicate devices cited above without raising new safety and/or effectiveness issues.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the top half of the circle. Inside the circle is a stylized image of an eagle with its wings spread.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAY 1 9 2000
Ms. Shoshana Friedman Medical Compression Systems Ltd. C/O Push-med Ltd. 117 Ahuza Street Ra'ananna 43373 Israel
Re: K993758 ProAir 3000 Compression System Requlatory Class: II Product Code: JOW Dated: March 5, 2000 Received: March 7, 2000
Dear Ms. Friedman:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the bubbancearly of analy marketed predicate devices marketed in interstate encrobare) 80 rogal the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to A substantially equivalent determination assumes compliance with 895. the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Please note: this response to your premarket Federal Register. notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.
{4}------------------------------------------------
Page 2 - Ms. Shoshana Friedman
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4639. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Brim E. Haneym
James E. Dillard III
Director
Division of Cardiovascular and
Division of Cardiovascular and Respiratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# INDICATIONS FOR USE
# 510(k) Number (if known):_____________________________________________________________________________________________________________________________________________________
ProAir 3000 Compression System Device Name:
#### Indications for Use:
The ProAir 3000 is a prescriptive device that induces controlled compression of the calf, the foot or combined compression of both.
The ProAir 3000 is intended for use by patients and medical professionals in treating many conditions, such as:
- prevention of deep vein thrombosis (DVT) .
- enhancement of blood circulation .
- reduction of post-operative pain and swelling .
- reduction of wound-healing time .
- stasis dermatitis .
- treatment and assist healing of cutaneous ulceration .
- venous stasis ulcers .
- leg ulcers .
- . chronic venous insufficiencies
- reduction of edema .
(PLEASE DO NOT WRITE BELOW THIS LINE -CONTINUE ON ANOTHER PAGE IF NEEDED)
M. Loree K. Buxerle
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.