← Product Code [MXJ](/productcode/MXJ) · K993754

# HSV-1 ELISA IGG, MODEL EL0910G (K993754)

_Mrl Diagnostics · MXJ · Feb 9, 2000 · Microbiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K993754

## Device Facts

- **Applicant:** Mrl Diagnostics
- **Product Code:** [MXJ](/productcode/MXJ.md)
- **Decision Date:** Feb 9, 2000
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.3305
- **Device Class:** Class 2
- **Review Panel:** Microbiology

## Indications for Use

MRL Diagnostics' HSV-1 ELISA IgG test is intended for qualitatively detecting the presence or absence of human IgG class antibodies to HSV-1 in human sera. In conjunction with the MRL HSV-2 ELISA IgG, the test is indicated for 1) testing sexually active adults, with or without a clinical history of herpes, for aiding in the presumptive diagnosis of HSV infection to identify persons who are at risk for transmitting or acquiring HSV so they may be counseled, and 2) testing expectant mothers for aiding in the presumptive assessment of the risk for acquiring and/or transmitting HSV to their child, so they may be counseled.

## Device Story

HSV-1 ELISA IgG is an in vitro diagnostic assay for qualitative detection of human IgG antibodies to HSV-1 in serum. Polystyrene microwells are coated with recombinant gG-1 antigen; patient serum and controls are incubated in wells. Specific IgG antibodies bind to antigen; nonspecific reactants are washed away. Peroxidase-conjugated anti-human IgG is added, followed by enzyme substrate and chromogen. Color development is stopped and quantified via spectrophotometric optical density (OD) reading. OD is directly proportional to antigen-specific IgG concentration. Results are determined by comparing sample OD to reference cut-off OD. Used in clinical laboratory settings by trained technicians. Output assists clinicians in presumptive diagnosis and risk assessment for HSV transmission, facilitating patient counseling.

## Clinical Evidence

Clinical performance evaluated using masked, archived sera from sexually active adults (n=246) and expectant mothers (n=241). Serological truth defined by type-specific Western blot. In sexually active adults, sensitivity was 91.2% (95% CI: 85.2-95.4%) and specificity 92.3% (95% CI: 85.4-96.6%). In expectant mothers, sensitivity was 96.0% (95% CI: 92.0-98.4%) and specificity 95.2% (95% CI: 86.5-99.0%). Additional evaluation included CDC HSV/CMV panel (n=99) and cross-reactivity testing against CMV, EBV, HHV6, and VZV, showing no cross-reactivity.

## Technological Characteristics

Polystyrene microwell plate coated with recombinant gG-1 antigen. Enzyme-linked immunosorbent assay (ELISA) format. Detection via peroxidase-conjugated anti-human IgG and spectrophotometric measurement of optical density. Manual or automated plate reader required. No specific software algorithm class or connectivity described; standalone diagnostic test.

## Regulatory Identification

Herpes simplex virus serological assays are devices that consist of antigens and antisera used in various serological tests to identify antibodies to herpes simplex virus in serum. Additionally, some of the assays consist of herpes simplex virus antisera conjugated with a fluorescent dye (immunofluorescent assays) used to identify herpes simplex virus directly from clinical specimens or tissue culture isolates derived from clinical specimens. The identification aids in the diagnosis of diseases caused by herpes simplex viruses and provides epidemiological information on these diseases. Herpes simplex viral infections range from common and mild lesions of the skin and mucous membranes to a severe form of encephalitis (inflammation of the brain). Neonatal herpes virus infections range from a mild infection to a severe generalized disease with a fatal outcome.

## Special Controls

*Classification.* Class II (special controls). The device is classified as class II (special controls). The special control for the device is FDA's revised guidance document entitled “Class II Special Controls Guidance Document: Herpes Simplex Virus Types 1 and 2 Serological Assays.” For availability of the guidance revised document, see § 866.1(e).

## Predicate Devices

- HSV-1 ELISA Test System, Zeus Scientific, Inc.
- HSV-1 Western Blot, University of Washington
- HSV-1 Culture, University of Washington

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

#### 9 2000 FEB

## MRL DIAGNOSTICS

993754

## 510(k) Summary of Safety and Effectiveness (Page 1 of 6)

| Applicant:                        | MRL Diagnostics<br>a Focus/MRL Inc. Company<br>10703 Progress Way<br>Cypress, California 90630                                                     |
|-----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|
| Establishment<br>Registration No: | 2023365                                                                                                                                            |
| Contact Person:                   | Michael J. Wagner, Esq.                                                                                                                            |
| Phone:                            | (714) 220-1900                                                                                                                                     |
| Telefax:                          | (714) 220-1182                                                                                                                                     |
| E-mail:                           | mwagner@mrlinfo.com                                                                                                                                |
| Summary Date:                     | October 31, 1999                                                                                                                                   |
| Device Name:                      | HSV-1 ELISA IgG                                                                                                                                    |
| Classification:                   | Herpes Simplex Virus Serological Reagents<br>21 CFR §866.3305<br>Class III                                                                         |
| Predicate<br>Device:              | 1) HSV-1 ELISA Test System, Zeus Scientific, Inc.<br>2) HSV-1 Western Blot, University of Washington<br>3) HSV-1 Culture, University of Washington |

{1}------------------------------------------------

### 510(k) Summary of Safety and Effectiveness (Page 2 of 6)

- In the MRL Diagnostics HSV-1 ELISA IgG assay, the polystyrene microwells are coated Device with recombinant gG-1 antigen. Diluted serum samples and controls are incubated in the Description: wells to allow specific antibody present in the samples to react with the antigen. Nonspecific reactants are removed by washing, and peroxidase-conjugated anti-human IgG is added and reacts with specific IgG. Excess conjugate is removed by washing. Enzyme substrate and chromogen are added, and the color is allowed to develop. After adding the Stop Reagent, the resultant color change is quantified by a spectrophotometric reading of optical density (OD) which is directly proportional to the amount of antigenspecific IgG present in the sample. Sample optical density readings are compared with reference cut-off OD readings to determine results.
- MRL Diagnostics' HSV-1 ELISA IgG test is intended for qualitatively detecting the Intended Use: presence or absence of human IgG class antibodies to HSV-1 in human sera. In conjunction with the MRL HSV-2 ELISA IgG, the test is indicated for 1) testing sexually active adults, with or without a clinical history of herpes, for aiding in the presumptive diagnosis of HSV infection to identify persons who are at risk for transmitting or acquiring HSV so they may be counseled, and 2) testing expectant mothers for aiding in the presumptive assessment of the risk for acquiring and/or transmitting HSV to their child, so they may be counseled.

{2}------------------------------------------------

### 510(k) Summary of Safety and Effectiveness (Page 3 of 6)

An outside investigator assessed the device with masked, archived and unselected sera Expected from 1) sexually active adults over the age of 14 (n = 246), and 2) from expectant mothers Values: (n = 241). Serological "truth" was defined using a type specific Western blot from a major university located in the Northwestern United States. Excluding four atypical Western blots and three ELISA equivocals, the observed vs expected prevalences and the positive predictive values (PPV) and negative predictive values (NPV) for the two populations are as follows:

| Population                    | HSV-1<br>Sero-status | Expected Prevalence |            | Observed Prevalence |              |
|-------------------------------|----------------------|---------------------|------------|---------------------|--------------|
|                               |                      | %                   | 95% CI     | WB                  | MRL<br>ELISA |
| Sexually<br>Active<br>Adults* | neg                  | 59.0%               | 56.3-61.8% | 55.7%               | 54.1%        |
|                               | +                    | 41.0%               | 38.2-43.7% | 42.3%               | 43.9%        |
| Expectant<br>Mothers**        | neg                  | 65.2%               | 64.2-66.2% | 71.0%               | 71.8%        |
|                               | +                    | 34.8%               | 33.8-35.8% | 28.6%               | 24.5%        |

#### Expected vs Observed Prevalence with Sexually Active Adults & Expectant Mothers

* Approximate prevalences for sexually active adult population estimated from unpublished values received from investigator at a major university in the Northwestern United States.

** Approximate prevalences for expected mother population estimated from published values.

#### Predictive Values vs Prevalence with Sexually Active Adults & Expectant Mothers

| Pos<br>Prev | Neg<br>Prev | Sexually Active Adults* |              |       |              | Expectant Mothers** |              |       |              |
|-------------|-------------|-------------------------|--------------|-------|--------------|---------------------|--------------|-------|--------------|
|             |             | PPV                     | PPV<br>95%CI | NPV   | NPV<br>95%CI | PPV                 | PPV<br>95%CI | NPV   | NPV<br>95%CI |
| 90%         | 10%         | 99.1%                   | 98.1-99.6%   | 57.1% | 39.4-75.9%   | 99.4%               | 98.4-99.9%   | 68.8% | 41.6-91.7%   |
| 80%         | 20%         | 97.9%                   | 95.9-99.1%   | 75.0% | 59.4-87.7%   | 98.8%               | 96.5-99.7%   | 83.2% | 61.6-96.1%   |
| 70%         | 30%         | 96.5%                   | 93.2-98.5%   | 83.7% | 71.5-92.4%   | 97.9%               | 94.1-99.6%   | 89.5% | 73.3-97.7%   |
| 60%         | 40%         | 94.7%                   | 89.7-97.7%   | 88.9% | 79.6-95.0%   | 96.8%               | 91.1-99.3%   | 93.0% | 81.0-98.5%   |
| 50%         | 50%         | 92.2%                   | 85.4-96.6%   | 92.3% | 85.4-96.6%   | 95.2%               | 87.2-99.0%   | 95.2% | 86.5-99.0%   |
| 40%         | 60%         | 88.8%                   | 79.6-94.9%   | 94.7% | 89.8-97.7%   | 93.0%               | 82.0-98.5%   | 96.7% | 90.6-99.3%   |
| 30%         | 70%         | 83.5%                   | 71.4-92.3%   | 96.5% | 93.2-98.5%   | 89.6%               | 74.5-97.7%   | 97.9% | 93.7-99.6%   |
| 20%         | 80%         | 74.8%                   | 59.3-87.5%   | 98.0% | 95.9-99.1%   | 83.3%               | 63.0-96.1%   | 98.8% | 96.2-99.7%   |
| 10%         | 90%         | 56.8%                   | 39.3-75.7%   | 99.1% | 98.1-99.6%   | 69.0%               | 43.1-91.6%   | 99.4% | 98.3-99.9%   |

* MRL HSV-1 ELISA IgG having 91.2% sensitivity & 92.3% specificity with sexually active adults. ** MRL HSV-1 ELISA IgG having 96.0% sensitivity & 95.2% specificity with expectant mothers.

{3}------------------------------------------------

### 510(k) Summary of Safety and Effectiveness (Page 4 of 6)

An outside investigator assessed the device's sensitivity and specificity with masked, Sensitivity and archived and unselected sera from expectant mothers (n = 241). Serological "truth" was Specificity with defined using a type specific Western blot from a major university located in the Expectant Northwestern United States. Excluding one atypical Western blots and one ELISA Mothers equivocal, the results are:

| Attribute            | %     | Ratio (n/N) | 95% CI     |
|----------------------|-------|-------------|------------|
| Sensitivity          | 96.0% | 170/177     | 92.0-98.4% |
| Specificity          | 95.2% | 59/62       | 86.5-99.0% |
| Pos Predictive Value | 98.3% | 170/173     | 95.0-99.6% |
| Neg Predictive Value | 88.9% | 59/69       | 75.0-92.8% |

#### Sensitivity, Specificity & Predictive Values with Expectant Mothers

Sensitivity. Specificity & Predictive Values with Sexually Active Adults

CDC's

Panel

An outside investigator assessed the device's sensitivity and specificity with masked, archived and unselected sera from sexually active adults over the age of 14 (n = 246). Serological "truth" was defined using a type specific Western blot from a major university located in the Northwestern United States. Excluding three atypical Western blots and two ELISA equivocal, the results are:

#### Sensitivity, Specificity & Predictive Values with Sexually Active Adults

| Attribute            | %     | Ratio (n/N) | 95% CI     |  |
|----------------------|-------|-------------|------------|--|
| Sensitivity          | 91.2% | 125/137     | 85.2-95.4% |  |
| Specificity          | 92.3% | 96/104      | 85.4-96.6% |  |
| Pos Predictive Value | 94.0% | 125/133     | 88.5-97.4% |  |
| Neg Predictive Value | 88.9% | 96/108      | 81.4-94.1% |  |

Sensitivity with An outside investigator assessed the device's sensitivity using sera from culture positive patients (n = 38). Disease state was defined using culture positivity and a type specific Culture Western blot from a major university located in the Northwestern United States. The Positives results are:

| Comparison         | %     | Ratio (n/N) | 95% CI     |
|--------------------|-------|-------------|------------|
| MRL EL vs. Culture | 78.9% | 30/38       | 62.7-90.4% |
| MRL EL vs. WB      | 81.1% | 30/37       | 64.8-92.0% |

An internal investigator evaluated the CDC's HSV/CMV serum panel (n = 99). The CDC Evaluation of panel results are presented as a means to convey further information on the performance HSV/CMV of this assay with a masked, characterized serum panel. This does not imply an endorsement of the assay by the CDC. The results of the CDC panel evaluation were:

| Interpretation |              |              | %     | Ratio<br>(n/N) | 95% CI     |
|----------------|--------------|--------------|-------|----------------|------------|
| CDC<br>HSV-1   | CDC<br>HSV-1 | MRL<br>ELISA |       |                |            |
| +              | + or neg     | +            | 93.1% | 54/58          | 83.3-98.1% |
| +              | neg          | neg          | 100%  | 13/13          | 75.3-100%  |
| neg            | neg          | neg          | 100%  | 28/28          | 87.7-100%  |

#### Evaluation of CDC's HSV/CMV Panel

714 220-1900 FAX • 714 220-1182

{4}------------------------------------------------

## MRL DIAGNOSTICS

### 510/k) Summary of Safety and Effectiveness (Page 5 of 6)

An outside investigator assessed the device's specificity using sera from a low prevalence Specificity with population of college students. Serological "truth" was defined using a type specific a Low Western blot from a major university located in the Northwestern United States. Prevalence Population Excluding one ELISA equivocal, the results are:

| Interpretation |           | %     | Ratio<br>(n/N) | 95% CI     |
|----------------|-----------|-------|----------------|------------|
| WB-1           | MRL ELISA |       |                |            |
| neg            | neg       | 98.2% | 55/56          | 90.5-100%  |
| +              | +         | 75.0% | 18/24          | 53.3-90.2% |

#### Specificity with a Low Prevalence Population

Type Specificity with HSV-2 Western Blot Positives

An outside investigator assessed the device's type specificity using HSV-1 Western blot negative and HSV-2 Western blot positive sera from the above described populations (n = 90): expectant mothers, sexually active adults, low prevalence persons, and HSV-1 culture positives. Serological "truth" was defined using a type specific Western blot from a major university located in the Northwestern United States. The results are:

| Type Specificity with HSV-2 Western Blot Positives |  |  |
|----------------------------------------------------|--|--|
|                                                    |  |  |

| Group |      |        | %     | Ratio (n/N) | 95% CI     |
|-------|------|--------|-------|-------------|------------|
| WB-1  | WB-2 | MRL EL |       |             |            |
| neg   | +    | neg    | 91.1% | 82/90       | 83.2-96.1% |
| neg   | +    | +      | 8.9%  | 8/90        | 3.9-16.8%  |

Cross-reactivity with Taxonomically Related Viruses

MRL assessed the device's cross-reactivity using sera (n=26) from 1) HSV sero-negative by another manufacturer's FDA cleared HSV ELISAs, and 2) IFA IgG positive for taxonomically similar viruses including CMV, EBV VCA, HHV6 and VZV. Discrepants between the FDA cleared HSV ELISAs and the MRL device were analyzed using a type specific Western blot from a major university located in the Northwestern United States. Excluding one ELISA equivocal that was not analyzed with the Western blot because of insufficient volume, the results are:

| IFA IgG Pos | % MRL EL Neg | Ratio (n/N) | 95% CI    |
|-------------|--------------|-------------|-----------|
| CMV         | 100%         | 12/12       | 73.5-100% |
| EBV VCA     | 100%         | 24/24       | 85.5-100% |
| HHV6        | 100%         | 24/24       | 85.5-100% |
| VZV         | 100%         | 23/23       | 85.2-100% |
| Total       | 100%         | 83/83       | 95.7-100% |

{5}------------------------------------------------

## 510(k) Summary of Safety and Effectiveness (Page 6 of 6)

MRL assessed the device's intra-assay and inter-assay reproducibility using a slightly Intra-assay and modified version of "Evaluation of Precision Performance of Clinical Chemistry Devices", Inter-assay NCCLS Document EP5-T2 (March 1992). Briefly, seven samples were run in duplicate, Reproducibility twice a day, for twenty days, for a total of forty runs, for a total of forty data points for Study each sample. Two sets of samples were masked duplicates.

| Sample | Index Mean | Intra-assay<br>% CV | Inter-assay<br>% CV |
|--------|------------|---------------------|---------------------|
| 11*    | 0.10       | 40.0%               | 50.4%               |
| 16*    | 0.10       | 0.0%                | 11.6%               |
| 12**   | 1.58       | 3.8%                | 7.0%                |
| 17**   | 1.55       | 40.0%               | 5.4%                |
| 13     | 2.67       | 3.6%                | 4.9%                |
| 14     | 3.08       | 3.9%                | 6.5%                |
| 15     | 11.96      | 2.2%                | 5.6%                |

#### Intra-assay & Inter-assay Reproducibility

* #11 & #16 are masked duplicates. ** #12 & #17 are masked duplicates.

Inter-lot Reproducibility An internal investigator assessed the device's inter-lot reproducibility. Five samples were run on three separate days with three separate lots. For one lot, the samples were run in triplicate, and run in duplicate with the other two lots. Each of the three lots had at least a different lot of Antigen Wells. The results are:

| Sample | Index Mean | % CV  |
|--------|------------|-------|
| 11*    | 0.09       | 26.8% |
| 16*    | 0.09       | 20.2% |
| 12**   | 1.47       | 8.2%  |
| 17**   | 1.41       | 2.9%  |
| 13     | 2.48       | 11.1% |
| 14     | 2.64       | 9.4%  |
| 15     | 10.85      | 25.1% |

#### Inter-lot Reproducibility

* #11 & #16 are masked duplicates. ** #12 & #17 are masked duplicates.

An internal investigator and two off site laboratories assessed the device's inter-laboratory Inter-laboratory Reproducibility reproducibility. Each of the three laboratories ran seven samples in triplicate on three different days.

| Inter-lab Reproducibility |            |                     |                     |
|---------------------------|------------|---------------------|---------------------|
| Sample                    | Mean Index | %CV of<br>Lab Means | Mean of<br>Lab %CVs |
| 11*                       | 0.15       | 6.6%                | 70.9%               |
| 16*                       | 0.15       | 23.6%               | 63.1%               |
| 12**                      | 1.44       | 13.5%               | 7.0%                |
| 17**                      | 1.48       | 13.0%               | 4.3%                |
| 13                        | 2.37       | 10.8%               | 5.7%                |
| 14                        | 2.72       | 10.7%               | 7.5%                |
| 15                        | 14.24      | 24.3%               | 64.6%               |

* #11 & #16 are masked duplicates. ** #12 & #17 are masked duplicates.

800 445-0185

{6}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus, which is a symbol often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus.

#### FEB 9 2000

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Mr. Michael J. Wagner, Esq. Senior Regulatory Affairs Specialist MRL Diagnostics 10703 Progress Way Cypress, California 90630

K993754 Re: Trade Name: HSV-1 ELISA IgG Test Regulatory Class: III Product Code: MXJ Dated: January 31, 2000 Received: February 2, 2000

#### Dear Mr. Wagner:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Ouality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{7}------------------------------------------------

Page 2

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Steven Butman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{8}------------------------------------------------

8993754 510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________

Device Name:

HSV-1 ELISA IgG

Indications for Use:

MRL Diagnostics' HSV-1 ELISA IgG test is intended for qualitatively detecting the presence or absence of human IgG class antibodies to HSV-1 in human sera. In conjunction with the MRL HSV-2 ELISA IgG, the test is indicated for 1) testing sexually active adults, with or without a clinical history of herpes, for aiding in the presumptive diagnosis of HSV infection to identify persons who are at risk for transmitting or acquiring HSV so they may be counseled, and 2) testing expectant mothers for aiding in the presumptive assessment of the risk for acquiring and/or transmitting HSV to their child, so they may be counseled.

(PLEASE DO NOT WRITE BELOW THIS LINE CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Woody Dubois

510(k) Number

Prescription Use
(Per 21 CFR 801.109)

OR

Over-The Counter Use

(Optional Format 1-2-96)

---

**Source:** [https://fda-staging.innolitics.com/device/K993754](https://fda-staging.innolitics.com/device/K993754)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
