# GASTRO-INTESTINAL SYSTEM EN-410WM (K993704)

_Fujinon, Inc. · FDA · Feb 11, 2000 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K993704

## Device Facts

- **Applicant:** Fujinon, Inc.
- **Product Code:** [FDA](/productcode/FDA.md)
- **Decision Date:** Feb 11, 2000
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

This video enteroscope is for the optical visualization of the upper gastrointestinal tract. This includes the Esophagus, Stomach, Duodeum and Small Bowel. It is intended for observation, diagnosis and endoscopic treatment.

## Device Story

The Fujinon EN-410WM is a video enteroscope designed for optical visualization of the upper gastrointestinal tract. It functions as an endoscopic imaging tool used by physicians in clinical settings to observe, diagnose, and perform endoscopic treatments within the esophagus, stomach, duodenum, and small bowel. The device captures visual data of the internal anatomy, which is transmitted to a video processor for display on a monitor, allowing the clinician to navigate the tract and identify pathologies or perform interventions. It facilitates minimally invasive diagnostic and therapeutic procedures, potentially reducing the need for more invasive surgical interventions.

## Clinical Evidence

No clinical data provided; bench testing only.

## Technological Characteristics

Video enteroscope for upper GI tract visualization. Consists of an insertion tube with an integrated imaging sensor (video endoscope). Operates via optical visualization transmitted to a compatible video processor. Sterilization is performed according to manufacturer instructions for reusable endoscopic equipment.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around the edge. Inside the circle is a stylized image of an eagle with three stripes across its body. The eagle appears to be in flight or poised for flight, with its wings slightly raised. The image is in black and white and has a somewhat grainy quality.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

FEB 1 1 2000

Fujinon, Inc. c/o Mr. Gary Adler Bingham Dana Murase Counselors at Law 399 Park Avenue New York, NY 10022-4689 Re: K993704 Gastro-Intestinal System EN-410WM Dated: January 12, 2000 Received: January 13, 2000 Regulatory Class: II Regulatory of 1500/Procode: 78 FDA

Dear Mr. Adler:

Bour and wed your Section 510(k) notification of intent to market the device referenced above and we have have have We have reviewed your Section 510(k) noticeations for use stated in the endosure) to legally
delemined the device is substantially equivals (for the indications for use the e delemined the device is substantialy equivalent (or toe states , to rast see and many , and the enation , and of the of the marketed predicate devices markets in high in they 20, 1070, movel the provisions of the provisions of the
Medical Device Amendments, or to devices that have been reviews, su Medical Device Amendments, or to devices hat have been estimation in the more of the consisted on the collections.
Federal Food, Drum and waters and the ecting market for ann Federal Food, Drug, and Cosmelic Act (Ad). You may, the Act include requirements for annual registration, listing of
provisions of the Act. The general controls provised mi provisions of the Act. The general controls provisions of the Accorner of the management of the manufaction.

If your device is classified (see above) into either class III (Special Controls) or class III (Premarket Approval), it may
n the Code of lf your device is classified (see above) into einer class in (species on be love (v r che invel in the Code of the Code of the Code of be subject to such additional controls. Existing major regulation assumes ompliance with
Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalion (OS) fo Fegulations, Title 21, Parts out o oss. A substantely cyterialth Cuality System Regulation (OS) for the Current Good Manufacturing Practice requirements as through periodic CS inspections, the Food and Medical Devices: General regulation (CFR Fall Dell the Comply with the GM regulation may result in
Drug Administration (FDA) will verify such assumply with the GM regulation Drug Administration (FDA) will vehily such assumplier in annumer oncerning your device in the Eederal
regulatory action. In addition, FDA may publish further announcement of regulatory action. In addition, FDA nier announcements on the onlines on offect any obligation your a
Register. Please note: this response to your premarke notiscination d Register. Please note: this response to your premarket nomically dos model the election of the may be of the provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 500(k) premately notification for I his letter will allow your nativeling your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your labeling regulation (21 CFR Pat 801 and additionally for not in il you desire specific dovice for your device of our aboung rance at (301) 554-4613. Additionally for questions on viro diagnosic devices), please contact the Office of Compliance at (301) 594-4639. Also, Also, Also, Also, Also, Also, Allas, the promotion and adventising of your defices, problem of the contriction" (21 CFR 807.97). Other please note tre regulation entiled, "Misfrancing of relating to potained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours.

CAPT Daniel G. Schultz, M.D. Acting Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{1}------------------------------------------------

Page 3/20

S10(k) NUMBER (IF KNOWN): K993704

DEVICE NAME: GASTRO INTES TINALSYSTEM

INDICATIONS FOR USE: EN-410WM

This video enteroscope is for the optical visualization of the upper gastrointestinal tract. This includes the Esophagus, Stomach, Duodeum and Small Bowel. It is intended for observation, diagnosis and endoscopic treatment.

> (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER IF NEEDED.)

Concurrence of CDRH, Office of Device Evaluation (OI

Prescription Use
(Per 21 CFR 801.109)
OR

Over-The-Counte (Optional For

r

(Division Sign-Off)
Division of Reproductive, Abdominal, ENT,
and Radiological Devices
510(k) Number K993704

---

**Source:** [https://fda-staging.innolitics.com/device/K993704](https://fda-staging.innolitics.com/device/K993704)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).