# MICROSTREP PLUS PANEL - HAEMOPHILUS/AMPICILLIN (K993679)

_Dade Microscan, Inc. · JWY · Jan 14, 2000 · Microbiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K993679

## Device Facts

- **Applicant:** Dade Microscan, Inc.
- **Product Code:** [JWY](/productcode/JWY.md)
- **Decision Date:** Jan 14, 2000
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.1640
- **Device Class:** Class 2
- **Review Panel:** Microbiology

## Indications for Use

The intended use of this panel is for determining quantitative and/or qualitative antimicrobial agent susceptibility of colonies grown on solid media of Haemophilus spp. After incubation in a non-CO2 incubator for 20 - 24 hours, the minimum inhibitory concentration (MIC) for the test organism is determined by observing the lowest antimicrobial concentration showing inhibition of growth. The two-fold dilutions of AMPICILLIN in the panel range from 0.015 to 16 mcg/ml for testing Haemophilus spp.

## Device Story

MicroScan MICroSTREP plus Panel is an in vitro diagnostic device used in clinical laboratories to determine the minimum inhibitory concentration (MIC) of Ampicillin for Haemophilus spp. isolates. The device consists of panels containing two-fold dilutions of the antimicrobial agent (0.015 to 16 mcg/ml). Clinical laboratory personnel inoculate the panel with bacterial colonies grown on solid media and incubate in a non-CO2 incubator for 20-24 hours. The MIC is determined by observing the lowest concentration of Ampicillin that inhibits bacterial growth. This quantitative result assists clinicians in selecting appropriate antibiotic therapy for patients with Haemophilus infections.

## Clinical Evidence

External evaluation conducted using fresh and stock clinical isolates and stock challenge strains of Haemophilus spp. Performance compared against NCCLS frozen reference panels. Results demonstrated an overall Essential Agreement of 91.5%. Reproducibility and quality control testing yielded acceptable results.

## Technological Characteristics

Microdilution panel containing two-fold dilutions of Ampicillin (0.015 to 16 mcg/ml). Principle of operation is broth microdilution for antimicrobial susceptibility testing. Designed for use with Haemophilus spp. isolates grown on solid media. Requires non-CO2 incubation for 20-24 hours.

## Regulatory Identification

An antimicrobial susceptibility test powder is a device that consists of an antimicrobial drug powder packaged in vials in specified amounts and intended for use in clinical laboratories for determining in vitro susceptibility of bacterial pathogens to these therapeutic agents. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.

## Predicate Devices

- MicroScan® Streptococcus MIC Panel ([K963641](/device/K963641.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# K993679

JAN 1 4 2000

#### 510(k) Summary Information:

|                     | Device Manufacturer: Dade MicroScan Inc.                    |
|---------------------|-------------------------------------------------------------|
| Contact name:       | Trevor Wall, Regulatory Affairs Manager                     |
| Fax:                | 916-374-3144                                                |
| Phone:              | 916-374-3332                                                |
| Date prepared:      | October 20, 1999                                            |
| Product Name:       | Microdilution Minimum Inhibitory Concentration (MIC) Panels |
| Trade Name:         | MicroScan® MICroSTREP plus™ Panel - Ampicillin              |
| Intended Use:       | To determine bacterial susceptibility to AMPICILLIN         |
| Indication for Use: | For use with Haemophilus spp.                               |
| Predicate device:   | MicroScan® Streptococcus MIC Panel (K963641).               |

#### 510(k) Summary:

The proposed MicroScan® MICroSTREP plus™ : AMPICILLIN, demonstrated substantially equivalent performance with Haemophilus isolates when compared with an NCCLS frozen Reference, as defined in the FDA DRAFT document "Review Criteria for Assessment of Antimicrobial Susceptibility Devices" (dated May 31, 1991).

The Premarket Notification (510[k]) presents data in support of the new MICroSTREP plus™ Panel with AMPICILLIN ( 0.015 to 16 mcg/ml).

The external evaluation was conducted with fresh and stock Efficacy isolates and stock Challenge strains of Haemophilus spp. The external evaluations were designed to confirm the acceptability of the proposed MICroSTREP plus™ Panel - AMPICILLIN, by comparing its performance with an NCCLS frozen Reference. The MICroSTREP plus™ Panel - AMPICILLIN demonstrated acceptable performance with an overall Essential Agreement of 91.5% when compared with the frozen Reference Panel.

Reproducibility testing demonstrated acceptable results with AMPICILLIN.

Quality Control testing demonstrated acceptable results with AMPICILLIN.

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized image of an eagle with three lines forming its body and head. The logo is surrounded by text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement.

## JAN 1 4 2000

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Mr. Trevor Wall Regulatory Affairs Manager Dade MicroScan, Inc. 1584 Enterprise Boulevard West Sacramento, California 95691

Re: K993679

> Trade Name: MicroScan® MICroSTREP plus™ Panel (Haemophilus/Ampicillin) Regulatory Class: II Product Code: JWY Dated: October 29, 1999 Received: November 1, 1999

Dear Mr. Wall:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{2}------------------------------------------------

Page 2

Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html"

Sincerely yours,

Steven Autman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Page of

## 510(k) Number (if known):_K99 3679

Device Name: MicroScau MicroScau Micro STREA plus

Indications For Use:

The intended use of this panel is for determining quantitative and/or qualitative antimicrobial agent susceptibility of colonies grown on solid media of Haemophilus spp.

After incubation in a non-CO2 incubator for 20 - 24 hours, the minimum inhibitory concentration (MIC) for the test organism is determined by observing the lowest antimicrobial concentration showing inhibition of growth.

The two-fold dilutions of AMPICILLIN in the panel range from 0.015 to 16 mcg/ml for testing Haemophilus spp.

### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

# Concurrence of CDRH, Office of Device Evaluation (ODE)

Wooley Dubocs

ion of Clinical Laboratory Devices 4611 510(k) Number

Prescription Use (Per 21 CFR 801.109)

OR

Over-The-Counter Use

(Optional Format 1-2-96

---

**Source:** [https://fda-staging.innolitics.com/device/K993679](https://fda-staging.innolitics.com/device/K993679)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com