NEXSYSTEM AMBULATORY PATIENT MONITOR MODEL NEX 100
Applicant
Nexan Telemed , Ltd.
Product Code
MWJ · Cardiovascular
Decision Date
Dec 28, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2800
Device Class
Class 2
Indications for Use
The Nexsystem ambulatory patient monitoring system is intended for use in the home or alternate care settings. It consists of a patient-worn sensor (Nexisensor), signal transfer unit (Nexender), communications module (Nextation), and a telemonitoring station computer based display and storage system (TMS) located at the health care professional's facility. The device stores and transmits ECG data, respiratory data, skin temperature, systolic and diastolic blood pressure (Non-invasive), and PEF and FEV1. The Nexsystem is not for use as a critical care monitoring system. The Nexsystem is not an apnea monitor.
Device Story
Nexystem is a non-invasive ambulatory monitoring system for home/alternate care settings. Input: physiological signals (ECG, respiration, skin temperature) via chest-worn Nexisensor; auxiliary inputs (blood pressure, spirometry) via Nextation base station. Nexisensor transmits data wirelessly to Nexender (Signal Transfer Unit), which forwards data to Nextation via UHF radio link. Nextation records/stores data and transmits to remote Telemonitoring Station (TMS) via telephone line. HCP uses Nexoft software on TMS to display, analyze, and print reports. System facilitates remote patient status checks and scheduled data review. Benefits: enables remote physiological monitoring, reducing need for in-person clinical visits; provides HCPs with trend data for clinical decision-making.
Clinical Evidence
No clinical data. Bench testing only. Compliance with ANSI/AAMI EC12 (disposable ECG electrodes) and ANSI/AAMI EC38 (ambulatory electrocardiographs) confirmed via test reports.
Technological Characteristics
System components: Nexisensor (disposable, Ag/AgCl electrodes), Nexender (UHF radio transmitter), Nextation (base station, serial ports), TMS (Windows NT PC). Connectivity: Wireless (UHF) between Nexisensor/Nexender and Nexender/Nextation; telephone line (modem) between Nextation/TMS. Power: Battery (Nexisensor/Nexender), AC line (Nextation). Software: Nexoft application. Standards: ANSI/AAMI EC12, EC38, IEC 60601-1.
Indications for Use
Indicated for patients in home or alternate care settings requiring ambulatory monitoring of ECG, respiration, skin temperature, non-invasive blood pressure, PEF, and FEV1. Not for critical care or apnea monitoring.
Regulatory Classification
Identification
A medical magnetic tape recorder is a device used to record and play back signals from, for example, physiological amplifiers, signal conditioners, or computers.
OMRON IC fully automatic blood pressure monitor (K913523)
Submission Summary (Full Text)
{0}------------------------------------------------
K993643
# 14 Safety and Efficacy - Premarket Notification 510(k) Summary
510(k) Summary as required per 807.92.
#### 14.1 Submitter Details
Nexan Telemed Ltd The Quorum Barnwell Road Cambridge CB5 8RE United Kingdom
Contact: Dr J. D. Place - Operations Director Phone: +44 1223 713500 Fax: +44 1223 713501
Date of submission: 1st September 1999
## 14.2 Device Name and Classification
- 14.2.1 Device name
Nexystem
#### 14.2.2 Device Common Name
Ambulatory Patient Monitor
## 14.2.3 Classification, product code
| Class: | II |
|----------------|---------------------------------|
| Product Codes: | ECG 74DPS CFR: 870.2340 |
| | Respiration 73BZQ CFR: 868.2375 |
| | Temperature 80FLL CFR: 880.2910 |
| | Transmitter 74DRG CFR: 870.2910 |
#### 14.3 Predicate Device Information
The following predicate devices are considered:
| Company | Device | 510(k) |
|------------------|------------------------------------|---------|
| Sabratek | Patient Home Monitoring System | K980619 |
| Protocol Systems | Propaq Encore | K951246 |
| Mortara | Ambulatory X-12 Telemetry Module | K974149 |
| Reynolds | Tracker Holter Recorder | K943278 |
| Reynolds | Pathfinder 7000 Holter Analyser | K951902 |
| Instromedix | Poseidon Cardiac Monitoring System | K964408 |
{1}------------------------------------------------
## 14.4 Device Description
## 14.4.1 System Overview
The Nexystem is a non-invasive ambulatory patient monitoring system for recording multiple physiological parameters from patients who may be located at home or in an alternate care setting. The Nexystem continuously gathers physiological data from a sensor band attached to the patient and transmits the data wirelessly via a Signal Transfer Unit (Nexender) to a Base Station Unit (Nextation) where the data are recorded and stored. Additionally, the Base Station Unit has interfaces for auxiliary sensors - spirometer and blood pressure monitor - for recording point in time lung function and blood pressure measurements. A Call Discriminator Unit enables incoming telephone calls to be correctly routed to either the Nextation or a telephone handset.
The data are transmitted for display, monitoring and storage on a computer (Telemonitoring Station TMS running the Nexoft application software) at a distant location (Telemonitoring Centre). This data transfer is under the control of the Health Care Professional (HCP) at the TMS. Data may be transferred in real time to enable the HCP to check on the quality of the physiological data being recorded and/or the status of the patient. Normally data is transferred at a scheduled time after the end of a patient data recording session. Once transferred to the TMS the data can subsequently be displayed for analysis by the HCP. The Nexystem enables the HCP to print reports of raw or summary data.
Image /page/1/Figure/5 description: This image shows a diagram of a telemonitoring system. The system includes a Nexisensor (sensor band) attached to a person's chest, a Nexender (signal transfer unit) attached to the person's waist, and a Nextation (base station unit). The system also includes a telemonitoring center with a telemonitoring station (TMS), a patient's telephone line, and a call discriminator unit. Auxiliary connections for blood pressure monitor and spirometer are also shown.
The Nexystem comprises a number of system components identified below.
{2}------------------------------------------------
#### 14.4.2 Identification of System Components
The Nexystem comprises several distinct components which together form a complete monitoring system. In this section each of these components is separately identified
#### 14.4.2.1 Nexisensor
The Nexisensor is a disposable, battery powered, adhesive band containing electrodes and sensors, which is attached to a patient's chest. It captures physiological signals, processes and transmits them, via a wireless link, to a Nexender.
#### 14.4.2.2 Nexender
The Nexender is a portable, battery powered, wireless Signal Transfer Unit either worn by, or in close proximity (within 1.5m) to, the patient. The Nexender receives the physiological data from the Nexisensor and forwards it to the Nextation via a UHF radio link when within operational range (30m).
#### 14.4.2.3 Nextation
The Nextation is a table-top device, requiring connection to line voltage and direct connection to a hard-wired telephone line. The Nextation acts as a Base Station Unit receiving and recording data from the Nexender. The Nextation has two serial ports for the direct connection of a spirometer and a blood pressure monitor for the recording of point in time measurements.
#### 14.4.2.4 Spirometer
The spirometer interfaced with the Nexystem is the Micro Medical MicroPlus Spirometer MS03.
#### 14.4.2.5 Blood Pressure Monitor
The Blood Pressure Monitor interfaced with the Nexystem is the OMRON IC fully automatic blood pressure monitor with inflation pre-set.
#### 14.4.2.6 Call Discriminator
In situations where a dedicated telephone line is not available at the patient location for use of the Nexystem a telephone call discriminator unit may optionally be utilised to differentiate between incoming voice calls for the patient and data calls from the TMS for the Nextation. The unit utilised is the Viking FAXJ-300 Fax Jack Phone/Fax Switch.
#### 14.4.2.7 Telemonitoring Station (TMS) Computer/Nexoft Software
The TMS Computer is a standard Windows NT PC located at the HCP's office running the Nexoft application software that is used to download patient data from the Nextation for display, recording and printout of reports for analysis by the Health Care Professional.
#### 14.5 Intended Use
The Nexystem is an ambulatory patient monitoring system intended for use in the home or alternate care settings. It consists of a patient worn sensor (Nexisensor), signal transfer unit (Nexender), communications module (Nextation), and a telemonitoring station computer based display and storage system (TMS) located at the health care professional's facility. The device stores and transmits ECG data, respiration data, skin temperature, systolic and diastolic blood pressure (non-invasive), and PEF and FEVI.
{3}------------------------------------------------
## 14.6 Non-clinical performance data for equivalence
### 14.6.1 ANSI/AAMI EC12
Compliance testing of the Nexisensor to ANSVAAMI EC12 Disposable ECG Electrodes has been conducted and the detailed Compliance Test Report 100-TR-080 is located in Appendix R.
The report concludes that the Nexisensor is compliant with ANSUAAMI EC12. This compliance is subject to the understanding that the Nexisensor electrodes are integral and not wired, and that the Nexisensor is not designed to withstand a defibrillator as it will be removed by paramedics prior to the application of the defibrillator electrodes.
## 14.6.2 ANSI/AAMI EC38
Compliance testing of the Nexystem to ANSVAAMI EC38 Ambulatory Electrocardiographs has been conducted and the detailed Compliance Test Report 100-TR-079 is located in Appendix S.
The report concludes that the Nexystem is a Type I ambulatory electrocardiograph and is compliant with ANSUAAMI EC38. This compliance is subject to the understanding that the Nexystem provides only a single lead of ECG and that as ECG analysis is not provided there is only a partial application of the standard.
### 14.7 Clinical performance data for equivalence
Not applicable
## 14.8 Predicate Device Comparison
The comparison of intended use and technological features of the Nexystem with other legally marketed devices taken together with the validation results, performance tests and other information in this submission indicate the Nexystem is substantially equivalent to legally marketed devices in safety, effectiveness and intended use.
{4}------------------------------------------------
のも、その他、その他のアイテムには、日本の中に、日本の中には、日本の中国の中国の中国の中国の中国の中国の中国の中国の中国の中国の中国の
## Predicate Devices:
Protocol Systems Propaq Encore
## 510(K)#: K980619 510(K)#: K951246
## Parameters being compared:
Physiological parameters, auxiliary devices, general equipment specification
Sabratek Patient Home Monitoring System (PHMS)
| Parameter | Nexystem | Sabratek PHMS | PROPAQ Encore | Explanation of Differences |
|---------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------|-------------------------------------------------------------------------|
| Electrocardiogram | | | | |
| Records and stores extended monitor<br>bandwidth electrocardiogram. | Yes, 0.05 - 85 Hz | No | Yes, 0.05 - 40 Hz | 85 Hz gives improved signal fidelity |
| Single lead (2 active electrodes plus a<br>'ground' electrode) | Yes | Yes | Yes | |
| QRS or arrhythmia detection | QRS detection | No | QRS detection | |
| Heart rate range | 30 - 250 BPM | n/a | 25-350 BPM | Nexystem range complies with EC38 |
| Heart rate accuracy | $\pm$ 3 BPM | n/a | $\pm$ 3 BPM | |
| Detects an electrode lead off condition | No | Yes | Yes | Signal quality can be checked via<br>Nexoft at any time |
| ECG sensing electrodes | Ag / AgCl custom | n/a | Any Ag / AgCl electrode is<br>recommended. | |
| Alarms on low / high heart rate | No | No | Yes | Events outside pre-set thresholds can<br>be logged for review. |
| Wireless operation | Yes | No | No | Provides greater operational patient<br>convenience |
| Respiration | | | | |
| Measure respiration rate by impedance | Yes | No | Yes | |
| Range | 0 - 72 BPM | n/a | 0 (apnoea), 2-150 BPM | Nexystem is not an apnoea monitor. |
| Accuracy | $\pm$ 2 BPM | n/a | $\pm$ 2 BPM or 2% | Validation against clinical trial data |
| Sensor | Ag / AgCl custom | n/a | Any Ag / AgCl electrode is<br>recommended | |
| Temperature | | | | |
| Thermistor | Alpha Sensors: medical 400 series<br>for skin surface (disposable) | n/a | Arbo: 400 series for skin<br>surface (disposable) | |
| Range | 25.0 - 45.0 °C | n/a | 30.0 - 42.0 °C | Nexystem has extended range. |
| Thermistor Accuracy | $\pm 0.2 °C$ | n/a | $\pm 0.1 °C$ | $\pm0.2$ °C is sufficiently accurate for<br>trend analysis. |
| Non Invasive Blood Pressure | Omron IC (HEM725CIC) | AND Medical | Unit meets ANSI-AAMI SP10-<br>1992 performance standard | Nexystem uses a legally marketed<br>device |
| 510K # | K913523 | K871720 | n/a | |
| Spirometer | Micromedical MicroPlus | n/a | n/a | Nexystem uses a legally marketed<br>device |
| 510K# | K963035 | n/a | n/a | |
| Nextation General | | | | |
| Communications capability | up to 2 external devices at any one<br>time | up to 10 external devices | n/a | 2 auxiliary sensors considered<br>adequate for any one patient. |
| Display | 2 line x 20 character LCD | 4 line x 20 character LCD | n/a | Adequate for patient messages. |
| Display back light | Yes | Yes | n/a | |
| Keypad | 3 push buttons | 5 row by 5 column custom | n/a | Keeps system operation simple. |
| Patient alerts / notification | Messages displayed to prompt<br>patient auxiliary measurements<br>and communicate system faults | Audio and visual patient<br>alarm notification | n/a | Nexystem not an alarm monitor. |
| Size | approx. 24 x 34 x 9 cm | Approx. 30 x 30 x 10cm | 17 x 21 x 13 cm (basic unit) | |
| Weight | 3.6 kg / 7.9 lbs. | 4.5 kg / 10 lbs. | 2.8 kg / 6.25 lbs (basic unit) | |
| Operating temperature | 0 to 45°C | 0 to 60°C | 0 - 40 °C | |
| Storage humidity | 5 to 95% non-condensing | 0 to 95% non-condensing | 15 to 95% non-condensing | |
| Shipping/storage temperature | -20 to 65°C | -20 to 60°C | -20 to 65°C | |
| Drip proof | Tested to IEC 60601-1 | IEC 529 IPX1 | IEC 529 IPX1 (monitor) | IEC 60601-1 test offers similar level<br>of protection as IEC 529 IPX 1 |
| Nextation Power | | | | |
| Operation | 110-120V 220-240V 50/60Hz | 115/230 VAC 50/60 Hz | 100-120V 220-240 V 50/60 Hz | |
| Current consumption | 250 mA continuous | Not known | 250 mA continuous | |
| Operation from internal battery pack | No | Yes, > 1 hour | Yes, > 2.5 hour | 1 hour data buffer in Nexender. |
| Double insulation | No | Yes | n/a | Nextation is class I device and earthed |
| Power input | < 100 Watts | < 40 Watts. | n/a | |
| Ambulatory sensor power | 3 x 1.4 V675AT Zinc Air cells | n/a | n/a | |
| Nexender power | 4 x 1.5 V Alkaline cells | n/a | n/a | |
| Parameter | Nexystem | Reynolds Holter Recorders and Analyser | Explanation of Differences | |
| Event Buttons | | | | |
| Number of event buttons | 2 | 1 | Provides more information. | |
| Location of event button(s) | On Nexender: worn by patient or placed close by. | On Holter recorder, always worn by patient. | | |
| Display of event button presses | In Nexoft, a list of all patient events can be seen;<br>users can view raw data for each event (or<br>download data for any event). | In Pathfinder, a summary table includes the total<br>number of patient events; user can tab through<br>raw ECG data for each event in turn. | | |
Page 121 of 125
{5}------------------------------------------------
Page 122 of 125
{6}------------------------------------------------
510(K)#: K974149
#### Mortara Ambulatory X-12 Telemetry Module Predicate: Parameters being compared: Nexender to Nextation radio link parameters
#### Explanation of Differences Nexender to Nextation Mortara X-12 Telemetry Module Parameter Radio Link Digital, frequency-shift keying Digital, frequency-shift keying Transmission scheme 80 kHz (915 MHz) and 320 kHz (2.15 GHz) Nexystem is a single channel system Radio channel spacing n/a 50 mV/m at 3 m Power output 50 mV/m at 3 m 433 MHz is European frequency band Frequency 433 MHz or 916 MHz 915 MHz or 2.45 GHz 915/916 are the same approved band.
### Predicate:
## Reynolds Tracker Holter Recorders Reynolds Pathfinder 7000 Holter Analyser
## 510(K)#: K943278 510(K)#: K951902
Event buttons Parameters being compared:
{7}------------------------------------------------
#### 510(K)#: K964408 Poseidon™ Cardiac Monitoring System Predicate: Nextation to TMS communications and Nexoft features Parameters being compared:
| Parameter | Nexystem | Poseidon™ Cardiac Monitoring System | Explanation of Differences |
|------------------------------------|--------------------------------------------------------------|-----------------------------------------------------------------|--------------------------------------------|
| Nextation to Nexoft Communications | | | |
| Modem | Meets V34 fax/data modem speeds and standards. | Meets V34 fax/data modem speeds and standards. | |
| Protocol | TCP/IP and PPP (including error detection and<br>correction) | CEN 231, SCP Digital Transmission<br>Error detection/correction | Nexystem uses standard Windows dial<br>up. |
| Nexoft Features | | | |
| Operating system | Windows NT | Windows NT | |
| Print out | Yes | Yes | |
| Storage | Magnetic hard drive | Magnetic disk | |
| Back up | Yes - Optical disc | Yes | |
| Alarms | No | No | |
| Data triggers | Yes | Yes | |
Page 124 of 125
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus, a symbol often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus. The logo is black and white.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DEC 2 8 1999
Nexan Telemed Ltd. c/o Quintiles, Inc. Kenneth A. Palmer, Ph.D. 15825 Shady Grove Road, Suite 130 Rockville, MD 20850-4008
Re: K993643 Nexsystem Abulatory_Parient Monitor Model Nex Regulatory Class: II (two) Product Code: MWJ October 28, 1999 Dated: Received: October 28, 1999
Dear Dr. Palmer:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 A substantially equivalent determination assumes compliance to 895. with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.
{9}------------------------------------------------
Page 2 - Kenneth A. Palmer, Ph.D.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
elia M., Witten, Ph.D., M.D. Acting Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological·Health
Enclosure
{10}------------------------------------------------
Page 1 of 1
510(k) Number: K993643
Device Name: Nexsystem Ambulatory Patient Monitor
Indications For Use:
The Nexsystem ambulatory patient monitoring system is intended for use in the home or alternate care settings. It consists of a patient-worn sensor (Nexisensor), signal transfer unit (Nexender), communications module (Nextation), and a telemonitoring station computer based display and storage system (TMS) located at the health care professional's facility. The device stores and transmits ECG data, respiratory data, skin temperature, systolic and diastolic blood pressure (Non-invasive), and PEF and FEV1.
The Nexsystem is not for use as a critical care monitoring system. The Nexsystem is not an apnea monitor.
なんて
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use X (Per 21 CFR 801.109) OR
Over-The-Counter Use
(Optional Format 1-2-96)
Walter Tikh
(Division Sign-Division of Cardiovascular. Respiratory. and Neurological Devic 510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.