The RX VIATRAC™ 18 Peripheral Dilatation Catheter is intended: - To dilate stenoses in the peripheral arteries (iliac, femoral, ilio-femoral, . popliteal, infra popliteal, renal arteries) - For the treatment of obstructive lesions of native or synthetic . arteriovenous dialysis fistulae. - For post deployment optimization of the 28 mm GUIDANT . MEGALINK™ Biliary Stent (6.0 to 10.0 mm diameters). The OTW VIATRAC™ 18 Peripheral Dilatation Catheter is intended: - . To dilate stenoses in the peripheral arteries (iliac, femoral, ilio-femoral, popliteal, infra popliteal, renal arteries) - For the treatment of obstructive lesions of native or synthetic . arteriovenous dialysis fistulae. - For post deployment optimization of the 28 mm GUIDANT . MEGALINK™ Biliary Stent (6.0 to 10.0 mm diameters).
Device Story
RX and OTW VIATRAC 18 Peripheral Dilatation Catheters are balloon angioplasty devices. RX platform features integrated shaft for rapid exchange; OTW platform features coaxial design with full-length inner member. Both utilize XCELON balloon with two radiopaque markers for positioning. Operated by physicians in clinical settings to dilate peripheral artery stenoses, treat arteriovenous dialysis fistula lesions, or optimize biliary stent deployment. Balloon provides expandable segment of known diameter/length at specific pressures. Output is mechanical dilation of target vessel/lesion. Benefits include restoration of blood flow or stent apposition.
Clinical Evidence
Bench testing only. Tests included accelerated aging, catheter preparation, crossing/collapsed profile, balloon compliance, inflation/deflation times, balloon fatigue, balloon rupture, soft tip tensile, catheter tensile, bend integrity, shaft pressure, inner member collapse, wall/mandrel penetration, support mandrel pull, and in-stent balloon rupture/fatigue. Results demonstrated performance equivalent to predicate devices.
Technological Characteristics
PTA catheters with XCELON balloon; 75 and 135 cm lengths; 6.0-10.0 mm diameters. RX platform (integrated shaft) and OTW platform (coaxial). Features two radiopaque markers. Materials, biocompatibility, and sterilization are equivalent to predicates.
Indications for Use
Indicated for patients requiring dilation of stenoses in peripheral arteries (iliac, femoral, ilio-femoral, popliteal, infra-popliteal, renal) or treatment of obstructive lesions in native/synthetic arteriovenous dialysis fistulae; also indicated for post-deployment optimization of 28 mm GUIDANT MEGALINK Biliary Stents (6.0-10.0 mm diameters).
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}------------------------------------------------
JAN 2 7 2000
510(k) SUMMARY
K993639
This 510(k) Summary is submitted in accordance with 21 CFR Part 807, Section 807.92.
| 1. | Submitter's Name: | Guidant Corporation<br>Advanced Cardiovascular Systems, Inc. |
|----|-----------------------------|-------------------------------------------------------------------------------|
| | Submitter's Address: | 3200 Lakeside Drive<br>Santa Clara, CA 95054 |
| | Telephone: | 408-235-3995 |
| | Fax: | 408-235-3743 |
| | Contact Person: | Susan Silavin, Ph.D. |
| | Date Prepared: | August 28, 1998 |
| 2. | Device Trade Name: | RX VIATRAC™ 14 Peripheral Dilatation Catheter |
| | Device Common Name: | Percutaneous Transluminal Angioplasty Catheter |
| | Device Classification Name: | LIT |
| 3. | | Predicate Devices: RX VIATRAC™ 14<br>Peripheral Dilatation Catheter (K983055) |
#### 4. Device Description:
The RX VIATRACTM 18 and the OTW VIATRACTM 18 Peripheral Dilatation Catheters were developed in parallel as two platforms, a Rapid Exchange (RX) platform and an Over-the-Wire (OTW) platform. The distal 20 cm of both catheters are identical however, the proximal portions are modified to accommodate either the RX or OTW platform.
Cordis OPTA 5TM PTA Catheter (K972825)
The RX catheter is a rapid exchange catheter with an integrated shaft system. The design is based upon the RX VIATRACTM 14 Peripheral Dilatation Catheter, manufactured by Guidant (K983055, 12/30/98). The RX VIATRAC™ 18 Peripheral Dilatation Catheter has catheter lengths of 75 and 135 cm, with balloon diameters of 6.0, 7.0, 8.0, 9.0 and 10.0 mm.
The OTW platform is an over-the-wire, coaxial design catheter that is similar to that of the RX catheter. However, the OTW catheter has an inner member extending the entire length of the catheter, and therefore has no mid-catheter junction. The OTW VIATRAC™ 18 Peripheral Dilatation Catheter has catheter lengths of 75 and 135 cm, with balloon diameters of 6.0, 7.0, 8.0, 9.0 and 10.0 mm.
{1}------------------------------------------------
The XCELON™ balloon, which is identical on both RX and OTW VIATRAC™ 18 Peripheral Dilatation Catheters has 2 radiopaque markers to aid in positioning the balloon in the stenosis, is designed to provide an expandable segment of know diameter and length at specific pressures.
### 5. Intended Use:
The RX VIATRACTM 18 Peripheral Dilatation Catheter is intended:
- To dilate stenoses in the peripheral arteries (iliac, femoral, ilio-femoral, . popliteal, infra popliteal, renal arteries)
- For the treatment of obstructive lesions of native or synthetic . arteriovenous dialysis fistulae.
- For post deployment optimization of the 28 mm GUIDANT . MEGALINK™ Biliary Stent (6.0 to 10.0 mm diameters).
The OTW VIATRAC™ 18 Peripheral Dilatation Catheter is intended:
- . To dilate stenoses in the peripheral arteries (iliac, femoral, ilio-femoral, popliteal, infra popliteal, renal arteries)
- For the treatment of obstructive lesions of native or synthetic . arteriovenous dialysis fistulae.
- For post deployment optimization of the 28 mm GUIDANT . MEGALINK™ Biliary Stent (6.0 to 10.0 mm diameters).
- 6. Technological Characteristics:
Comparisons of the new and predicate devices show that technological characteristics such as materials, biocompatibility, performance properties (see below), sterilization, and packaging are identical or substantially equivalent to the currently marketed predicate devices.
#### 7. Performance Data:
Bench testing was performed to demonstrate that the met the acceptance criteria and performed similar to the predicate RX VIATRAC™ 14 Peripheral Dilatation Catheter and the OPTA 5TM PTA Catheter. The following tests were performed:
| Accelerated Aging Testing |
|------------------------------------|
| Catheter Preparation Test |
| Crossing Profile/Collapsed Profile |
| Balloon Compliance Test |
| Balloon Inflation/Deflation Times |
| Balloon Fatigue Test |
| Balloon Rupture Test |
| Catheter Soft Tip Tensile Test |
{2}------------------------------------------------
| Catheter Tensile Test |
|----------------------------------------|
| Catheter Bend Integrity Test |
| Catheter Shaft Pressure Test |
| Inner Member Collapse Test |
| Catheter Wall/Mandrel Penetration Test |
| Support Mandrel Pull Test |
| In-Stent Balloon Rupture Test |
| In-Stent Balloon Fatigue Test |
The results from the bench tests showed that the new RX and OTW VIATRAC™ 18 Peripheral Dilatation Catheters met acceptance criteria and performed in a manner equivalent to the predicate RX VIATRAC™ 14 Peripheral Dilatation Catheter and the Cordis OPTA 5TM PTA Catheter. No new safety or effectiveness issues were raised during the testing program.
- 8. Conclusions:
Since the new RX and OTW VIATRAC™ 18 Peripheral Dilatation Catheters have the same intended use, similar design and technological characteristics, equivalent performance properties, identical sterilization and packaging, and no new safety or effectiveness issues, the RX VIATRAC™ 18 Peripheral Dilatation Catheter and the OTW VIATRACTM 18 Peripheral Dilatation Catheter may be considered substantially equivalent to the predicate RX VIATRAC™ 14 Peripheral Dilatation Catheter and Cordis OPTA 5TM PTA Catheter.
{3}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an emblem that resembles a stylized eagle or bird-like figure with three curved lines representing its body and wings. The image is in black and white.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JAN 2 7 2000
Susan Silavin, Ph.D. Senior Regulatory Affairs Coordinator Guidant Corporation P.O. Box 58167 Santa Clara, CA 95052-8167
кадзеза Re: RX and OTW VIATRAC 18 Peripheral Dilatation Trade Name: Catheter Requlatory Class: II Product Code: LIT October 27, 1999 Dated: October 28, 1999 Received:
Dear Dr. Silavin:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 A substantially equivalent determination assumes compliance to 895. with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in requlatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}------------------------------------------------
Page 2 - Susan Silavin, Ph.D.
This letter will allow you to begin marketing your device as described ' in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at Also, please note the regulation entitled, (301) 594-4639. "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Christopher
Celia M. Witten, Ph.D., M.D. Acting Director Division of Cardiovascular, Respiratory and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Indications for Use Statement
510(k) Number (if known): K993639
# Device Name: RX VIATRAC™ 18 Peripheral Dilatation Catheter OTW VIATRACTM 18 Peripheral Dilatation Catheter
Indications for Use:
The RX VIATRAC™ 18 Peripheral Dilatation Catheter is intended:
- To dilate stenoses in the peripheral arteries (iliac, femoral, ilio-femoral, . popliteal, infra popliteal, renal arteries)
- For the treatment of obstructive lesions of native or synthetic arteriovenous . dialysis fistulae.
- For post deployment optimization of the 28 mm GUIDANT MEGALINK™ . Biliary Stent (6.0 to 10.0 mm diameters).
The OTW VIATRAC™ 18 Peripheral Dilatation Catheter is intended:
- To dilate stenoses in the peripheral arteries (iliac, femoral, ilio-femoral, . popliteal, infra popliteal, renal arteries)
- For the treatment of obstructive lesions of native or synthetic arteriovenous . dialysis fistulae.
- For post deployment optimization of the 28 mm GUIDANT MEGALINK™ . Biliary Stent (6.0 to 10.0 mm diameters).
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use | <u>X</u> | OR | Over-The-Counter | <u></u> |
|----------------------|----------|----|------------------|--------------------------|
| (Per 21 CFR 801.109) | | | | (Optional Format 1-1-96) |
Christopher for Witten
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.