The BEBIG IsoSeeds® Pd-103 are intended for use in the treatment of cancer with radioactive sources in close proximity to or within the tumor.
Device Story
BEBIG IsoSeeds® Pd-103 are cylindrical, sealed brachytherapy sources containing Pd-103 radioactivity; used for localized tumor treatment. Sources consist of metallic Pd-103 deposited within a porous ceramic tube, encased in a titanium capsule, and sealed via laser weld. A gold wire serves as a radiopaque marker. Sources are provided non-sterile and require sterilization before use. Clinicians implant these sources in close proximity to or within tumors to deliver localized radiation therapy. The device provides a therapeutic benefit by targeting malignant tissue while minimizing exposure to surrounding healthy structures.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by bench testing, adherence to international standards for sealed radioactive sources (ISO 2919, ISO 9978, ANSI/HPS N43.6-1997), and quality system compliance.
Technological Characteristics
Cylindrical sealed source, 4.5 mm length, 0.8 mm diameter. Materials: Titanium (ASTM F67-95 grade 2) capsule, porous ceramic core, gold wire radiopaque marker. Energy source: Pd-103 radionuclide (half-life 16.97 days). Sterilization: Steam sterilization required. Standards: ISO 2919, ISO 9978, ANSI/HPS N43.6-1997, DIN EN 25426.
Indications for Use
Indicated for treatment of selected localized tumors, including superficial, intraabdominal, and intrathoracic tumors (e.g., head, neck, lung, pancreas, prostate). May be used alone or in combination with external beam radiation.
Regulatory Classification
Identification
A radionuclide brachytherapy source is a device that consists of a radionuclide which may be enclosed in a sealed container made of gold, titanium, stainless steel, or platinum and intended for medical purposes to be placed onto a body surface or into a body cavity or tissue as a source of nuclear radiation for therapy.
Special Controls
*Classification.* Class II (special controls). A prostate seeding kit intended for use with a radionuclide brachytherapy source only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
K993552
# HAR 3 1 2006
# 510(k) Summary of Safety and Effectiveness
#### BEBIG IsoSeeds® Pd-103
#### Company Name
BEBIG Isotopentechnik und Umweltdiagnostik GmbH Robert Rössle Str. 10 13125 Berlin, Germany
#### Official Contact
Sven Langer Regulatory Affairs
#### Device Name
| Proprietary Name: | BEBIG IsoSeeds® Pd-103 |
|----------------------|---------------------------------------------------|
| Common Name: | Brachytherapy Palladium-103 Sources |
| Classification Name: | 21 CFR 892.5730 Radionuclide Brachytherapy Source |
## Predicate Devices used for Substantial Equivalence
| Device | Premarket # |
|----------------------------------------|-------------|
| UroMed Brachytherapy Iodine-103 Source | K982226 |
| Theragenics Modified Palladium Seed | K874787 |
## Intended Use
The BEBIG IsoSeeds® Pd-103 are intended for use in the treatment of cancer with radioactive sources in close proximity to or within the tumor.
#### Indications for use
The BEBIG Pd-103 Radionuclide Brachtherapy Sources are indicated for the treatment of selected localized tumors. These sources are commonly used to treat superficial, intraabdominal and intrathoracic tumors. Tumors of the head, neck, lung, pancreas and prostate are commonly treated. They may be used alone, or in combination with external beam radiation.
#### Description
The BEBIG IsoSeeds® Pd-103 are cylindrical sealead sources containing Pd-103 radioactivity. The sources are 4.5 mm long and 0.8 mm in diameter. The outer capsule of the source is composed of titanium being closed at one end and sealed by a laser weld at the other end in a final step. The metallic palladium-103 is deposited within a porous ceramic tube. A radiopaque marker is located in the center of the ceramic tube to serve as an x-ray marker. The radiopaque marker is composed of gold wire.
Palladium seeds have a half live of 16,97 days and are available in a range of activity levels. The BEBIG IsoSeeds® Pd-103 are provided non-sterile and must be sterilized prior to use.
{1}------------------------------------------------
#### Summary of Standards Achieved
Medizinproduktegesetz (German Law concerning medical devices)
Directive 90/385/EEC for active implants
DIN EN 1441 Medical devices- Risk analysis
DIN EN 25426 Sealed radioactive sources - Classification, requirements, test methods
ISO 2919 sealed radioactive sources - General requirements and classifications
ISO 9978: 1992(E), Radiation Protection-Sealed radioactive sources- Leakage test Methods
ANSI/HPS N43.6-1997-Sealed Radioactive Sources Classification
ASTM Standard for Titanium: F67-95 rev. 95 grade 2
DIN EN 554 Sterilization of medical devices - Validation and routine control of steam sterilization
DIN EN 980 Graphic Symbols used for labeling of medical devices
DIN EN 1041 Terminology, symbols and information associated with medical devices
DIN EN 25430 Radiation safety labeling
ISO 90001: Quality Management Systems
ISO 46001: Medical devices quality system
FDA Quality System Regulation 21 CFR Part 820 Good Manufacturing Practice.
DIN EN 6804 Radiation safety regulations for handling sealed radiation sources
DIN EN 25407 Shielding from ionizing radiation
DIN EN 25422 Storing of radioactive devices - Protection from radiation fire and theft
DIN EN 25425 Radionuclide labs - Design, radiation safety and fire protection
#### Summary
In Summary, the BEBIG IsoSeeds® Pd-103 are substantially equivalent to legally marketed devices. Quality System Controls assure the device is substantially equivalent to predicate devices with respect to its performance, safety, and effectiveness.
{2}------------------------------------------------
#### Certificates 10.
- TÜV Certificate for quality system according to 46001
- Certificate of ISO-Classification with protocol
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an emblem featuring a stylized depiction of three horizontal lines curving upwards, resembling a bird in flight or a symbolic representation of human services.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAR 3 1 2000
Dear Mr. Langer:
Sven Langer Regulatory Affairs BEBIG Isotopentechnik und Unweltdiagnostik GMBH Robert-Rossle/Str.10 D-13125 Berlin-Buch, Germany
Re:
K993552 IsoSeed Pd-103 Dated: January 10, 2000 Received: January 19, 2000 Regulatory class: II 21 CFR 892.5730/Procode: 90 KXK
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce pror to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class III (Premarket Aproval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalion assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notfication submission does not affect any obligation you micht have under sections 531 through 542 of the Act for devices under Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compiance at (301) 594-4591. Additionally, for question and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-5597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Daniel G. Schultz, M.D.
Daniel G. Schultz, M.D. Captain, USPHS Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure(s)
{4}------------------------------------------------
1 of 1 510(k) Premarket Notification Indication for Use BEBIG Brachytherapy Palladium-103 Sources
510(k) Number (if known): K
Device Name: BEBIG IsoSeed® Pd-103
Indications for Use:
The BEBIG Pd-103 Radionuclide Brachtherapy Sources are indicated for the treatment of The DEDIS I a 105 rainers. These sources are commonly used to treat superficial, intrasereotico roundou tumors. There box. Tumors of the head, neck, lung, pancreas and prostate aboommal and marations may be used alone, or in combination with external beam radiation.
Elvin A. Segears
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT and Radiological Devices 510(k) ivumber .
Prescription Use _ (Per 21 CFR 801.109)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.