K993535 · Riley Medical, Inc. · FRG · Dec 7, 2001 · General Hospital
Device Facts
Record ID
K993535
Device Name
METAPAK MULTI-PURPOSE INSTRUMENT TRAY
Applicant
Riley Medical, Inc.
Product Code
FRG · General Hospital
Decision Date
Dec 7, 2001
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6850
Device Class
Class 2
Indications for Use
The MetaPak Instrument Tray is used for loading surgical instruments in order to conveniently organize, sterilize, transport, and store the instruments between uses. The MetaPak Instrument Tray can be used in pre-vacuum steam and ethylene oxide sterilization cycles.
Device Story
MetaPak Multi-Purpose Instrument Tray is a container system for surgical instruments; consists of rectangular base and cover with latches; includes optional insert trays, protective mats, and slotted instrument bars/holders for stabilization. Used in clinical settings to organize, sterilize, transport, and store instruments between uses. Compatible with pre-vacuum steam and ethylene oxide sterilization cycles. Device protects instruments during processing and storage; facilitates efficient handling by surgical staff.
Clinical Evidence
Bench testing only. Validation testing confirmed that sterilization conditions were achieved in fully loaded MetaPak Instrument Trays of all three container base material compositions (aluminum, stainless steel, and aluminum/Radel) containing inserts, protective mats, and accessories in pre-vacuum steam and ethylene oxide sterilization processes.
Technological Characteristics
Container system consisting of rectangular base and lid with latches. Materials: Aluminum, stainless steel, and aluminum/Radel plastic combination. Accessories: Insert trays, protective mats, slotted instrument bars/holders. Compatible with pre-vacuum steam and ethylene oxide sterilization. Dimensions: Various sizes (e.g., 22"x10"x6", 26"x10"x6").
Indications for Use
Indicated for use in healthcare settings to organize, sterilize, transport, and store surgical instruments during pre-vacuum steam and ethylene oxide sterilization cycles.
Regulatory Classification
Identification
A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
{0}------------------------------------------------
DEC 0 7 2001
510(k) Summary
# Kg93535
## for
# Riley Medical, Inc. MetaPak Multi-Purpose Instrument Tray
- SPONSOR 1.
Riley Medical, Inc. 27 Wrights Landing Auburn, ME 04210
Contact Person: Mr. Martin M. Moore 207-786-2775 Telephone:
November 28, 2001 Date Prepared:
#### DEVICE NAME 2.
Riley Modical, Inc. MetaPak Multi-Purpose Instrument Proprietary Name:
Tray
Tray Instrument sterilization tray Common/Usual Name: Accessory to sterilization wrap Classification Name:
#### PREDICATE DEVICES 3.
- Riley MultiPak Surgical Instrument Tray .
- Sklar Instruments Instrument Tray u
#### DEVICE DESCRIPTION 4.
The Riley Medical MetaPak Multi-Purpose Instrument Tray (MetaPak Instrument Tray) is an instrument container consisting of a rectangular box base with a cover that fastens to the base using latches on the short ends. The following sizes of bases and lids are available:
- Tall hase and lid: 22" x 10" x 6" .
- Short base and lid: 22" x 10" x 4" .
- Extra long base and lid: 26" x 10" x 6" .
Riley Medical, Inc. Additional Information for K993535
ा -
November 28, 2001
Page C-1
{1}------------------------------------------------
Insert trays can be stacked in the container base to increase the instrument-carrying capacity of the container. The following insert sizes are available:
- . Full insert: 21" x 9.5" x 2.5"
- Small insert: 10" x 9.5" x 2.5" t
Protective mats are available that can be placed on the insert trays and in the bottom of the container base to stabilize and protect the instruments. Slotted instrument bars and holders are also available to organize, stabilize, and protect the instruments in the container.
#### INTENDED USE ર્ણ
The MetaPak Instrument Tray is used for loading surgical instruments in order to conveniently organize, sterilize, transport, and store the instruments between uses.
The MetaPak Instrument Tray can be used in pre-vacuum steam and ethylene oxide sterilization cycles.
#### TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE 6.
The proposed MetaPak Instrument Tray and the predicate MultiPak and Sklar Instrument Trays are all used for storage, transport, and sterilization of surgical instruments between uses. Both the proposed and predicate devices are suitable for use in pre-vacuum steam and ethylene oxide sterilization processes.
The MetaPak Instrument Tray is identical to the predicate MultiPak Instrument Tray, cleared for marketing as K944025, with the exception of a modification to the materials for the container base. The original MultiPak Instrument Tray container base was constructed entirely of Radel plastic. The MetaPak Instrument Tray container base will be offered in aluminum, stainless steel, and an aluminum/Radel combination with the long sides made of aluminum and short sides made of Radel plastic.
#### PERFORMANCE TESTING 7.
Validation testing was conducted which confirmed that sterilization conditions were achieved in fully loaded MetaPak Instrument Trays of all three container base material compositions containing one full insert and two half inserts, protective mats, and accessories in prc-vacuum steam and ethylene oxide sterilization processes.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the top half of the circle. Inside the circle is an abstract symbol that resembles three human profiles facing to the right, with flowing lines beneath them.
### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# DEC 0 7 2001
Riley Medical, Incorporated C/O Ms. Mary McNamara-Cullinane Staff Consultant Medical Device Consultants, Incorporated 49 Plain Street North Attleboro, Massachusetts 02760
Re: K993535
Trade/Device Name: Metapak Multi-Purpose Instrument Tray Regulation Number: 880.6850 Regulation Name: Instrument Sterilization Tray Regulatory Class: II Product Code: FRG Dated: October 9, 2001 Received: October 10, 2001
Dear Ms. Cullinane:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{3}------------------------------------------------
### Page 2 - Ms. Cullinane
You must comply with all the Act's requirements, including, but not limited to: registration You must comply with an all the Process of CFR Part 801); good manufacturing practice and fishing (21 CFR Part 001), abouting systems (QS) regulation (21 CFR Part 820); and if requirements as set form in are quady ion control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section This letter will and w you to oegal marketing of substantial equivalence of your device to 910(K) promative nowledge device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 If you desire specific as now as a more diagnostic devices), please contact the and additionally 27 OF FF & 1 301) 594-4618 . Additionally, for questions on the promotion and Office of Coffigur device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under nother (11 Cr Cr Cr or a from the Division of Small Manufacturers, International and the Not may be obtained its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
A. Ulatowski Timothy Director Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
510(k) Number (if known): K993535
Device Name: Riley Medical MetaPak Multi-Purpose Instrument Tray
Indications For Use:
The Riley Medical MetaPak Multi-Putpose Instrument Tray is used for loading surgical I he Teney Medical Weaks weaks organize, sterilize, transport, and store the instruments between uses.
The Riley Medical MetaPak Multi-Purpose Instrument Tray can be used in pre-vacuum steam and ethylene oxide sterilization cycles.
PLEASE DO NOT WRITE BELOW THIS LINE · CONTINUE ON ANOTHER PAGE IF NECESSARY)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Olin S. Lim
(Division Sign-Off) Division of Dental, Infection Control, and General Hospital Dev 510(k) Number
Prescription Use (Pcr 21 CFR 801.109) ાર
Over - The - Counter Use _____________________________________________________________________________________________________________________________________________________
(Optional Format 1-2-96)
Riley Medical, Inc. Additional Information for K993535 November 28, 2001
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.