K993530 · Math Solutions, LLC · LLZ · Dec 15, 1999 · Radiology
Device Facts
Record ID
K993530
Device Name
SYSTEM 2100
Applicant
Math Solutions, LLC
Product Code
LLZ · Radiology
Decision Date
Dec 15, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
The system is intended for use by any health practitioner who has the need for any of the below functions: 1. The software may be used to display medical images. The user may adjust the contrast of the images and image process the images with an edge enhancement algorithm. The user can zoom in on a region of an image, enlarging that region on the display screen. Other image processing functions may be added in the future to allow the user to enhance some aspect of an image, such as histogram processing, spatial filtering, frequency filtering. 2. Images from the same series can be stacked to form a three dimensional model. This will allow the use of the system to reformat images in other planes, such as coronal and sagittal planes. 3. Outlining tools are provided to use the system to outline regions of interest. These outlines may be combined to define three dimensional shapes which can be viewed in three dimensional perspective views. 4. The region of interest outlines may be written back to DICOM formatted image files for export to other systems. 5. The system can be used to determine the stereotactic coordinates of a point located on an image when a stereotactic system is employed. 6. The system may be used to provide support for any stereotactic system that consists of fixed rods which can be uniquely located on CT and MRI images. 7. The system may be used to solve for the transformation between two images sets, referred to as image fusion. 8. A region of interested outlined on one image set may be transferred to a fused image set. 9. The system may be used as a front end to other systems that do not provide image fusion. An example would be outlining a region of interest on MRI scans, transferring the outlines to CT scans, and writing the outlines back into DICOM image files. Those files can be available to other systems such as a radiation therapy treatment planning system. 10. The system may be used to print hardcopy images of selected images, reformatted images, and three dimensional perspective views. 11. Image segmentation tools are provided to create three dimensional structures for viewing in three dimensional perspective views. The intention is also to provide these basic functions as a C++ object library to any party that may wish to use the library to build their own applications. Those customers will be completely responsible for the integration, testing, use of the end product that they produce, and compliance with federal and state regulations.
Device Story
System 2100 is a software-based radiological image display and processing system. It accepts medical images (CT/MRI) as input. Core functions include image enhancement (contrast, edge, spatial/frequency filtering), 3D model reconstruction from stacked image series, image segmentation, and outlining of regions of interest (ROI). It supports stereotactic frame coordinate determination and image fusion (transformation between image sets) using stereotactic frames, common points, or common surfaces. Output includes displayed images, 3D perspective views, and DICOM-formatted files containing ROI contours for export to external systems like radiation therapy treatment planners. Used in clinical settings by health practitioners. It also functions as a C++ object library for third-party developers to integrate into other applications. The system aids clinicians in visualization and treatment planning by enabling multi-planar reformatting and cross-modality image registration.
Clinical Evidence
Bench testing only. No clinical data provided.
Technological Characteristics
Software-based radiological display system. Implemented as a C++ object library. Supports DICOM standard for image file input/output. Features include edge enhancement, histogram processing, spatial/frequency filtering, 3D reconstruction, and image fusion algorithms.
Indications for Use
Indicated for use by health practitioners requiring medical image display, processing, enhancement, 3D modeling, segmentation, stereotactic coordinate determination, and image fusion for diagnostic or treatment planning purposes.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
Predicate Devices
Precision View (Elekta Oncology)
Reference Devices
Xknife (Radionics)
Gamma Plan (Elekta Instruments, Inc.)
Render Plan 3-D (Elekta Oncology)
Prowess Systems CT simulation systems
CMS Focus (Computerized Medical Systems, Inc.)
Penicle (ADAC Laboratories)
Plato (Nucletron Corp.)
Submission Summary (Full Text)
{0}------------------------------------------------
## DEC 1 5 1999
General Purpose Radiological Image Display System File: 510ksum.doc Date: Dec. 7, 1999
## 510(k) Summary
System 2100 is a software program intended as a medical image display system that performs the following functions:
1. Displays medical images.
2. Can provide image processing and enhancements, such as contrast control, edge enhancement.
3. Can stack images from a series to form a three dimensional model.
4. Provides outlining and segmentation tools.
5. Can display outlined and segmented structures in three dimensional perspective views.
6. Provides generic support for stereotactic frames with the determination of the stereotactic coordinates of a point.
7. Provides three methods for solving for the transformation of coordinates between two image series, referred to as image fusion. The three methods are using stereotactic frames, using common points, and using common surfaces.
8. Can serve as a front end to other systems that do not have image fusion or segmentation methods.
9. Can be used as a means for printing hardcopy of images, reformatted images, and 3d perspective views.
10. Can serve as an object library for university research centers and other vendors (who are then responsible for the use and complying with federal and state regulations).
The system is intended for use by any health practitioner who may have need of the functions provided.
The capabilities provided in System 2100 can be found in a wide variety of present medical devices. Most CT and MRI consoles (General Electric Medical Systems, Rochester. NY) provide the means to view the images generated by those devices and will allow the reformatting of images in other planes.
Image fusion and support for stereotactic frames is provided by the product Xknife sold by Radionics, 2 Brookline Place, Suite 205, Brookline, MA 02146. Another product that supports a stereotactic frame and provides image fusion is Gamma Plan from Elekta Instruments, Inc., 3155 Northwoods Parkway, Norcross, Georgia, 30071.
System 2100 is similar to a predicate image display system call Precision View sold by Elekta Oncology (at the same address above for Elekta Instruments). Precision View has outlining and segmentation methods, image fusion, and may down load region of interest contours in DICOM files to a radiation therapy treatment planning system.
{1}------------------------------------------------
Outlining and solid model displays can be found in most therapy planning systems today, such as Render Plan 3-D, sold by Elekta Oncology (address above); Treatment planning and CT simulation systems sold by Prowess Systems, 1370 Ridgewood Dr., Suite 20, Chico, CA 95973; CMS Focus sold by Computerized Medical Systems, Inc., 1195 Corporate Lake Dr., St. Louis, MO 63132; Penicle sold by ADAC Laboratories, 540 Alder Dr., Milpitas, CA 905035; Plato sold by Nucletron Corp., 7080 Columbia Gateway Dr., Columbia, MD 21046.
Document approved by: Wendel Dean Renner
ﮩ
Winell Dean Kenner
Title: President
Date: Dec. 7, 1999
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol consisting of three stylized human profiles facing to the right, stacked one above the other.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DEC 1 5 1999
Wendel Dean Renner Math Resolutions LLC 5975 Gales Lane Columbia, MD 21045
Re: K993530 System 2100 (Radiological Display Software) Dated: October 15, 1999 Received: October 18, 1999 Regulatory class: II 21 CFR 892.2050/Procode: 90 LLZ
Dear Mr. Renner:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act). You may, therefore, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumotions. Failure to comply with the GMP requiration may result in requlatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your labeling requlation (21 CFR Part 801 and additionally 809.10 for in yitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
CAPT Daniel G. Schultz, M.D. Acting Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
l of Page
| 510(k) NUMBER (IF KNOWN): | K993530 |
|---------------------------|-------------|
| DEVICE NAME: | System 2100 |
INDICATIONS FOR USE:
see attached statement
.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PI IF NEEDED.) • .
Concurrence of CDRH, Office of Device Evaluation (ODE)
.
Prescription Use (Per 21 CFR 801.109)
:
OR
Over-The-Counter-U (Optional Format
Teriel C. Hyam
(Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological Devic
510(k) Number K943530
{4}------------------------------------------------
. .
file: induse2.doc
## 510(k) Number: K993530
Device Name: System 2100
Indications For Use:
The system is intended for use by any health practitioner who has the need for any of the below functions:
1. The software may be used to display medical images. The user may adjust the contrast of the images and image process the images with an edge enhancement algorithm. The user can zoom in on a region of an image, enlarging that region on the display screen. Other image processing functions may be added in the future to allow the user to enhance some aspect of an image, such as histogram processing, spatial filtering, frequency filtering.
2. Images from the same series can be stacked to form a three dimensional model. This will allow the use of the system to reformat images in other planes, such as coronal and sagittal planes.
3. Outlining tools are provided to use the system to outline regions of interest. These outlines may be combined to define three dimensional shapes which can be viewed in three dimensional perspective views.
4. The region of interest outlines may be written back to DICOM formatted image files for export to other systems.
5. The system can be used to determine the stereotactic coordinates of a point located on an image when a stereotactic system is employed.
6. The system may be used to provide support for any stereotactic system that consists of fixed rods which can be uniquely located on CT and MRI images.
7. The system may be used to solve for the transformation between two images sets, referred to as image fusion.
8. A region of interested outlined on one image set may be transferred to a fused image set.
9. The system may be used as a front end to other systems that do not provide image fusion. An example would be outlining a region of interest on MRI scans, transferring the outlines to CT scans, and writing the outlines back into DICOM image files. Those files can be available to other systems such as a radiation therapy treatment planning system.
{5}------------------------------------------------
10. The system may be used to print hardcopy images of selected images, reformatted images, and three dimensional perspective views.
11. Image segmentation tools are provided to create three dimensional structures for viewing in three dimensional perspective views.
The intention is also to provide these basic functions as a C++ object library to any party that may wish to use the library to build their own applications. Those customers will be completely responsible for the integration, testing, use of the end product that they produce, and compliance with federal and state regulations.
Document approved by: Wendel Dean Renner
Wendell Dean Kennon
Date: Dec. 7, 1999 Title: President
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.