VISUALINE V DRUGS OF ABUSE DIPSTRIP PANEL FOR QUALITATIVE DETERMINATION OF COCAINE, CANNABINOIDS, MORPHINE, METHAMPE
K993485 · Sun Biomedical Laboratories, Inc. · DIO · Dec 27, 1999 · Clinical Toxicology
Device Facts
Record ID
K993485
Device Name
VISUALINE V DRUGS OF ABUSE DIPSTRIP PANEL FOR QUALITATIVE DETERMINATION OF COCAINE, CANNABINOIDS, MORPHINE, METHAMPE
Applicant
Sun Biomedical Laboratories, Inc.
Product Code
DIO · Clinical Toxicology
Decision Date
Dec 27, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.3250
Device Class
Class 2
Indications for Use
Simultaneous qualitative visual tests for five (5) drugs of abuse: Cocaine, Morphine, Cannabinoids, Methamphetamine and Phencyclidine and their metabolites in urine sample.
Device Story
Visualine V Drugs of Abuse Dipstrip Panel is an in-vitro diagnostic test for qualitative detection of five drugs of abuse in human urine. Device utilizes competitive immunochromatographic technique; urine sample wicks through porous membrane; drug in sample competes with drug-labeled microspheres for limited antibody binding sites. Presence of drug inhibits formation of colored precipitin at probe site. Test performed by dipping panel into urine sample; results interpreted via visual color comparison after 5 minutes. Used in clinical or laboratory settings for rapid screening. Provides qualitative results to healthcare providers to assist in identifying potential drug use; requires confirmation by alternative chemical methods for definitive identification.
Clinical Evidence
Clinical correlation studies conducted at Redwood Toxicology Laboratory and Sun Biomedical Laboratories. Compared Visualine V Dipstrip Panel against Hitachi 717 EIA analyzer using blinded urine samples. Results: Cocaine (Sensitivity 98%, Specificity >99%, Efficiency >99%); Cannabinoids, Morphine, Methamphetamine, and Phencyclidine (Sensitivity >99%, Specificity >99%, Efficiency >99%). Reproducibility studies (within-run, run-to-run, day-to-day, lot-to-lot) all >99%.
Technological Characteristics
Competitive immunochromatographic assay; utilizes polyclonal and monoclonal antibodies; tracer is Ab colloidal complex; porous membrane test strip; visual color precipitin detection; storage 2-30°C; no ancillary equipment required.
Indications for Use
Indicated for the qualitative detection of Cocaine, Morphine, Cannabinoids, Methamphetamine, and Phencyclidine and their metabolites in human urine samples.
Regulatory Classification
Identification
A cocaine and cocaine metabolite test system is a device intended to measure cocaine and a cocaine metabolite (benzoylecgonine) in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of cocaine use or overdose.
Special Controls
*Classification.* Class II (special controls). A cocaine and cocaine metabolite test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).
Reference Devices
Hitachi 717 instrument (using Diagnostic Reagents, Inc. EIA)
Submission Summary (Full Text)
{0}------------------------------------------------
DEC 27 1999
K993485
SUN BIOMEDICAL LABORATORIES, INC.
604 VPR CENTER, 1001 LOWER LANDING ROAD, BLACKWOOD, NJ 0801 TEL: (609) 401-1080 • FAX: (609) 401-1090 • E-mail: sunlabs@juscom.c
## 510(k) CONTENT SUMMARY
1. Name of Manufacturer:
Sun Biomedical Laboratories, Inc. 604 VPR Center 1001 Lower Landing Road Blackwood, NJ 08012
- 2. Trade Name Visualine®V Drugs of Abuse Dipstrip Panel for Qualitative Determination of Cocaine, Cannabinoids, Morphine, Methamphetamine, and Phencyclidine and their metabolites in Human Urine Samples.
- 3. Common Name
An in-virro immunoassay test by visual color comparison for the detection of Cocaine. Carnabinoids, Morphine, Methamphetamine, and Phencvclidine in human urine samples,
### 4. Regulation # and Classification:
Reg. #862-3170, Class II Device
- 5. Test Description:
The Visualine® V Drugs of Abuse Dipstrip Panel test is based on the principle of antigen-antibody complexation and is used for the analysis of Cocaine, Cannabinoids, Morphine, Methamphetamine, and Phencyclidine and their corresponding metabolites in human urine samples. The assay utilizes a competitive immunochromatographic technique involving a sample of test urine delivered through a wicking action as the panel (holding the porous membrane) is dipped into the urine sample. The drug in the sample compeics for the limited antibody sites on the colored microspheres. When an adequate amount of drug is present, it will fill the limited antibody binding sites. This will prevent attachment of the colored microspheres to the probe site on the membrane. Therefore, a positive urine sample will inhibit the formation of precipitin at the probe site.
{1}------------------------------------------------
# 6. Comparison of Two Test Systems for Correlation Studies:
The Visualine@V Drugs of Abuse Dipstrip Panel for Qualitative Determination of Cocaine, Cannabinoids, Morphine, Methamphetamine, and Phencyclidine assay is correlated to the individual Hitachi 717 instrument using Diagnostic Reagents, Inc. material. The following table illustrates the similarities and differences between the two assays.
| | Hitachi 717<br>(Diagnostic Reagents, Inc EIA) | Visualine®V Drugs of Abuse<br>Dipstrip Panel for Qualitative<br>Determination of Cocaine,<br>Cannabinoids, Morphine,<br>Methamphetamine, and<br>Phencyclidine |
|---------------------|-----------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Test Principle | Homogenous enzyme<br>immunoassay | Competitive binding<br>immunoassay |
| Sample/Sample Size | 200 uL urine | Approx. 200 ul urine |
| Antibody | Polyclonal and Monoclonal | Polyclonal and Monoclonal |
| Tracer | Drug-Glucose-6-Phosphate<br>Dehydrogenase | Ab Colloidal Complex |
| Detection Method | Change in absorbance (ΔΑ) value<br>detected spectrophotometrically | Visual color<br>precipitin formation |
| Test Run Time | 10-20 minutes, dependent on test | 5 minutes |
| Storage Requirement | 2-8°C (36-46°F) | 2-30°C (36-86°F) |
| Detection Level | Cocaine<br>300 ng/ml<br>Cannabinoids<br>50 ng/ml<br>Morphine<br>300 ng/ml<br>Methamphetamine<br>1000 ng/ml<br>Phencyclidine<br>25 ng/ml | 300 ng/ml<br>50 ng/ml<br>300 ng/ml<br>1000 ng/ml<br>25 ng/ml |
| Ancillary Equipment | 717 Hitachi Analyzer, EIA<br>Calibrators | none |
# 7. Visualine®V Dipstrip Panel Performance Characteristics
- A. Correlation studies between Diagnostic Reagents, Inc. EIA individual assays and Visualine® V dipstrip Panel for Cocaine and it's metabolites, Cannabinoids and it's metabolites, Morphine and it's metabolites, Methamphetamine, and Phencyclidine were conducted at Redwood Toxicology Laboratory, Santa Rosa California and Sun Biomedical Laboratories. Presumptive positive samples (tested on the Hitachi) for each drug were provided by Redwood and tested at Sun Biomedical Labs where blinded studies were conducted to determine the correlation between the two methodologies. Negative samples consisted of UTAK Laboratories Drug Free Urine
{2}------------------------------------------------
pool, Lot# 2647, and in-house negative urines tested with Sun Biomedical's Visualine® II product. Correlation with Hitachi 717 yielded the following data:
| Cocaine: | Sensitivity | 49 / 50 | 98 % |
|------------------|-------------|-----------|-------|
| | Specificity | 247 / 247 | >99 % |
| | Efficiency | 296 / 297 | >99 % |
| Cannabinoids: | Sensitivity | 50 / 50 | >99 % |
| | Specificity | 253 / 253 | >99 % |
| | Efficiency | 303 / 303 | >99 % |
| Morphine: | Sensitivity | 50 / 50 | >99 % |
| | Specificity | 257 / 257 | >99 % |
| | Efficiency | 307 / 307 | >99 % |
| Methamphetamine: | Sensitivity | 50 / 50 | >99 % |
| | Specificity | 255 / 255 | >99 % |
| | Efficiency | 305 / 305 | >99 % |
| Phencyclidine: | Sensitivity | 50 / 50 | >99 % |
| | Specificity | 255 / 255 | >99 % |
| | Efficiency | 305 / 305 | >99 % |
#### B. Specificity and Substances Detected:
The individual tests are specific to the labeled drug of abuse or structurally related The test detects Cocaine and its metabolites at 300 ng/ml, compounds. Cannabinoids and its metabolites at 50 ng/ml, Morphine and its metabolites at 300 ng/ml, Methamphetamine at 1000 ng/ml and Phencyclidine at 25 ng/ml.
- C. Reproducibility studies for Cocaine, Cannabinoids, Morphine, Precision: Methamphetamine and PCP indicate:
| Within run and run to run | > 99 % |
|---------------------------|--------|
| Within day and day to day | > 99 % |
| Within lot and lot to lot | > 99 % |
{3}------------------------------------------------
| 8. Attachments:<br>Correlation Studies | (Accuracy) | Section A |
|-----------------------------------------------|---------------------------------------------------------------------------------|-----------|
| Specificity Studies | (Interference substances and cutoff levels of<br>Each individual Drug of Abuse) | Section B |
| Sensitivity Studies | (Analytical studies) | Section C |
| Reproducibility Studies | (Precision) | Section D |
| Visualine® V Dipstrip Panel<br>Package Insert | | Section E |
| DRI Inserts for Hitachi 717 | | Section F |
| Label Copies | | Section G |
| Stability Studies | | Section H |
:
:
"
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular arrangement of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES USA". Inside the circle is a stylized image of an eagle with three human profiles incorporated into the design of the eagle's neck and head. The eagle is facing to the left.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
DEC 2 7 1999
Ming Sun, Ph.D. President Sun Biomedical Laboratories, Inc. 604 VPR Center 1001 Lower Landing Road Blackwood, New Jersey 08012
Re: K993485
> Trade Name: Visualine@ V Drugs of Abuse Dipstrip Panel Regulatory Class: II Product Code: DIO, LDJ, DJG, DKZ, LCM Dated: October 11. 1999 Received: October 14, 1999
Dear Dr. Sun:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{5}------------------------------------------------
#### Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Steven Putman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
1 of
Page__________________________________________________________________________________________________________________________________________________________________________
K 993485 S10(k) Number (if known): _
Visualine V Drugs of Abuse Dipstrip Panel crice Name:_
Indications For Use:
Simultaneous qualitative visual tests for five (5) drugs of abuse: Cocaine, Morphine, Cannabinoids, Methamphetamine and Phencyclidine and their metabolites in urine sample.
Den Cooper
(Division Sign-Off)
(Division Sign-Off)
Division of Clinical Laboratory Devices
510(k) Number K993485
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH. Office of Device Evaluation (ODE)
**Prescription Use**
(Per 21 CFR 801.109)
OR
Over-The-Counter Use_
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.