K993464 · C.R. Bard, Inc. · MJC · Dec 10, 1999 · Gastroenterology, Urology
Device Facts
Record ID
K993464
Device Name
BARDEX I.C. PEDIATRIC FOLEY CATHETER
Applicant
C.R. Bard, Inc.
Product Code
MJC · Gastroenterology, Urology
Decision Date
Dec 10, 1999
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 876.5130
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Bardex I. C. Pediatric Foley Catheter is intended for use in the drainage and/or collection and/or measurement of urine. Generally, drainage is accomplished by inserting the catheter through the urethra and into the bladder. However, drainage is sometimes accomplished by suprapubic or other placement of the catheter, such as a nephrostomy tract.
Device Story
Bardex I. C. Pediatric Foley Catheter is a two-way latex urinary catheter; features silver and hydrogel coatings. Device inserted through urethra, suprapubic, or nephrostomy tract into bladder for urine drainage, collection, or measurement. Used in clinical settings by healthcare professionals. Silver coating provides antimicrobial properties; hydrogel coating enhances lubricity for patient comfort during insertion and indwelling use. Output is physical urine drainage; aids clinical management of urinary retention or monitoring. Benefits include reduced risk of catheter-associated complications via antimicrobial surface and improved patient comfort.
Clinical Evidence
Bench testing only. Device held to same design, manufacture, and performance specifications as currently marketed Foley catheters. Testing conducted in accordance with FDA draft 'Guidance for the Content of Premarket Notifications for Conventional and Antimicrobial Foley Catheters' (September 12, 1994).
Technological Characteristics
Two-way latex Foley catheter; 8 and 10 French sizes; 3cc balloon. Surface features metallic silver coating and hydrogel hydrophilic polymer coating applied from bifurcation to tip (internal/external, including balloon). Sterile, disposable. No electronic or software components.
Indications for Use
Indicated for pediatric patients requiring urinary drainage, collection, or measurement via urethral, suprapubic, or nephrostomy tract placement.
Regulatory Classification
Identification
A urological catheter and accessories is a flexible tubular device that is inserted through the urethra and used to pass fluids to or from the urinary tract. This generic type of device includes radiopaque urological catheters, ureteral catheters, urethral catheters, coudé catheters, balloon retention type catheters, straight catheters, upper urinary tract catheters, double lumen female urethrographic catheters, disposable ureteral catheters, male urethrographic catheters, and urological catheter accessories including ureteral catheter stylets, ureteral catheter adapters, ureteral catheter holders, ureteral catheter stylets, ureteral catheterization trays, and the gastro-urological irrigation tray (for urological use).
Predicate Devices
Bard Hydrogel/Silver-Coated Foley Catheter
Bard Latex Urinary Catheters
Bard Hydrogel-Coated Foley Catheters
Submission Summary (Full Text)
{0}------------------------------------------------
K993464 Euge 1of2
Bard Medical Division C. R. Bard, Inc. 8195 Industrial Blvd. Covington, GA 30209-2695
DEC I 0 1999
BARID
### SECTION VI
## 510 (k) SUMMARY OF SAFETY AND EFFECTIVENESS INFORMATION
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990.
A. Submitter's Information:
| Submitter's Name: | C. R. Bard, Inc., Medical Division |
|-------------------------|---------------------------------------------------|
| Address: | 8195 Industrial Blvd.<br>Covington, Georgia 30014 |
| Contact Person: | Georgia C. Abernathy |
| Contact Person's Phone: | (770) 784-6454 |
| Contact Person's Fax: | (770) 784-6419 |
| Date of Preparation: | October 8, 1999 |
#### B. Device Name:
| Trade Name: | Bardex® I. C. Pediatric Foley Catheter |
|----------------------|-------------------------------------------------------------------|
| Common / Usual Name: | Latex Pediatric Foley Catheter with lubricious and silver coating |
| Classification Name: | Urological catheter (antimicrobial) and accessories |
#### C. Predicate Device Names:
| Trade Name: | Bard Hydrogel/Silver-Coated Foley Catheter |
|-------------|--------------------------------------------|
| Trade Name: | Bard Latex Urinary Catheters |
| Trade Name: | Bard Hydrogel-Coated Foley Catheters |
#### D. Device Description:
The Bardex I. C. Pediatric Foley Catheter is a two-way latex Foley catheter with silver and hydrogel coatings.
#### E. Intended Use:
The Bardex I. C. Pediatric Foley Catheter is intended for use in the drainage and/or collection and/or measurement of urine. Generally, drainage is accomplished by inserting the catheter through the urethra and into the bladder. However, drainage is sometimes accomplished by suprapubic or other placement of the catheter, such as a nephrostomy tract.
#### F. Technological Characteristics Summary:
Table VI-1 provides a tabulated comparison summary of the technological characteristics of the Bardex I.C. Pediatric Foley Catheter versus the predicate devices.
{1}------------------------------------------------
K993464 Page Zixi
| Component/<br>Characteristic | Bardex I. C.<br>Pediatric Foley Catheter<br>(Hydrogel Impregnated) | Bard Hydrogel/<br>Silver-Coated Foley<br>Catheter | Bard Catheters | Bard Hydrogel-Coated<br>Foley Catheters | Ultragen |
|-------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------|
| Indications or<br>Intended Use | The Bardex I. C.<br>Pediatric Foley Catheter<br>is intended for use in<br>the drainage and/or<br>collection and/or<br>measurement of urine.<br>Generally, drainage is<br>accomplished by<br>inserting the catheter<br>through the urethra and<br>into the bladder.<br>However, drainage is<br>sometimes<br>accomplished by<br>suprapubic or other<br>placement of the<br>catheter, such as a<br>nephrostomy tract. | The Bard Hydrogel/<br>Silver-Coated Foley<br>Catheter is intended for<br>use in the drainage<br>and/or collection and/or<br>measurement of urine.<br>Generally, drainage is<br>accomplished by<br>inserting the catheter<br>through the urethra and<br>into the bladder.<br>However, drainage is<br>sometimes<br>accomplished by<br>suprapubic or other<br>placement of the<br>catheter, such as a<br>nephrostomy tract. | Bard catheters are<br>intended for use in the<br>drainage and/or<br>collection and/or<br>measurement of urine.<br>Generally, drainage is<br>accomplished by<br>inserting the<br>catheter/drain through<br>the urethra and into the<br>bladder. However,<br>drainage is sometimes<br>accomplished by<br>suprapubic or other<br>placement of the<br>catheter/drain, such as a<br>nephrostomy tract. | Bard Hydrogel-Coated<br>Foley catheters are<br>intended for use in<br>drainage of urine.<br>Generally, drainage is<br>accomplished by<br>inserting the catheter<br>through the urethra and<br>into the bladder.<br>However, drainage is<br>sometimes<br>accomplished by<br>suprapubic or other<br>placement of the<br>catheter, such as a<br>nephrostomy tract. | No<br>substantial<br>difference |
| Disposable | Yes | Yes | Yes | Yes | None |
| Sterile | Yes | Yes | Yes | Yes | None |
| Catheter Base<br>Material | latex | latex | latex | latex | None |
| French sizes<br>Available** | 8 and 10 Fr. | 12-30 Fr. | Includes 8 and 10 Fr. | Includes 8 and 10 Fr. | Pediatric<br>sizes added<br>to hydrogel/<br>silver-coated<br>catheter line |
| Balloon sizes | 3cc | Smallest is 5cc | Includes 3cc | Includes 3cc | #1 identical<br>to #3 and #4 |
| Silver Coating<br>Form | Metallic | Metallic | N/A | N/A | #1 and #2<br>identical |
| Lubricious<br>Coating | Hydrogel hydrophilic<br>polymer | Hydrogel hydrophilic<br>polymer | Hydrogel hydrophilic<br>polymer | Hydrogel hydrophilic<br>polymer | None |
| Catheter<br>Surface<br>Hydrogel<br>Coated | From bifurcation to tip,<br>internal and external<br>including balloon | From bifurcation to tip,<br>internal and external<br>including balloon | From bifurcation to tip,<br>internal and external<br>including balloon | From bifurcation to tip,<br>internal and external<br>including balloon | None |
| Catheter<br>Surface Silver<br>Coated | From bifurcation to tip,<br>internal and external<br>including balloon | From bifurcation to tip,<br>internal and external<br>including balloon | N/A | N/A | #1 and #2<br>identical |
Table VI-1
** New feature(s) or a change in this 510(k)
Performance Data Summary: G.
> The Bardex I.C. Pediatric Folcy Catheter referenced in this submission is held to the same design, manufacture, and performance specifications as those Foley catheters currently manufactured. Performance and functional testing standards are based on the FDA draft "Guidance for the Content of Premarket Notifications for Conventional and Antimicrobial Foley Catheters'' dated September 12, 1994.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features a stylized image of an eagle with its wings spread, symbolizing protection and care. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle, indicating the department's name and national affiliation. The seal is simple, using only black and white, which gives it a clean and official appearance.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DEC I 0 1999
Ms. Georgia C. Abernathy Regulatory Affairs Associate Bard Medical Division C. R. Bard, Inc. 8195 Industrial Blvd. Covington, GA 30014
Re: K993464 Bardex® I. C. Pediatric Foley Catheter Dated: October 8, 1999 Received: October 13, 1999 Regulatory Class: II 21 CFR 876.5130/Procode: 78 KOD and MJC
Dear Ms. Abernathy:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have have have have have have have have have have have headly vve have reviewed your section 310(i) notifications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactings of the marketed predicate devices marketing in merse phor to high as with the provisions of the provisions of the Medical Device Antendhens, or to device that have both realer market the device, subject to the general control Federal FOOD, DTQ, and Cosment Act (Act (Act include requirements for annual reguirements for annual registration, listing of provisions of the Act. The general obtitled provibitions against misbranding and adulteration.
If your device is classified (see above) into either class III (Special Controls) or class III (Premarket Approval), it may if your device is classlited (Sec above) into entrel sides in (Special on affecting your device can be found in the Code of be subject to such additional oonly in John Ally equivalent determination assumes compliance with r ederal Regulations, Thic Pr, This 600 to Society System Regulation (GS) for the Current Good Manufacturing Practions) as a that, through periodic QS inspections, the Food and Medical Devices: "Ocheral rogulation" (2) Stssumptions. Failure to comply with the GMP regulation may result in Drug Action (1 DA) will ron) Seen blish further announcements concerning your device in the Federal register. In addition, von may pour premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA r ma reter will and you to bogin marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for ஸ் with diagnostic devices), please contact the Office of Complance at (301) 594-4613. Additionally for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, the promotion and advertising by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
CAPT Daniel G. Schultz, M.D. Acting Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# SECTION I - D
# INDICATIONS FOR USE STATEMENT
5993444 and the same of the same of the same of the same of the same of the same of the same of the same of the states of the states of the states of the states of the states of the 510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________ Device Name: _________________________________________________________________________________________________________________________________________________________________
Device Name: Bardex® I. C. Pediatric Foley Catheter
Indications for Use:
The Bardex I. C. Pediatric Foley Catheter is intended for use in the drainage is I he Bardex T. C. Pediativ Poloy Catholor and of urine. Generally, drainage is and/or collection and/or measurement of the urethra and into the bladder.
accomplished by inserting the catheter through the or other placement accomplished by inserting the catherer through the around.
However, drainage is sometimes accomplished by suprapubic or other placement frowover, ateming
of the catheter, such as a nephrostomy tract.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
CONCURRENCE OF CDRH, OFFICE OF DEVICE EVALUATION (ODE)
| Prescription Use<br>(Per 21 CFR 801.109) | OR Over-The-Counter Use |
|------------------------------------------|-------------------------|
|------------------------------------------|-------------------------|
(Optional Format 1/2/96)
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT,
and Radiological Devices
| 510(k) Number | K993464 |
|---------------|---------|
|---------------|---------|
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.