← Product Code [ITI](/productcode/ITI) · K993406

# TRAX, POWERED WHEELCHAIR, MODEL 308686-00-0 (K993406)

_Permobil AB · ITI · Dec 18, 2000 · Physical Medicine · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K993406

## Device Facts

- **Applicant:** Permobil AB
- **Product Code:** [ITI](/productcode/ITI.md)
- **Decision Date:** Dec 18, 2000
- **Decision:** SESE
- **Submission Type:** Abbreviated
- **Regulation:** 21 CFR 890.3860
- **Device Class:** Class 2
- **Review Panel:** Physical Medicine
- **Attributes:** Therapeutic

## Indications for Use

The intended use of TRAX is to provide indoor and outdoor mobility to persons restricted to a sitting position.

## Device Story

Permobil TRAX is a rear-wheel-driven, battery-powered wheelchair. It utilizes two 12V 97Ah gel batteries and two 12V DC permanent magnet motors with integrated parking brakes. The user operates the device via a joystick controller (SAFEGATE) for proportional speed and steering. The base is constructed of welded steel with laser-cut sheet details, featuring a suspension system with stabilizers and an extendable, pivoted front section. The device provides mobility for users restricted to a sitting position in indoor and outdoor environments. It does not incorporate complex software algorithms or AI; control is achieved through standard electronic motor management.

## Clinical Evidence

No clinical data; bench testing only.

## Technological Characteristics

Welded steel base with laser-cut sheet details; two 12V DC permanent magnet motors; two 12V 97Ah gel batteries; joystick-based proportional control; SAFEGATE power electronics; rear-wheel drive; suspension with stabilizer; teleflex wire steering.

## Regulatory Identification

A powered wheelchair is a battery-operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.

## Predicate Devices

- Max90 ([K870224](/device/K870224.md))
- Exterior ([K870225](/device/K870225.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the word "permobil" in a bold, sans-serif font. The "o" in "permobil" is slightly larger than the other letters. The word is enclosed in a thin, black oval. The image is in black and white.

K99 3906

# 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS

# PERMOBIL POWERED WHEELCHAIR TRAX

8 2000

| Submitter           | Permobil AB<br>Box 120<br>S-861 23 Timrå<br>Sweden | DEC 1 |
|---------------------|----------------------------------------------------|-------|
| Phone               | +46 60 59 59 00                                    |       |
| Fax                 | +46 60 57 52 50                                    |       |
| Contact person      | Bengt Persson                                      |       |
| E-mail address      | bengt.persson@permobil.se                          |       |
| Date Prepared       | September 30 1999                                  |       |
| Device name         | TRAX                                               |       |
| Classification name | Powered wheelchair                                 |       |

### ldentification of predicate devices Max90 (K870224) Exterior (K870225)

#### Intended use

The intended use of TRAX is to provide indoor and outdoor mobility to persons restricted to a sitting position.

#### Description of the device

The Permobil powered wheelchair TRAX is rear wheel driven, battery powered, motor driven and is controlled by the Permobil wheelchair controller SAFEGATE. The user interface is a joystick manufactured by Sakai for proportional control of the speed and the steering.

The wheelchair is powered by two 12V 97 Ah gel batteries.

Theoretical driving range on the fully charged batteries is 30 - 50 km depending of the terrain the chair is driven on.

The wheelchair consists of the following basic sub-sections:

- · Base with two direct-drive units with integrated parking brakes, two 10"x3" rear wheels, suspension with stabilizer and two 12V 97Ah batteries.
- Extendable and pivoted front, length adjustable 0-200 mm (0-8"), teleflex . wire steering, suspension and two 8"x2,5" front wheels.
- 
- SAFEGATE TRAX Power electronics. .
- · Seating system.

The base is of welded steel construction with laser cut sheet details. The direct-drive units consist of two 12V DC permanent magnet motors with gears and integrated parking brakes.

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the word "permobil" in a bold, sans-serif font. The word is enclosed in a black oval. The letters are all lowercase except for the "o" in "permobil", which is subscripted.

## 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS

## PERMOBIL POWERED WHEELCHAIR TRAX

### Substantial equivalence

The Permobil powered wheelchair TRAX is substantial equivalent with respect to, intended use, energy source and materials, because those are identical to the predicate device(s).

#### Safety and effectiveness

The Permobil powered wheelchair ITRAX has in substantials the same technological characteristics and the same safety and effectiveness as the predicate device(s) and the design changes declared in this submission do not raise new questions of safety and effectiveness.

#### Indication for Use

The intended use is to provide indoor and outdoor mobility to persons restricted to a sitting position.

#### 510(k) number

Not assigned at the writing of this submission

#### Device name

TRAX

### Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription use (per 21 CFR 801.109) []

or

#### Over the counter use

(Division Sign-Off) Division of Ggeneral Restorative Devices

510(k) number

{2}------------------------------------------------

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird with three stylized wing-like shapes. The symbol is rendered in black, and the text is also in black against a white background.

#### Public Health Service

DEC 1 8 2000

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Mr. Bengt Persson Quality Manager Permobil AB Engelbrekts vag 101 S-191 62 Sollentuna Sweden

Re: K993406

Trade Name: Trax, Powered Wheelchair, Model 308686-00-0 Regulatory Class: II Product Code: ITI Dated: November 22, 2000 Received: November 22, 2000

Dear Mr. Persson:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

{3}------------------------------------------------

### Page 2 - Mr. Bengt Persson

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

Mark N Millikan

Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## 510(k) notification K993406

FDA's request for additional information dated, January 6, 2000.

Enclosure D

# 510(k) Number (if known):

## K9930406

Permobil TRAX

Device Name:

Indication For Use:

Powered wheelchair, battery-operated device with wheels that is intended for medical purposes to provide indoor and outdoor mobility to persons restricted to a sitting position. Class II. The indoor use is depending of available space.

# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109) OR

Over-The-Counter Use (Optional Format 1-2-96)

Mark N. Mulkison

(Division Sign-Off)
Division of General Restorative Devices
K993906

510(k) Number ***_***

---

**Source:** [https://fda-staging.innolitics.com/device/K993406](https://fda-staging.innolitics.com/device/K993406)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com