The subject devices are single use components, intended for use in conjunction with associated acetabular shells, femoral bearings, and femoral hip stems as part of a cemented or cementless total hip replacement procedure. Indications for use, in keeping with those of other commercially available Class II total hip devices, are as follows: - noninflammatory degenerative joint disease including osteoarthritis and avascular necrosis, - rheumatoid arthritis, - correction of functional deformity, - revision procedures where other treatments or devices have failed, and - treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.
Device Story
Crossfire™ System 12® Acetabular Inserts are UHMWPE hip replacement components; intended for use in cemented or cementless total hip arthroplasty. Device functions as a bearing surface within an acetabular shell; mates with femoral bearings and stems. Crossfire™ process involves crosslinking polyethylene to reduce wear rates. Used in clinical settings by orthopedic surgeons. Output is a mechanical bearing surface; clinical decision-making involves restoring joint function and mobility. Benefits include reduced wear compared to standard polyethylene, potentially extending implant longevity.
Clinical Evidence
Bench testing only. Fatigue loading of the smallest insert size (P1) assembled in metal backings for 10^7 cycles showed no structural compromise. Axial distraction testing confirmed sufficient locking strength pre- and post-fatigue. Wear performance was characterized via in vitro hip joint simulation (5 million cycles) using bovine calf serum, demonstrating 78-90% reduction in gravimetric wear compared to standard polyethylene. Note: in vitro results do not quantitatively predict clinical wear.
Technological Characteristics
UHMWPE per ASTM-F648; crosslinked via process cleared in K974685. Components are acetabular inserts for total hip replacement. Mechanical locking mechanism within acetabular shells. Single-use, sterile.
Indications for Use
Indicated for patients requiring total hip replacement due to noninflammatory degenerative joint disease (osteoarthritis, avascular necrosis), rheumatoid arthritis, functional deformity, revision needs, or proximal femur fractures (nonunion, femoral neck, trochanteric) with head involvement.
Regulatory Classification
Identification
A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.
{0}------------------------------------------------
(993352
NOV 1 9 1999
Additional Crossfire™Acetabular Components
## 510(k) Summary
# Additional Crossfire™ Acetabular Components
Submission Information
| Name and Address of the Sponsor: | Howmedica Osteonics Corp. |
|----------------------------------|---------------------------|
| | 59 Route 17 |
| | Allendale, NJ 07401-1677 |
| Contact Person: | Terry Sheridan Powell |
| Date of Summary Preparation: | November 19, 1999 |
**Device Identification**
| Proprietary Name: | Crossfire™ System 12® Acetabular Inserts |
|-------------------|--------------------------------------------|
| Common Name: | Artificial Total Hip Replacement Component |
| | (Acetabular Inserts) |
| Classification Name and Reference: | 888.3358: Prosthesis, hip, semi-constrained, metal- |
|------------------------------------|-----------------------------------------------------|
| | polymer, porous, uncemented |
#### Predicate Device Identification
- Osteonics Polyethylene Acetabular Inserts, fabricated from crosslinked polyethylene . called "Crossfire", were determined Substantially Equivalent via 510(k) #K974685 on 12/16/97 (sponsor: Osteonics Corp.).
{1}------------------------------------------------
.
- Howmedica System 120 Acetabular Inserts (non-Crossfire) were determined
Substantially Equivalent as follows:
- via 510(k) #K903362. These were room air irradiated versions, previously called -Osteolock inserts (sponsor: Howmedica Inc..).
- via 510(k) #K951114. These were room air irradiated versions, specifically for the . additional P1 26mm size.
- via 510(k) #K951115. These were room air irradiated versions, specifically for the additional 22mm sizes.
- via 510(k) #K934060. These were Duration® Stabilized versions. -
- via 510(k) #K963612. This submission allowed additional wear claims for the -Duration® Stabilized versions.
#### Device Description
The subject Crossfire™System 12® Acetabular Inserts are the same as the predicate non-Crossfire™System 12® Acetabular Inserts, except that the subject devices are fabricated from polyethylene that has been crosslinked via the procedure described in, and found Substantially Equivalent in, 510(k) #K974685.
#### Intended Use:
The subject devices are single use components, intended for use in conjunction with associated acetabular shells, femoral bearings, and femoral hip stems as part of a cemented or cementless total hip replacement procedure. Indications for use, in keeping with those of other commercially available Class II total hip devices, are as follows:
{2}------------------------------------------------
- noninflammatory degenerative joint disease including osteoarthritis and avascular . necrosis,
Necrosis,
- rheumatoid arthritis, .
- correction of functional deformity, .
- . revision procedures where other treatments or devices have failed, and
- treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur . with head involvement that are unmanageable using other techniques.
#### Statement of Technological Comparison:
#### Materials
The subject and predicate Crossfire™ System 120 Acctabular Inserts are manufactured from UHMWPE per ASTM-F648. The UNMWPE material to be used for the subject System 120 Acetabular Inserts will be crosslinked in exactly the same manner used for the acetabular inserts determined Substantially Equivalent via K974685.
### Indications for Use
The indications for use of the subject and predicate System 12@ Acetabular Inserts are identical.
#### Design
The subject (Crossfire) and predicate (non-Crossfire) System 12@ Acetabular Inserts are identical in design. Additional testing has been performed on Crossfire™ versions of the System 12@ Acetabular Inserts to validate that these inserts are not adversely affected, with regard to locking strength or fatigue strength, as a result of the Crossfire™ process.
{3}------------------------------------------------
510(k) Summary
#### Additional Crossfire™Acetabular Components Performance Data:
Crossfire polyethylene has been well characterized previously under 510(k) #K974685. Testing of the subject acetabular inserts, therefore, focused on the structural aspects of System 12 inserts assembled within Howmedica-design (e.g., Osteolock) acetabular shells.
The smallest insert size (P1) was assembled within metal backings and fatigue loaded. After 107 cycles, three inserts were axially distracted (pushed-out) and two were carefully removed by sectioning their metal shells. All inserts were inspected for signs of structural compromise with no gross evidence of impending catastrophic failure. Other surface features, e.g. screw hole impressions, were consistent with those observed in an earlier study used to originally qualify Osteonics-design Crossfire inserts. Axial distraction test results demonstrated sufficient locking strength pre-and post- fatigue testing.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized image of a bird with three heads facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the bird image.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV 1 9 1999
Ms. Elizabeth Staub ris. Birzabeen boaan
Vice President, Quality Assurance/Regulatory Affairs/Clinical Research Howmedica Osteonics Corporation 59 Route 17 07401-1677 Allendale, New Jersey
K993352 Re: Additional Crossfire™ Acetabular Components Trade Name: Regulatory Class: II Product Codes: LPH and JDI Dated: November 4, 1999 Received: November 8, 1999
Dear Ms. Staub:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. ਜੋ substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{5}------------------------------------------------
#### Page 2 - Ms. Elizabeth Staub
This letter will allow you to beqin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Russell J. Tayer
for James E. Dillard III Acting Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## 510(k) Number (if known): K 993352
Device Name: Additional Crossfire™ Acetabular Components
Indications For Use:
The subject devices are single use components, intended for use in conjunction with associated acetabular shells, femoral bearings, and femoral hip stems as part of a cemented or cementless total hip replacement procedure. Indications for use, in keeping with those of other commercially available Class II total hip devices, are as follows:
- noninflammatory degenerative joint disease including osteoarthritis and avascular necrosis, .
- . rheumatoid arthritis,
- . correction of functional deformity,
- revision procedures where other treatments or devices have failed, and .
- treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur with . head involvement that are unmanageable using other techniques.
The additional Crossfire™ Acetabular Inserts include the following devices:
- System 12® Acetabular Inserts .
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Russell Lozano SUTTO
Division Sign-Off Restorative Devices Division of Gene 79587 510(k) Number.
Prescription Use a (Per 21 CFR 801.109)
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
OR
Over-The-Counter Use
(Optional Format 1-2-96)
{7}------------------------------------------------
#### Wear Claims:
The Howmedica Osteonics Crossfire Polyethylene Acetabular Inserts, 2041C-2850, show a 90% reduction in gravimetric wear rate versus the same acetabular inserts fabricated from standard polyethylene, 2041-2850. These inserts mate with a 50mm acetabular shell, have a 10° elevated rim, a 28mm inner diameter, and a 9,4mm bearing thickness. Testing was performed under multiaxial hip ioint simulation for 5 million cycles, using a 28mm CoCr articulating counterface and a bovine calf serum lubricant. The results of in vitro hip wear simulator tests have not been shown to quantitatively predict clinical wear performance.
The Howmedica Osteonics Crossfire Polyethylene Acetabular Inserts, 2041C-2850, show an 88% reduction in gravimetric wear rate versus the same acetabular inserts fabricated from standard polyethylene, 2041-2850, when evaluated following an accelerated aging cycle. These inserts mate with a 50mm acetabular shell, have a 10° elevated rim, a 28mm inner diameter, a 9.4mm bearing thickness, and were aged under 80°C in air for 14 days. Testing was performed under multiaxial hip joint simulation for 5 million cycles, using a 28mm CoCr articulating counterface and a bovine calf serum lubricant. The results of in vitro hip wear simulator tests have not been shown to quantitatively predict clinical wear performance.
The Howmedica Osteonics Crossfire Polyethylene Acetabular Inserts, 2041C-2850, show a 78% reduction in gravimetric wear rate versus the same acetabular inserts fabricated from standard polyethylene, 2041-2850, when evaluated under abrasive wear conditions. These inserts mate with a 50mm acetabular shell, have a 10° elevated rim, a 28mm inner diameter, and a 9.4mm bearing thickness. Testing was performed under multiaxial hip joint simulation for 5 million cycles, using a 28mm CoCr articulating counterface, a bovine calf serum lubricant, and an abrasive media of bone cement particulate. The results of in vitro hip wear simulator tests have not been shown to quantitatively predict clinical wear performance.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.