← Product Code [LZA](/productcode/LZA) · K993311

# COMFIT POWDER FREE NITRILE EXAMINATION GLOVE, STERILE (K993311)

_Wrp Asia Pacific Sdn. Bhd. · LZA · Nov 19, 1999 · General Hospital · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K993311

## Device Facts

- **Applicant:** Wrp Asia Pacific Sdn. Bhd.
- **Product Code:** [LZA](/productcode/LZA.md)
- **Decision Date:** Nov 19, 1999
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.6250
- **Device Class:** Class 1
- **Review Panel:** General Hospital

## Indications for Use

The gloves are intended to be worn on the hand of healthcare personnel and similar personnel to prevent contamination between the healthcare or similar personnel and the patient's body, fluids, waste, or environment.

## Device Story

Sterile, powder-free nitrile examination glove; intended for use by healthcare personnel to prevent cross-contamination between examiner and patient/environment. Device acts as a physical barrier. Used in clinical settings. Provides protection against fluids and waste. Benefits patient and clinician by reducing risk of contamination.

## Clinical Evidence

Bench testing only. Performance data includes 1000 ml watertight test (passing G-I, AQL 2.5 and G-II, AQL 4.0), dimensions (length, palm width), wall thickness, and physical properties (tensile strength and elongation before/after aging). Residual powder content measured at <2mg/glove.

## Technological Characteristics

Nitrile latex material; powder-free; sterile; single-wall thickness (finger 0.21mm, palm 0.14mm); tensile strength (before aging 23.88 MPa, after aging 20.48 MPa). Complies with ASTM D 3578-99 standards.

## Regulatory Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of WRP Asia Pacific Sdn Bhd, formerly known as Wembley Rubber Products. The image also contains the number K993311 and the date NOV 19 1999. The logo consists of the letters WRP in a black oval shape. The text "WRP Asia Pacific Sdn Bhd" is written in a bold, sans-serif font.

# ATTACHMENT 3 (Revised)

Lot 1, Jalan 3, Kawasan Perusahaan Bandar Baru Salak Tinggi, 43900 Sepang, Selangor Darul Ehsan, MALAYSIA

TEL +60-3-846 1486 FAX +60-3-846 1485/1557 EML mktgwrp@ibm.net URL www.wrpworld.com

CONTACT PERSON : Y. W. CHOW

## 510(k) SUMMARY

- COMFIT POWDER FREE NITRILE EXAMINATION 1. Trade Name : GLOVE, STERILE
- Examination Gloves 2. Common Name :
- Patient Examination Glove 3. Classification Name :

#### Substantial Equivalence : 4.

Class I nitrile latex patient examination's glove, 80 LZA, powder free. It meets ASTM Standard D 3578-99 for all properties with the exception of ultimate elongation before the aging test.

## 5. Intended Use of Device :

The gloves are intended to be worn on the hand of healthcare personnel and similar personnel to prevent contamination between the healthcare or similar personnel and the patient's body, fluids, waste, or environment.

Page 1 of 3

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the logo for WRP Asia Pacific Sdn Bhd. The logo includes the letters "WRP" in bold, white font inside of a black oval. Below the company name is the text "FORMERLY KNOWN AS WEMBLEY RUBBER PRODUCTS (M) SDN BHD" and the number "1 4 7 8 1 7 V".

## 6. Summary of Performance Data :

Performance data of gloves to ASTM D 3578-99 standard and FDA 1000 ml watertight test.

| TEST                             | ASTM D 3578-99         | COMFIT Powder Free Nitrile Examination Glove, Sterile - refer to Attachment 8 of Device Test Report of Compliance    |
|----------------------------------|------------------------|----------------------------------------------------------------------------------------------------------------------|
| 1. Watertight (1000 ml)          | G-I, AQL 2.5           | Pass based on<br>1) Single Sampling Plan, G-I, AQL 2.5,<br>2) Multiple Sampling Plan, G-II, AQL 4.0                  |
| 2. Length (mm)                   |                        |                                                                                                                      |
| Size                             |                        |                                                                                                                      |
| XS                               | min 220                | -                                                                                                                    |
| S                                | min 220                | 248                                                                                                                  |
| M                                | min 230                | 250                                                                                                                  |
| L                                | min 230                | 260                                                                                                                  |
| XL                               | -                      | -                                                                                                                    |
| 3. Palm Width (mm)               |                        |                                                                                                                      |
| Size                             |                        |                                                                                                                      |
| XS                               | $70 \pm 10$            | -                                                                                                                    |
| S                                | $80 \pm 10$            | 84                                                                                                                   |
| M                                | $95 \pm 10$            | 95                                                                                                                   |
| L                                | $111 \pm 10$           | 104                                                                                                                  |
| XL                               | -                      | -                                                                                                                    |
| 4. Single Wall Thickness<br>(mm) |                        |                                                                                                                      |
| Finger                           | min 0.08               | 0.21                                                                                                                 |
| Palm                             | min 0.08               | 0.14                                                                                                                 |
| TEST                             | ASTM D 3578-99         | COMFIT Powder Free Nitrile Examination Glove, Sterile -<br>refer to Attachment 8 of Device Test Report of Compliance |
| 5. Physical Properties           |                        |                                                                                                                      |
| Before Aging :                   |                        |                                                                                                                      |
| Tensile Strength (MPa)           | min 14.0               | 23.88                                                                                                                |
| Ultimate Elongation (%)          | min 700                | 669                                                                                                                  |
| After Aging :                    |                        |                                                                                                                      |
| Tensile Strength (MPa)           | min 14.0               | 20.48                                                                                                                |
| Ultimate Elongation (%)          | min 500                | 668                                                                                                                  |
| 6. Residual Powder               | Less than<br>4mg/glove | Less than 2mg/glove                                                                                                  |

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for WRP Asia Pacific Sdn Bhd. The logo consists of the letters "WRP" in a bold, sans-serif font, with the words "WRP Asia Pacific Sdn Bhd" to the right of the letters. Below the company name, it says "FORMERLY KNOWN AS WEMBLEY RUBBER PRODUCTS (M) SDN BHD" in a smaller font.

## 7. Substantial Equivalence based on Assessment of Non-Clinical Performance Data

The performance test data of device as shown above indicate that these nitrile gloves meet requirements of ASTM D 3578-99 Standard Specification for Rubber Examination Gloves with the exception of ultimate elongation before the aging test.

## 8. Conclusion

Based on the test results, these nitrile gloves meet the requirement of FDA's 1000 ml Watertight Test and ASTM D 3578-99 Standard Specification for tensile strength. Our inhouse specifications for tensile strength is minimum 15.5 MPa and test results exceed the in-house specifications and ASTM D 3578-99 requirement for tensile strength.

Date Summary Prepared : September 22, 1999 (Amended on October 27, 1999)

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the circumference. Inside the circle is a stylized symbol that resembles a human figure or a flame, composed of three curved lines that overlap and flow together.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

NOV 1 9 1999

Mr. Yue Wah Chow Head of RA/QA WRP Asia Pacific Sdn. Bhd. Lot 1, Jalan 3, Kawasan Perusahaan Bandar Baru Salak Tinggi 43900 Sepang Selangor Darul Ehsan, Malaysia

Re: K993311 Powder-Free Nitrile Examination Glove, Trade Name: Sterile Regulatory Class: I Product Code: LZA September 29, 1999 Dated: October 4, 1999 Received:

Dear Mr. Chow:

We have reviewed your Section 510(k) notification of intent to we have referenced above and we have determined the market the acreatially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in

{4}------------------------------------------------

Page 2 - Mr. Chow

the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4692. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to
premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fdaf.gov/cdrh/dsmamain.html".

Sincerely yours,

Timothy A. Ulatowski

Timothy A. Ulatowsk Director Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radioloqical Health

Enclosure

{5}------------------------------------------------

## ATTACHMENT 2

Applicant : WRP Asia Pacific Sdn. Bhd.

K993311 510(k) Number (if known) :

#### COMFIT POWDER FREE NITRILE EXAMINATION Device Name : GLOVE, STERILE

## Indications For Use :

- The patient examination glove is a disposable device intended for medical 1. purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109)

Over-The-Counter Use OR

Olin S. Lin

َ ivision Sign-Off) Colsion of Dental, Infection Control, · General Hospital D ಿ(k) Number _

---

**Source:** [https://fda-staging.innolitics.com/device/K993311](https://fda-staging.innolitics.com/device/K993311)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).