← Product Code [LKJ](/productcode/LKJ) · K993294

# DIAMEDIX IS-ANA ELISA SCREEN TEST SYSTEM (K993294)

_Diamedix Corp. · LKJ · Oct 25, 1999 · Immunology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K993294

## Device Facts

- **Applicant:** Diamedix Corp.
- **Product Code:** [LKJ](/productcode/LKJ.md)
- **Decision Date:** Oct 25, 1999
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.5100
- **Device Class:** Class 2
- **Review Panel:** Immunology

## Indications for Use

The Diamedix Immunosimplicity (Is) ANA ELISA ScreenTest Kit is a qualitative enzyme immunoassay intended to screen for the presence of antinuclear antibodies (ANAs) in human serum as an aid in the diagnosis of certain systemic rheumatic diseases. This assay collectively detects in one well, total ANAs against double-stranded DNA (dsDNA and nDNA), Histones, SSA, SSB, Sm, Sm/RNP, Scl-70, Jo-1 and centromeric antigens along with sera positive for Immunofluorescent (IF) HEp-2 ANAs. These reagents can be used either manually or in conjunction with the MAGO Plus Automated EIA Processor.

## Device Story

The Is-ANA ELISA Screen Test System is an enzyme-linked immunosorbent assay (ELISA) for detecting antinuclear antibodies (ANAs) in human serum. The device utilizes microtiter wells coated with antigens extracted from the HEp-2 nucleus (dsDNA, Histones, SSA, SSB, Sm, Sm/RNP, Scl-70, Jo-1, and centromeric antigens). Patient serum is added to the wells; if ANAs are present, they bind to the immobilized antigens. After washing, an enzyme-labeled anti-human immunoglobulin conjugate is added, followed by a substrate solution. The resulting color intensity, measured photometrically, is proportional to the concentration of specific antibodies in the sample. The test is performed either manually or using the MAGO Plus Automated EIA Processor in a clinical laboratory setting. Results assist clinicians in the diagnosis of autoimmune disorders by identifying the presence of ANAs, which may then be further evaluated with specific tests.

## Clinical Evidence

Clinical evaluation included 331 characterized sera samples (normal donors, monospecific sera, IFA-ANA positive sera, and autoimmune disease sera). Performance was assessed manually and via the MAGO Plus processor. Relative sensitivity ranged from 95.9% to 97.9% and relative specificity from 93.8% to 95.7% compared to other commercial ELISA methods. Clinical sensitivity for autoimmune disease sera was 98.5%-100%. Precision studies (intra-assay and inter-assay) showed CVs generally below 15%. Bench testing confirmed performance equivalence between manual and automated workflows.

## Technological Characteristics

Enzyme-linked immunosorbent assay (ELISA) using microtiter wells coated with HEp-2 nuclear antigens. Photometric detection of colorimetric reaction. Compatible with manual processing or the MAGO Plus Automated EIA Processor. In vitro diagnostic device.

## Regulatory Identification

An antinuclear antibody immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoimmune antibodies in serum, other body fluids, and tissues that react with cellular nuclear constituents (molecules present in the nucleus of a cell, such as ribonucleic acid, deoxyribonucleic acid, or nuclear proteins). The measurements aid in the diagnosis of systemic lupus erythematosus (a multisystem autoimmune disease in which antibodies attack the victim's own tissues), hepatitis (a liver disease), rheumatoid arthritis, Sjögren's syndrome (arthritis with inflammation of the eye, eyelid, and salivary glands), and systemic sclerosis (chronic hardening and shrinking of many body tissues).

## Predicate Devices

- Diamedix Immunosimplicity ANA (Is-ANA) Screen Test

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

OCT 2 5 1999

# Appendix E. 510(k) Summary of Safety and Effectiveness

## The following section is included as required by the Safe Medical Device Act (SMDA) of 1990.

Name: Address:

Diamedix Corporation 2140 N. Miami Avenue Miami, FL 33127

Contact Person: Dr. Lynne Stirling Phone Number: 305-324-2354 Fax Number: 305-324-2388

000197

{1}------------------------------------------------

# 510(k) Summary of Safety and Effectiveness

This summary of 510(k) safety and effectiveness information is being submitted in accordance I ins summary of SMDA 1990 and 21 CFR 807.92.

The assigned 510(k) number is: _ 1993299

#### Applicant Information:

| Date Prepared: | September 29, 1999   |
|----------------|----------------------|
| Name:          | Diamedix Corporation |
| Address:       | 2140 N. Miami Avenue |
|                | Miami, FL 33127      |

| Contact Person: | Dr. Lynne Stirling |
|-----------------|--------------------|
| Phone Number:   | 305-324-2354       |
| Fax Number:     | 305-324-2388       |

#### Device Information:

| Trade Name:          | Is-ANA ELISA Screen Test System                                             |
|----------------------|-----------------------------------------------------------------------------|
| Common Name:         | ANA Screening Test                                                          |
| Classification Name: | Antinuclear Antibody Immunological Test System (866.5100), product code LLL |

### Equivalent Device:

Diamedix Immunosimplicity ANA (Is-ANA) Screen Test

Device Description: The Is-ANA ELISA ScreenTest System is an enzyme-linked immuno-Device Description: The 15 Privated of anti-nuclear antibodies (ANA) in human serum.

Intended Use: The assay is intended for use in detecting ANAs in human serum. The assay collectively detects in one well total ANAs against dsDNA, Histones, SSA, SSB, Sm, Sm/ contectively delects in one won tour in the agentive for IFA Hep-2 ANAs. The results of the assay can be used as an aid in the diagnosis of certain autoimmune disorders.

#### Principle of the Procedure:

The Is-ANA ELISA Screen Test System is an enzyme-linked immunosorbent assay to detect THE IS-ANA ELISTE SCIEON For Dysterns and other antigens extracted from the HEP-2 nucleus ANAS in human solid phase microtiter well. Diluted test sera are added to each well. If antiare attached to a sona prisos missens are present in the patient sample they will bind to the oodies winch recognize the unegels and wells are washed to remove unbound antibody. An enzyme labeled anti-human immunoglobulin (conjugate) is added to each test well. If antibody enzyme labeled anti-haman minutegrooming on the After incubation, the wells are washed to is present the only mo mates answer solution is then added to each well. If enzyme is present from prior step, the reaction is stopped and the color intensity is measured photometriprosons from pror essy are of the specific antibodies present in the patient sample.

000198

{2}------------------------------------------------

# Performance Characteristics

## A. Comparison Testing: Relative Sensitivity and Specificity

The Diamedix Is-ANA ELISA Screen Test Kit was evaluated relative to another commercially available ANA Screen test. One hundred and forty-six sera from normal blood donors and one hundred and eighty-five sera from clinical patients were tested by the Is-ANA Screen ELISA Test Kit and the comparative method. Testing by both methods was performed both manually and using the MAGO PLUS Automated EIA Processor. The results shown in Table 1 are the comparison of the Is-ANA ELISA Screen performed manually compared to the comparative method, both manual and automated. The results shown in Table 2 are the comparison of the the comparative method, ocal the MAGO Plus compared to the comparative method, both manual and automated. TABLE 1

| Is-ANA ELISA Screen<br>Manual | Other ELISA : Manual |       |           | Other ELISA :  MAGO Plus |        |           |
|-------------------------------|----------------------|-------|-----------|--------------------------|--------|-----------|
| # of Sera                     | %                    | 95%CI | # of Sera | %                        | 95% CI |           |
| Relative Sensitivity          | 188/196              | 95.9  | 92.1-98.2 | 185/191                  | 96.9   | 93.3-98.8 |
| Relative Specificity          | 107/112              | 95.5  | 89.9-98.5 | 110/115                  | 95.7   | 90.1-98.6 |
| Overall Agreement             | 295/308*             | 95.8  | 92.7-97.7 | 295/306**                | 96.4   | 93.7-98.2 |

* Twenty-two samples equivocal in either or both methods were excluded from calculations; ** Twenty-one samples, equivocal in either or both methods were excluded from calculations.

| TABL<br>ﺎ<br>- 5 | 2 |
|------------------|---|
|------------------|---|

| Is-ANA ELISA Screen<br>MAGO Plus | Other ELISA : Manual |      |           | Other ELISA : MAGO Plus |      |           |
|----------------------------------|----------------------|------|-----------|-------------------------|------|-----------|
|                                  | # of Sera            | %    | 95%CI     | # of Sera               | %    | 95% CI    |
| Relative Sensitivity             | 183/188              | 97.3 | 93.9-99.1 | 183/187                 | 97.9 | 94.6-99.4 |
| Relative Specificity             | 105/112              | 93.8 | 87.5-97.5 | 109/116                 | 94.0 | 88.0-97.5 |
| Overall Agreement                | 288/300*             | 96.0 | 93.1-97.9 | 292/303**               | 96.4 | 93.6-98.2 |

* Twenty-two samples equivocal in either or both methods were excluded from calculations; ** Twenty-five samples, equivocal in either or both methods were excluded from calculations.

## B. Clinical Sensitivity and Specificity using Characterized Sera

A total of three hundred and thirty-one characterized sera were assayed using the Is-ANA ELISA Screen test Kit. These consisted of a number of sera of known ANA reactivity and a number of samples with no known apparent ANA reactivity. Samples tested were as follows:

#### a, NORMAL BLOOD DONOR SERA

One hundred and forty-six samples from normal blood donors were tested in the Is-ANA ELISA Screen. Results are shown in Table 3.

#### b. MONOSPECIFIC SERA

Forty-five sera obtained from a variety of sources and shown to contain monospecific antibodies of clinical significance were tested in the Is-ANA ELISA Screeen Test Kit both manually and using the MAGO Plus Automated Processor. The results are summarized in Table 3.

{3}------------------------------------------------

#### b. IFA-ANA POSITIVE SERA

Seventy sera shown to be positive by the IFA- ANA method were tested in the Is-ANA ELISA Screen Test Kit. These seventy sera consisted of thirty-five sera with IFA-ANA titers between 1:40 and 1:320 and thirty-five sera with titlers in excess of 1: 320. The results obtained are summarized in Table 3.

#### d. AUTOIMMUNE DISEASE STATE SERA

Seventy sera from a variety of sources from patients diagnosed with an autoimmune disorder were tested both manually and on the MAGO Plus using the Is-ANA ELISA Screen. Results are shown in Table 3.

| TABLE 3                                                |          |            |            |             |
|--------------------------------------------------------|----------|------------|------------|-------------|
| Patient group                                          | Positive | Negative   | Equivocal* | Total       |
| a. Normal Sera<br>Manual<br>MAGO Plus                  | 14<br>14 | 119<br>116 | 13<br>16   | 146<br>146  |
| b. Monospecific<br>Sera<br>Manual &<br>MAGO Plus       | 45<br>44 | 0<br>0     | 0<br>0     | 45<br>44**  |
| c. IFA-ANA<br>Low Titer Sera<br>Manual<br>MAGO Plus    | 30<br>34 | 4<br>0     | 1<br>1     | 35<br>35    |
| High Titer Sera<br>Manual &<br>MAGO Plus               | 35<br>35 | 0<br>0     | 0<br>0     | 35<br>35    |
| d. Autoimmune<br>Disease Sera<br>Manual &<br>MAGO Plus | 70<br>64 | 0<br>1     | 0<br>0     | 70<br>65*** |

* Equivocal results were excluded from calculations ** one sample QNS for MAGO Plus *** Five samples QNS for MAGO Plus

| Clinical Specificity    | Manual          | 95% CI     | MAGO Plus       | 95% CI     |
|-------------------------|-----------------|------------|-----------------|------------|
| Normals                 | 119/133 = 89.5% | 84.3-94.7  | 116/130 = 89.2% | 83.9-94.6  |
| Clinical Sensitivity    |                 |            |                 |            |
| Low Titer ANA+ Sera     | 30/34 = 88.2%   | 72.6-96.7  | 34/34 = 100.0%  | 89.7-100.0 |
| High Titer ANA+ Sera    | 35/35 = 100.0%  | 90.0-100.0 | 35/35 = 100.0%  | 90.0-100.0 |
| Monospecific Sera       | 45/45 = 100.0%  | 92.1-100.0 | 44/44 = 100.0%  | 92.0-100.0 |
| Autoimmune Disease Sera | 70/70 - 100.0%  | 94.9-100.0 | 64/65 = 98.5%   | 91.7-100.0 |

#### C. Precision

The precision of the Is-ANA ELISA Screen Test Kit was determined by testing six different sera (2 negative and 4 positive) plus the kit calibrator and controls in two different runs on three different days. Precision was evaluated manally and using the MAGO PLUS Processor. The introl and internas on the manual procedure is shown in Table 4 and for the MAGO Plus Automated EIA Processor in Table S.

{4}------------------------------------------------

| SERUM   | INTRA-ASSAY DAY 1 |       |      | INTRA-ASSAY DAY 2 |       |      | INTRA-ASSAY DAY 3 |       |      | INTERASSAY |       |      |
|---------|-------------------|-------|------|-------------------|-------|------|-------------------|-------|------|------------|-------|------|
|         | Mean Index        | SD    | CV%  | Mean Index        | SD    | CV%  | Mean Index        | SD    | CV%  | Mean Index | SD    | CV%  |
| A (NEG) | 0.262             | 0.007 | 2.7  | 0.223             | 0.024 | 10.8 | 0.280             | 0.019 | 6.8  | 0.255      | 0.030 | 11.6 |
| B (NEG) | 0.292             | 0.009 | 3.1  | 0.220             | 0.049 | 22.3 | 0.247             | 0.019 | 7.7  | 0.253      | 0.042 | 16.7 |
| C (POS) | 1.677             | 0.221 | 13.2 | 1.731             | 0.148 | 8.5  | 1.682             | 0.030 | 1.8  | 1.697      | 0.147 | 8.7  |
| D (POS) | 1.696             | 0.098 | 5.8  | 1.521             | 0.093 | 6.1  | 1.586             | 0.050 | 3.2  | 1.601      | 0.108 | 6.7  |
| E (POS) | 3.310             | 0.192 | 5.8  | 3.232             | 0.338 | 10.5 | 3.179             | 0.114 | 3.6  | 3.305      | 0.261 | 7.9  |
| F (POS) | 3.482             | 0.299 | 8.6  | 3.365             | 0.294 | 8.7  | 2.963             | 0.132 | 4.5  | 3.270      | 0.330 | 10.1 |
| c/o CAL | 1.022             | 0.068 | 6.7  | 0.992             | 0.110 | 11.1 | 0.999             | 0.121 | 12.1 | 1.004      | 0.097 | 9.6  |
| POS     | 2.938             | 0.176 | 6.0  | 3.022             | 0.256 | 8.5  | 2.766             | 0.122 | 4.4  | 2.909      | 0.211 | 7.3  |
| NEG     | 0.247             | 0.039 | 15.8 | 0.210             | 0.036 | 17.1 | 0.264             | 0.030 | 11.4 | 0.240      | 0.041 | 16.9 |

TABLE 4: Manual Precision

TABLE 5 : MAGO Plus Precision

| SERUM   | INTRA-ASSAY DAY 1 |       |     | INTRA-ASSAY DAY 2 |       |      | INTRA-ASSAY DAY 3 |       |      | INTERASSAY |       |      |
|---------|-------------------|-------|-----|-------------------|-------|------|-------------------|-------|------|------------|-------|------|
|         | Mean Index        | SD    | CV% | Mean Index        | SD    | CV%  | Mean Index        | SD    | CV%  | Mean Index | SD    | CV%  |
| A (NEG) | 0.377             | 0.028 | 7.4 | 0.407             | 0.052 | 12.8 | 0.395             | 0.068 | 17.2 | 0.39       | 0.051 | 12.9 |
| B (NEG) | 0.322             | 0.008 | 2.5 | 0.320             | 0.021 | 6.6  | 0.322             | 0.025 | 7.8  | 0.32       | 0.018 | 5.6  |
| C (POS) | 2.402             | 0.059 | 2.5 | 2.540             | 0.239 | 9.4  | 2.625             | 0.100 | 3.8  | 2.52       | 0.173 | 6.8  |
| D (POS) | 2.215             | 0.106 | 4.8 | 2.218             | 0.120 | 5.4  | 2.437             | 0.158 | 6.5  | 2.29       | 0.163 | 7.1  |
| E (POS) | 4.382             | 0.171 | 3.9 | 4.528             | 0.183 | 4.0  | 4.835             | 0.305 | 6.3  | 4.58       | 0.291 | 6.4  |
| F (POS) | 5.385             | 0.071 | 1.3 | 5.470             | 0.137 | 2.5  | 5.903             | 0.246 | 4.2  | 5.59       | 0.281 | 5.0  |
| c/o CAL | 0.968             | 0.053 | 5.5 | 1.025             | 0.078 | 7.6  | 1.113             | 0.055 | 4.9  | 1.04       | 0.085 | 8.2  |
| POS     | 4.407             | 0.238 | 5.4 | 4.310             | 0.270 | 6.3  | 5.050             | 0.167 | 3.3  | 4.59       | 0.401 | 8.7  |
| NEG     | 0.350             | 0.031 | 8.9 | 0.343             | 0.035 | 10.2 | 0.388             | 0.030 | 7.7  | 0.36       | 0.040 | 11.1 |

## D. Correlation of Manual and MAGO PLUS Results

The Is-ANA ELISA Screen Test Kit has been developed for automated as well as manual use. To demonstrate the equivalence of the manual and MAGO Plus procedures, the results of three hundred and twenty-five serum samples tested by both methods were plotted. A scattergram and regression line of the results obtained with 95% confidence intervals is shown in FIGURE 1. The data indicate a good correlation with a correlation coefficient (r) of 0.9712.

Image /page/4/Figure/6 description: This image is a scatter plot that shows the relationship between MAGO Plus Index Values and Manual Index Values. The x-axis represents the Manual Index Values, ranging from 0 to 10, while the y-axis represents the MAGO Plus Index Values, ranging from -2 to 16. The plot shows a positive correlation between the two variables, as the MAGO Plus Index Values tend to increase with the Manual Index Values. There is a line of best fit through the data.

Image /page/4/Figure/7 description: The image is titled "FIGURE 1. Manual vs MAGO PLUS Correlation". The title is written in a bold, sans-serif font. The title is centered horizontally and vertically in the image. The image is a simple title for a figure.

000201

{5}------------------------------------------------

# Expected Values

The expected value for a normal patient is a negative result. However, positive ANA results may be found in apparently healthy individuals. In a recent study 12.4% of sera from normal healthy donors gave a detectable ANA result . Patient sera containing autoantibodies to those antigens represented in the Is-ANA ELISA Screen Test Kit will give positive results which can be further evaluated in specific tests. The number of positive samples detected is dependent upon the populations being tested.

The expected values in a normal S. Florida blood donor population were evaluated by assaying one hundred and forty-six sera both manually and using the MAGO Plus Automated EIA Processor. Figures 2 and 4 show the distribution of results in this normal population. For manual and MAGO PLUS testing 9.6% (14/146) gave positive results. Thirteen samples (8.9%) gave equivocal results manually and sixteen samples (11.4%) gave equivocal results on the MAGO I'lus. The remainder of the samples gave negative results. Of the positive samples, three were found to contain specific autoantibodies and an additional two samples gave weakly positive IFA-ANA results.

In the present studies one-hundred and eighty-five clinical sera obtained from patients with an autoimmune disease or with a known autoantibody reactivity were also evaluated in the Is-ANA ELISA Screen Test Kit. (for MAGO Plus testing one hundred and seventy-nine were available, six being QNS). Figures 3 and 5 show the distribution of results for this positive population. For manual testing 97.2% (180/185) of samples were positive. For MAGO Plus testing 98.9% (177/179) were positive.

Image /page/5/Figure/4 description: The image contains four plots comparing expected values for normal and clinical samples using manual and MAGO Plus methods. Figure 2 shows expected values for normal samples using the manual method, with a few outliers at the end. Figure 3 shows expected values for clinical samples using the manual method, with a more gradual increase in values. Figures 4 and 5 show the same comparison using MAGO Plus, with Figure 4 showing normal samples and Figure 5 showing clinical samples.

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, symbolizing the department's mission to protect the health of all Americans.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

OCT 2 5 1999

Dr. Lynne Stirling Vice President, Regulatory Affairs DiaMcdix Corporation 2140 North Miami Avenue Miami, Florida 33127

Re: K993294

> Trade Name: Diamedix Is-ANA ELISA Screen Test System Regulatory Class: II Product Code: LKJ Dated: September 29, 1999 Received: October 1, 1999

Dear Dr. Stirling:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{7}------------------------------------------------

#### Page 2

Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Steven Sutman

Steven I. Gutman, M.D, M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{8}------------------------------------------------

Page **_ of _**

10(k) **Number (if known):** K9932394

(Per 21 CFR 801.109)

## DEVICE NAME : Is-ANA ELISA Screen Test System

Indications for Use : The Diamedix Immunosimplicity (Is) ANA ELISA ScreenTest Kit is a qualitative enzyme immunoassay intended to screen for the presence of antinuclear antibodies (ANAs) in human serum as an aid in the diagnosis of certain systemic rheumatic diseases. This assay collectively detects in one well, total ANAs against double-stranded DNA (dsDNA and nDNA), Histones, SSA, SSB, Sm, Sm/RNP, Scl-70, Jo-1 and centromeric antigens along with sera positive for Immunofluorescent (IF) HEp-2 ANAs. These reagents can be used either manually or in conjunction with the MAGO Plus Automated EIA Processor.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

| (Division Sign-Off)                     |         |
|-----------------------------------------|---------|
| Division of Clinical Laboratory Devices |         |
| 510(k) Number                           | K983294 |

| Prescription Use | <div> <img alt="Checked" src="checkbox_checked.png"/> </div> | OR | Over-The-Counter Use |
|------------------|--------------------------------------------------------------|----|----------------------|
|------------------|--------------------------------------------------------------|----|----------------------|

(Optional Format 1-2-96)

---

**Source:** [https://fda-staging.innolitics.com/device/K993294](https://fda-staging.innolitics.com/device/K993294)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com