SAFETYTIP MULTI-SAMPLE NEEDLE, SAFETYTIP BLOOD COLLECTION SYSTEM
K993125 · Safety Medical Supply Int,., Inc. · FMI · Jun 20, 2000 · General Hospital
Device Facts
Record ID
K993125
Device Name
SAFETYTIP MULTI-SAMPLE NEEDLE, SAFETYTIP BLOOD COLLECTION SYSTEM
Applicant
Safety Medical Supply Int,., Inc.
Product Code
FMI · General Hospital
Decision Date
Jun 20, 2000
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5570
Device Class
Class 2
Indications for Use
The SafetyTip™ Multi-Sample Needle is a sterile, multisample single-use device for blood collection. The needle is designed with an attached safety sheath, which can be activated to sheath the needle immediately after venipuncture to provide protection from accidental needlesticks. The SafetyTip™ Multi-Sample Needle can be used with any tube holder to perform blood collection (phlebotomy) After use, immediately dispose of the Multi-Sample Needle in an approved Bio-Hazard container. The SafetyTip™ Blood Collection System is a sterile, multisample needle, permanently attached to a sterile tube holder. The needle is used to draw blood from a vein and the tube holder holds the test tubes while the blood is drawn into the tubes. After blood collection, the tubes are then placed in the phlebotomy carriers or refrigeration units and the disposable SafetyTip™ Blood collection System is then discarded into an approved Bio-Hazard container.
Device Story
SafetyTip™ Multi-Sample Needle and Blood Collection System; sterile, single-use phlebotomy devices. Components: cannula, hub, safety sheath, rubber sleeve, tube holder. Operation: venipuncture performed; safety sheath manually activated post-use by pressing needle against hard surface; two-stage closure allows temporary re-opening or permanent locking via tabs to prevent accidental needlesticks. Used in clinical settings by healthcare providers. Provides physical barrier between user and contaminated needle post-procedure; reduces risk of sharps injury. Designed for one-handed motion compliance with CAL-OSHA standards.
Clinical Evidence
No clinical data. Substantial equivalence determination based on non-clinical data and design comparison to predicate devices.
Technological Characteristics
Multi-sample needle (18G-22G, 1"-3.5" length) and tube holder system. Features hinged safety sheath with two-stage locking mechanism (temporary/permanent). Materials include medical surgical grade rubber sleeve. Manual operation; no energy source. Sterile, single-use. Mechanical safety feature for sharps protection.
Indications for Use
Indicated for venous blood collection in patients requiring phlebotomy. Used by healthcare professionals to draw blood samples while providing protection against accidental needlesticks via an integrated safety sheath.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
Predicate Devices
Punctur-Guard (Bio-Plexus)
Monoject (Sherwood)
Vacutainer (Becton Dickinson)
Standard Needle Holder (Bio-Plexus)
Blood Needle Holder System (Sage)
Safety-Lok Needle (Becton Dickinson)
Submission Summary (Full Text)
{0}------------------------------------------------
K993125
## JUN 2 0 2000
Image /page/0/Picture/2 description: The image is a black and white logo. The logo consists of a globe on top of three vertical bars. The globe is shaded with a dot pattern. The three vertical bars are solid black and are of equal width.
SAFETY MEDICAL S: PPLY INTERNATIONAL. INC. 135 Beaver Street 4™ Floor Wal Tham, MA 02154 (781) 642-7171 Fax (781) 642-7373
# "510 (k) Summary" as required by section 807.92 ( c )
#### 807.92(a) (1)
Submitter's Name: Yovette Mumford-President of Safety Medical Supply International, Inc. 135 Beaver Street 4th Floor Waltham, MA 02452-8424 (781) 642-7171 Fax (781) 642-7373 Contact Name: Yovette Mumford or Geneva Burchell Date Summary was prepared September 10, 1999
### 807.92 (a) (2)
Trade Name: SafetyTip™ Multi-Sample Needle Common Name: Blood Collection Needle Classification Name: Needle, Hollow Double Lumen, All types;80 FMI Class II Device
## Trade name: SafetyTip™ Blood Collection System
Common Name: Tube Holder Classification Name: Tube Holder JKA Class II Device
#### 807.92(a)(3)
The legally marketed devices to which Safety Medical Supply International, Inc. claims equivalence:
(For the SafetyTip™ Multi-Sample Needle) "Punctur-Guard" Bio-Plexus
"Monoject" Sherwood
Becton Dickinson "Vacutainer"
| (For the SafetyTip™ Blood Collection System) | |
|----------------------------------------------|------------------------------|
| Bio-Plexus | "Standard Needle Holder" |
| Sage | "Blood Needle Holder System" |
| Becton Dickinson | "Safety-Lok Needle" |
{1}------------------------------------------------
## 807.92(a)(4)
## Description of the devices:
The Multi-Sample needle consists of a Cannula, Hub, Safety Sheath, Rubber Sleeve, and lower cover.
The Multi-Sample needle consists of a hypodermic needle in sizes 18 Gauge to 22 gauge with lengths of 1" to 3 ½". It has a hinged cover/sheath that is pulled open by the user by pinching the sides to expose the needle. When closing there are two levels, first. Close the sheath over the needle by pressing the needle down against any hard surface. At this noint the needle may be re-opened by using the normal technique. To use the permanent locking system, close the sheath over the needle by pressing the needle down against any hard surface then, pressing the tip of the sheath with more force until the needle is lodged between the locking tabs, creating a permare rot lock.
The SafetyTip™ Blood Collection System consists of a Cannula, Hub, Safety Sheath, Medical Surgical Grade Rubber Sleeve Holder and The Tube Holder. This is a "safety" device in that by using the permanent-locking feature, It may prevent accidental needlesticks. Its one-handed loosing motion is in compliance with the CAL-OSHA standards. When cloning the Safety Tip™ Blood Collection System there are two levels, first. close the sheath over the needle by pressing the needle down against any hard surface. At this point the needle may be re-opened by using the normal technique. To use the permanent locking system, close the sheath over the needle by pressing the needle down against any hard surface then, pressing the tip of the sheath with more force until the needle is lodged between the locking tabs, creating a permanent lock.
** The Multi Sample Needle and The Blood Collection Unit are designed to be used as "one" product, or used separately.
## 807.92(a)(5)
Intended Use: The SafetyTip™ Multi Sample Needle is a sterile, multisample single-use device for blood collection. The needle is designed with an attached safety sheath, which can be activated to sheath the needle immediately after venipuncture to provide protection from accidental needlesticks. The SafetyTip™ Multi-Sample Needle can be used with any tube holder to perform blood collection (phlebotomy) After use,
{2}------------------------------------------------
immediately dispose of the Multi-Sample Needle in an approved Bio-Hazard container.
## 807.92(a)(5) continued
Intended Use: The SafetyTip™ Blood Collection System is a sterile, multisample needle, permanently attached to a sterile tube holder. The needle is used to draw blood from a vein and the tube holder holds the test tubes while the blood is drawn into the tubes. After blood collection, the tubes are then placed in the phlebotomy carriers or refrigeration units and the disposable SafetyTip™ Blood collection System is then discarded into an approved Bio-Hazard container.
#### 807.92(a)(6)
There is one significant technological characteristic of our device compared to predicate devices in that the SafetyTip™ Multi-Sample Needle and the SafetyTip™ Blood Collection System contain a safety "sheath" that is intended to cover the needle after use and before disposal for healthcare worker protection.
## 807.92(b)(1)
The determination of substantial comparison of the device features, materials, and intended use. Please refer to the substantial equivalence table in this 510(k) submission for more specific details.
#### 807.92(b)(2)
The determination of substantial equivalence is not based on a assessment of clinical performance data. We are planning to conduct clinical trials at medical facilities and institutions as soon as possible, and will submit any relevant information in the form of an addendum if needed. (# 3 of the 510(k) content and format states, "Clinical data is not needed for not t devices cleared by the 510(k) process." )
#### 807.92(b)(3)
The conclusions drawn from the non-clinical data is that the SafetyTip™ Multi-Sample Needle and the SafetyTip™ Blood Collection System are substantially equivalent to the commercially available devices on the market todav.
{3}------------------------------------------------
:
807.92(d) The FDA has made no requests at this point in the process of the application.
1977 - 1997 - 1997 - 1997
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image is a black and white circular logo for the Department of Health & Human Services - USA. The logo features the department's name encircling a symbol. The symbol consists of three stylized human profiles facing right, stacked on top of each other, with flowing lines representing hair or clothing.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUN 2 0 2000
Ms. Geneva Burchell Safety Medical Supply International, Incorporated 69 Hickory Drive, 1st Floor Waltham, Massachusetts 02451-1011
Re : K993125 Trade Name: SafetyTip® Multi-Sample Needle and SafetyTip® Blood Collection System Regulatory Class: II Product Code: FMI Dated: April 5, 2000 Received: April 14, 2000
Dear Ms. Burchell:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.
{5}------------------------------------------------
Page 2 - Ms. Burchell
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4692. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html" a
Sincerely yours,
Timothy A. Ulatowski
Director Division of Dental, Infection Control, and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
510(k) Number if known):993125
Device Name: SafetyTip™ Multi-Sample Needle and SafetyTip™ Blood Collection System
Indications for use:
The SafetyTip™ Multi-Sample Needle and SafetyTing Blood Collection The Sarely In -- Mun-Sample thood collection. The safety sheath helps to System is to be used for vehous blood conceasing the needle before and after use.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of device evaluation (ODE)
Patricio Cucenite
(Division Sign-Off) (Division of Dental, Infection Control, and General Hospital Devices and General Hospida Hospida 2933
(Optional Format 3-10-98)
510(k) Number_
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.