EXACTECH ACUMATCH INTEGRATED HIP SYSTEM A-SERIES POROUS COATED ACETABULAR COMPONENT
K993082 · Exactech, Inc. · LPH · Nov 19, 1999 · Orthopedic
Device Facts
Record ID
K993082
Device Name
EXACTECH ACUMATCH INTEGRATED HIP SYSTEM A-SERIES POROUS COATED ACETABULAR COMPONENT
Applicant
Exactech, Inc.
Product Code
LPH · Orthopedic
Decision Date
Nov 19, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3358
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
All Exactech Hip Systems are indicated for use in skeletally mature individuals undergoing primary surgery for total hip replacement due to osteoarthritis, rheumatoid arthritis, osteonecrosis, ankylosing spondylitis, congenital hip dysplasia, post-traumatic degenerative problems of the hip, and for treatment of proximal femoral fractures where prosthetic replacement is determined by the surgeon as the preferred treatment. Components of Exactech Hip Systems are also potentially indicated for revision of failed previous reconstructions where sufficient bone stock is present and to restore mobility resulting from previous fusion. The AcuMatch A-Series Porous Coated Acetabular Components are indicated for pressfit and cemented applications.
Device Story
AcuMatch A-Series is a two-piece acetabular component for total hip arthroplasty; consists of porous-coated titanium alloy shell and UHMWPE liner. Shells available in screwless, cluster, or multi-hole designs; optional hydroxyapatite (HA) coating. Liner options include various face angles and lateralized configurations. Used in OR by orthopedic surgeons for press-fit or cemented fixation. Instrumentation prepares acetabular cavity; shell impacted into bone. Locking mechanism secures liner to shell. Device restores hip joint function, mobility, and stability in patients with degenerative hip disease or fractures.
Clinical Evidence
Bench testing only. Engineering evaluations compared torsional strength and liner disengagement force against predicate MCS components. Results demonstrated A-Series locking mechanism is stronger and requires greater force for disassembly than MCS. Clearance modeling performed for shell/liner fit. No clinical data provided.
Technological Characteristics
Materials: Titanium alloy (Ti-6Al-4V, ASTM F136-96) shell; UHMWPE (ASTM F648-98) liner; chemically pure titanium beads (ASTM F67-95, grade 2) for porous coating. Optional HA coating (MAF 339). Design: Hemispherical shell, 14 sizes (44-70mm). Locking: Apical hole mechanism with 12-tab anti-rotational indexing. Sterilization: Gamma irradiation (SAL 10-6) per AAMI guidelines.
Indications for Use
Indicated for skeletally mature patients requiring primary total hip replacement or revision of failed reconstructions due to osteoarthritis, rheumatoid arthritis, osteonecrosis, ankylosing spondylitis, congenital hip dysplasia, post-traumatic degeneration, or proximal femoral fractures. Contraindicated in patients with active infection, insufficient bone stock, neuromuscular disorders preventing joint control, or conditions where weight/age/activity level suggest early implant failure.
Regulatory Classification
Identification
A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.
{0}------------------------------------------------
## 510(k) Summary of Safety and Effectiveness
| Trade Name: | Exactech® AcuMatch A-Series<br>Porous Coated Acetabular Component |
|----------------------|-----------------------------------------------------------------------------------------------|
| Common Name: | Total Hip Prosthesis Acetabular Component |
| Classification Name: | Prothesis, Hip, Semi-Constrained, Metal/Polymer,<br>Porous, Uncemented (Acetabular Component) |
### Legally Marketed Devices for Substantial Equivalence Comparison:
The AcuMatch A-Series Acetabular System (referred to as A-Series from this point forward) is made of similar materials and is of a similar design to other legally marketed acetabular components. Most notably, the A-Series is equivalent in materials and design to the Exactech MCS Porous Cup and Exactech MCS/HA Porous Cup:
| Product Code | Model | Manufacturer | 510(k) Number |
|--------------|--------|--------------|---------------|
| LPH | MCS | Exactech | K921114 |
| LPH | MCS/HA | Exactech | K990197 |
Additionally, the A-Series System is of a similar design to other acetabular components on the market:
| Product Code | Model | Manufacturer | 510(k) Number |
|--------------|------------|----------------|---------------|
| LPH | Reflection | Smith & Nephew | K932755 |
| LPH | Omnifit | Osteonics | -------- |
| LPH | Trilogy | Zimmer | K934765 |
{1}------------------------------------------------
### 510(k) Summary of Safety and Effectiveness
#### Description:
The A-Series acetabular component is a two piece device consisting of a porous metal acetabular shell and a mating polyethylene liner. The metal component is machined from titanium alloy (Ti-6A1-4V) conforming to ASTM F136-96. Three layers of chemically pure titanium beads conforming to ASTM F67-95, grade 2, are sintered to the outer surface of the metal shell to produce the porous coating.
There are three basic shell designs, each of which is available in 14 sizes. Each shell type is available with and without a hydroxyapatite (HA) coating on the beaded surface. The HA coating characteristics are defined in FDA Master File MAF 339 (BioCoat, Inc.).
The acetabular shell design and size options are summarized in Table 1 below.
| Metal Acetabular Shell Options | | | |
|----------------------------------------------------------------------------------------------|---------|-----------------|-----------------------|
| Design | # Sizes | O.D. Size Range | HA Coating<br>Option? |
| Screwless<br>(no holes) | 14 | 44 - 70 mm | Yes |
| Cluster*<br>(3 dome holes) | 14 | 44 - 70 mm | Yes |
| Multi-Hole*<br>(4 rim holes, 8 dome<br>holes) | 14 | 44 - 70 mm | Yes |
| *These designs allow for the use of Ti-6Al-4V bone screws to supplement fixation if desired. | | | |
Table 1
The A-Series cups are intended for press-fit applications. Components without the HA coating may also be used in cemented applications. Instrumentation is used to prepare the cavity of the acetabulum for implanting of the shell. The A-Series shells are designed such that an interference condition is created between the prepared acetabulum and the rim diameter of the porous coated shell. Instrumentation is used to impact the shell into the bone for a secure fit. The cluster and multi-hole designs allow for use of Ti-6Al-4V bone screws to supplement fixation if desired.
{2}------------------------------------------------
### 510(k) Summary of Safety and Effectiveness
The A-Series acetabular liner is composed of ultra-high-molecular-weight polyethylene (UHMWPE) as called out in ASTM F648-98. There are five primary liner designs with various sizes having the ability to accept 22, 26, 28, and 32 mm diameter femoral heads. The minimum bearing thickness for each liner type is greater than 6 mm.
| UHMWPE Liner<br>Design | Sizes | I.D. | Characteristics |
|------------------------|---------|----------------|---------------------------------------------|
| 15° Liner | 7 sizes | 22,26 mm | 15° Face Angle |
| | 6 sizes | 28 mm | |
| | 4 sizes | 32 mm | |
| 0° Liner | 7 sizes | 22,26 mm | 0° Face Angle |
| | 6 sizes | 28 mm | |
| | 4 sizes | 32 mm | |
| Extended Coverage | 7 sizes | 22,26 mm | 5° / 20° Face Angle |
| | 6 sizes | 28 mm | |
| | 4 sizes | 32 mm | |
| 15° Lateralized + 4 mm | 7 sizes | 22,26,28,32 mm | Center rotation displaced<br>4mm laterally. |
| 0° Lateralized + 4 mm | 7 sizes | 22,26,28,32 mm | Center rotation displaced<br>4mm laterally. |
The acetabular liner options are summarized in Table 2.
l'able 2
The A-Series has an apical hole locking mechanism and a 12-tab anti-rotational indexing system at the rim of the cup.
### Packaging and Sterilization:
A-Series Acetabular components are double packed in thermoformed trays and sealed with Tyvek® lids. The UHMWPE liners are vacuum sealed in two pouches. The bone screw packaging includes two heat sealed pouches (non-vacuum). Inner packaging units are placed in cardboard containers and sealed with shrink wrap. The A-Series packaging also includes various product identification labels, sterility labels, product tracing labels and sterility indicators.
The AcuMatch A-Series is sterilized by gamma irradiation to a Sterility Assurance Level (SAL) of 106. Exactech utilizes Method 3, Protocol B from the "AAMI Guideline for gamma radiation sterilization" for the sterility dose setting and validation procedures.
### Section 4
{3}------------------------------------------------
## 510(k) Summary of Safety and Effectiveness
#### Intended Use:
All Exactech Hip Systems are indicated for use in skeletally mature individuals undergoing primary surgery for total hip replacement due to osteoarthritis, rheumatoid arthritis, osteonecrosis, ankylosing spondylitis, congenital hip dysplasia, post-traumatic degenerative problems of the hip, and for treatment of proximal femoral fractures where prosthetic replacement is determined by the surgeon as the preferred treatment. Components of Exactech Hip Systems are also potentially indicated for revision of failed previous reconstructions where sufficient bone stock is present and to restore mobility resulting from previous fusion.
The AcuMatch A-Series Porous Coated Acetabular Components are indicated for pressfit and cemented applications.
#### Contraindications:
Exactech Hip Systems are contraindicated in patients with active infection, patients without sufficient bone stock to allow appropriate insertion and fixation of the prosthesis, in neuromuscular disorders that do not allow control of the hip joint, and in patients whose weight, age, or activity level would cause the surgeon to expect early failure of the system.
#### Substantial Equivalence Information:
The Exactech AcuMatch A-Series Porous Coated Acetabular Component has the same intended uses and similar technological features as Exactech's MCS line of acetabular components.
The A-Series and MCS devices have identical material specifications and surface treatments. Both designs consist of a titanium alloy shell (Ti-6A1-4V, ASTM F136-96) shell and a mating ultra-high-molecular-weight-polyethylene (UHMWPE, ASTM F648-98) liner. Both have hemispherical outside geometries and identical surface coatings made of three uniform layers of chemically pure titanium beads. In addition, both devices offer an optional hydroxyapatite (HA) coating for the beaded surface. Detailed surface characterizations, including relevant performance data is found in Exactech's MCS and MCS/HA premarket notifications, #K921114 and #K990197.
{4}------------------------------------------------
### 510(k) Summary of Safety and Effectiveness
There are more liner and shell options with the A-Series line. While the MCS series offers a multi-hole and a peripheral screw design, the A-Series also includes no-hole, multi-hole and cluster hole configurations. In addition to the standard 0° and standard 15° options of the MCS line, the A-Series components include extended coverage, 0° +4mm lateralized and 15° +4mm lateralized liners.
Both the MCS and A-Series have apical hole locking mechanisms. The antirotational indexing system at the rim of the cups differ in that the A-Series has twelve tabs and the MCS is octagonal in configuration. Tests were performed to determine the torsional strength of the proposed A-Series system in comparison to Exactech's MCS cup. The torsional study shows that the locking mechanism on the A-Series is stronger than the same size MCS component. Another engineering evaluation shows that the force required to disengage an A-Series liner from its shell exceeds the force required to disassemble the MCS component. An engineering report modeling the clearance between the spherical radius of the A-Series shell and liner shows a 0.008 inch difference for the least material condition and 0.001 inch difference for the maximum material condition.
In addition to similarities with the Exactech MCS components, the A-Series has a similar geometry, material composition, surface characterization and locking mechanism to other devices legally marketed by other manufacturers. These include but are not limited to the Reflection by Smith and Nephew, the Omnifit by Osteonics, the Trilogy by Zimmer and the Inter-Op by Sulzer. Each of these devices, like the Exactech A-Series, is indicated for press-fit and cemented applications and is supplied sterile. All the predicate devices are made of the same materials as the A-Series, a Ti-6Al-4V alloy shell with a mating UHMWPE liner. These materials have historically proven to be successful in orthopedic implant applications. The Reflection model by Smith and Nephew and the Omnifit model by Osteonics have a titanium beaded porous surface similar to the A-Series. Sulzer's Inter-OP acetabular components have an option for an additional hydroxyapatite (HA) coating as does the Exactech A-Series. All the predicate devices have hemispherical geometries, save the Omnifit which has a dual radius design. The minimum thickness of the polyethylene liner for the A-Series is 6.3 mm compared to a 3.9mm minimum thickness for the MCS. The other predicate devices have minimum liner thicknesses ranging from 5.0 to 7.6 mm.
Lever out torque analysis of the new Exactech A-Series component places the strength of the locking mechanism in the range of other legally marketed devices.
{5}------------------------------------------------
Image /page/5/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular arrangement of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA". Inside the circle is a stylized symbol that resembles an abstract caduceus, with three lines forming a shape similar to a bird or a stylized human figure.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Nov 1 9 1999
Ms. Lisa Simpson Regulatory Representative Exactech, Inc. 2320 NW 66th Court Gainesville, Florida 32653
K993082 Re: Exactech Integrated Hip System Trade Name: Requlatory Class: II Product Codes: LPH and JDI September 14, 1999 Dated: September 15, 1999 Received:
Dear Ms. Simpson:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{6}------------------------------------------------
#### Page 2 - Ms. Lisa Simpson
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to
premarket notification" (21 CFR 807.97). Other gener Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Sincerely yours,
[signature]
James E. Dillard III Acting Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
#### Indications for Use
K99500
510(k) Number:
**Device Name:**
- - -
Exactech® AcuMatch™ Integrated Hip System A-Series Porous Coated Acetabular Component
#### Indications for Use:
All Exactech Hip Systems are indicated for use in skeletally mature individuals undergoing primary surgery for total hip replacement due to osteoarthritis, rheumatoid arthritis, osteonecrosis, ankylosing spondylitis, congenital hip dysplasia, post-traumatic degenerative problems of the hip, and for treatment of proximal femoral fractures where prosthetic replacement is determined by the surgeon as the preferred treatment. Components of Exactech Hip Systems are also potentially indicated for revision of failed previous reconstructions where sufficient bone stock is present and to restore mobility resulting from previous fusion.
The AcuMatch A-Series Porous Coated Acetabular Components are intended to be used in press fit and cemented applications.
#### Contraindications:
Exactech Hip Systems are contraindicated in patients with active infection, patients without sufficient bone stock to allow appropriate insertion and fixation of the prosthesis, in neuromuscular disorders that do not allow control of the hip joint, and in patients whose weight, age, or activity level would cause the surgeon to expect early failure of the system.
Please do not write below this line - use another page if needed.
| Concurrence of CDRH, Office of Device Evaluation (ODE) | |
|--------------------------------------------------------|---------|
| (Division Sign-Off) | |
| Division of General Restorative Devices | |
| 510(k) Number | K993082 |
| Prescription Use | | or | Over the Counter Use | No |
|------------------|--|----|----------------------|----|
|------------------|--|----|----------------------|----|
Section 3
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.