K993069 · General Electric Co. · IYO · Sep 22, 1999 · Radiology
Device Facts
Record ID
K993069
Device Name
GE DIASONICS GATEWAY AND GATEWAY FX
Applicant
General Electric Co.
Product Code
IYO · Radiology
Decision Date
Sep 22, 1999
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1560
Device Class
Class 2
Attributes
Pediatric
Indications for Use
Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: Ophthalmic, Fetal, Abdominal, Intraoperative, Small Organ, Neonatal Cephalic, Adult Cephalic, Cardiac, TR, TV, PV, and Musculo-skeletal Conventional & Superficial.
Device Story
Mobile diagnostic ultrasound console; utilizes transducers to acquire acoustic signals from human tissue; processes signals via digital image processing to generate B-mode, M-mode, Pulsed Doppler, Color Doppler, and Power Doppler images. Harmonic imaging feature enhances visualization of nonlinear tissue characteristics and contrast media. Operated by clinicians in clinical settings; output displayed on integrated monitors for real-time visualization. Assists in evaluation of soft tissue and vascular disease; supports clinical decision-making through diagnostic imaging. Benefits patients by providing non-invasive diagnostic assessment.
Clinical Evidence
No clinical data required. Device evaluated via bench testing for acoustic output, biocompatibility, and thermal, electrical, and mechanical safety. Conformance to 21 CFR 820 and ISO 9001/EN 46001 quality standards verified.
Technological Characteristics
Mobile ultrasound console; keyboard control panel; B&W and color monitors; assorted transducers (e.g., CLA/3.5MI/40, CLA/3.5MI/50-2D). Harmonic imaging implemented via conventional digital image processing. Operates in A, B, M, PWD, CWD, Color Doppler, Amplitude Doppler, and combined modes. Conforms to medical device safety standards for acoustic output, biocompatibility, and electrical/mechanical safety.
Indications for Use
Indicated for diagnostic ultrasound imaging or fluid flow analysis in ophthalmic, fetal, abdominal, intraoperative, small organ (breast, thyroid, testes, penis), neonatal/adult cephalic, cardiac, transrectal, transvaginal, peripheral vascular, and musculoskeletal applications. Prescription use only.
Regulatory Classification
Identification
An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
K993069
Special 510(k) Premarket Notification GE Diasonics Gateway with Harmonic Imaging Modification September 10, 1999
SEP 2 2 1999
## Attachment B:
Summary of Safety and Effectiveness Prepared in accordance with 21 CFR Part 807.92(c).
Image /page/0/Picture/5 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" intertwined within a circle. The letters and the circle are black, while the background is white.
GE Medical Systems
General Electric Company P.O. Box 414, Milwaukee, WI 53201
#### Section a):
- GE Medical Systems 1. Submitter: PO Box 414 Milwaukee, WI 53201
- Allen Schuh, Contact Person: Manager, Safety and Regulatory Engineering Telephone: 414-647-4385; Fax: 414-647-4090
September 10, 1999 Date Prepared:
- GE Diasonics Gateway & Gateway FX Diagnostic Ultrasound with Harmonic Imaging 2. Device Names: Modification. Ultrasonic Pulsed Doppler Imaging System, 21 CFR 892.1550, 90-IYN
3. Marketed Device: Diasonics Gateway & Gateway FX diagnostic ultrasound system, 510(k) Numbers K951998 and K961002, currently in commercial distribution.
4. Device Description: The GE Diasonics Gateway FX with Harmonic Imaging are mobile consoles approximately 69 cm wide, 107 cm deep and 132 cm high equipped with a keyboard control panel, black and white video display monitor, color video display monitor, assorted transducers and optional image storage or hard-copy devices. Harmonic imaging enhances or highlights the imaging of nonlinear tissue characteristics and contrast media.
5. Indications for Use: The GE Diasonics Gateway & Gateway FX with Harmonic Imaging are general purpose ultrasound imaging systems intended for use in the evaluation of soft tissue and vascular disease in the head, neck, chest, abdomen, pelvis, male and female reproductive organs, limbs and pregnant uterus. Specific indications are: Ophthalmic, Fetal, Abdominal, Intraoperative, Small Organ, Neonatal Cephalic, Adult Cephalic, Cardiac, TR, TV, PV, and Musculo-skeletal Conventional & Superficial.
6. Comparison with Predicate Device: The GE Diasonics Gateway FX with Harmonic Imaging are of a comparable type and substantially equivalent to the currently marketed Gateway FX. They have the same technological characteristics, use the same design, construction and materials, are comparable in key safety and effectiveness features, and have the same intended uses, operating modes and transducers as the predicate device.
### Section b):
1. Non-clinical Tests: The device has been evaluated for acoustic output, biocompatibility, and thermal, electrical and mechanical safety, and has been found to conform with applicable medical device safety standards. Harmonic imaging is implemented with conventional digital image processing technology.
2. Clinical Tests: None required.
3. Conclusion: Intended uses and other key features are consistent with traditional clinical practice, FDA guidelines, and established methods of patient examination. The design and development process of the manufacturer conforms with 21 CFR 820 and ISO 9001/EN 46001 quality system standards. The product is designed to conform with applicable medical device safety standards and compliance is verified through independent evaluation with ongoing factory surveillance. Diagnostic ultrasound has accumulated a long history of safe and effective performance. Therefore, it is the opinion of GE Medical Systems that the GE Diasonics Gateway & Gateway FX with harmonic imaging is substantially equivalent with respect to safety and effectiveness to devices currently cleared for market.
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image is a black and white seal for the Department of Health & Human Services - USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of an eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
SEP 2 2 1999
Mr. Allen Schuh Manager, GE Ultrasound Safety and Regulatory Engineering GE Medical Systems General Electric Company P.O. Box 414 Milwaukee, WI 53201
Re: K993069
GE Diagnostics Gateway and Gateway FX Diagnostic Ultrasound System Regulatory Class: II (TWO) Product Code: 90-IYO and 90-ITX 21 CFR 892.1560 21 CFR 892.1570 Dated: September 10, 1999 Received: September 13, 1999
Dear Mr. Schuh:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranion.
This determination of substantial equivalence applies to the following transducers intended for use with the GE Diagnostics Gateway and Gateway FX Diagnostic Ultrasound System, as described in your premarket notification:
> Transducer Model Numbers: CLA/3.5MI/40 CLA/3.5MI/50-2D
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval) it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic QS inspections, the FDA will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification does not affect any obligation you may have under sections 531 and 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{2}------------------------------------------------
Page 2 - Mr. Allen Schuh
This determination of substantial equivalence is granted on the condition that prior to shipping the first device, you submit a postclearance special report. This report should contain complete information, including acoustic output measurements based on production line devices, requested in Appendix G, (enclosed) of the Center's September 30, 1997 "Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers." If the special report is incomplete or contains unacceptable values (e.g., acoustic output greater than approved levels), then the 510(k) clearance may not apply to the production units which as a result may be considered adulterated or misbranded.
The special report should reference the manufacturer's 510(k) number. It should be clearly and prominently marked "ADD-TO-FILE" and should be submitted in duplicate to:
> Food and Drug Administration Center for Devices and Radiological Health Document Mail Center (HFZ-401) 9200 Corporate Boulevard Rockville, Maryland 20850
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4591. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
If you have any questions regarding the content of this letter, please contact Robert A. Phillips, Ph.D. at (301) 594-1212.
Sincerely yours,
Kinnel C. Sgmm
CAPT Daniel C. Schultz, M.J.
CAPT Daniel G. Schultz, M.D. Acting Director, Division of Reproductive. Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure(s)
{3}------------------------------------------------
#### Diagnostic Ultrasound Indications for Use Form
## GE Diasonics Gateway and Gateway FX Systems
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | |
|-------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) |
| Ophthalmic | P | P | P | P | | P | P | | P | |
| Fetal | P | P | P | P | | P | P | | P | |
| Abdominal | P | P | P | P | | P | P | | P | |
| Intraoperative (specify) | P | P | P | P | | P | P | | P | |
| Intraoperative Neurological | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ (specify) | P | P | P | P | | P | P | | P | |
| Neonatal Cephalic | P | P | P | P | | P | P | | P | |
| Adult Cephalic | P | P | P | P | | P | P | | P | |
| Cardiac | P | P | P | P | P | P | P | | P | |
| Transesophageal | | | | | | | | | | |
| Transrectal | P | P | P | P | | P | P | | P | |
| Transvaginal | P | P | P | P | | P | P | | P | |
| Transuretheral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular | P | P | P | P | P | P | P | | P | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal Conventional | P | P | P | P | | P | P | | P | |
| Musculo-skeletal Superficial | P | P | P | P | | P | P | | P | |
| Other (specify) | N | | | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
-
Additional Comments: Cardiac is Adult and Pediatric. Small Organ includes: breast, thyroid, testes and penis. Combined modes include B or M and Pulsed, Color or Power Doppler. Clearance via K951998.
Intraoperative (K961002) includes abdominal organs.
Harmonic Imaging (other feature) applies to CLA 3.5/MI-50-2D probes in B mode only.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Timmills Flynn
Division Sign-Off
Division Sign-Off) Division of Reproductive, Abdominal, EN and Radiological D 510(k) Number
Prescription User (Per 21 CFR 801.109)
{4}------------------------------------------------
### Diagnostic Ultrasound Indications for Use Form
## GE Diasonics Gateway and Gateway FX with CLA/3.5MI/40 Probe
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | |
|-------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | P | P | P | | P | P | | P | |
| Abdominal | | P | P | P | | P | P | | P | |
| Intraoperative (specify) | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ (specify) | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transuretheral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal Conventional | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | |
| Other (specify) | | N | | | | | | | | |
N= new indication; P= previously cleared by FDA via K951998; E= added under Appendix E
Additional Comments: Combined modes include B or M and Pulsed, Color or Power Doppler.
Harmonic Imaging (other feature) applies to B mode only.
(PLEASE DO NOT WRITE BELOW THIS LINE · CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
David A. Hegemann
(Division Sign-Off) Division of Reproductive, Abdominal, ENT and Radiological D 510(k) Number
Prescription User (Per 21 CFR 801.109)
{5}------------------------------------------------
Special 510(k) Premarket Notification GE Diasonics Gateway with Harmonic Imaging Modification September 10, 1999
## Diagnostic Ultrasound Indications for Use Form
# GE Diasonics Gateway FX with CLA/3.5Ml/50-2D Probe
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | |
|-------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|----------------------|--------------------|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify | Other<br>(specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | E | E | E | | E | E | | E | |
| Abdominal | | E | E | E | | E | E | | E | |
| Intraoperative (specify) | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ (specify) | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transuretheral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal Conventional | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | |
| Other (specify) | | N | | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E (24 OCT 1996)
Additional Comments: Combined modes include B or M and Pulsed, Color or Power Doppler.
Harmonic Imaging (other feature) applies to B mode only.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Arind C. Ingram
Division Sign Off
(Division Sign-Off) Division of Reproductive, Abdominal, EN and Radiological D 510(k) Number
Prescription User (Per 21 CFR 801.109)
E-4
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.