K992985 · Heraeus Kulzer, Inc. · KLE · Oct 14, 1999 · Dental
Device Facts
Record ID
K992985
Device Name
GLUMA COMFORT BOND
Applicant
Heraeus Kulzer, Inc.
Product Code
KLE · Dental
Decision Date
Oct 14, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3200
Device Class
Class 2
Indications for Use
GLUMA® Comfort Bond is a light curing single component bonding agent use in restorative adhesive dentistry specifically developed for bonding resin-based filling materials (e.g. composites, compomers, Polyglas®) to hard dental tissues. Other indications include bonding of amalgam and laboratory-produced restorations. GLUMA® Comfort Bond permits priming and bonding to be carried out in single step.
Device Story
Gluma Comfort Bond is a one-bottle, light-cured dental adhesive system. It functions as a priming and bonding agent for restorative dentistry. The device is applied by a clinician to conditioned, moist tooth surfaces using consecutive strokes, followed by air-drying and light curing for 20 seconds. It utilizes a chemical composition of HEMA for wetting/penetration, 4-META for bonding strength to composites and non-precious metal alloys (e.g., CoCr-based casting alloys, amalgam), and urethane di-methacrylate for cross-linked polymer network formation. Ethanol serves as a carrier and water-chaser. The device is used in a clinical setting to improve the retention of restorative materials to dental hard tissue.
Clinical Evidence
Bench testing only. Shear bond strength testing on dentin (25.1 ± 3.5 MPa) and enamel (30.8 ± 6.1 MPa). Gap-free sealing performance evaluated in laboratory settings. No clinical human trials presented.
Technological Characteristics
One-bottle, light-cured dental adhesive. Composition: HEMA (wetting/penetration), 4-META (bonding to composites/alloys), urethane di-methacrylate (cross-linking), and ethanol (solvent/carrier). Application: manual application to conditioned tooth surface, air-dry, 20-second light cure. Class II device.
Indications for Use
Indicated for use in restorative adhesive dentistry to bond resin-based filling materials (composites, compomers, Polyglas®) to hard dental tissues, and for bonding amalgam and laboratory-produced restorations.
Regulatory Classification
Identification
A resin tooth bonding agent is a device material, such as methylmethacrylate, intended to be painted on the interior of a prepared cavity of a tooth to improve retention of a restoration, such as a filling.
Predicate Devices
Bond -1 (Jeneric/Penetron Incorporated)
ALL Bond 2 (BISCO)
3M Single Bond Dental Adhesive (3M)
One Step Dental Adhesive
Reference Devices
Prime & Bond 2.1
Submission Summary (Full Text)
{0}------------------------------------------------
K992985
OCT 1 4 1999
## 510(K) Summary
Date: August 19, 1999
- Address: Heraeus Kulzer, Inc. 4315 S. Lafayette Blvd.
- Contact: Cheryl V. Zimmerman 219-299-6662
- Gluma® Comfort Bond Device:
## Substantially Equivalent to:
Bond -1 (Jeneric/Penetron Incorporated, Wallingford, CT) ALL Bond 2 (BISCO) 3M Single Bond Dental Adhesive (3M) One Step Dental Adhesive
## Device Description:
Gluma One Bond, available since October, 1998, has been proven to be an excellent, onebottle bonding system. In independent testing, Gluma One Bond had shear bond strength scores of 25.4 MPa to dentin and 35.1 MPa to enamel. In addition, all six fillings in the test were gap-free. The REALITY evaluation organization gave Gluma One Bond four stars, making it a REALITY'S Choice, calling it a "highly effective adhesive," due to its "proven ingredients." Their tests "found its bond strengths are quite comparable to Prime & Bond 2.1."2
Of particular interest in Gluma One Bond is 4-META. 4-META is a well known and proven chemical substance used in dentistry for many years. 4-META, besides adding to the strength of the bond to composites, also allows bonding to amalgam and other nonprecious metal alloys.
{1}------------------------------------------------
The acetone in the Gluma One Bond has been replaced with ethanol and the product name changed to Gluma Comfort Bond . Gluma Comfort Bonda is a one bottle adhesive, designed to bond resinous restarative matirlas to dental hard tissue. This product will provide the strength of 4-META and the ability to do a traditional 2-3 second drying after etching.
The composition of Gluma Comfort Bond is based on three monomers which individually have been used in marketed dental products for decades. The hydrophilic hydroxyethylmethacrylate (HEMA) is essential for perfect wetting of and penetration into the conditioned tooth structure. Similarly, 4-MET(A) has hydrophilic and moieties for wetting and polymerization, whereas the urethane di-methacrylate monomer is responsible for formation of a cross-linked polymer network. The ethanol has the function of a carrier for the monomers while being an effective water chaser at the same time.
An early shear bond strength test of Gluma Comfort Bond indicates scores of 25.1 ± 3.5 MPa to dentin and 30.8 ± 6.1 MPa to enamel. In addition, five of the six fillings in the test were gap-free.
The application procedure is very easy. The solution is applied in small amounts to the conditioned moist tooth surface with two to three consecutive strokes, the water and ethanol is eliminated by a gentle air blast and finally, the resin is light cured for 20 seconds. GLUMA COMFORT BOND is also proven to bond to non-precious dental alloys, such as CoCr-based casting alloy and amalgam.
ﺍﻟﺨﻤﺮ ﺍﻟ
Testing method described in "Laboratory evaluation of one-component enamel/dentin bonding agents," Finger W.J., Fritz U., Am J Dent 1996; 9:206-210
<sup>2</sup> REALITY. Volume 13. Houston, TX: REALITY Publishing; 1999
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized representation of a human figure in profile, with three overlapping faces suggesting a sense of community or interconnectedness. The figure is positioned within a circular border, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the perimeter of the circle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
OCT 1 4 1999
Ms. Cheryl V. Zimmerman Manager, Quality Operations & Regulatory Affairs Heraeus Kulzer, Inc. 4315 South Lafayette Boulevard South Bend, Indiana 46614-2517
K992985 Re : Gluma® Comfort Bond Trade Name: Requlatory Class: II Product Code: KLE Dated: September 2, 1999 Received: September 3, 1999
Dear Ms. Zimmerman:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. ਕ substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) requlation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any
{3}------------------------------------------------
Page 2 - Ms. Zimmerman
obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4692. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Patrux Cincentiffer
Timothy A. Ulatowski Director Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Page 1 of
510(k) Number (if Known): 长99298S
Device Name: Gluma® Comfort Bond
Indications For Use:
GLUMA® Comfort Bond is a light curing single component bonding agent use in restorative adhesive dentistry specifically developed for bonding resin-based filling materials (e.g. composites, compomers, Polyglas®) to hard dental tissues. Other indications include bonding of amalgam and laboratory-produced restorations. GLUMA® Comfort Bond permits priming and bonding to be carried out in single step.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device evaluation (ODE)
| Prescription Use<br>(Per 21 CFR 801.109) | <div style="display:inline-block; vertical-align:middle;">✓</div> |
|------------------------------------------|-------------------------------------------------------------------|
|------------------------------------------|-------------------------------------------------------------------|
OR
| Over-The-Counter Use | |
|----------------------|--|
|----------------------|--|
| | (Optional Format 1-2-96) |
|--|--------------------------|
|--|--------------------------|
| (Division Sign-Off) | <div style="text-decoration:overline;">Susan Reang</div> |
|------------------------------------------------------------------------|----------------------------------------------------------|
| Division of Dental, Infection Control,<br>and General Hospital Devices | |
| 510(k) Number | <div style="text-decoration:overline;">Icandars</div> |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.