← Product Code [JIX](/productcode/JIX) · K992845

# HICHEM CALIBRATOR 1 KIT, HICHEM CALIBRATOR 2 KIT, HICHEM CALIBRATOR 3 KIT, CALIBRATION STANDARDS KIT (K992845)

_Elan Holdings, Inc. · JIX · Oct 18, 1999 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K992845

## Device Facts

- **Applicant:** Elan Holdings, Inc.
- **Product Code:** [JIX](/productcode/JIX.md)
- **Decision Date:** Oct 18, 1999
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.1150
- **Device Class:** Class 2
- **Review Panel:** Clinical Chemistry

## Indications for Use

HiChem® Calibrators 1, 2 and 3 are intended to calibrate the Beckman® SYNCHRON CX® DELTA and CX® CE® Systems for the quantitative determination of sodium, potassium, chloride, urea, giucose, creatinine, calcium and total CO2. HiChem® Calibrators 1, 2 and 3 for the SYNCHRON CX® DELTA and CX® CE® Systems are intended to establish points of reference that are used in the determination of values in the measurement of substances in specimens. These calibrators are intended for professional use only.

## Device Story

HiChem Calibrators 1, 2, and 3 serve as reference standards for Beckman SYNCHRON CX DELTA and CX CE clinical chemistry systems. Used in professional clinical laboratory settings, these calibrators establish reference points for quantitative analysis of analytes (sodium, potassium, chloride, urea, glucose, creatinine, calcium, total CO2) in patient serum, plasma, urine, and CSF. By providing known values to the analyzer, the device enables accurate calibration of the system's measurement process. This ensures the reliability of subsequent patient sample testing, supporting healthcare providers in clinical decision-making regarding patient metabolic and electrolyte status.

## Clinical Evidence

Bench testing only. Method comparison studies performed on the SYNCHRON CX DELTA System using serum, plasma, urine, and CSF samples. Comparison against Beckman calibrators yielded linear regression results with r-values ranging from 0.935 to 1.000 across all analytes (Calcium, Chloride, Potassium, Sodium, Total CO2, BUN, Glucose, Creatinine). Sample sizes ranged from 44 to 160 per analyte/specimen type.

## Technological Characteristics

In vitro diagnostic calibrator kits. Designed for use with Beckman SYNCHRON CX DELTA and CX CE systems. Provides reference points for quantitative clinical chemistry assays. No specific materials of construction or software algorithms described; device functions as a chemical reference standard.

## Regulatory Identification

A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)

## Special Controls

*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

## Predicate Devices

- SYNCHRON® CX® Calibrators 1, 2 & 3 (product nos. 465908, 465909 and 465910)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K99 28415

élan diagnostics

Image /page/0/Picture/2 description: The image is a black and white logo for "élan". The logo features a stylized letter "e" formed by multiple parallel lines that curve around to create the shape. To the left of the "e", there are several short, horizontal lines that resemble a stylized wing or speed lines. Below the graphic element, the word "élan" is written in a bold, sans-serif font, with an acute accent over the "e".

Summary of 510(k) Safety and Effectiveness Information

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

HiChem® Calibrators 1, 2 and 3 are intended to calibrate the Beckman® SYNCHRON CX® DELTA and CX® CE® Systems for the quantitative determination of sodium, chloride, urea, glucose, creatinine, calcium and total CO2. The HiChem® Calibrator 1 Kit, Calibrator 2 Kit and Calibrator 3 Kit are substantially equivalent to the SYNCHRON® CX® Calibrators 1, 2 & 3, product nos. 465908, 465909 and 465910, manufactured by Beckman Coulter, Inc.

The effectiveness of the HiChem® Calibrators 1, 2 and 3 is shown by the following method comparison studies.

## Method Comparison

Serum, plasma, urine and CSF were collected from adult patients and assayed using HiChem® and Beckman® calibrators, reagents and wash solutions on a SYNCHRON CX® DELTA System. Results were compared by least squares linear regression and the following statistics were obtained.

| Analyte    | Specimen     | Unit   | Regression Statistics |      |       |       | Summary Statistics |        |        |
|------------|--------------|--------|-----------------------|------|-------|-------|--------------------|--------|--------|
|            |              |        | n                     | a    | b     | r     | range              | mean X | mean Y |
| Calcium    | Serum/Plasma | mg/dL  | 160                   | 0.0  | 0.989 | 0.985 | 7.1 - 10.6         | 9.26   | 9.13   |
|            | Urine        | mg/dL  | 74                    | -0.2 | 1.007 | 0.998 | 2.4 - 15.2         | 8.45   | 8.3    |
| Chloride   | Serum/Plasma | mmol/L | 160                   | 1.0  | 0.988 | 0.935 | 98.2 - 127.5       | 107.3  | 107.0  |
|            | Urine        | mmol/L | 78                    | -5.1 | 1.049 | 0.999 | 22.4 - 289         | 126.8  | 127.9  |
|            | CSF          | mmol/L | 44                    | -3.4 | 1.024 | 0.985 | 113.8 - 152.4      | 126.5  | 126.1  |
| Potassium  | Serum/Plasma | meq/L  | 160                   | 0.13 | 0.969 | 1.000 | 3.20 - 10.82       | 5.02   | 5.00   |
|            | Urine        | meq/L  | 80                    | 0.01 | 0.993 | 1.000 | 3.48 - 136.0       | 50.5   | 50.2   |
| Sodium     | Serum/Plasma | meq/L  | 160                   | 9.1  | 0.930 | 0.938 | 131.8 - 159.1      | 141.0  | 140.3  |
|            | Urine        | meq/L  | 78                    | -0.3 | 1.000 | 1.000 | 16.9 - 288.1       | 118.2  | 117.8  |
| Total CO2  | Serum/Plasma | mmol/L | 160                   | 1.2  | 0.949 | 0.953 | 9.5 - 29.1         | 23.3   | 23.2   |
| BUN        | Serum/Plasma | mgN/dL | 160                   | -0.3 | 0.987 | 0.999 | 4 - 126            | 19.0   | 18.4   |
|            | Urine        | mgN/dL | 79                    | 0.9  | 0.979 | 1.000 | 6 - 142            | 76.2   | 75.4   |
| Glucose    | Serum/Plasma | mg/dL  | 159                   | -1.2 | 1.011 | 0.999 | 29 - 341           | 100.7  | 100.6  |
|            | Urine        | mg/dL  | 81                    | -2.1 | 1.012 | 0.999 | 1 - 359            | 197.3  | 197.4  |
|            | CSF          | mg/dL  | 45                    | 1.0  | 0.973 | 0.999 | 3 - 186            | 81.9   | 80.7   |
| Creatinine | Serum/Plasma | mg/dL  | 160                   | 0.0  | 0.991 | 0.997 | 0.4 - 5.4          | 1.09   | 1.08   |
|            | Urine        | mg/dL  | 79                    | -0.3 | 1.000 | 1.000 | 12.1 - 309.2       | 112.5  | 112.3  |

Wynn Stocking

Wynn Stocking Manager of Regulatory Affairs Elan Diagnostics

20 August, 1999

510(k) Notification, HiChem® Calibrators 1, 2 & 3, page 81 is a division of Elan Pharmaceuticals

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with three lines representing its wings.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

OCT 1 8 1999

Mr. Wynn Stocking Manager, Regulatory Affairs Elan Diagnostics 231 North Puente Street Brea. California 92821

Re: K992845

> Trade Name: HiChem® Calibrator 1 Kit, Calibrator 2 Kit and Calibrator 3 Kit Regulatory Class: II Product Code: JIX Dated: August 20, 1999 Received: August 24, 1999

Dear Mr. Stocking:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{2}------------------------------------------------

## Page 2

Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Steven Putman

Steven I. Gutman, M.D. M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

510(k) Number (if known):

K992845

Device Names:

HiChem® Calibrator 1 Kit, HiChem® Calibrator 2 Kit, and HiChem® Calibrator 3 Kit

Indications for Use:

HiChem® Calibrators 1, 2 and 3 are intended to calibrate the Beckman® SYNCHRON CX® DELTA and CX® CE® Systems for the quantitative determination of sodium, potassium, chloride, urea, giucose, creatinine, calcium and total CO2.

HiChem® Calibrators 1, 2 and 3 for the SYNCHRON CX® DELTA and CX® CE® Systems are intended to establish points of reference that are used in the determination of values in the measurement of substances in specimens.

These calibrators are intended for professional use only.

Respectfully,

Wynn Stocking

Wynn Stocking Regulatory Affairs Manager Elan Diagnostics

20 August, 1999

Sean Cooper

(Division Si
Division of Ciracal I shoratory Levices
510(k) Numb K992845

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use
(Per 21 CFR 801.109)

OR

Over-The-Counter Use

(Optional Format 1-2-96)

510(k) Notification, HiChem® Calibrators 1, 2 & 3 Elan Diagnostics, Brea, California

20 August, 1999 page 85 of 85

---

**Source:** [https://fda-staging.innolitics.com/device/K992845](https://fda-staging.innolitics.com/device/K992845)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
