NUVASIVE SPINAL ARTHROSCOPE, MODELS 0, 30 AND 70 DEGREES
K992782 · Nuvasive, Inc. · HRX · Oct 5, 1999 · Orthopedic
Device Facts
Record ID
K992782
Device Name
NUVASIVE SPINAL ARTHROSCOPE, MODELS 0, 30 AND 70 DEGREES
Applicant
Nuvasive, Inc.
Product Code
HRX · Orthopedic
Decision Date
Oct 5, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.1100
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The NuVasive Spinal Arthroscope, consisting of a rigid diagnostic arthroscope with outer sheath, is intended to achieve percutaneous vizualization of, and/or to assist in performing percutaneous surgical procedures on, the spinal nerve root, foramina, intervertebral disc, and the surrounding tissues of the spine via uniportal or biportal posterior or posterolateral approach, where anatomic restrictions permit percutaneous access. The device is intended for use in conjunction with the NuVasive Guided Spinal Arthroscopy System under real-time radiographic visualization via image-intensified C-arm fluoroscopy, but may also be employed independent of that system where it is compatible in diameter and length with other commercially available arthroscopic instruments, and with surgical need.
Device Story
Rigid diagnostic arthroscope; 0°, 30°, and 70° viewing angles; 400mm working length; 105° field of view. Employs rod lens imaging system and fiberoptic light guide for illumination. Includes outer sheath with 5.0mm diameter and luer lock stopcock for irrigation. Used in clinical settings by surgeons for percutaneous spinal procedures; typically used with real-time C-arm fluoroscopy. Connects to standard video systems and light sources via included adaptors. Provides direct visualization of spinal anatomy to assist surgical navigation and intervention.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Rigid arthroscope; rod lens imaging; fiberoptic illumination; 5.0mm outer diameter sheath; luer lock irrigation port; reusable; non-sterile supply. Compatible with standard video systems and light sources.
Indications for Use
Indicated for percutaneous visualization and assistance in surgical procedures on spinal nerve roots, foramina, intervertebral discs, and surrounding spinal tissues via posterior or posterolateral approach in patients where anatomic restrictions permit percutaneous access.
Regulatory Classification
Identification
An arthroscope is an electrically powered endoscope intended to make visible the interior of a joint. The arthroscope and accessories also is intended to perform surgery within a joint.
Predicate Devices
Arthroscope (Henke-Sass, Wolf, GmbH)
PercScope™ (Clarus Medical)
Yeung Endoscopic Spine System (Richard Wolf)
Panoview Arthroscope (Richard Wolf)
Spine Endoscope (Endius)
Angled Neuroendoscope (AESCULAP®)
Neurological Endoscope (Richard Wolf)
Olympus Neuro Endoscope (Olympus Optical)
INCLUSIVETM Endoscopic System (Sofamor Danek)
Submission Summary (Full Text)
{0}------------------------------------------------
### 510(k) Summary VII.
K992782
In accordance with the Safe Medical Devices Act (SMDA) of 1990 and Title 21 of the Code of Federal Regulations, Part 807), and in particular §807.92, the following summary of information is provided:
## Submitted by: A.
R. Stephen Reitzler Vice President, Regulatory Affairs and Quality Assurance NuVasive. Incorporated 10065 Old Grove Road, Suite A San Diego, California 92131 Telephone: (858) 271-7070 Telefacsimile: (858) 271-7101
### B. Device Name
Trade or Proprietary Name: NuVasive Spinal Arthroscope
Common or Usual Name: Arthroscope
Classification Name: Arthroscope
### C. Predicate Devices
The subject device is substantially equivalent, in whole or in part, to one or more of the following predicate devices, among others: .
Arthroscope (Henke-Sass, Wolf, GmbH) PercScope™ (Clarus Medical) Yeung Endoscopic Spine System (Richard Wolf) Panoview Arthroscope (Richard Wolf) Spine Endoscope (Endius) Angled Neuroendoscope (AESCULAP®) Neurological Endoscope (Richard Wolf) Olympus Neuro Endoscope (Olympus Optical) INCLUSIVETM Endoscopic System (Sofamor Danek)
## D. Device Description
The subject Spinal Arthroscope is a rigid diagnostic arthroscope with outer sheath. The diagnostic scope is available in three (3) models, consisting of arthroscopes with 0°, 30°, and 70° viewing angles, each of which has a working length of 400 +0.5 mm. The field of view for each is 105°.
{1}------------------------------------------------
The device employs a rod lens imaging system and a fiberoptic light guide for illumination. The standard eyepiece permits use of the subject device with various commercially-available video systems. While the device is manufactured with an ACMI-type illumination sidearm, it is supplied with both Richard Wolf-, and Storz-type adaptors which permit connection to many commercially-available light sources. The outer sheath has a 5.0 mm outer diameter and is equipped with a standard luer lock stopcock to permit irrigation through the sleeve. The subject Spinal Arthroscope is provided non-sterile, and is reusable.
#### E. Intended Use
The NuVasive Spinal Arthroscope, consisting of a rigid diagnostic arthroscope with an outer sheath, is intended to achieve percutaneous vizualization of, and/or to assist in performing percutaneous surgical procedures on, the spinal nerve root, foramina, intervertebral disc, and surrounding tissues of the spine via biportal posterior or posterolateral approach, where anatomic restrictions permit percutaneous access. The device is intended for use in conjunction with the NuVasive Guided Spinal Arthroscopy System under real-time radiographic visualization via imageintensified C-arm fluoroscopy, but may be employed independent of that system where it is compatible in diameter and length with commercially available arthroscopic instrumentation, and with surgical need.
#### F. Comparison to Predicate Devices
The subject device has indications for use which are substantially equivalent to those of one or more of the predicate devices, is composed of the same or equivalent materials as one or more of the predicate devices, has the same design features as one or more of the subject devices, and has functional characteristics which are the same or equivalent to those of one or more of the predicate devices. Due to the equivalency of indications for use, materials of composition, design features, method of use, and functional characteristics, the device raises no new safety or effectiveness issues.
#### Summary of Non-Clinical Tests G.
(Not applicable.)
## H. Summary of Clinical Tests
(Not applicable.)
## Conclusions of Non-Clinical and Clinical Tests I.
(Not applicable.)
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo is circular in shape, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter of the circle. In the center of the circle is an abstract image of a human figure, with the head and shoulders visible. The figure is facing to the right and appears to be in motion.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
OCT 5 1999
Mr. Steve Reitzler Vice President, Regulatory Affairs and Quality Assurance Nuvasive, Inc. 10065 Old Grove Road, Suite A San Diego, California 92131
Re: K992782
Trade Name: Spinal Arthroscope Regulatory Class: II Product Code: HRX Dated: August 17, 1999 Received: August 19, 1999
Dear Mr. Reitzler:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that. through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{3}------------------------------------------------
# Page 2 - Mr. Steve Reitzler
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Colie M. Witter, Ph.D., M.D.
Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
#### V. Draft Labeling
#### Indications for Use A.
510(k) Number (if known):
Device Name: NuVasive, Inc., Spinal Arthroscope
Indications for Use:
The NuVasive Spinal Arthroscope, consisting of a rigid diagnostic arthroscope with outer sheath, is intended to achieve percutaneous vizualization of, and/or to assist in performing percutaneous surgical procedures on, the spinal nerve root, foramina, intervertebral disc, and the surrounding tissues of the spine via uniportal or biportal posterior or posterolateral approach, where anatomic restrictions permit percutaneous access. The device is intended for use in conjunction with the NuVasive Guided Spinal Arthroscopy System under real-time radiographic visualization via image-intensified C-arm fluoroscopy, but may also be employed independent of that system where it is compatible in diameter and length with other commercially available arthroscopic instruments, and with surgical need.
| | Concurrence of CDRH Office of Device Evaluation (ODE) |
|------------------|-------------------------------------------------------|
| | (Division Sign-Off) |
| | Division of General Restorative Devices |
| 510(k) Number | K992782 |
| Prescription Use | <p>(Per 21 CFR 801.109)</p> |
| | OR |
| | Over-The-Counter Use |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.