CARDIOTRON EKG MULTI-PHASE INFORMATION ANALYSIS SYSTEM, MODELS 3800, 6800, 8800
K992703 · Premier Heart, LLC · DQK · Mar 21, 2000 · Cardiovascular
Device Facts
Record ID
K992703
Device Name
CARDIOTRON EKG MULTI-PHASE INFORMATION ANALYSIS SYSTEM, MODELS 3800, 6800, 8800
Applicant
Premier Heart, LLC
Product Code
DQK · Cardiovascular
Decision Date
Mar 21, 2000
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1425
Device Class
Class 2
Indications for Use
The Cardiotron Multi-Phase EKG Information Analysis System is intended to be used as an aid to diagnosis by means of analysis of the EKG waveform in the frequency domain (power spectral estimate).
Device Story
System collects EKG signals from leads II and V5 for 2-3 minutes; inputs processed via EKG amplifiers and A/D converter. Software transforms time-domain EKG data to frequency-domain using Fast Fourier Transform (FFT) power spectrum estimate equations. System provides ischemia indexes (positive/negative) to assist physician diagnosis. Operates in clinical settings; physician-operated. Remote model allows data transmission via internet to secondary terminal. Patient history (78 questions) stored in database for physician review; not used for automated diagnosis. Benefits include standardized frequency-domain analysis to aid in identifying ischemic conditions.
Clinical Evidence
No new clinical trials presented. Substantial equivalence based on identical algorithmic processing to predicate device. Reported performance metrics: sensitivity approximately 90%, specificity approximately 85%.
Technological Characteristics
System includes EKG amplifiers, A/D converter, computer, monitor, and printer. Signal processing via Fast Fourier Transform (FFT) power spectrum estimate. Connectivity includes local and remote (internet-based) data transmission. Software performs automated mathematical analysis of EKG waveforms. Components are off-the-shelf.
Indications for Use
Indicated for use as an aid to diagnosis by analyzing EKG waveforms in the frequency domain. Patient population includes individuals undergoing EKG assessment; no specific age or gender restrictions stated. Contraindications not specified.
Regulatory Classification
Identification
A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.
Predicate Devices
New York Heart Instrument Inc.'s EKG Multi-Phase Information Diagnosis System, Model FFF-IA
Submission Summary (Full Text)
{0}------------------------------------------------
## 510(k) SUMMARY K992703 for Cardiotron LLP Inc.'s Cardiotron EKG Multi-Phase Information Analysis System
Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared
Premier Heart LLP 14 Vanderventer Avenue, Suite 138 Port Washington, NY 11050 Phone: (516) 883-3838 Facsimile: (516) 883-5812
Contact person: Dr. Joseph Shen
Date Prepared: March 20, 2000
#### Name of Device
Trade Name: Cardiotron EKG Multi-Phase Information Analysis System
Common Names: Computerized EKG Analysis System
Classification Names: Computerized EKG and Programmable Diagnostic Computers
Product Code: DQK
#### Predicate Device
New York Heart Instrument Inc.'s EKG Multi-Phase Information Diagnosis System, Model FFF-IA
#### Intended Use
The Cardiotron EKG Multi-Phase Information Analysis System is intended to be used as an aid to diagnosis by means of analysis of the EKG waveform in the frequency domain (power spectral estimate).
### Technological Characteristics and Substantial Equivalence
The Cardiotron EKG Multi-Phase Information Analysis System will be available in two models, the basic model system and the remote model system. The basic model system consists of two standard EKG amplifiers and an off-the-shelf analog-to-digital ("A/D") converter, EKG lead wires, a computer, a monitor, and a printer. The remote model system additionally includes a dummy terminal or computer. The minor differences between the two devices is discussed in more detail below. The EKG amplifiers are supplied by the Fortune, Co. The additional components, which are off-the-shelf
{1}------------------------------------------------
products, are unmodified. The computer is programmed with the Company's multi-phase information analysis software.
To operate the system, six standard EKG leads are placed on the patient in lead positions II and V5. EKG data is collected from both leads simultaneously for approximately two to three minutes. In addition, the patient is asked, through prompts appearing on the display, approximately 78 questions related to his/her medical history, such as age, sex and smoker/non-smoker. These questions are stored in a database as a matter of convenience and as a record-keeping function and not subsequently used to provide a software-based diagnosis of the patient. For example, the Cardiotron's software has the ability to provide a partial medical history of the patient so that the physician can look up and review the answers to the questions in the software=s database. The data from the EKG leads is amplified and processed, using conventional techniques, in the EKG amplifier. The data is fed from the EKG amplifier to the A/D converter, which converts the data from an analog form to a digital form. The data, which has been converted to digital form, is fed to the computer where it is transformed from the time domain to the frequency domain through the application of Fast Fourier Transform power spectrum estimate equations which have been programmed in the multi-phase analysis software installed in the computer.
The Cardiotron System and the NYHI System both consist of two EKG amplifiers, an A/D Converter, EKG lead wires, a computer programmed with software, a monitor, and a printer. Moreover, both systems use the same algorithms in their software to process the data and provide a diagnosis. The same inventor developed the software for the NYHI System and the Cardiotron System. Both systems also provide the same power spectrum estimate used in the diagnosis. Both systems also contain the same approximately 78 questions that are asked of the patient, and the answers stored in a database. The answers to the questions also are used in the identical manner as recorded in a database for the convenience of the physician. In addition, both systems provide the same ischemia indexes in the form of a positive or negative index. They also have the same sensitivities (i.e., approximately 90%) and specificities (i.e., approximately 85%).
The only difference between the two systems is the selection of components, the actual software code itself, and the Cardiotron's remote use function and associated hardware. For example, the Cardiotron System uses EKG amplifiers supplied by Fortune, Co. whereas the NYHI System's EKG amplifiers are supplied by Minta, Co. This does not raise any new issues of safety or effectiveness because both amplifiers have very similar specifications. The Cardiotron System and NYHI System use different A/D converters, EKG lead wires, computers, printers, and monitors although they provide the same basic functions. These differences also do not raise new issues of safety or effectiveness because the components are all very similar. Where there are significant differences, those differences do not affect safety or effectiveness. For example, the NYHI System does not have a high resolution color monitor whereas the Cardiotron System has a high resolution color monitor. This difference is solely for marketing purposes.
{2}------------------------------------------------
As noted above, the Cardiotron System uses slightly different code. This difference, however does not raise new issues of safety or effectiveness. First, the software code merely automates the processing of mathematical algorithms, and the mathematical algorithms in the software code of both systems are the same. The code differs only in the manner in which data is stored and in the terms used in the code. Second, as described below, the Company has validated the software to demonstrate that the software functions as intended.
Third, the use of a dummy terminal or additional computer to collect and e-mail the data to a remote location does not raise new issues of safety or effectiveness. As described below, the Company has validated the software that performs these functions. Moreover, the transmission of digital data across the internet is a common operation performed by millions of people. The concern that the data may be lost has been addressed by the Cardiotron's software, which saves the data as well as transmits the data. Thus, if the data transmitted to the remote location is for some reason "lost", it can be resent.
In summary, the Cardiotron System and its predicate device, the NYHI System, have the same intended use and very similar, principles of operation, and technological characteristics. Although there are minor difference in the actual components and software code, these differences do not raise new questions of safety or efficacy.
40024448.doc
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle with three swooping lines forming its body and wings. The eagle is positioned to the right of a circular arrangement of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAR 2 1 2000
Mr. William D. Hare Cardiotron, LLC c/o Fish & Richardson, PC 601 13th Street, N.W. Suite 901 South Washington, DC 20005
Re: K992703 Multi-Phase EKG Information Analysis System Regulatory Class: II (two) Product Code: 74 DQK Dated: December 21, 1999 Received: December 22, 1999
Dear Mr. Hare:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to leqally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the qeneral controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}------------------------------------------------
MAR 2 1 2000
#### Page 2 - Mr. William D. Hare
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the requlation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Jeanne H. Weathers fer
James E. Dillard III Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## 510(k) Number (if known): K992703
# Device Name: Cardiotron Multi-Phase EKG Information Analysis System
Indications For Use:
The Cardiotron Multi-Phase EKG Information Analysis System is intended to be used as an aid to diagnosis by means of analysis of the EKG waveform in the frequency domain (power spectral estimate).
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Bvaluation (ODE)
K. Westerlund
(Division Sign-Off)
vision of Cardiovascular, Respiratory, and Neurological Devic 2703 510(k) Number
Prescription Use Per 21 CFR 801.109)
OR
Over-The-Counter Use
(Optional Format 1-2-96)
90428.W11
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.