The i-STAT Celite Activated Clotting Time (ACT) test cartridge is an in vitro diagnostic test used to monitor moderate- and high-level hevarin therapy through analysis of arterial and venous whole blood samples. The cartridge is to be used with the i-STAT System 200 model analyzer. As part of the i-STAT System, the Celite ACT test cartridge is to be used by trained and certified health care professionals in accordance with a facility's policies and procedures.
Device Story
i-STAT Celite ACT test cartridge; single-use; contains gold film sensor on silicon/silicon dioxide substrate. Input: ~40 microliters fresh whole blood. Operation: cartridge inserted into i-STAT Model 200 analyzer; automated heating to 37°C; blood oscillation mixes sample with reagent (Celite activator, thrombin substrate). Principle: thrombin-mediated conversion of synthetic substrate produces electroactive marker; electrochemical detection of marker indicates coagulation endpoint. Output: clot time in seconds. Used in clinical settings by trained healthcare professionals. Output informs heparin therapy management; benefits include rapid, point-of-care monitoring of anticoagulation status.
Clinical Evidence
Clinical studies at three sites compared i-STAT Celite ACT to Hemochron predicate using patient samples from cardiac catheterization and bypass (N=118, 74, 101). Correlation coefficients ranged 0.923-0.949; slopes 0.902-1.00. Additional comparison at two sites for hemodialysis/ECMO patients (N=52, 14) showed mean bias of -24 seconds. Bench testing established linearity (50-1000 seconds), sensitivity (77 seconds/U/ml heparin), and precision (within-sample imprecision 5.6%).
Technological Characteristics
Single-use cartridge; gold film sensor on silicon/silicon dioxide substrate. Reagents: Celite (diatomaceous earth) activator, thrombin substrate, inert matrix. Energy: thermally controlled analyzer (37°C). Connectivity: i-STAT Model 200 portable analyzer. Dimensions: 40 microliter sample volume. Sterilization: not specified.
Indications for Use
Indicated for monitoring heparin therapy in patients with pulmonary embolism or venous thrombosis, and for monitoring anticoagulation during medical procedures including catheterization, cardiac surgery, organ transplant, and dialysis. Contraindicated for patients receiving aprotinin.
Regulatory Classification
Identification
An activated whole blood clotting time tests is a device, used to monitor heparin therapy for the treatment of venous thrombosis or pulmonary embolism by measuring the coagulation time of whole blood.
Predicate Devices
Hemochron® Systems Activated Clotting Time
Submission Summary (Full Text)
{0}------------------------------------------------
## FEB 1 0 2000
Image /page/0/Picture/1 description: The image shows the logo for i-STAT Corporation. The logo is in black and white and features the text "i-STAT" in a bold, sans-serif font. Below the text "i-STAT" is the word "CORPORATION" in a smaller, sans-serif font. There is a line above the text "i-STAT" and below the word "CORPORATION".
# (992571
0000000000
-
### 510(k) Summary
This summary of the 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1900 and CFR 807.92.
The assigned 510(k) number is:
Summary prepared on: July 25, 1999
#### Submitted by:
i-STAT Corporation 104 Windsor Center Dr. E. Windsor, NJ 08520 Phone: 609-443-9300 FAX:609-443-9310
Contact: Paul VanDerWerf, PhD Vice President, Regulatory Affairs and Quality Assurance
#### Establishment Registration Number: 2234478
#### Identification of Device:
Device Name: Activated Clotting Time Test Proprietary/Trade Name: i-STAT® Celite® ACT Common Name: ACT Classification: Activated Clotting Time Tests Device Classification: II Regulation Number: CFR § 864.7140 Panel: Hematology Product Code: JBP
#### Identification of the Predicate Device:
Device Name: Hemochron® Systems Activated Clotting Time
#### Intended Use of the Device:
The i-STAT Celite Activated Clotting Time (ACT) test cartridge is an in vitro diagnostic test used to monitor moderate- and high-level hevarin therapy through analysis of arterial and venous whole blood samples. The cartridge is to be used with the i-STAT System 200 model analyzer.
As part of the i-STAT System, the Celite ACT test cartridge is to be used by trained and certified health care professionals in accordance with a facility's policies and procedures.
#### Description of the Device:
The i-STAT Celite ACT test is contained in a single test cartridge. In use, approximately 40 microliters of fresh whole blood are placed in the cartridge as described below. The cartridge is inserted into the thermally controlled i-STAT Model 200 Portable Clinical Analytical steps are performed automatically. Patient and user information may be entered into the analyzer via a keypad during the automated analysis cycle.
In the i-STAT ACT test the endpoint is indicated by the appearance of an electroactive marker generated by the thrombin-mediated conversion of a synthetic substrate included in the reagent. Detection of the marker indicates generation of thrombin and therefore complete activation of the coagulation cascade. The reported result is calculated from the time and rate of the substrate conversion and is given in seconds. The
104 Windsor Center Drive, East Windsor, NJ 08520 Tel: 609-43-9310 Web: http://www.i-STAT.com #
{1}------------------------------------------------
reported result correlates to the result of a traditional ACT in which the endpoint is indicated by physical clot formation.
The ACT cartridge is assembled from plastic components that provide the conduits for fluid handling and house the sensor chips. The coagulation test is identified to the user through the name and color code on the cartridge label and by the analyzer through features integral to the cartridge.
In the ACT cartridge the sensor comprises a gold film patterned on a silicon/silicon dioxide substrate.
During the test the blood sample is mixed with reagents which are coated on the cartridge cover in a segment of the sensor channel. The reagent layer includes an activating agent, a thrombin substrate, and inert matrix components. These reagents allow activation of the coagulation cascade and detection of clot formation.
Whole blood is introduced into the sample well of the cartridge at the sample port and the cartridge is closed and inserted into the analyzer. Insertion of the cartridge initiates a controlled and monitored sequence of steps in the instrument. These are:
- Electrical contact is made between the analyzer electronic input circuits and the cartridge. The . analyzer identifies the type of cartridge being used and the tests contained in the cartridge.
- The dry chips and sensor channel are heated to 37C. .
- The blood is then moved forward. Feedback from the fluid position sensor is used to allow . controlled oscillation of the blood segment resulting in dissolution of the reagent layer.
- . Following mixing, a count up time is displayed.
- During the course of testing, the position of the blood segment is actively controlled to . maintain the length of the blood containing the reagent coincident with the endpoint detector.
- . Calculation of sample clot time is performed and displayed.
#### Comparison to Technological Features of the Predicate Device:
The following technological differences between the i-STAT and the Hemochron Systems Activated Clotting Time tests are noteworthy.
- Endpoint detection in both the Hemochron and i-STAT Systems relies upon detecting the action of . thrombin, the final enzyme activated in the coagulation cascade, on a substrate within the sample. In the case of the Hemochron, thrombin converts its natural substrate, fibrin which then crosslinks and causes localized or extended clotting throughout the sample. The instrument detects clot formation as the resulting impedance of the magnet's motion through the sample. In the i-STAT test the generated thrombin converts an added substrate to a species which can be detected electrochemically. The signal for the appearance of this marker is used to assign the endpoint time. Because the traditional ACT test is not highly sensitive to fibrinogen, however, this does not cause a method-to-method bias that is a function of the fibrinogen values greater than 100 mg/dl. Below 100 mg/dl the Hemochron result may be prolonged relative to that of the i-STAT.
- . The volume of blood required for the Hemochron and i-STAT tests is significantly different. The Hemochron requires either 2.0 or 0.4 ml of blood depending on the tube type where the i-STAT test requires 0.040 ml. Although such a difference does not necessarily imply systematic differences in the results of test methods, in this case it is a contributor to a between method bias that is a function of the pre-analytical temperature of the blood. The relatively long time constant for sample heating in the Hemochron results in prolongation of results with hypothermic samples.
#### Summary of Non-Clinical Performance in Support of Substantial Equivalence:
- . Studies using ex-vivo heparinized whole blood samples establish that the i-STAT Celite ACT test responds linearly to the heparin concentration across its reportable range of 50 to 1000 seconds. The
{2}------------------------------------------------
average sensitivity across multiple donors is 77 seconds / U/ml heparin. This is equivalent to the sensitivity of the Hemochron System Activated Celite Clotting Time test.
- . From the heparin linearity experiments it can be seen that the within-sample reproducibility of the i-STAT test is 5.6% in samples spanning the reportable range. The Hemochron Celite ACT package insert indicates imprecision of 8.8% in whole blood samples
- . The imprecision of the i-STAT test in plasma controls was established using in-house and user studies. Overall the Level 1 and Level 2 Controls read 221 ± 19 seconds (8.4% C.V.) and 456 ± 22 seconds (4.8% C.V.), respectively. This includes within-lot, lot-to-lot, vial-to-vial, analyzer-to-analyzer, and operator-to-operator components of the imprecision. The total imprecision in whole blood controls reported for the Hemochron Celite ACT is 6.5% and 5.3% for Level 1 and Level 2 Controls, respectively.
#### Summary of Clinical Test Performance in Support of Substantial Equivalence Claims: Studies conducted at three external sites compared the results of the i-STAT Celite ACT (y) to those of the predicate device (x) for samples from cardiac catheterization and bypass procedures. The identical sample was tested on each instrument. The methods were compared using least square regression analysis. The results are summarized in the table below.
| Statistic | Definition | Site 1 | Site 2 | Site 3 |
|-------------|-----------------------------------------------------------------------------------------------|--------|--------|--------|
| N | The number of patient samples used in the<br>comparison | 118 | 74 | 101 |
| Mean | The average of the comparative method result<br>over the sample population | 272 | 396 | 371 |
| Range | The range of comparative method results<br>obtained over the sample population | 887 | 1182 | 1079 |
| Sx | The standard deviation of the comparative<br>method results across the sample population | 180 | 240 | 272 |
| Slope | The least squares linear regression estimate of<br>the slope | 1.00 | 0.951 | 0.902 |
| Intercept | The least squares linear regression estimate of<br>the intercept | 0 | 25 | 24 |
| Correlation | The correlation coefficient calculated from<br>linear regression | 0.949 | 0.923 | 0.949 |
| Syx | The standard error of the estimate of the<br>regression of y on x | 94 | 97 | 55 |
| Syx % | The relative standard error of the estimate | 15.9% | 15.7% | 12.6% |
| Sxx % | Relative within-sample imprecision of the<br>comparative method over the sample<br>population | 9.1% | 9.7% | 6.7% |
| Syy % | Relative within-sample imprecision of the test<br>method over the sample population | 5.6% | 4.4% | 4.7% |
A comparison of the i-STAT and comparative method was also performed at two sites using samples from patients undergoing hemodialysis and extra-corporeal membrane oxygenation. Because the limited range of the results precludes comparison using linear regression, the average bias of the methods was determined. The results are presented in the table below.
{3}------------------------------------------------
| Parameter | Definition | Site 1 | Site 2 |
|-----------|--------------------------------------------------------------------------------------|--------|---------|
| Number | Number of patient samples used in the comparison | 52 | 14 |
| Mean | The average of the comparative method result over<br>the sample population | 189 | 213 |
| Range | The range of comparative method results obtained<br>over the sample population | 93-270 | 199-233 |
| Mean bias | Average bias between the i-STAT and comparative<br>method | -24 | -24 |
| $s_x$ | Standard deviation of the comparative method<br>results across the sample population | 37 | 10 |
| $s_{xx}$ | Pooled imprecision of the x duplicates | 16 | 10 |
| $s_{yy}$ | Pooled imprecision of the y duplicates | 9 | 6 |
| $s_{yx}$ | The standard error of the estimate of the regression<br>of y on x | 19 | 8 |
#### Limitations:
The i-STAT Celite ACT test is to be used with fresh venous or arterial whole blood samples. The presence of exogenously added heparin, citrate, oxalate, or EDTA will interfere with test results. Poor technique in sample collection may also compromise the results. Samples drawn from insufficiently flushed catheters or from traumatic venipunctures may be contaminated with interfering substances. Samples should be collected into plastic syringes or tubes. Collection into glass may prematurely activate coagulation resulting in accelerated clotting times.
The i-STAT ACT test uses Celite brand diatomaceous earth as the activator of the intrinsic pathway. The result may, therefore, be prolonged in the presence of aprotinin. The test is not recommended for use with patients receiving aprotinin.
Hemodilution, platelet dysfunction, factor deficiencies, and dysprothrombinemias may affect the results of this test.
The i-STAT ACT test is not affected by hematocrit in the range of 20 - 70%, fibrinogen concentration in the range from 100 - 500 mg/dL, or sample temperature from 15 - 37℃.
#### Conclusions:
Based on the non-clinical data the i-STAT Celite ACT test responds linearly to heparin concentration in the range from 50-1000 seconds, is insensitive to sample temperature, hematocrit, and fibrinogen. Studies using plasma controls and whole blood indicate adequate precision for normal and prolonged clot times. Clinical data indicates acceptable correlation to the predicate device.
i-STAT is a registered trademark of i-STAT Corporation, East Windsor, NJ. Celite is a registered trademark of Celite Corporation, Santa Barbara, CA. Hemochron is a registered trademark of International Technidyne Corporation, Edison, NJ.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around the perimeter. Inside the circle is an image of three stylized human profiles facing to the right, with a stylized bird-like shape above them.
## FEB 1 0 2000
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Paul VanDerWerf, Ph.D. Vice President, Regulatory Affairs and Quality Assurance i-STAT Corporation 104 Windsor Center Drive E. Windsor, New Jersey 08520
Re: K992571 Trade Name: i-STAT® Celite® ACT Regulatory Class: II Product Code: JBP Dated: December 21, 1999 Received: December 22, 1999
Dear Dr. VanDerWerf:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{5}------------------------------------------------
Page 2
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Steven Butman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## 3 Indications for use
510(k) Number (if known): _ K99857)_
Device Name: Celite ACT test.
The i-STAT Celite ACT test is useful for monitoring patients receiving heparin for treatment of pulmonary embolism or venous thrombosis, and for monitoring anticoagulation therapy in patients undergoing medical procedures such as catheterization, cardiac surgery, surgery, organ transplant and dialysis.
Peter E. Maklin
(Division Sign-Off)
Division of Clinical Laboratory Devices k992591
(Please do not write below this line-continue on another page if needed.)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use _____________________________________________________________________________________________________________________________________________________________
OR
Over-The-Counter-Use__________________________________________________________________________________________________________________________________________________________
(Per 21 CFR 801.109)
(Optional Format 1-2-96)
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.