VANTEX CENTRAL VENOUS CATHETER WITH OLIGON MATERIAL WITH OR WITHOUT AMC THROMBOSHIELD COATING
K992532 · Baxter Healthcare Corp · FOZ · Jun 13, 2000 · General Hospital
Device Facts
Record ID
K992532
Device Name
VANTEX CENTRAL VENOUS CATHETER WITH OLIGON MATERIAL WITH OR WITHOUT AMC THROMBOSHIELD COATING
Applicant
Baxter Healthcare Corp
Product Code
FOZ · General Hospital
Decision Date
Jun 13, 2000
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5200
Device Class
Class 2
Indications for Use
The Vantex™ Central Venous Catheter with OLIGON™ with or without AMC™ Thromboshield™ coating are indicated for use in patients requiring pressure monitoring, infusion of solutions, and blood sampling in the central vein. All catheters are manufactured with a polyurethane-based OLIGONTM antimicrobial polymer. The OLIGON™ material uses silver as the antimicrobial agent.
Device Story
Triple lumen central venous catheter; constructed from polyurethane-based Oligon™ antimicrobial polymer containing silver; optional AMC™ Thromboshield™ coating. Features soft, flexible distal tip to reduce vessel perforation risk; includes injection caps, removable slide clamps, and suture loops. Used in clinical settings for pressure monitoring, solution infusion, and blood sampling. Clinicians insert catheter into central vein; radiopacity enhanced via barium sulfate for visualization on chest X-ray. Antimicrobial properties provide ≥ 3 log reduction of common pathogens within 48 hours. Device benefits include reduced infection risk and improved catheter patency.
Clinical Evidence
Clinical study conducted in Canada evaluated safety of silver, platinum, and benzalkonium chloride release from the catheter. Serum and urine levels were monitored before, during, and after use; results showed safe levels of all substances. Marketing evaluation in Europe assessed performance and radiopacity, leading to the addition of barium sulfate to the Oligon™ formulation.
Technological Characteristics
Polyurethane-based Oligon™ antimicrobial polymer (silver-based); optional AMC™ Thromboshield™ coating (benzalkonium chloride). Triple lumen configuration. Includes barium sulfate for radiopacity. Sterilized via 100% ethylene oxide. Complies with ISO 10555-1, ISO 10555-3, and ISO 10993-1 standards.
Indications for Use
Indicated for patients requiring pressure monitoring, infusion of solutions, and blood sampling in the central vein.
Regulatory Classification
Identification
An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.
Predicate Devices
Multi-Med Multi-Lumen Central Venous Catheter with AMC™ Thromboshield™
Arrow International's Central Venous Multi-Lumen Catheter with ARROWgard Blue
Submission Summary (Full Text)
{0}------------------------------------------------
| 510(k) Summary, Safety and Effectiveness | 1992532 |
|------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | Edwards Lifesciences, LLC<br>17221 Red Hill Avenue<br>Irvine, California 92614 USA |
| Contact: | Jason Smith<br>Phone: 949-250-2662<br>Fax: 949-756-4021 |
| Device Trade Name: | VantexTM Central Venous Catheter with OligonTM<br>material with or without AMCTM ThromboshieldTM<br>coating |
| Common Name: | Central Venous Catheter |
| Classification: | Class II (Reference 21 CFR 880.5200) |
| Predicate or Legally<br>Marketed Device: | Multi-Med Multi-Lumen Central Venous Catheter<br>with AMCTM ThromboshieldTM and Arrow<br>International's Central Venous Multi-Lumen<br>Catheter with ARROWgard Blue |
| Date prepared: | July 27, 1999 |
#### Device Description
The Vantex™ Central Venous Catheter with Oligon™ material with or without AMC™ Thromboshield™ coating is a triple lumen central venous catheter constructed primarily from Oligon™ material. A soft, flexible tip is formed onto the distal tip of the catheter to reduce the potential for vessel perforation. Each lumen is intended for pressure monitoring, solution infusion, and blood sampling. The Vantex™ Central Venous Catheter with Oligon™ material with or without AMC™ Thromboshield™ coating is provided with injection caps to assist in maintaining lumen sterility and patency, removable slide clamps for each lumen extension, and integral and moveable suture loops for securing the catheter at the insertion site. The device will be packaged in a Barex tray sealed with a Tyvek lid and sterilized using 100% ethylene oxide.
{1}------------------------------------------------
#### Indications for Use:
The Vantex™ Central Venous Catheter with Oligon™ material with or without AMC™ Thromboshield™ is indicated for use in patients requiring pressure monitoring, infusion of solutions, and blood sampling in the central vein. All catheters are manufactured with a polyurethane-based Oligon™ antimicrobial polymer. The Oligon™ material uses silver as the antimicrobial agent.
### Technology Comparison
The Vantex™ Central Venous Catheter with Oligon™ material with or without AMC™ Thromboshield™ is technologically comparable to the predicate devices in construction, materials, and physical specifications. Furthermore, design, manufacturing, and sterilization procedures are representative of current industry practices.
#### Test Summary, In-vitro
The non-clinical testing performed Vantex™ Central Venous Catheter with Oligon™ material with or without AMC™ Thromboshield™ included those set forth in the Draft Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters. The results of the in vitro testing demonstrate that the catheters meet the appropriate requirements of the ISO 10555-1 ("Sterile, single-use intravascular catheters -- Part 1: General requirements") and ISO 10555-3 ("Sterile, single-use intravascular catheters --Part 3: Central venous catheters) standards.
Functional testing was performed on the Vantex™ Central Venous Catheter with Oligon™ material with or without AMC™ Thromboshield™ to evaluate the integrity and performance of the device. Based upon the results of this testing, the Edwards Lifesciences LLC, has determined that the Vantex™ Central Venous Catheter with Oligon™ material with or without AMC™ Thromboshield™ is safe and effective and is acceptable in design and construction for its intended use.
{2}------------------------------------------------
Further in vitro testing was conducted to verify the antimicrobial properties of the Vantex™ Central Venous Catheters with Oligon™ material with or without AMC™ Thromboshield™ coating. The results of this testing demonstrate that the Oligon™ material provides antimicrobial effectiveness (≥ 3 log reduction within 48 hours) against the following organisms commonly associated with catheter related infections: Staphylococcus epidermidis, Staphylococcus aureus, Enterococcus faecalis, Candida albicans, Escherichia coli, Serratia marcescens, Acinetobacter calcoacelicus, Corynebacterium diptheriae, Enterobacter aerogenes, Klebsiella pneumoniae, GMRSa, and Pseudomonas aeruginosa.
#### Test Summary, In-vivo
Clinical testing has been conducted in Canada to acquire safety data regarding silver, platinum, and benzalkonium chloride released from the Vantex ™ Central Venous Catheters with Oligon™ material with or without AMCTM Thromboshield™ coating. The purpose of this study was to determine the safety of the Oligon™ material by evaluating the silver and platinum levels in blood serum prior to, during, and after catheter use and to determine the safety of the AMC™ Thromboshield™ coating by evaluating the bezalkonium chloride levels in blood serum prior to, during, and after use. Safe levels of silver, platinum, and benzalkonium were found in both blood serum and urine during this study.
- Prior to market release of the product in Europe, a Marketing Evaluation was conducted to collect information on customer satisfaction with general performance characteristics of the Vantex™ Catheter. The clinicians expressed input as to the visibility of the catheter on a chest X-Ray. As a result of this feedback, Edwards Lifesciences modified the formulation of the Oligon™ material to include barium sulfate, which provides additional radiopacity to the catheter. The barium sulfate makes the radiopacity characteristic of the VantexTM Catheters comparable to that of the predicate Baxter Multi-Med® Catheters.
{3}------------------------------------------------
## Test Summary, Biocompatibility:
Biocompanibility testing was performed on the Vantex™ Central Venous Catheter with Oligon™ material with or without AMC™ Thromboshicid™ (both Oligon™ formulations) in accordance with the requirements specified in International Standards Organization (ISO) 10993-1-1994 Biological Evaluation of Medical Devices - Part 1: Guidance on Selection of Tests and the FDA General Program Memorandum No. G95-1. The Vantex™ Central Venous Catheter with Oligon™ matcrial with or without AMC™ Thromboshield™ (both Oligon™ formulations) was found to be biocompatible and nontoxic and acceptable for its intended use.
# Rationale for Substantial Equivalence Determinations:
The battery of non-clinical and clinical tests discussed above demonstrates that Vantex™ Central Venous Catheter with Oligon™ material with or without AMC™ Thromboshield™ for both the original and barium-filled Oligon™ formulations exhibits comparable mechanical and functional specifications to the predicate devices in addition to being biocompatible and chemically acceptable. Based upon those characteristics, Vantex™ Central Venous Catheter with Oligon™ material with or without AMC™ Thromboshield™ is substantially equivalent to the predicate devices in safety and effectiveness in addition to being intended for the same uses.
for Smith
son Smith Regulatory Affairs Associate Edwards Lifesciences LLC
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features the department's name encircling a stylized eagle emblem. The eagle is depicted with three lines forming its body and wings, and wavy lines representing water beneath it. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the emblem.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUL 7 - 2000
Mr. Jason Smith Regulatory Affairs Associate Edwards Lifesciences LLC One Edwards Way Irvine, California 92614
Re : K992532 Trade Name: Vantex™ Central Venous Catheter with OLIGON Material with or without AMC™ Thromboshield™ Coating Requlatory Class: II Product Code: FOZ Dated: March 14, 2000 Received: March 8, 2000
Dear Mr. Smith:
This letter corrects our substantially equivalent letter of June 13, 2000, regarding the Regulatory Class.
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to leqally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Requlation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug
{5}------------------------------------------------
Page 2 - Mr. Smith
Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in requlatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note; this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4692. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address
"http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely.
Timothy A. Ulatowski Director Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
| Page | of ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ |
|------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
510(k) Number (if known); K992532
Vantex™ Central Venous Catheters with OLIGON™ material Device Name: with or without AMC™ Thromboshield™ coating
Indications For Use:
The Vantex™ Central Venous Catheter with OLIGON™ with or without AMC™ Thromboshield™ coating are indicated for use in patients requiring pressure monitoring, infusion of solutions, and blood sampling in the central vein. All catheters are manufactured with a polyurethane-based OLIGONTM antimicrobial polymer. The OLIGON™ material uses silver as the antimicrobial agent.
#### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office Of Device Evaluation (ODE)
Prescription Use
OR
Over-The-Counter Use -------
(Per 21 CFR 801.109)
(Optional Format 1-2-96)
Patteux Cruccete
(Division Sign-Off) Division of Dental, Infection Control, and General Hospital Devices 510(k) Number_1992532
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.