← Product Code [GEX](/productcode/GEX) · K992508

# TSX PROBE, OCULAR CONTACT PROBE;TSX PROBE,OCULAR CONTACT PROBE,SMA (K992508)

_Instruments For Medicine and Diagnostics · GEX · Aug 19, 1999 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K992508

## Device Facts

- **Applicant:** Instruments For Medicine and Diagnostics
- **Product Code:** [GEX](/productcode/GEX.md)
- **Decision Date:** Aug 19, 1999
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

Contact Transscleral Cyclophotocoagulation (CYC) for the reduction of intra ocular pressure (IOP) in glaucoma patients. This ocular contact probe to be used for this procedure. Ablation of the Ciliary Body in glaucoma patients.

## Device Story

TSX Probe is a sterile, single-use, disposable ocular contact probe; used for Transscleral Cyclophotocoagulation (CYC). Device consists of handgrip, tip, and focusing lens; tip made of clear amber medical plastic angled to match human eye curvature for placement near limbus. 600 um fiber optic cable connects to HGM SmartScan™ or other SMA-compatible infrared lasers. Physician operates probe by moving it in circular fashion around cornea; laser energy delivered at regular intervals to ciliary body. Procedure reduces IOP in advanced glaucoma patients; provides non-invasive alternative to cyclocryotherapy with reduced postoperative pain and inflammation.

## Clinical Evidence

No clinical data provided; substantial equivalence based on design, intended use, and technological similarity to the predicate device.

## Technological Characteristics

Materials: Medical-grade amber plastic tip. Sensing/Actuation: Fiber optic delivery of infrared laser energy. Dimensions: 600 um fiber, 3-meter length. Connectivity: SMA-compatible interface for infrared lasers. Sterilization: Ethylene Oxide (EtO). Form factor: Handheld disposable probe with angled distal tip.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- Iris G-Probe

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# K 992508 510(k) Summary per 21 CFR § 807.92 AUG 19 1999

Instruments for Medicine and Diagnostics, Inc. (a.k.a. I4MD) 3959 West 1820 South Salt Lake City, Utah 84104 (801) 972-0500 (801) 972-4884 (fax) Tracy S. Best, Regulatory Manager Preparation Date: July 19, 1999

#### Summary of Safety and Effectiveness for:

Trade Name: TSX Probe

Common Name: Laser Delivery Device, Ocular Contact Probe

Classification Name: Laser, Ophthalmic, Accessory

### Legally Marketed Predicate Devices for Substantial Equivalence:

* Iris G-Probe, Manufactured by Iriderm Division, Laser Instrument, Surgical, Powered

#### Rationale for SE:

The Iris G-Probe is a sterile single use delivery device for the reduction of intra ocular pressure (IOP) in glaucoma patients. A three meter, 600 um fiber delivers the light energy from the laser to the focusing lens. The g-probe is furnished as sterile to the customer. The treatment performed using this non-invasive contact probe allows for selective ablation of ciliary processes. The procedure, Transscleral Cyclophotocoagulation (CYC), causes less postoperative pain and inflammation and is an alternative treatment from Cyclocryotherapy. The device is compatible with infrared photocoagulators.

### Description of Submitted Device:

The TSX Probe consists of three basic parts; handgrip, tip and focusing lens and is a presterilized (Ethylene Oxide) disposable contact probe. Its tip is a clear amber medical plastic that is angled to match the curvature of the human eye and therefore allows for the placement of the probe vertically, closest to the limbus. The angle of the contact tip (distal end) conforms to the curvature of the eye. The three-meter 600 um fiber plugs into an HGM SmartScan™ laser or other SMA connection compatible device allows the physician to perform noninvasive contact CYC with marketed (approved) infrared lasers.

{1}------------------------------------------------

## Intended Uses of the TSX Probe:

See attachment "5" for a complete listing of indicated uses.

### Technological Characteristics and Substantial Equivalence:

The TSX Probe is similar to the G-Probe, both have a 600 um fiber which delivers the treatment light to the tip of the contact probe. The TSX Probe is sleeker having a contact tip which is smaller than the G-Probe. Both offer treatment modalities for laser light delivery to ocular tissue, for various types of glaucoma in it's advanced stages. The probes are moved in a circular fashion around the cornea using the limbus as a circular guide. Both probes allow laser energy to be fired at regular intervals into the ciliary body. The two probes are similar in design, intended use, and compatible with lasers widely in use.

### Conclusion:

The TSX Probe is substantially equivalent to other existing laser surgical delivery devices already in commercial distribution. The labeling of the TSX Probe as a single-use pre-sterilized delivery device for treatment of (CYC) is an addition to a large field of Ophthalmic treatment methods with medical lasers.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an emblem featuring a stylized depiction of three human profiles facing to the right, with flowing lines suggesting movement or progress.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

AUG 9 1999

Mr. Tracy S. Best Regulatory Affairs Instruments for Medicine and Diagnostics, Inc. 3959 West 1820 South Salt Lake City. Utah 84104

Re: K992508 Trade Name: TSX Probe Regulatory Class: II Product Code: GEX Dated: July 19, 1999 Received: July 27, 1999

Dear Mr. Best:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{3}------------------------------------------------

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

Russell Sage

he Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

510(k) Number (if known): _ K 992908

Device Name: _ TSX Probe

Indications For Use:

Contact Transscleral Cyclophotocoagulation (CYC) for the reduction of intra
ment of the same was in the saylar contact on the to be Contact Transscieral Cyclopholocoagnation. This ocular contact probe to be
ocular pressure (IOP) in glaucoma patients. This ocular contact probe to be ocular pressure (1017 in glausema partices for this procedure.

Ablation of the Ciliary Body in glaucoma patients.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PROED IS NEEDED IS NEEDED IS Concurrence of CDRH, Office of Device Evaluation (ODE)

Russell Payne for JPD

Division of General Restorative

Prescription Use . (Per 21 CFR 801.109)

OR

Over-The-Counter Use -

---

**Source:** [https://fda-staging.innolitics.com/device/K992508](https://fda-staging.innolitics.com/device/K992508)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).