ULTRAMARK 400C/SONOACE 6000 ULTRASOUND SYSTEM

K992470 · Advanced Technology Laboratories, Inc. · IYO · Aug 23, 1999 · Radiology

Device Facts

Record IDK992470
Device NameULTRAMARK 400C/SONOACE 6000 ULTRASOUND SYSTEM
ApplicantAdvanced Technology Laboratories, Inc.
Product CodeIYO · Radiology
Decision DateAug 23, 1999
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1560
Device ClassClass 2
AttributesPediatric

Indications for Use

Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: Fetal (includes infertility monitoring of follicle development) - Abdominal - Intra-operative (abdominal organs and peripheral vessel, neurological) - Pediatric Small Organ - Neonatal Cephalic . Adult Cephalic - Cardiac - Trans-Rectal - Trans-Vaginal - Peripheral-Vascular - Muscular-Skeletal (conventional, superficial) -

Device Story

Mobile, software-controlled diagnostic ultrasound system; acquires ultrasound data via piezoelectric transducers (2.0-9.0 MHz). Transforms reflected sound waves into B-mode, M-mode, Color-Flow/Power Doppler, Pulsed Doppler, 3D, and Tissue Harmonic Imaging (THI) outputs. Operated by healthcare professionals in clinical settings. Features digital beamforming, digital dynamic receive focusing, and cine loop (64 images). Provides anatomical measurements (distances, areas, volumes) and obstetric/cardiac/vascular analysis packages. Displays real-time mechanical and thermal indices. Supports biopsy guidance via optional adapters. Output viewed on monitor; aids clinical diagnosis and tissue sampling. DICOM 3.0 compatible for remote viewing.

Clinical Evidence

Bench testing only. Acoustic output measurements provided to demonstrate compliance with FDA Track 3 limits. No clinical data presented.

Technological Characteristics

Piezoelectric transducer technology; digital beamforming; digital dynamic receive focusing. Supports 2.0-9.0 MHz probes. Connectivity: DICOM 3.0. Safety standards: UL 2601-1, CSA C22.2 No. 601.1, EN 60601-1-2. Acoustic output: MI 1.9, ISPTA 720 mW/cm2. Software-controlled.

Indications for Use

Indicated for diagnostic ultrasound imaging and fluid flow analysis in fetal, abdominal, intra-operative, pediatric, small organ, neonatal/adult cephalic, cardiac, trans-rectal, trans-vaginal, peripheral-vascular, and musculoskeletal applications. Includes biopsy guidance and infertility monitoring. Prescription use only.

Regulatory Classification

Identification

An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Special Controls

*Classification.* Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ UM 400C/ SA 6000C Ultrasound System K992470 ## 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS This summary of safety and effectiveness is provided as part of this Premarket Notification in compliance with 21 CFR, Part 807, Subpart E, Section 807.92. 1) Submitter's name, address, telephone number, contact person: Advanced Technology Laboratories, Inc. P.O. Box 3003 Bothell, WA 98031-3003 Vice President, Worldwide Quality & Regulatory Affairs Telephone: (425) 487-7602 Prepared: August 23, 1999 Name of the device, including the trade or proprietary name if applicable, the common or 2) usual name, and the classification name, if known: > Common/Usual Name: Diagnostic Ultrasound System and Accessories Proprietary Name: Ultramark® 400C Diagnostic Ultrasound System and Transducers. Also called: SonoAce 6000C Diagnostic Ultrasound System and Transducers. | Classification Names: | FR Number | Product Code | |------------------------------------------|-----------|--------------| | Ultrasound Pulsed Echo Imaging System | 892.1560 | 90-IYO | | Ultrasonic Pulsed Doppler Imaging System | 892.1550 | 90-IYN | | Diagnostic Ultrasound Transducer | 892.1570 | 90-ITX | #### Identification of the predicate or legally marketed device: 3) Advanced Technology Laboratories, Inc. believes that the UM 400C/ SA 6000C Ultrasound System is substantially equivalent to the currently marketed SA 8800/HDI® 1500 system (K974269). {1}------------------------------------------------ 510(k) Premarket Notification Tissue Harmonic Imaging ## UM 400C/ SA 6000C Ultrasound System #### Device Description: 4) The UM 400C/SA 6000C system is a general purpose, mobile, software controlled, diagnostic ultrasound system. Its function is to acquire ultrasound data and to display the data as B-mode, M-mode, Color-Flow Doppler, Pulsed (PW) Doppler, Power Doppler, 3D, Tissue Harmonic Imaging or in a combination of these modes. M-mode uses the sweep display method which has its images flow from the left to the right on the monitor The UM 400C/SA 6000C also gives the operator the ability to measure anatomical structures and offers analysis packages that provide information that is used to make a diagnosis by competent health care professionals. The system has real time acoustic output display with two basic indices, a mechanical index and a thermal index, which are both automatically displayed. Nine different models of transducers are available and any two may be connected at the same time. In addition to the initial operational settings for each transducer preprogrammed in the system, user-customized parameter settings for each transducer may be inserted by the operator and stored for recall as needed via the system control panel. Customization includes transmit focusing, filtering, image enhancement processing, dynamic window curve selection. Controls are also provided to select display format (single and various combinations), to activate zoom features, and to utilize the cine loop function. The UM 400C/SA 6000C system uses digital beamforming technology, and supports a variety of Linear and Convex probes for a wide variety of applications. It is a diagnostic ultrasound scanner, which provides high resolution, high penetration performance, and various measurement functions. Probes are supported in frequencies from 2.0 MHz to 9.0 MHz. The system can be used to measure distances and calculate areas, circumferences and volumes, as well as calculate the date of delivery by using BPD (biparietal diameter), OFD (occipitofrontal diameter), HC (head circumference), AC (abdominal circumference), AD (abdominal diameter), FL (femur length), CRL (crown rump length), APTD (anteroposterior trunk diameter), TTD (transverse trunk diameter), GS (gestational sac), LMP (last menstural period.), Cardiac Analysis and Vascular Analysis. Biopsy guidelines are provided on screen to assist in the collection of tissue samples, using biopsy guide adapters offered as an optional accessory. The UM 400C/SA 6000C supports the Cine function (capable of storing up to 64 sequential images) and real-time zoom function to the region-of-interest. The system provides the ability to perform remote viewing of images, without compression, via a Dicom 3.0 compatible output. Management of patient history is possible by {2}------------------------------------------------ image-filing function. High-resolution images are provided by utilizing a technology called digital dynamic receive focusing. The UM 400C/SA 6000C has been designed to meet the following electromechanical safety standards: - -UL 2601-1, Underwriters Laboratories Standards, Medical Electrical Equipment - -C22.2 No. 601.1, Canadian Standards Association, Medical Electrical Equipment - declaration of the acoustic output of medical diagnostic ultrasonic equipment - EN 60601-1-2 (IEC 601-1-2,) European Norm, Collateral Standard: Electromagnetic -Compatibility #### Intended Use: 5) Ultramark 400C/ SonoAce 6000C intended uses as defined FDA guidance documents are: - Fetal (includes infertility monitoring of follicle development) - - Abdominal - - Intra-operative (abdominal organs and peripheral vessel, neurological) - - -Pediatric - Small Organ - - Neonatal Cephalic . - Adult Cephalic - - Cardiac - - Trans-Rectal - - Trans-Vaginal - - Peripheral-Vascular - - Muscular-Skeletal (conventional, superficial) - Typical examinations performed using the system are: - General abdominal and pelvic studies including organ surveys, assessment, and retro-peritoneal cavity studies. - Study of small parts including breasts, shoulders, thyroid, and the abdominal wall. . - Pediatric scans of organs and bony structures. - - -Peripheral vascular applications including carotid arteries, legs, arms, feet, and penile artery. - -Monitoring procedures for infertility studies (other than in vitro fertilization). - -First, second and third trimester pregnancy studies. - Prostate, prostate biopsy guidance, and rectal wall studies. - - -Neonatal head studies. - Transcranial studies of middle cerebral arteries, internal carotid artery, and vertebral arteries. - - -Cardiac studies in adults and children. - Biopsy guidance for tissue or fluid sampling. - - Conventional podiatry scans. - - Intraoperative application including soft tissue structures. . {3}------------------------------------------------ K992470 p. 34,4 UM 400C/ SA 6000C Ultrasound System #### 6) Technological Characteristics: This device operates identical to the predicate devices in that piezoelectric material in the transducer is used as an ultrasound source to transmit sound waves into the body. Sound waves are reflected back to the transducer and converted to electrical signals that are processed and displayed as a 2D and M-mode, Spectral Doppler, Color Doppler, Power Doppler, 3D, or Tissue Harmonic Imaging images. Transducer patient contact materials are biocompatible. The device's acoustic output limits are: All Applications: | TIS/TIB/TIC | 0.0 - 5.0 | (Range) | |-------------|------------|-----------| | ISPTA | 720 mW/cm2 | (Maximum) | | MI | 1.9 | (Maximum) | The limits are the same as predicate Track 3 devices. {4}------------------------------------------------ ## DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/4/Picture/1 description: The image shows a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three curved lines above it, representing the department's mission to protect and promote the health of all Americans. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle. The logo is simple and recognizable, conveying the department's authority and purpose. ### Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 # AUG 7 3 1999 Mr. Terrence J. Sweeney Vice President, Worldwide Quality and Regulatory Affairs ATL Ultrasound Post Office Box 3003 · Bothel, Washington 98041-3303 K992470 Re: > Trade Name: Ultramark® 400C/Sonoace 6000C Diagnostic Ultrasound System with Tissue Harmonic Imaging (THI) Regulatory Class: II 21 CFR §892.1560/Procode: 90 IYO Dated: July 23, 1999 Received: July 26, 1999 Dear Mr. Sweeney: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. This determination of substantial equivalence applies to the following transducer intended for use with the Ultramark® 400C/Sonoace 6000C Diagnostic Ultrasound System with Tissue Harmonic Imaging (THI), as described in your premarket notification: ## Transducer Model Number ## C3-7 Curved Linear Array 4.5 MHz/60R/60D/128 elements If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval) it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing {5}------------------------------------------------ ### Page 2 - Mr. Terrence J. Sweeney Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic QS inspections, the FDA will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification does not affect any obligation you may have under sections 531 and 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. This determination of substantial equivalence is granted on the condition that prior to shipping the first device, you submit a postclearance special report. This report should contain complete information, including acoustic output measurements based on production line devices, requested in Appendix G, (enclosed) of the Center's September 30, 1997 "Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers." If the special report is incomplete or contains unacceptable values (e.g., acoustic output greater than approved levels), then the 510(k) clearance may not apply to the production units which as a result may be considered adulterated or misbranded. The special report should reference the manufacturer's 510(k) number. It should be clearly and prominently marked "ADD-TO-FILE" and should be submitted in duplicate to: > Food and Drug Administration Center for Devices and Radiological Health Document Mail Center (HFZ-401) 9200 Corporate Boulevard Rockville, Maryland 20850 This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4591. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597 or at its Internet address: "http://www.fda.gov/cdrh/dsmamain.html". {6}------------------------------------------------ # Page 3 - Mr. Terrence J. Sweeney l : If you have any questions regarding the content of this letter, please contact Paul M. Gammell, Ph.D. at (301) 594-1212. Sincerely yours, CAPT Daniel G. Schultz, M.D. Acting Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure(s) {7}------------------------------------------------ ## 4.3 INDICATIONS FOR USE ### STIC ULTRASOUND INDICATIONS STATEMENT DIA 510(k) Number: Ultramark® 400C/SonoAce 6000C Ultrasound System System: Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | Clinical Application | | Mode of Operation (*includes simultaneous B-mode) | | | | | | | |---------------------------|------------------------------------------|---------------------------------------------------|---|-----|-----|-------------------|----------------------|------------------| | General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) | | Ophthalmic | Ophthalmic | | | | | | | | | | Fetal (See Note 3) | P | P | P | P | P | Note 1 | Notes 2, 5, 6 | | | Abdominal | P | P | P | P | P | Note 1 | Notes 2, 5, 6 | | | Intra-operative (Abdominal,<br>vascular) | P | P | P | P | P | Note 1 | Notes 2, 5 | | | Intra-operative (Neuro.) | P | P | P | | P | Note 1 | Note 5 | | Fetal Imaging<br>& Other | Laparoscopic | | | | | | | | | | Pediatric | P | P | P | P | P | Note 1 | Notes 2, 5, 6 | | | Small Organ (See Note 4) | P | P | P | P | P | Note 1 | Notes 2, 5, 6 | | | Neonatal Cephalic | P | P | P | | P | Note 1 | Note 5 | | | Adult Cephalic | P | P | P | | P | Note 1 | Note 5 | | | Trans-rectal | P | P | P | P | P | Note 1 | Notes 2, 5 | | | Trans-vaginal | P | P | P | P | P | Note 1 | Notes 2, 5 | | | Trans-urethral | | | | | | | | | | Trans-esoph. (non-Card.) | | | | | | | | | | Musculo-skel. (Convent.) | P | P | P | | P | Note 1 | Notes 2, 5 | | | Musculo-skel. (Superfic.) | P | P | P | | P | Note 1 | Notes 2, 5 | | | Intra-luminal | | | | | | | | | | Other (spec.) | | | | | | | | | | Other (spec.) | | | | | | | | | | Cardiac Adult | P | P | P | P | P | Note 1 | Note 5 | | Cardiac | Cardiac Pediatric | P | P | P | P | P | Note 1 | Note 5 | | | Trans-esophageal (card) | | | | | | | | | | Other (spec.) | | | | | | | | | Peripheral<br>Vessel | Peripheral vessel | P | P | P | | P | Note 1 | Note 5 | | | Other (spec.) | | | | | | | | N=new indication; P= previously cleared by FDA in K981510 & K990970; E= added under Appendix E Additional Comments: Color Doppler includes Color Amplitude Doppler (P) Note 1: PWD/Color Doppler, PWD/Power Doppler (P) Note 2: Includes imaging for guidance of biopsy (P) Note 3: Includes infertility monitoring of follicle development (P) Note 4: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients (P) Note 5: 3D Imaging (P) Note 6: Tissue Harmonic Imaging (N) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109) Vinil h. Kym Division Sign-Off) Division of Reproductive, Abdominal, E and Radiological I 510(k) Number {8}------------------------------------------------ # DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE STATEMENT : K992470 510(k) Number: System: Ultramark® 400C/ SonoAce 6000C Ultrasound System C3-7 Curved Linear Array 4.5 MHz/60R/60D/128 elements Scanhead: Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | Clinical Application | | Mode of Operation (*includes simultaneous B-mode) | | | | | | | |---------------------------|------------------------------------------|---------------------------------------------------|---|-----|-----|-------------------|----------------------|------------------| | General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) | | Ophthalmic | Ophthalmic | | | | | | | | | | Fetal (See Note 3) | N | N | N | | N | Note 1 | Notes 2, 5, | | | Abdominal | N | N | N | | N | Note 1 | Notes 2, 5, | | | Intra-operative (Abdominal,<br>vascular) | | | | | | | | | | Intra-operative (Neuro.) | | | | | | | | | Fetal Imaging<br>& Other | Laparoscopic | | | | | | | | | | Pediatric | N | N | N | | N | Note 1 | Notes 2, 5, | | | Small Organ (See Note 4) | N | N | N | | N | Note 1 | Notes 2, 5, | | | Neonatal Cephalic | | | | | | | | | | Adult Cephalic | | | | | | | | | | Trans-rectal | | | | | | | | | | Trans-vaginal | | | | | | | | | | Trans-urethral | | | | | | | | | | Trans-esoph. (non-Card.) | | | | | | | | | | Musculo-skel. (Convent.) | | | | | | | | | | Musculo-skel. (Superfic.) | | | | | | | | | | Intra-luminal | | | | | | | | | | Other (spec.) | | | | | | | | | | Cardiac Adult | | | | | | | | | Cardiac | Cardiac Pediatric | | | | | | | | | | Trans-esophageal (card.) | | | | | | | | | | Other (spec.) | | | | | | | | | Peripheral<br>Vessel | Peripheral vessel | | | | | | | | | | Other (spec.) | | | | | | | | N= new indication; P= previously cleared by FDA; E= added under Appendix E Additional Comments: Color Doppler includes Color Amplitude Doppler Note 1: PWD/Color Doppler, PWD/Power Doppler (N) Note 2: Includes imaging for guidance of biopsy (N) Note 3: Includes infertility monitoring of follicle development (N) Note 4: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients (N) Note 5: 3D Imaging (N) Note 6: Tissue Harmonic Imaging (N) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109) Chid A. Segmen (Division Sign-Off) Division of Reproductive, Abdominal, EN and Radiological Devi 510(k) Number K992470
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...