← Product Code [FYA](/productcode/FYA) · K992436

# GOWN BACK, MODEL 13-403N; 13-404N; 13-405N (K992436)

_Techstyles · FYA · Jan 20, 2000 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K992436

## Device Facts

- **Applicant:** Techstyles
- **Product Code:** [FYA](/productcode/FYA.md)
- **Decision Date:** Jan 20, 2000
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4040
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery

## Indications for Use

The Gown Back is to be worn over a surgeon's gown during procedures and in situation when the surgeon's back needs to be considered sterile. This accessory device is used to protect both operating room personnel and the surgical patient from the transfer of microorganisms, body fluids, and particulate matter.

## Device Story

Gown Back is a single-use, fluid-repellent vest with a front snap opening; worn over standard surgical gowns by surgeons or surgical team members in the operating room. Device provides supplementary protection for chest and back in light to moderate fluid situations; not a primary protective garment or fluid barrier. Design intended to maintain sterility of surgeon's back during procedures. Benefits include reduced risk of transfer of microorganisms, body fluids, and particulate matter between personnel and patients.

## Clinical Evidence

No clinical data; bench testing only.

## Technological Characteristics

Single-use vest; nonwoven fabric; treated for fluid repellency; front snap opening; non-sterile barrier design.

## Regulatory Identification

Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.

## Special Controls

*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.

## Predicate Devices

- Optima® Surgical Gowns
- Orex Surgeon's Vest
- White Knight Surgeon's Vest

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

## TECHStules

K992436

PREMARKET NOTIFICATION 510(k) SUMMARY (As Required by 21 CFR 807.92(c))

June 30, 1999 Contact Person: Stephanie Howard

Device

Trade name - GOWN BACK Common name - Surgeon's vest or gown back Classification name - Surgical gown - 21 CFR 878.4040

Legally marketed predicate device Optima® Surgical Gowns Orex Surgeon's Vest White Knight Surgeon's Vest

Description and Conclusions

The device is a vest, commonly called a Gown Back, with a front snap opening. The Gown Back is fluid repellent; it is Single Use. The primary intended use for the Gown Back is as a sterile vest to be worn over a surgeon's gown during procedures or in situations when the surgeon's back needs to be considered sterile. It provides a supplementary layer of protection for the chest and back in light to moderate fluid situations. The Gown Back is not a primary protective garment. The vest design does not provide complete coverage, and the fabric is fluid repellent, not a fluid barrier.

Comparison

Optima® Surgical Gowns and Techstyles, Inc. Gown Back are made of the same nonwoven fabric, the same color and have been treated with the same fluid repellency treatment. The Orex Surgeon's Vest and the White Knight Surgeon's Vest are functionally the same as Techstyles, Inc. Gown Back, though they are of different fabrics.

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. Inside the circle is an emblem featuring a stylized caduceus, which is a symbol often associated with healthcare and medicine. The caduceus in the logo is depicted with three intertwined strands.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JAN 2 0 2000

Ms. Stephanie T. Howard Techstyles, Incorporated 16415 Addison Road, Suite 850 Addison, Texas 75001

K992436 Re: Gown Back, Model 13-403N; 13-404N; 13-405N Trade Name: Regulatory Class: II Product Code: FYA Dated: December 1, 1999 Received: December 6, 1999

Dear Ms. Howard:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Druq, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions aqainst misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. ਚ substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any

{2}------------------------------------------------

Page 2 - Ms. Howard

obligation you might have under sections 531 through 542 of obligation you magno maer the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4692. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note ene organis on entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

Timothy A. Ulatowski

Timothy A. Ulatowski Director Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Ver/ 3 – 4/24/96

Applicant: Techstyles, Inc.

510(k) Number (if known):

Device Name: GOWN BACK, Surgeon's Vest - Single Use/Disposable

## Indications For Use:

Target population: Surgeons or other members of the surgical team.

The Gown Back is to be worn over a surgeon's gown during procedures and in situation when the surgeon's back needs to be considered sterile. This accessory device is used to protect both operating room personnel and the surgical patient from the transfer of microorganisms, body fluids, and particulate matter.

Concurrence of CDHR, Office of Device Evaluation (ODE)

OVER TINE COUNTER DEVICE

(Per 21 CFR 801.109)

(Optional Format 1-2-96)

(Division Sign-Off)
Division of Dental, Infection Control,
and General Hospital Devices

510(k) Confidential

510(k) Number _

---

**Source:** [https://fda-staging.innolitics.com/device/K992436](https://fda-staging.innolitics.com/device/K992436)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com