NEO-2 METER

K992431 · Automata Instrumentation, Inc. · FIZ · Oct 13, 1999 · Gastroenterology, Urology

Device Facts

Record IDK992431
Device NameNEO-2 METER
ApplicantAutomata Instrumentation, Inc.
Product CodeFIZ · Gastroenterology, Urology
Decision DateOct 13, 1999
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.5820
Device ClassClass 2

Indications for Use

The NEO-2 Meter may be used by hemodialysis personnel to test conductivity, temperature, pressure, and pH of the dialysate solution used with hemodialysis delivery systems. The NEO-2 Meter may also be used to test the conductivity/temperature and pH of acid and sodium bicarbonate dialysate concentrates and water used in hemodialysis applications.

Device Story

NEO-2 Meter is a handheld diagnostic tool for hemodialysis environments; operated by hemodialysis personnel. Device measures conductivity, temperature, pressure, and pH of dialysate solutions, concentrates, and water. Provides real-time monitoring of critical parameters for hemodialysis delivery systems. Assists clinicians in verifying solution quality and system performance; ensures adherence to safety standards for dialysate preparation and delivery. Benefits patient by ensuring dialysate composition and system pressure remain within therapeutic ranges during hemodialysis treatment.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Handheld meter for measuring conductivity, temperature, pressure, and pH. Designed for use with hemodialysis delivery systems and associated fluids. Operates as a diagnostic monitoring tool. No specific materials, software algorithms, or connectivity standards detailed in the provided documentation.

Indications for Use

Indicated for use by hemodialysis personnel to monitor conductivity, temperature, pressure, and pH of dialysate solutions, concentrates (acid and sodium bicarbonate), and water in hemodialysis applications.

Regulatory Classification

Identification

A hemodialysis system and accessories is a device that is used as an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and that consists of an extracorporeal blood system, a conventional dialyzer, a dialysate delivery system, and accessories. Blood from a patient flows through the tubing of the extracorporeal blood system and accessories to the blood compartment of the dialyzer, then returns through further tubing of the extracorporeal blood system to the patient. The dialyzer has two compartments that are separated by a semipermeable membrane. While the blood is in the blood compartment, undesirable substances in the blood pass through the semipermeable membrane into the dialysate in the dialysate compartment. The dialysate delivery system controls and monitors the dialysate circulating through the dialysate compartment of the dialyzer.(1) The extracorporeal blood system and accessories consists of tubing, pumps, pressure monitors, air foam or bubble detectors, and alarms to keep blood moving safely from the blood access device and accessories for hemodialysis (§ 876.5540) to the blood compartment of the dialyzer and back to the patient. (2) The conventional dialyzer allows a transfer of water and solutes between the blood and the dialysate through the semipermeable membrane. The semipermeable membrane of the conventional dialyzer has a sufficiently low permeability to water that an ultrafiltration controller is not required to prevent excessive loss of water from the patient's blood. This conventional dialyzer does not include hemodialyzers with the disposable inserts (Kiil type) (§ 876.5830) or dialyzers of high permeability (§ 876.5860). (3) The dialysate delivery system consists of mechanisms that monitor and control the temperature, conductivity, flow rate, and pressure of the dialysate and circulates dialysate through the dialysate compartment of the dialyzer. The dialysate delivery system includes the dialysate concentrate for hemodialysis (liquid or powder) and alarms to indicate abnormal dialysate conditions. This dialysate delivery system does not include the sorbent regenerated dialysate delivery system for hemodialysis (§ 876.5600), the dialysate delivery system of the peritoneal dialysis system and accessories (§ 876.5630), or the controlled dialysate delivery system of the high permeability hemodialysis system § 876.5860). (4) Remote accessories to the hemodialysis system include the unpowered dialysis chair without a scale, the powered dialysis chair without a scale, the dialyzer holder set, dialysis tie gun and ties, and hemodialysis start/stop tray.

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/2 description: The image shows the address for the Food and Drug Administration. The address is 9200 Corporate Boulevard, Rockville MD 20850. The text is black and the background is white. OCT 1 3 1999 Ms. Linda V. Masano President AUTOMATA Instrumentation, Inc. 15600 North 78th Street Scottsdale, AZ 85260-1735 Re: K992431 NEO - 2 Meter Dated: July 20, 1999 Received: July 21, 1999 Regulatory Class: II 21 CFR §876.5820/Product Code: 78 FIZ Dear Ms. Masano: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html". Sincerely yours, CAPT Daniel G. Schultz, M.D. Acting Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {1}------------------------------------------------ AUTOMATA NEO-2 Premarket Notification ## INDICATIONS FOR USE STATEMENT G. ## Indications for Use Statement K992431 510(k) Number (if Known): _ Device Name: NEO-2 Meter Indications for Use: The NEO-2 Meter may be used by hemodialysis personnel to test conductivity, temperature, pressure, and pH of the dialysate solution used with hemodialysis delivery systems. The NEO-2 Meter may also be used to test the conductivity/temperature and pH of acid and sodium bicarbonate dialysate concentrates and water used in hemodialysis applications. ## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED.) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109) OR Over-the-Counter Use David A. Stypmann (Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological Devi 510(k) Number
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...