The Web Clinical System is intended for the management and displaying of x-ray images, other radiological objects, and information. It can manage images from different modalities, single and multiple file servers, and interfaces to various Radiological Information Systems (RIS), image storage and printing devices using DICOM or similar interface standards within the system or across computer networks at distributed locations.
Device Story
RAYPAX Web Clinical System (WCS) is a digital imaging workstation for managing/displaying radiological images and data. Inputs include radiological images from various modalities, file servers, and RIS via DICOM or similar standards. WCS retrieves and browses these images over the Internet, Intranet, or computer networks. Used in clinical settings by trained professionals (physicians, radiologists, nurses, technicians). Output is visual display of medical images for clinical review. Device does not contact patients or control life-sustaining equipment; interpretation remains with the human clinician. Benefits include remote access to radiological data across distributed locations to facilitate clinical decision-making.
Clinical Evidence
No clinical data. Bench testing only; hazard analysis performed.
Technological Characteristics
Digital imaging workstation software. Connectivity via DICOM or similar interface standards for integration with RIS, image servers, and printing devices. Operates over Internet/Intranet/computer networks. Non-patient contacting. No energy source or specific materials of construction described.
Indications for Use
Indicated for management and display of radiological images and data by trained professionals including physicians, radiologists, nurses, medical technicians, and assistants.
Regulatory Classification
Identification
A medical image communications device provides electronic transfer of medical image data between medical devices. It may include a physical communications medium, modems, and interfaces. It may provide simple image review software functionality for medical image processing and manipulation, such as grayscale window and level, zoom and pan, user delineated geometric measurements, compression, or user added image annotations. The device does not perform advanced image processing or complex quantitative functions. This does not include electronic transfer of medical image software functions.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows a logo or wordmark that appears to be for a company or brand named "REMIX". The logo is rendered in a style that gives it a textured or speckled appearance, possibly to create a distressed or vintage effect. The letter "R" is more prominent and solid compared to the rest of the word, which seems to fade or become more sparse as it moves to the right.
#### 510(k) Summary of Safety and Effectiveness OCT 1 3 1999
K992426
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990.
### Date Prepared:
July 16, 1999
# Submitter's Information: 21 CFR 807.92(a){1)
SAMSUNG SDS CO., LTD 707-19, Yoksam-Dong, Kangnam-Gu, Seoul, South Korea, 135-080
# Trade Name, Common Name and Classification: 21 CFR 807.92(a)(2)
| Trade Name: | Samsung RAYPAX™ Web Clinical System |
|------------------------|-----------------------------------------------|
| Common Name: | Digital Imaging Workstation |
| Classification Name: | System, Digital Image Communication Accessory |
| Device Classification: | 21 CFR 892.2010 |
# Predicate Device: 21 CFR 807. 92(a)(3)
The Samsung SDS Co. Ltd. RAYPAX™ WCS is substantially equivalent to the: Manufacturer: SECTRA IMTEC AB Device Name: Wise II Image Management System 510(k) Number: K983447 9/30/98 Date Received: Decision Date: 10/29/98 Decision: Substantially Equivalent Panel Code device reviewed by: Radiology Panel Code device classified by: Radiology Product Code: 90 LMB Classification: Class I
### Device Description: 21 CFR 807 92(a)(4)
The Web-Clinical System (WCS) is a component of the RAYPAX System, a PACS solution by Samsung or can be a separate device for other manufacturer's PACS. The WCS is a system for managing radiological images and data by retrieving and browsing medical images through the World Wide Web, (Internet), a clients Intranet or over computer networks.
### Indications for Use: 21 CFR 807 92(a)(5)
The Web Clinical System is intended for the management and displaying of x-ray images, other radiological objects, and information. It can manage images from different modalities, single and multiple file servers, and interfaces to various Radiological Information Systems (RIS), image storage and printing devices using DICOM or similar interface standards within the system or across computer networks at distributed locations.
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows a stylized text that appears to read "RIVAGE". The text is rendered in a speckled or grainy texture, giving it a distressed or vintage look. The letter "R" is the most prominent and solid, while the remaining letters gradually fade out, creating a sense of depth or movement.
Typical users of this system are trained professionals, including but not limited to physicians, radiologists, nurses, medical technicians, and assistants.
# Technological Characteristics: 21 CFR 807 92(a)(6)
The device does not contact the patient, nor does it control any life sustaining devices. A physician, providing ample opportunity for competent human interprets images and information being printed.
# Conclusion: 21 CFR 807 92(b)(1)
The 510(k) Pre-Market Notification for the WCS contains adequate information and data to enable FDA - CDRH to determine substantial equivalence to the predicate device.
- The RAYPAX™ WCS system has been designed and will be manufactured in accordance 1. with the voluntary standards listed in the enclosed voluntary standard survey.
- 2. The submission contains the results of a hazard analysis and all potential hazards have been classified as Minor.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized symbol featuring three horizontal bars above a wavy line, representing the department's mission related to health and human services.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
OCT 1 3 1999
Carl Alletto Samsung SDS Co. Ltd. C/O Otech, Inc. 2001 East Oakshores Drive Aubrey, Texas 76227
Re:
K992426
RAYPAX™ Web Clinical System (WCS) Dated: July 16, 1999 Received: July 21, 1999 Regulatory class: I 21 CFR 892.2020/Procode: 90 LMD
Dear Mr. Alletto:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in . regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notfication"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Capt. Daniel G. Schultz, M.D. Acting Director. Division of Reproductive. Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "RAX" in a stylized font. The letter "R" is large and bold, while the letters "A" and "X" are smaller and more faded. The letters appear to be made up of small dots, giving them a textured look.
Page 1 of -1
(Indications for Use Form)
**510(k) Number:** K992426
# Device Name:
Samsung SDS Co. Ltd. RAYPAX™ Web Clinical System
# Indications for Use:
- The Web Clinical System is intended for the management and displaying of xray images, other radiological objects, and information. It can manage images from different modalities, single and multiple file servers, and interfaces to various Radiological Information Systems (RIS), image storage and printing devices using DICOM or similar interface standards within the system or across computer networks at distributed locations.
Typical users of this system are trained professionals, including but not limited to physicians, radiologists, nurses, medical technicians, and assistants.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
David G. Hanson
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT,
and Radiological Devices
510(k) Number K992426
OR
Prescription Use
(Per 21 CFR 801.109)
Over-The-Counter Use
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.