The MRP-Titan system is intended to be used as the proximal part of a hip stem arthoplasty. The indications for use of the MRP-Titan system are revision operations and femur fractures of primary hip operated patients as well as tumour situations at the femur.
Device Story
MRP-Titan is a modular revision hip system designed for femoral canal fixation via a star-shaped stem design. System components include curved or straight stems, necks (with/without fins/trochanter), locking screws, and extension sleeves. Surgeons select components based on required length and anatomical needs. Used in clinical settings by orthopedic surgeons during revision hip arthroplasty, fracture repair, or tumor resection. The device acts as the proximal portion of a hip stem prosthesis, providing structural support and fixation within the femur. It is compatible with Biolox forte hip heads. Benefits include restoration of hip function, leg length, and antetorsion. Components are single-use and provided sterile.
Clinical Evidence
No clinical data provided. Evidence consists of bench testing, specifically fatigue testing conducted in accordance with ISO 7206-4 and ISO 7206-8. The device with a 12/14 taper successfully survived more than 5 x 10^6 cycles without failure under specified loading conditions.
Technological Characteristics
Modular hip system components made of Ti6Al4V (ASTM F136) and Titanium. Features a star-shaped stem design for femoral canal fixation. Employs a 12/14 taper for hip head connection. Sterilized via Gamma-Radiation (> 25 kGy).
Indications for Use
Indicated for primary and revision hip arthroplasty, revision of primary hip stem prostheses, proximal femur fractures, proximal femur tumors, reconstruction of the upper proximal femur following primary hip surgery or trauma, leg-lengthening, and improving antetorsion after failed primary hip stem surgery.
Regulatory Classification
Identification
A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.
{0}------------------------------------------------
K992403
Manufacturers name: ROTEC MEDIZINTECHNIK GmbH Adress: Am Mühlberg 31 D - 91085 Weisendorf Telephone: *49 - 9135 - 32 15 Contact Person: Dr.-Ing. Ulrich Holzwarth Telephone: *49 - 9135 - 71 03 38
## Revised Summary of Safety & Effectiveness Information Supplied as part of the 510 (k) Application for the ROTEC MEDIZINTECHNIK GmbH MRP-Titan System
Proprietary Name: ROTEC, MRP-Titan System
Complete device description:
The patientmatched modular revision hip system MRP-Titan was developed to achieve best fixation of the stem in the femur canal by using a star-shaped design. Depending on the needed length the surgeon is able to decide using the sleeve or not. All necks are available in three sizes to achieve the total length needed. The use of the instruments and the single step to assemble the system is shown very clearly in the enclosed pictures.
The MRP-Titan components are single-use devices.
Identification of all device components, sizes and materials:
| Name | Article No. | Diameter/size | Materials |
|--------------------------------|-------------|---------------|-----------|
| Stem 200 mm, curved | 61013-03 | 13 mm | Ti6Al4V |
| Stem 200 mm, curved | 61014-03 | 14 mm | Ti6Al4V |
| Stem 200 mm, curved | 61015-03 | 15 mm | Ti6Al4V |
| Stem 200 mm, curved | 61016-03 | 16 mm | Ti6Al4V |
| Stem 200 mm, curved | 61017-03 | 17 mm | Ti6Al4V |
| Stem 200 mm, curved | 61018-03 | 18 mm | Ti6Al4V |
| Stem 200 mm, curved | 61019-03 | 19 mm | Ti6Al4V |
| Stem 200 mm, curved | 61020-03 | 20 mm | Ti6Al4V |
| Stem 200 mm, curved | 61021-03 | 21 mm | Ti6Al4V |
| Stem 200 mm, curved | 61022-03 | 22 mm | Ti6Al4V |
| Stem 140 mm, straight | 61013-01 | 13 mm | Ti6Al4V |
| Stem 140 mm, straight | 61014-01 | 14 mm | Ti6Al4V |
| Stem 140 mm, straight | 61015-01 | 15 mm | Ti6Al4V |
| Stem 140 mm, straight | 61016-01 | 16 mm | Ti6Al4V |
| Stem 140 mm, straight | 61017-01 | 17 mm | Ti6Al4V |
| Stem 140 mm, straight | 61018-01 | 18 mm | Ti6Al4V |
| Stem 140 mm, straight | 61019-01 | 19 mm | Ti6Al4V |
| Stem 140 mm, straight | 61020-01 | 20 mm | Ti6Al4V |
| Stem 140 mm, straight | 61021-01 | 21 mm | Ti6Al4V |
| Stem 140 mm, straight | 61022-01 | 22 mm | Ti6Al4V |
| Special Implant for Trochanter | 61000-04 | small | Ti6Al4V |
| Special Implant for Trochanter | 61000-05 | large | Ti6Al4V |
{1}------------------------------------------------
| Neck with Trochanter | 61000-11 | K | Ti6Al4V |
|----------------------|----------|-------|----------|
| Neck with Trochanter | 61000-12 | M | Ti6Al4V |
| Neck with Trochanter | 61000-13 | L | Ti6Al4V |
| Locking screw | 61000-06 | | Ti6Al4V |
| Neck without fin | 61002-11 | K | Ti6Al4V |
| Neck without fin | 61002-12 | M | Ti6Al4V |
| Neck without fin | 61002-13 | L | Ti6Al4V |
| Neck with fin | 61001-11 | K | Ti6Al4V |
| Neck with fin | 61001-12 | M | Ti6Al4V |
| Neck with fin | 61001-13 | L | Ti6Al4V |
| Screw | 60901-04 | small | Titanium |
| Screw | 60901-07 | large | Titanium |
| Extension sleeve | 61016-04 | 16 mm | Ti6Al4V |
| Extension sleeve | 61018-04 | 18 mm | Ti6Al4V |
| Extension sleeve | 61020-04 | 20 mm | Ti6Al4V |
| Extension sleeve | 61022-04 | 22 mm | Ti6Al4V |
Included are Confidential Detailed Drawings of the MRP-Titan System
Implant Materials: Medical Alloy Ti6Al4V, ASTM F136 Biolox® forte, (Medical Alumina Al2O3), ISO 6474
Statement of Safety: We have reviewed all prior literature on this type of implant device and can not find a substantial body of information on the adverse effects with this device.
## Potential Risks:
The potential risks associated with this device are the same as with any joint replacement device. These include, but are not limited to:
- Reaction to bone cement Fracture of the component Cardiovascular disorders Implant loosening/migration Soft tissue imbalance Deformity of the ioint Dislocation Metal sensitivity
- Bone fracture Hematoma Blood vessel damage Nerve damage Excessive wear Infection Delayed wound healing
Substantial equivalency: ROTEC MRP-Titan functions in a similar manner as predicted commercially available devices and are equivalent to systems cited in this application (table 2-2 new) especially to Link, Modular Revision Hip Stem, K 955296 or Johnson & Johnson, System S-ROM, K 961 939.
Indications for use: The MRP-Titan system is intended to be used as the proximal part of a hip stem arthoplasty. The indications for use of the MRP-Titan system are revision operations and femur fractures of primary hip operated patients as well as tumour situations at the femur.
{2}------------------------------------------------
The MRP-Titan System is designed and proven to match with the following hip head: Bioloxforte 12/14, manufactured by Ceramtec CE 0044, FDA Master File. Fatigue testing acc. to ISO 7206-4, -8 has demonstrated that the device with 12/14 taper survived more than 5 x 10 cycles without any failure while loaded acc. to ISO 7206.
Sterilization: The MRP-Titan system metal elements and Alumina balls will be shipped in sterile package by Gamma-Radiation, > 25 kGy, by the process shown and validated in chapter 15, truthful and accurate statement is included.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, symbolizing the department's mission to protect and promote the health and well-being of the nation.
Public Health Service
JAN 21 2000
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Dr. -Ing. Ulrich Holzwarth Head of Quality Assurance ROTEC MEDIZINTECHNIK GmbH Am Muhlberg 31 D-91084 Weisendorf Germany
Re: K992403 Trade Name: MRP-Titan Regulatory Class: II Product Code: LZO Dated: November 12, 1999 Received: November 15, 1999
Dear Dr. -Ing. Ulrich Holzwarth:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895, A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}------------------------------------------------
Page 2- Dr. -Ing. Ulrich Holzwarth
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Henrici Jagan
James E. Dillard III Acting Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
510(k) Number (if known): K992403
Device Name: MRP-Titan
Indications for use:
- Primary and revision arthroplasty: Revision of primary hip stem prostheses ●
- Fractures of the proximal femur ●
- Tumor situation of the proximal femur ●
- Reconstructions of the upper area of the proximal femur after both primary hip ● operation or traumafracture
- Recovery of leg-lengthening ●
- Improving antetorsion after failed primary hip stem operation ●
Date: 99-11-05
## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
Rumell Lyons
(Division Sign-Off) Division of General Restorative Device 510(k) Number _
Prescription Use X
OR
Over-The=Counter Use
1-3
(Optional Format 1-2-96)
CE
19
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.