The COLORSOFT PROSTHETIC Tinted Soft Contact lens is indicated for daily wear to enhance or alter the apparent eye color, including ocular masking, either in sighted or non-sighted eyes that require a prosthetic contact lens for cosmetic management of conditions such as comeal, iris or lens abnormalities. The lens may also be prescribed for occlusive therapy for conditions such as diplopia, amblyopia or extreme photophobia. The lens may be disinfected using a chemical disinfection system only.
Device Story
Prosthetic soft contact lens; modifies existing clear contact lenses via tinting process. Reactive color additives (e.g., reactive black 5, blue 19, etc.) affixed to anterior surface corresponding to iris; additives reacted with poly hydroxyethyl methacrylate. Available in various patterns: clear lens with black pupil, black occluder, black annular with clear pupil, or tinted lens with black pupil. Used by eyecare professionals to manage cosmetic ocular conditions or provide occlusive therapy. Lens handling and cleaning do not remove additives; optical performance remains unchanged. Patient benefits include improved cosmetic appearance and management of light sensitivity or double vision.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on descriptive characteristics, material safety of color additives, and performance comparisons to predicate devices.
Technological Characteristics
Soft contact lens material (poly hydroxyethyl methacrylate); reactive color additives (e.g., reactive black 5, blue 19, yellow 15) affixed to anterior surface. Dimensions: pupil sizes 2.0-12.5 mm; annular diameters 7.5 mm to full lens diameter. Chemical disinfection only. No software or electronic components.
Indications for Use
Indicated for daily wear in sighted or non-sighted patients requiring prosthetic contact lenses for cosmetic management of corneal, iris, or lens abnormalities; also indicated for occlusive therapy for diplopia, amblyopia, or extreme photophobia.
Regulatory Classification
Identification
A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.
Predicate Devices
Alden Classic Prosthetic Tinted Soft Contact Lens
COLORSOFT Color Enhanced Tinted Soft Contact Lens
Submission Summary (Full Text)
{0}------------------------------------------------
# 510 (k) SUMMARY OF SAFETY AND EFFECTIVENESS
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
K992339 The assigned 510(k) number is:
#### Applicant information:
| Date Prepared: | July 12, 1999 |
|------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name: | Colorsoft Laboratories |
| Address | 623 Glacier Drive<br>Grand Junction, CO 81503 |
| Contact Persons: | Mrs. Deanna Werber |
| Phone Number: | (970) 248-9445 |
| USA Consultant: | Martin Dalsing,<br>Med-Vice Consulting, Inc.<br>Consultant for Colorsoft Laboratories, Inc<br>623 Glacier Drive<br>Grand Junction, CO 81503<br>(970) 243-5490<br>Fax #: (970) 243-5501<br>E-mail: mdalsing@gj.net |
### Device Information:
| Device Classification: | Class II |
|------------------------|--------------------------------------------------|
| Classification Number: | LPL |
| Trade Name: | COLORSOFT PROSTHETIC<br>Tinted Soft Contact lens |
| Classification Name: | Lenses, Soft Contact, Daily Wear |
{1}------------------------------------------------
## Substantially Equivalent Devices:
The "COLORSOFT PROSTHETIC" Tinted Soft Contact Lens is substantially equivalent to the "Alden Classic Prosthetic Tinted Soft Contact Lens" and the Colorsoft Laboratories "COLORSOFT Color Enhanced Tinted Soft Contact Lens", the predicate devices.
### Device Descriptive Characteristics:
The COLORSOFT PROSTHETIC Tinted Soft Contact Lens are lens that have been previously prescribed for a specific patient. They have been supplied to Colorsoft Laboratories to be modified by a tinting process using color additives that have been listed as safe for contact lenses in accordance with the FDA's color additive regulations. The color additives are used in amounts not to exceed the minimum reasonably required to accomplish the intended coloring effect. As part of the manufacturing process, the lenses containing the color additives are thoroughly washed to remove unbound reactive color additives. The manufacturing process alters and/or changes the specifications to the clear version of a contact lens by affixing a listed reactive color additive on that portion of the anterior (front) surface of the lens that corresponds to the iris. The COLORSOFT PROSTHETIC contact lenses are tinted to the eyecare professional instructions.
The prosthetic color additive effect is formed by reacting one or more of the reactive color additives listed in this paragraph with (poly hydroxyethyl methacrylate). The reactive color additives that may be used either alone or in combination are: reactive black 5, reactive blue 19, reactive blue 19, reactive blue 4, reactive blue 163. reactive red 11. reactive red 180. reactive vellow 15, reactive yellow 86, or reactive orange 78. The color additives used are not removed by lens handling or approved cleaning/disinfecting procedures. The prosthetic lens tinting process does not alter the optical and/or performance characteristics of the finished tinted soft contact lens.
Tint Patterns Available:
- 1. Clear lens with Black Pupil. Pupil sizes available in 2.0 mm to 12.5 mm.
- 2. Black Occluder Lens. A Central Black area that occludes light. Available to full lens diameter in 0.5 mm increments.
- 3. Black Annular with clear pupil. Black Annular diameter range 7.5 mm to full lens diameter in 0.5 mm increments. Clear Pupil diameter range 2.0 mm to 7.5 mm in 0.5 mm increments.
- 4. Tinted lens with Black Pupil. Uses the Colorsoft Color Enhanced tinted contact lens with black pupil. Pupil sizes available in 2.0 mm to 12.5 mm diameter.
{2}------------------------------------------------
## INDICATIONS FOR USE:
The COLORSOFT PROSTHETIC Tinted Soft Contact lens is indicated for daily wear to enhance or alter the apparent eye color, including ocular masking, either in sighted or non-sighted eyes that require a prosthetic contact lens for cosmetic management of conditions such as corneal, iris or lens abnormalities. The lens may also be prescribed for occlusive therapy for conditions such as diplopia, amblyopia or extreme photophobia. The lens may be disinfected using a chemical disinfection system only.
The following table summarizes Colorsoft Laboratories claim of substantial equivalency in terms of safety and efficacy to the predicate devices previously mentioned.
| | Characteristic | COLORSOFT PROSTHETIC<br>(Colorsoft Laboratories) | ALDEN CLASSIC PROSTHETIC<br>(Predicate Device) |
|-----|-----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1.) | INTENDED<br>USE | Cosmetic Management of conditions such as<br>corneal, iris, or lens abnormalities. | Cosmetic Management of conditions such as<br>corneal, iris, or lens abnormalities. |
| 2.) | INDICATION | The CANTOR & SILVER 5X<br>PROSTHETIC Tinted Soft Contact Lens are<br>indicated for daily wear to enhance or alter<br>the apparent eye color, including ocular<br>masking, either in sighted or non-sighted<br>eyes that require a prosthetic contact lens<br>for cosmetic management of conditions such<br>as corneal, iris or lens abnormalities. The<br>lens may also be prescribed for the<br>correction of refractive ametropia (myopia,<br>hyperopia and astigmatism) in aphakic and<br>not aphakic persons or for occlusive therapy<br>for conditions such as diplopia, amblyopia<br>or extreme photophobia. The lens are<br>disinfected using a hydrogen peroxide lens<br>care system only and are available in a<br>frequent replacement program . | The ALDEN CLASSIC PROSTHETIC<br>(polymacon) Tinted, Soft Contact Lens is<br>indicated for daily wear to enhance or alter<br>the apparent eye color, including ocular<br>masking, either in sighted or non-sighted<br>eyes that require a prosthetic contact lenses<br>for cosmetic management of conditions such<br>as corneal, iris or lens abnormalities. The<br>lens may also be prescribed for the<br>correction of refractive ametropia (myopia,<br>hyperopia and astigmatism) in aphakic and<br>not aphakic persons or for occlusive therapy<br>for conditions such as diplopia, amblyopia<br>or extreme photophobia. The lens may be<br>disinfected with either a chemical or a heat<br>disinfection system. |
| 3.) | FDA "listed"<br>colored additives | The reactive colored additives consist of<br>reactive black 5, reactive blue 4, reactive<br>blue 19, reactive 21, reactive blue 163,<br>reactive yellow 15, reactive yellow 86,<br>reactive orange 78, reactive red 11 and<br>reactive red 180. | Vat Green 1, Vat Brown 1,<br>Vat Blue 6, |
| 4.) | Uses and<br>restrictions. | The color additives listed above may be used<br>to color contact lenses in amounts not to<br>exceed the minimum reasonably required to<br>accomplish the intended prosthetic effect. | The color additives listed above may be used<br>to color contact lenses in amounts not to<br>exceed the minimum reasonably required to<br>accomplish the intended prosthetic effect. |
| 5.) | Color Additive<br>Characteristics | The color additives used are not removed by<br>lens handling and approved<br>cleaning/disinfecting procedures. The<br>optical and perform ance characteristics are<br>not altered by the lens coloring process. | The color additives used are not removed by<br>lens handling and approved<br>cleaning/disinfecting procedures. The<br>optical and performance characteristics are<br>not altered by the lens coloring process. |
Table #1 - Substantial Equivalence
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
AUG 16 1999
Colorsoft Laboratories Corporation c/o Mr. Martin Dalsing 623 Glacier Drive Grand Junction, CO 81503
Re: K992339 Trade Name: COLORSOFT PROSTHETIC Tinted Soft Contact lens Regulatory Class: II Product Code: 86 LPL Dated: July 12, 1999 Received: July 13, 1999
Dear Mr. Dalsing:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations .
{4}------------------------------------------------
Page 2 - Mr. Martin Dalsing
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
A. Ralph Rosenthal
A. Ralph Rosenthal, M.D. Director Division of Ophthalmic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## INDICATIONS FOR USE STATEMENT
Device Name:
## COLORSOFT PROSTHETIC Tinted Soft Contact lens
#### INDICATIONS FOR USE:
The COLORSOFT PROSTHETIC Tinted Soft Contact lens is indicated for daily wear to enhance or alter the apparent eye color, including ocular masking, either in sighted or non-sighted eyes that require a prosthetic contact lens for cosmetic management of conditions such as comeal, iris or lens abnormalities. The lens may also be prescribed for occlusive therapy for conditions such as diplopia, amblyopia or extreme photophobia. The lens may be disinfected using a chemical disinfection system only.
い
## (PLEASE DO OT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
t. Daniel W.C. Brown, PH.D
(Division Sign-Off)
Division of Ophthalmic Devices
510(k) Number K992339
\$\\$
Prescription Use X (Per 21 CFR 801.109) Or
Over-The-Counter Use
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.