Clinical laboratory specimens; Routine clinical practice settings (point-of-care and centralized laboratories)
Performance of the reagent strips was evaluated by typical users in clinical settings to demonstrate that results are comparable to current laboratory methods.
Clinical setting performance evaluation by typical users
Patients in point-of-care and centralized laboratory locations
MULTISTIX® 10 SG Reagent Strips and commonly used laboratory methods
Clinical test result comparability
Indications for Use
Reagent Strips are intended for use in at-risk patient groups to assist diagnosis in the following areas: kidney function, carbohydrate metabolism (e.g., diabetes mellitus), urinary tract infections and liver function. The reagent strips also measure physical characteristics, including acid-base balance and urine concentration. The test results can be used along with other diagnostic information to rule out certain disease states.
Device Story
Firm plastic reagent strips for urinalysis; contains pads for low-level protein (albumin), high-level protein, creatinine, blood, glucose, ketone, nitrite, pH, specific gravity, bilirubin, and urobilinogen. Input: urine specimen (random, overnight, or timed). Operation: manual dip and visual color chart comparison or instrumental reading via CLINITEK family of analyzers. Output: semi-quantitative results for analytes and protein-to-creatinine ratio. Used in point-of-care (clinics, physician offices) and centralized labs by professionals. Results assist clinicians in diagnosing kidney, metabolic, liver, and urinary tract conditions; used alongside other diagnostic data to rule out disease states.
Clinical Evidence
Clinical studies conducted in point-of-care and centralized laboratory settings by typical users. Performance of visual and instrumental (CLINITEK analyzers) readings compared to current MULTISTIX 10 SG Reagent Strips and standard laboratory methods. Results demonstrated that the device provides clinical results comparable to current test methods.
Technological Characteristics
Firm plastic strips with multiple reagent pads. Low-level protein assay uses dye-binding (high affinity dye: Bis(3',3"-diiodo-4',4"-dihydroxy-5',5"-dinitrophenyl)-3,4,5,6-tetrabromosulfonephthalein). Creatinine assay uses peroxidase-like activity of copper creatinine complexes catalyzing diisopropylbenzene dihydroperoxide and 3,3',5,5'-tetramethyl-benzidine. Instrumental reading via CLINITEK series analyzers.
Indications for Use
Indicated for at-risk patient groups to assist in diagnosis of kidney function, carbohydrate metabolism (e.g., diabetes), urinary tract infections, and liver function; measures urine physical characteristics. For professional use in point-of-care (physician offices, clinics) and centralized laboratory settings.
Regulatory Classification
Identification
A urinary protein or albumin (nonquantitative) test system is a device intended to identify proteins or albumin in urine. Identification of urinary protein or albumin (nonquantitative) is used in the diagnosis and treatment of disease conditions such as renal or heart diseases or thyroid disorders, which are characterized by proteinuria or albuminuria.
Predicate Devices
MULTISTIX® 10 SG Reagent Strips
MICROALBUSTIX™ Reagent Strips
CLINITEK® Microalbumin Reagent Strips
Submission Summary (Full Text)
{0}------------------------------------------------
## NOV 2 4 1999
Page 1 of 2
# 510(k) Safety and Effectiveness Summary
| Prepared: | July, 1999 |
|----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | Bayer Corporation, Business Group Diagnostics |
| Address: | 1884 Miles Avenue, P. O. Box 70<br>Elkhart, Indiana 46515-0070 |
| Contact: | Syed Ataullah<br>Manager, Regulatory Compliance<br>219-262-6865 |
| Device: | Trade/Proprietary Name: "NEW MULTIPLES" Reagent Strips<br>Common/Usual Name: Multiple Reagent Strips for Urinalysis<br>Document Control Number: K99 2257 |
| Classification: | Division of Clinical Laboratory Devices<br>Panel - Clinical Chemistry and Toxicology<br>Classification Code - 75 JIR (Urinary protein or albumin) Class 1<br>75 JFY (Creatinine, enzymatic) Class 2 |
| Predicate Device(s): | MULTISTIX® 10 SG Reagent Strips<br>MICROALBUSTIX™ Reagent Strips<br>CLINITEK® Microalbumin Reagent Strips |
Device Description:
Bayer's "NEW MULTIPLES" Reagent Strips are firm plastic strips for urinalysis that contain reagent areas for low level protein (15 mg/dL albumin), high level protein (>30 mg/dL protein), creatinine, blood, glucose, ketone (acetoaceticacid), nitrite, pH, specific gravity, bilirubin, and urobilinogen. A protein-to-creatinine ratio is also determined. New Multiples Reagent Strips are manually dipped into a urine specimen and "read" visually using a color chart. "NEW MULTIPLES" Reagent Strips may also be read instrumentally, using the CIINITEK® family of Urine Chemistry Analyzers. The method is designed for use with random, overnight or timed specimens. Semi-quantitative results are available within one minute.
{1}------------------------------------------------
Intended Use:
Reagent Strips are intended for use in at-risk patient groups to assist diagnosis in the following areas: kidney function, carbohydrate metabolism (e.g., diabetes mellitus), urinary tract infections and liver function. The reagent strips also measure physical characteristics, including acid-base balance and urine concentration. The test results can be used along with other diagnostic information to rule out certain disease states.
### Technological Characteristics:
Bayer's "NEW MULTIPLES" Reagent Strips include new reagent areas for determination of proteinuria. A new protein reagent allows "low level" detection of albumin at 15 mg/dL in addition to the current protein reagent used for detection of "high level" protein at 30 mg/dL. A new creatinine reagent is used that measures creatinine levels from 10 to 300 mg/dL. The protein-to-creatinine ratio is reported as Normal and Abnormal.
The assay for low level protein (albumin) is based on dye binding using the high affinity dye. Bis(3',3"-diiodo-4',4"-dihydroxy-5',5"-dinitrophenyl)-3,4,5,6tetrabromosulfonephthalein). At a constant pH, the development of a blue color is owing to the presence of albumin; the intensity of the color is proportional to the albumin concentration. The assay for creatinine is based on the peroxidase-like activity of copper creatinine complexes that catalyze the reaction of diisopropylbenzene dihydroperoxide and 3,3 ',5 '-tetramethyl-benzidine.
### Assessment of Performance:
The performance of Bayer's "NEW MULTIPLES" Reagent Strips was studied (visually and instumentally using a family of CLINITEK® Analyzers) in clinical setting by typical users. The results were compared to current MULTISTIX® 10 SG Reagent Strips and commonly used laboratory methods. The studies demonstrated that typical users in centralized and point-of-care laboratories can obtain clinical test results that are comparable to commonly used laboratory methods.
#### Conclusion:
Bayer's "NEW MULTIPLES" Reagent Strips have been developed for urinalysis with the currently used reagent areas and new protein-low, protein-high, and creatinine reagent pads. The "NEW MULTIPLES" Reagent Strips are for use in near-patient (point-of-care) and centralized laboratory locations. Studies show that the product provides clinical results comparable to other test methods in current clinical practice.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and a circle of text around the symbol. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".
NOV 2 4 1999
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Mr. Sved Ataullah Manager, Regulatory Compliance Bayer Corporation Business Group Diagnostics 1884 Miles Avenue, P.O. Box 70 Elkhart, Indiana 46515-0070
Re: K992257
> Trade Name: "NEW MULTIPLES" Reagent Strips Regulatory Class: II Product Code: JFY, JIL Regulatory Class: I Product Code: JIR, CEN, JRE, JJB, JIO, JIN, JMT, CDM, KOO Dated: October 18, 1999 Received: October 20, 1999
Dear Mr. Ataullah:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic OS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{3}------------------------------------------------
#### Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours.
Steven Sutman
Steven I. Gutman, M.D, M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
510(k) Number (if known): 长 992257
Device Name:
"NEW MULTIPLES" Reagent Strips
Indications for Use:
Bayer's "NEW MULTIPLES" Reagent Strips are firm plastic strips that contain reagent areas to test for low level protein (15 mg/dL albumin), high level protein (>30 mg/dL protein), creatinine, occult blood, glucose, ketone (acetoacetic acid), leukocytes, nitrite, pH, specific gravity, bilirubin, and urobilinogen. A protein to creatinine ratio is also determined. The strips are read visually by comparison to a color chart on the bottle label. The "NEW MULTIPLES" Reagent Strips can also be read instrumentally on the CLINITEK®50, CLINITEK® 100, CLINITEK® 200+ and CLINITEK® 500 Urine Chemistry Analyzers.
Bayer's "NEW MULTIPLES" Reagent Strips are intended for use in at-risk patient groups to assist diagnosis in the following areas: kidney function, carbohydrate metabolism (e.g., diabetes mellitus), urinary tract infections and liver function. The strips also measure physical characteristics, including acid-base balance and urine concentration. Test results can be used along with other diagnostic information to rule out certain disease states. Bayer's "NEW MULTIPLES" Reagent Strips are for professional use in Point -Of-Care locations such as physicians offices or clinics and centralized laboratory locations such as in hospitals.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| | Jean Coopy |
|---------------------------------------|------------|
| (Division Sign-Off) | |
| Division of Clinical Latora y Devices | |
| 510(k) Number | K992957 |
Prescription Use V (Per 21 CFR 801.109)
OR
Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________
(Optional Format 1-2-96)
Page 0
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.