← Product Code [NCW](/productcode/NCW) · K992228

# BAYER IMMUNO 1 HER-2/NEU ASSAY (K992228)

_Bayer Corp. · NCW · Sep 29, 2000 · Immunology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K992228

## Device Facts

- **Applicant:** Bayer Corp.
- **Product Code:** [NCW](/productcode/NCW.md)
- **Decision Date:** Sep 29, 2000
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.6010
- **Device Class:** Class 2
- **Review Panel:** Immunology
- **Attributes:** Real-World Evidence

## Real-World Evidence

| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
| --- | --- | --- | --- | --- | --- |
| K992228 · Sep 29, 2000 | BAYER IMMUNO 1 HER-2/NEU ASSAY | Bayer Corp. | Retrospective clinical serum samples | Retrospective clinical serum samples from 60 patients were used to correlate serial HER-2/neu measurements with clinical status (progression vs. lack of progression) to validate the assay's utility in monitoring metastatic breast cancer. | Retrospective study; Metastatic breast cancer; Clinical monitoring; Disease progression |

### Clinical Evidence

| Study Design | Population | Comparator | Key Endpoints |
| --- | --- | --- | --- |
| Retrospective clinical evaluation; Follow-up/Duration: 6-12 month period | Patients with metastatic breast cancer; Sample Size: 60 patients; Number of Sites: 3 | Not applicable for this study | Correlation of serial HER-2/neu changes with clinical status (progression vs. lack of progression) |

## Indications for Use

The Bayer Immuno 1™ Her-2/neu Assay is an in vitro, diagnostic device intended for use in the quantitative determination of HER-2/neu in human serum. HER-2/neu values obtained may be used in the follow-up and monitoring of patients with metastatic breast cancer. HER-2/neu values should be used in conjunction with information available from clinical and other diagnostic procedures in the management of breast cancer. The clinical utility of the serum measurement of HER-2/neu as a prognostic indicator for early detection of recurrence and in the management of patients on immunotherapy regimens has not been fully established.

## Device Story

Assay measures HER-2/neu protein in human serum samples; utilizes immunoassay technology with monoclonal antibodies (NB-3 conjugated to fluorescein; Fab' fragment of TA-1 conjugated to alkaline phosphatase). Samples, calibrators, or controls incubated at 37°C with conjugates; magnetic particles coated with anti-fluorescein antibody added for binding; complex washed to remove unbound molecules; colorimetric substrate added. Absorbance measured at 405 and 450 nm; rate of conversion proportional to HER-2/neu concentration. Used in clinical laboratory settings by trained personnel. Output provides quantitative HER-2/neu levels; clinicians use data to monitor disease progression in metastatic breast cancer patients. Benefits include objective tracking of clinical status over time.

## Clinical Evidence

Retrospective clinical study of 60 metastatic breast cancer patients over 6-12 months. Primary endpoint: correlation between serial serum HER-2/neu changes and clinical status (progression vs. lack of progression). A ≥15% increase in HER-2/neu was used as the threshold for progression. Results showed 66/111 progression events correctly identified and 109/142 non-progression events correctly identified. Total CV <3%. Analytical sensitivity (MDC) 0.11 ng/mL.

## Technological Characteristics

Immunoassay using monoclonal antibodies (NB-3, TA-1). Reagents: fluorescein-conjugated antibody and alkaline phosphatase-conjugated Fab' fragment. Separation via anti-fluorescein coated magnetic particles. Detection: colorimetric substrate, absorbance at 405/450 nm. Range: 0-250 ng/mL. Calibration: 6-level set (0-250 ng/mL). Controls: 3-level set. Software: cubic-through-zero curve-fitting algorithm.

## Regulatory Identification

A tumor-associated antigen immunological test system is a device that consists of reagents used to qualitatively or quantitatively measure, by immunochemical techniques, tumor-associated antigens in serum, plasma, urine, or other body fluids. This device is intended as an aid in monitoring patients for disease progress or response to therapy or for the detection of recurrent or residual disease.

## Special Controls

*Classification.* Class II (special controls). Tumor markers must comply with the following special controls: (1) A guidance document entitled “Guidance Document for the Submission of Tumor Associated Antigen Premarket Notifications (510(k)s) to FDA,” and (2) voluntary assay performance standards issued by the National Committee on Clinical Laboratory Standards.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# SEP 2 9 2000

# 510(K) SUMMARY

## FOR THE

# BAYER IMMUNO 1™ HER-2/neu ASSAY

This summary of 510(k) safety and effectiveness information is being submitted in Fills summary of a requirements of SMDA 1990 and 21 CFR 807.92.

The assigned 510(k) number is: K992228

#### GENERAL INFORMATION 1.

Trade Name:

Bayer Immuno 1™ HER-2/neu Assay

Classification Name:

Tumor-Associated Antigen Immunological Test Systems

Fredric Cline

Fredrick Clerie Director Regulatory Affairs Bayer Corporation Business Group Diagnostics 511 Benedict Avenue Tarrytown, NY 10591 Phone 914-524-2954 FAX 914-524-2500

9-22-2000

Date

{1}------------------------------------------------

This premarket notification is to add the Bayer Immuno 1™HER-2/neu Assay to the intended use of the Bayer Immuno 1™System. In this 510(k) application, the performance and clinical safety and effectiveness of the Bayer Immuno 1 HER-2/neu Assay for the management (monitoring) of metastatic breast cancer patients has been established by external clinical studies in the target population of longitudinal metastatic breast cancer patients and by comparison to accepted diagnostic procedures in accordance with the "Guidance Document For Submission of Tumor Associated Antigen Premarket Notifications, 510(k), to the FDA." Clinical evaluations of the Bayer Immuno 1 HER-2/neu Assay at three US clinical trial sites demonstrated clinical safety and effectiveness. These studies validated clinical performance characteristics and the comparison to accepted diagnostic procedures.

#### INDICATIONS FOR USE 2.

The Bayer Immuno 1™ HER-2/neu Assay is an in vitro diagnostic assay intended to quantitatively measure HER-2/neu protein in human serum on the Bayer Immuno 1 M System. HER-2/neu values obtained may be used in the follow-up and monitoring of patients with metastatic breast cancer. HER-2/neu values should be used in conjunction with information available from clinical and other diagnostic procedures in the management of breast cancer. The clinical utility of serum measurement of HER-2/neu as a prognostic indicator for early detection of recurrence and in the management of patients on immunotherapy regimens has not been fully established.

#### DEVICE DESCRIPTION 3.

The Bayer Immuno I™ HER-2/neu Assay utilizes a well-established immunoassay technology in which one monoclonal HER-2/neu antibody (NB-3) is conjugated to fluorescein (designated Reagent 1, or R1) and the Fab' fragment of another monoclonal HER-2/neu antibody (TA-1) is conjugated to alkaline phosphatase (Reagent 2, or R2). The R1 and R2 conjugates are reacted with patient sample, calibrator, or control and are incubated at 37℃ on the system. Immuno 1 Magnetic Particles coated with an antifluorescein antibody (mIMP™ Reagent) are then added and a second incubation occurs during which the antibody complex is bound. The magnetic particles complexed with the 

{2}------------------------------------------------

immunological sandwich are then washed to separate unbound molecules, and a colorimetric substrate is added. The rate of conversion of substrate to a compound with absorbance at 405 and 450 nm is measured is proportional to the concentration of HER-2/neu in the sample. A cubic-through-zero curve-fitting algorithm is used to generate standard curves.

The assay has a range of zero to 250 ng/mL. The Bayer SETpoint™ HER-2/neu Calibrators consist of a set of six calibrator levels at 0, 10, 25, 60, 125, and 250 ng/mL. The Bayer TESTpoint™ HER-2/neu Controls consist of a set of three control levels at approximately 15, 50 and 100 ng/mL.

#### SUMMARY OF STUDIES 4.

Non-clinical studies were performed to validate the performance of the method according to the protocol entitled "Non-clinical Testing Protocol for the Evaluation of the of the Bayer Immuno 1™ HER-2/neu Assay." Protocols were performed at the Bayer Corporation laboratories in Tarrytown, NY and Elkhart, IN. These studies included evaluation of interfering substances, cross-reactivity, heterophilic antibodies, calibration linearity, sample linearity, parallelism (sample dilution), hook effect, reproducibility, and reagent lot-to-lot variation.

The clinical evaluation of Immuno 1 HER-2/neu Assay as an aid in the management of breast cancer patients with Stage IV metastatic disease, during the course of disease and therapy, was conducted at three US clinical trial sites.

#### Characterization of the Antigen. 4.1

The calibrator antigen used in the Bayer Immuno 1 HER-2/neu assay is derived from a recombinant 3T3 mouse cell line 3-30. HER-2/neu p105 antigen is harvested from the tissue culture media and concentrated 10-fold. Western blot analysis shows a single dominant band with a molecular weight of 105,000 Daltons consistent with the HER-2/neu extracellular domain.

{3}------------------------------------------------

#### Characterization of the Antibodies 4.2

The NB-3 monoclonal (part no. 7591MR) and TA-1 monoclonal anti-HER-2/neu (part no. 7590MR) are used in the preparation of the R1 and R2 reagents for the Bayer Immuno 1m HER-2/neu Assay. The procedures for the preparation of the NB-3 R1 Reagent fluorescein and TA-1 R2 Reagent alkaline phosphatase conjugates are standard protocols similar to those used for conjugate preparation for other 510(k) Bayer Immuno 1 assays. Bayer Corporation (Business Group Diagnostics) in Elkhart is the approved supplier for the antibodies and antibody conjugates.

#### Assay Performance 4.3

#### Specificity: Interference 4.3.1

The recovery of HER-2/neu from patient samples was studied before and after spiking the serum samples with the potentially interfering substance. Each potential interferent was tested at a maximum concentration.

The Immuno 1 HER-2/neu Assay was performed on serum samples or pools of serum to which were added various concentrations of triglycerides, hemoglobin, immunoglobulin, bilirubin, albumin or cholesterol. HER-2/neu values were also measured in serum samples after spiking with either an individual chemotherapeutic drug, "Over the Counter" (OTC) drug, vitamin or HERCEPTIN® (Trastuzumab), trademark of Genetech BioOncology, South San Francisco, CA. None of the potential endogenous or exogenous interferents demonstrated any significant interfering effects on HER-2/neu recovery.

#### Cross-Reactivity 4.3.2

Possible cross-reactions in the Immuno 1 HER-2/neu assay were studied by comparing HER-2/neu recoveries in patient samples with Human Epidermal Growth Factor. The maximum effect seen with this crossreactant was not significant (<1%).

{4}------------------------------------------------

#### Heterophilic Antibodies 4.3.3

To investigate the effectiveness of the assay's reagent formulation in minimizing heterophilic antibody interferences, patient samples with HAMA, RF titers, or autoimmune diseases were tested for possible interference in the HER-2/neu assay. The observed HER-2/neu recoveries indicated a lack of significant heterophilic interference in the assay and demonstrated the effectiveness of the reagent formulation in minimizing these interferences.

#### Linearity 4.3.4

To determine the linearity of this assay, five individual serum samples from breast cancer patients were diluted (100%, 75%, 50%, 25%, and 0%) with a pool of human sera. Recoveries of the intermediate dilutions were all between 95 and 102 percent of the expected values. These results demonstrate the linearity of HER-2/neu recoveries over the entire calibration range.

#### Hook Effect (Antigen Excess) 4.3.5

Extremely high concentrations of HER-2/neu seen in some malignant conditions may cause a "hook effect" in an assay. An excess of analyte saturates both label and capture antibody and causes the reported concentration to "hook" back into the assay range rather than be flagged as above range. HER-2/neu antigen was diluted in Level 1 Calibrator at concentrations of 300 ng/mL to 10,000 ng/mL. Results clearly demonstrated the lack of a hook effect in the Immuno 1 HER-2/neu Assay at Her-2/neu values ≤10,000 ng/mL.

#### Parallelism (Dilution Studies) 4.3.6

As a further verification of assay linearity, and to qualify the Level 1 Calibrator as a sample diluent, five patient serum samples containing a high level of HER-2/neu were diluted (100%, 75%, 50%, 25%, and 0%) with Level 1 Calibrator. Linear regression analysis for the determination of deviations from linearity for each of these clinical samples showed no deviation from linearity. The recovery of HER-2/neu assay values ranged from 90% to 103%. These data demonstrate that Level 1 calibrator is an acceptable diluent for high samples, with accurate

{5}------------------------------------------------

recovery of diluted values. This study was repeated using the same samples to qualify Immuno 1 Sample Diluent B as a sample diluent. The recovery of HER-2/neu assay values ranged from 100% to 110%. These data demonstrate that Immuno 1 Sample Diluent B is also an acceptable diluent for high samples, with accurate recovery of diluted values.

#### 4.3.7 Reproducibility

Intra- and inter-assay reproducibility were evaluated at three clinical trial sites for Bayer SETpoint Complexed Her-2/neu Calibrators, Controls and a human serum pool with HER-2/neu concentration approximately 15 ng/mL. Imprecision data pooled across Immuno 1 HER-2/neu reagent lots and systems(sites) showed maximal total coefficients of variation (%CV) of 2.4% over the range of the assay method. This is well within acceptable limits for an assay of this type.

|              | Number of<br>Runs | Number of<br>Replicates | Mean  | Within Run |     | Total   |     |
|--------------|-------------------|-------------------------|-------|------------|-----|---------|-----|
|              |                   |                         | ng/mL | STD DEV    | %CV | STD DEV | %CV |
| Serum Pool   | 80                | 400                     | 15.6  | 0.27       | 1.7 | 0.28    | 1.8 |
| Calibrator 2 | 80                | 640                     | 10.0  | 0.24       | 2.4 | 0.24    | 2.4 |
| Calibrator 3 | 80                | 640                     | 24.9  | 0.51       | 2.0 | 0.52    | 2.1 |
| Calibrator 4 | 80                | 640                     | 59.3  | 1.09       | 1.8 | 1.12    | 1.9 |
| Calibrator 5 | 80                | 640                     | 123.2 | 2.33       | 1.9 | 2.38    | 1.9 |
| Calibrator 6 | 80                | 490                     | 245.2 | 4.16       | 1.7 | 4.26    | 1.7 |
| Control L3   | 80                | 400                     | 108.6 | 1.77       | 1.6 | 1.79    | 1.7 |

#### 4.3.8 Sensitivity (Detection Limit)

Sensitivity of the Immuno 1 HER-2/neu Assay was evaluated at three clinical trial sites by determining the Minimum Detectable Concentration (MDC). An MDC of 0.11 ng/mL was observed when assaying 480 replicates of the Immuno 1 HER-2/neu zero calibrator using two Immuno 1 HER-2/neu calibrator lots and three Immuno 1 HER-2/neu reagent lots.

{6}------------------------------------------------

#### 4.4 CLINICAL STUDIES

The clinical utility of the Bayer Immuno I HER-2/neu assay in monitoring patients with metastatic breast cancer was evaluated using retrospective serum samples from three clinical sites in the United States. Serum HER-2/neu values were measured in 60 patients with metastatic breast cancer over a 6-12 month period. These results were then separated into groups that either showed HER-2/neu values that paralleled the clinical course of disease, or HER-2/neu values that did not parallel the clinical course as follows: For all patients whose pretreatment serum HER-2/neu values exceeded 15 ng/mL, all serial measurements were analyzed visit-to-visit and serial changes in serum HER-2/neu were correlated with changes in clinical status. For each pair of serial measurements, an increase of equal or greater than 15% was considered to indicate progression, and a change of less than 15% increase was considered to indicate a lack of progression. Results presented in Table 1 show the overall correspondence of the serial HER-2/neu changes and changes in clinical status.

### Table 1: Correspondence of Serial HER-2/neu Changes and

|                        | Change In Clinical Status |                        |       |
|------------------------|---------------------------|------------------------|-------|
| Change In<br>HER-2/Neu | Progression               | Lack Of<br>Progression | Total |
| ≥ 15% increased        | 66                        | 33                     | 99    |
| < 15% increase         | 44                        | 109                    | 154   |
| Total                  | 111                       | 142                    | 253   |

|  | Clinical Status |
|--|-----------------|
|  |                 |
|  |                 |

{7}------------------------------------------------

# Distribution of HER-2/neu Concentrations; Sensitivity and Specificity

The Immuno 1 HER-2/neu Assay was used to estimate the clinical (crosssectional) sensitivity in patients with breast cancer and characterize the frequency distribution of Immuno 1 HER-2/neu assay values in a population of breast cancer patients by stage of disease.

Specificity of the Immuno 1 HER-2/neu Assay was determined in patients with benign breast diseases, other non-malignant diseases, and in normal healthy individuals.

The Upper Limit of 15 ng/ml is comparable to the Upper Limit seen with manual kits used for research purposes, and is also equivalent to the 15 ng/ml Upper Limit found by Bayer in a previous study.

## Conclusions from the Clinical Studies

The results of this retrospective clinical trial demonstrate that the Bayer Immuno 1™ HER-2/neu assay is reproducible, and is safe and effective for the management and follow-up of patients with metastatic breast cancer.

HER-2/neu is the first cellular oncogene, which has been shown useful in this clinical setting. Data collected from this study show that changes in serum HER-2/neu concentrations over time in metastatic breast cancer patients reflect changes in clinical status such as progression of disease.

The reproducibility of the Immuno 1 HER-2/neu assay is outstanding with total CVs of less than 3%. The detection limit of 0.1 ng/mL is acceptable for the intended use of this assay. This demonstrates that this assay should provide reliable and reproducible results when tested by different laboratories using different manufactured lots of reagents at different times.

#### રું. CONCLUSIONS DRAWN FROM ALL THE STUDIES

## Valid Scientific Evidence

The conclusions drawn from these studies are based upon valid scientific evidence. Data were gathered following a well-designed protocol, in a 

{8}------------------------------------------------

research laboratory operating under the principles of Good Clinical Clinical data were gathered during well controlled Practices. investigations conducted by qualified experts. Patient case histories were well documented. The results of this study are comparable to literature reports of experiences with HER-2/neu assays.

## Method Performance

Immuno 1 HER-2/neu results are highly reproducible with a maximum inter-assay %CV pooled over reagent lots and clinical sites of 2.4% over the range of the assay. Other performance characteristics including analytical sensitivity and specificity, cross-reactivity, linearity, antigen excess hook effect meet the accepted specifications set for an assay of this type.

### Safety and Effectiveness

These clinical studies confirm the safety and effectiveness of the Immuno I HER-2/neu Assay as an aid in the follow-up and management of metastatic breast cancer patients. The correspondence between Immuno 1 HER-2/neu concentrations and the patients' clinical course of disease demonstrate that the Immuno 1 HER-2/neu Assay may be used in conjunction with other clinical indicators to confirm disease progression in metastatic breast cancer patients.

{9}------------------------------------------------

Image /page/9/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with three curved lines representing its wings.

SEP 2 9 2000

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Mr. Fredrick Clerie Director Regulatory Affairs Bayer Corporation Business Group Diagnostics 511 Benedict Avenue Tarrytown, New York 10591

Re: K992228

> Trade Name: Bayer Immuno 1™ HER-2/neu Assay Regulatory Class: II Product Code: NCW Dated: July 6, 2000 Received: July 10, 2000

Dear Mr. Clerie:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{10}------------------------------------------------

Page 2

This letter will allow you to begin marketing your device as described in your 510(k) premarket This letter will anow you to oogin manating your device to a legally marketed nouncation. The I D7X Intamg of bassification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and IT you desire specific actrice for your assic devices), please contact the Office of Compliance at additionally 807.10 for mi vitio diagnestions on the promotion and advertising of your device, (301) 594-4560. Practiceliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Information on your rosponsionness and its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Steven Butman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{11}------------------------------------------------

# 001001

Page 1 of 1

510(k) Number (if known):

Device Name: Bayer Immuno 1TM HER-2/neu Assay

Indications For Use:

The Bayer Immuno 1™ Her-2/neu Assay is an in vitro, diagnostic device intended for use in the quantitative determination of HER-2/neu in human serum. HER-2/neu values obtained may be used in the follow-up and monitoring of patients with metastatic breast cancer. HER-2/neu values should be used in conjunction with information available from clinical and other diagnostic procedures in the management of breast cancer. The clinical utility of the serum measurement of HER-2/neu as a prognostic indicator for early atine) of and in the management of patients on immunotherapy regimens has not been fully established.

# (PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)
Division of Clinical Laboratory Devices
510(k) Number K992228

Prescription Use ✓
OR
Over-the-counter Use

Prescription Use (Per 21 CFR 801.109)

(Optional Format 1-2-96)

---

**Source:** [https://fda-staging.innolitics.com/device/K992228](https://fda-staging.innolitics.com/device/K992228)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
