SHIELDMASTER POWDERED PURPLE NITRILE EXAMINATION GLOVES

K992210 · Safeskin Corp. · LZA · Sep 10, 1999 · General Hospital

Device Facts

Record IDK992210
Device NameSHIELDMASTER POWDERED PURPLE NITRILE EXAMINATION GLOVES
ApplicantSafeskin Corp.
Product CodeLZA · General Hospital
Decision DateSep 10, 1999
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.6250
Device ClassClass 1

Indications for Use

A medical glove intended to be worn on the hands of healthcare and similar personnel to prevent contamination between such personnel and the patient.

Device Story

ShieldMaster Powdered Purple Nitrile Gloves are disposable patient examination gloves; worn by healthcare personnel to provide barrier protection; prevent contamination between personnel and patients. Device manufactured from nitrile material; meets ASTM D 3578-99 and ASTM D 6319-99 standards for dimensions, physical properties, and freedom from pinholes. Used in clinical settings; provides physical barrier; reduces risk of cross-contamination.

Clinical Evidence

Bench testing only. Device meets ASTM D 3578-99 and ASTM D 6319-99 standards for dimensions, physical properties, and freedom from pinholes (ASTM D 5151). Biocompatibility testing included primary skin irritation in rabbits and guinea pig sensitization, both of which passed. No clinical data required for medical gloves.

Technological Characteristics

Disposable nitrile examination gloves. Materials meet ASTM D 3578-99 and ASTM D 6319-99. Testing includes freedom from pinholes per ASTM D 5151. Biocompatibility confirmed via rabbit skin irritation and guinea pig sensitization tests.

Indications for Use

Indicated for use by healthcare and similar personnel as a protective barrier to prevent cross-contamination between personnel and patients.

Regulatory Classification

Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ SEP 1 0 1999 August 25, 1999 - 510(k) Summary of Safety and Effectiveness Information [1] - [2] Safeskin Corporation 12671 High Bluff Drive San Diego, CA 92130 Telephone: 619-794-8111 619-350-2382 Fax: Eugene V. Goorchenko Contact: 619-509-7010 Telephone: 619-350-2382 Fax: - (3) Trade Name: ShieldMaster Powdered Purple Nitrile Gloves Medical Gloves Common Name: Classification Name: Patient Examination Gloves - The predicate is a Safeskin blue Nitrile Glove which meets all of the requirements [4] of ASTM D 3578-99, Standard Specification for Rubber Examination Gloves (with the exception of elongation). - The SheildMaster Powdered Purple Nitrile Glove will meet all the current [ર] specifications for ASTM D 3578-99 (with the exception of elongation). - ShieldMaster Powdered Purple Nitrile Gloves are disposable devices intended to [6] be worn by healthcare and similar personnel to prevent contamination between such personnel and the patient. - ShieldMaster Powdered Purple Nitrile Gloves possess the following technological [7] characteristics (as compared to ASTM or equivalent standards): | Characteristics | Standards | |-----------------------|--------------------------------------------| | Dimensions | Meets ASTM D 3578-99 | | Physical Properties | Meets ASTM D 3578-99 and<br>ASTM D 6319-99 | | Freedom from pinholes | Meets ASTM D 3578-99 | Meets ASTM D 5151 992210 {1}------------------------------------------------ Image /page/1/Picture/0 description: The image shows a sequence of handwritten characters. The characters appear to be 'K992210'. The numbers are written in a simple, slightly rounded style, and the letter 'K' is in uppercase. Safeskin Corporation Biocompatability | Primary Skin Irritation in Rabbits | Passes | |------------------------------------|--------| | Guinea Pig Sensitization | Passes | - The performance test data that support a determination of substantial equivalence [8] are described above. - Clinical data are not needed for medical gloves. [୨] - [10] It can be concluded that the ShieldMaster Powdered Purple Nitrile Glove will perform according to the glove performance standards referenced in Section 7 above and therefore will meet ASTM standards, FDA requirements, and the labeling claims for the product. Consequently, this device is substantially equivalent to currently marketed devices. {2}------------------------------------------------ Public Health Service Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles three overlapping human figures or abstract shapes, creating a sense of movement or connection. SEP 1 0 1999 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Mr. Euqene V. Goorchenko Director of Requlatory Affairs Safeskin Corporation 12671 High Bluff Drive San Diego, California 92130 Re : K992210 Shieldmaster Powdered Purple Nitrile Trade Name: Examination Gloves Requlatory Class: I Product Code: LZA Dated: June 30, 1999 Received: July 1, 1999 Dear Mr. Goorchenko: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. ਮੁ substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) requlation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of {3}------------------------------------------------ ## Page 2 — Mr. Goorchenko the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4690. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda,gov/cdrh/dsmamain.html". Sincerely yours, Timothy A. Ulatowski Direc or Division of Dental, Infection Control, and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ K992210 Safeskin Corporation ## INDICATIONS FOR USE Safeskin Corporation 510(k) Number: Applicant: Device Name: Examination Powdered Purple Nitrile Medical Glove Indications for Use: A medical glove intended to be worn on the hands of healthcare and similar personnel to prevent contamination between such personnel and the patient. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH Office of Device Evaluation (ODE) Qlin S. Lin Over-The-Counter (Division Sign-Off) Division of Dental, In and General Hosp 510(k) Number 8
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...