The Pacific Hemostasis ThromboScreen® 400C is a photo-optical instrument used for the performance of in-vitro diagnostic coagulation testing of citrated plasma samples in the clinical laboratory. The ThromboScreen® 400C has both clot and chromogenic testing capabilities. Assays performed on the instrument include routine clotting tests such as Prothrombin Time (PT), Activated Partial Thromboplastin Time (APTT), Fibrinogen (Clauss and Derived methods), and PT and APTT-based factor assays. Chromogenic tests include assays such as Antithrombin III, Protein C and Heparin Xa.
Device Story
ThromboScreen® 400C (TS400C) is a manual photo-optical coagulation instrument for clinical laboratories. Input: citrated plasma samples and reagents manually pipetted by user. Operation: halogen lamp light source; temperature-regulated incubator block (36.5-37.5°C) with 4 measuring positions, 5 reagent positions, and 24 cuvette prewarming positions. Device uses 405 nm filter for both clotting and chromogenic assays. Output: optical density changes measured to determine clotting times or chromogenic activity. Healthcare providers use results to assess coagulation status, monitor therapy, or diagnose clotting disorders. Benefits: provides standardized, objective measurement of coagulation parameters compared to manual visual methods.
Clinical Evidence
Bench testing only. Precision studies (within-run and between-run) performed using Pacific Hemostasis reagents. Method comparison studies conducted against MLA-900C/1000C using patient samples (healthy and pathological). Correlation coefficients (r) ranged from 0.88 to 0.99 across various assays (PT, APTT, Fibrinogen, Factors V/VIII, ATIII).
Indicated for in-vitro diagnostic coagulation testing of citrated plasma samples in clinical laboratory settings. Includes routine clotting tests (PT, APTT, Fibrinogen) and chromogenic assays (Antithrombin III, Protein C, Heparin Xa).
Regulatory Classification
Identification
A coagulation instrument is an automated or semiautomated device used to determine the onset of clot formation for in vitro coagulation studies.
Special Controls
*Classification.* Class II (special controls). A fibrometer or coagulation timer intended for use with a coagulation instrument is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.
{0}------------------------------------------------
#### PREMARKET NOTIFICATION 510(K) SUMMARY 10.0
| Applicant: | Laura A. Worfolk, Ph.D.<br>Pacific Hemostasis<br>11515 Vanstory Drive<br>Huntersville, NC 28078<br>704-875-0494<br>Fax # 704-875-2092 |
|----------------------|---------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Laura A. Worfolk<br>Phone: 704-875-0494<br>Fax #: 704-875-2092 |
| Date: | June 21, 1999 |
| Trade Name: | Pacific Hemostasis ThromboScreen® 400C |
| Common Name: | Manual Coagulation Instrument |
| Classification Name: | Instrument, Coagulation (per 21 CFR section 864.5400) |
| Equivalent Devices: | MLA-900C, #K884863 & the MLA-1000C, #K894052 |
# Description of the ThromboScreen® 400C
The ThromboScreen® 400C (TS400C) is a photo-optical instrument used for the performance of in-vitro diagnostic clotting and chromogenic procedures in the clinical laboratory. The instrument utilizes photo-optical principles for both clotting and chromogenic assays. The ThromboScreen® 400C light source is provided by a halogen lamp. The incubator block is temperature regulated to 36.5 - 37.5℃ and contains four measuring positions, five reagent and 24 cuvette prewarming positions. A detailed description of the device, including an explanation of how it functions, is described in the ThromboScreen® 400C Operator's Manual, section 1, Introduction.
### Intended Use of the ThromboScreen® 400C
The Pacific Hemostasis ThromboScreen® 400C is a photo-optical instrument used for the performance of in-vitro diagnostic coagulation testing of citrated plasma samples in the clinical Coagulation testing capabilities of the device include routine clotting tests such as laboratory. Prothrombin Time (PT), Activated Partial Thromboplastin Time (APTT), Fibrinogen (Clauss and Derived methods), as well as PT and APTT-based factor assays. Chromogenic tests include assays such as Antithrombin III, Protein C and Heparin Xa.
{1}------------------------------------------------
### Summary of Substantial Equivalence Comparisons
The ThromboScreen® 400C (TS400C) was compared to the MLA-900C and the MLA-1000C (K884863 & K894052, respectively). All three instruments have a similar intended use: for in-vitro diagnostic coagulation testing in the clinical laboratory. Further, the proposed device and the predicate devices utilize photo-optical measurement principles for both clot and chromogenic based assays,
The TS400C is a "manual" coagulation instrument, in that the user must pipet both sample and test reagent. In contrast, the MLA-900C is semi-automated, and the MLA-1000C is a fully-automated instrument. The MLA-900C requires manual sample addition, but has an automatic pipet for reagent addition. The MLA-1000C has an automatic pipetting system, which adds both sample and test reagent. The light source for the MLA instruments is a halogen lamp and the wavelength is set at 550 nm for clotting assays, and 405 nm for chromogenic assays. A halogen lamp is also utilized by the TS400C, however a 405 nm filter is used for both clotting and chromogenic assays. Although the TS400C utilizes a different wavelength for clotting assays compared to the MLA, it has been optimized for this specific light source/filter combination. The performance data generated support this statement (Tables 1-3).
Within-run and between-run precision studies were performed using Pacific Hemostasis (PH) brand reagents. Tables 1 and 2 summarize the data obtained from these studies. Comparison testing performed is shown in Table 3. Specimens were evaluated from apparently healthy individuals and from patients with different pathological conditions, which are expected to affect the results for a particular assav. For derived Fibrinogen comparison testing, extrapolated data was included in the Note: analysis. This was done so that a wide range of Fibrinogen results could be compared between the two instruments. In a normal clinical testing situation, patient values above the highest standard curve point are not reported. Rather, the user is advised to report the result as greater than the highest standard curve value, and use the Clauss methodology for Fibrinogen testing (page 48 of the Operator's Manual).]
{2}------------------------------------------------
| | Test | TS400C | | | MLA | | |
|-----------------------|--------|--------|--------|------|------|--------|------|
| | | Low | Normal | High | Low | Normal | High |
| PT | Site 1 | | 1.9 | 2.5 | | 1.1 | 3.8 |
| | Site 2 | | 4.3 | 5.2 | | 1.5 | 2.0 |
| APTT | Site 1 | | 5.4 | 3.1 | | 0.8 | 0.9 |
| | Site 2 | | 2.4 | 4.1 | | 3.3 | 2.2 |
| Clauss Fib. * | | 7.2 | 6.7 | 6.4 | 2.8 | 1.2 | 3.8 |
| Derived Fib. * | | 7.5 | 2.6 | 2.3 | 4.4 | 3.4 | 2.1 |
| Factor V* | | 4.0 | 4.5 | | 1.2 | 2.0 | |
| Factor VIII* | | 9.9 | 9.7 | | 5.3 | 4.7 | |
| Chromogenic<br>ATIII* | | 6.9 | 4.9 | | 10.1 | 3.2 | |
Table 1. Summary of Within-run Precision Studies, %CV
*Testing at Site 1 only. (Shaded areas, no testing performed. Only clinically significant ranges tested.) Note: MLA-1000C used at Site 1, MLA-900C used at Site 2.
| | | TS400C | | | MLA-1000C | | |
|-------------|-----|--------|------|-----|-----------|------|--|
| Test | Low | Normal | High | Low | Normal | High | |
| PT | | 3.4 | 6.0 | | 1.4 | 4.5 | |
| APTT | | 2.5 | 1.8 | | 1.4 | 1.2 | |
| Clauss Fib. | 4.1 | 4.0 | 2.9 | 5.5 | 4.5 | 4.9 | |
| ATIII | 9.4 | 4.5 | | 9.9 | 5.1 | | |
# Table 2. Summary of Between-run Precision Testing*
*Testing performed at Site 1 only.
{3}------------------------------------------------
| Test | Site & Sample # | Correlation<br>Coefficient, r | Regression Equation |
|------------------------------------------|-----------------|-------------------------------|----------------------|
| (Reagent, Unit) | | | |
| Prothrombin Time (PT) | Site #1 - 94 | 0.99 | y = 1.045x - 0.3358 |
| (Thromboplastin DS,<br>seconds) | Site #2 - 139 | 0.99 | y = 1.141x - 1.5248 |
| Prothrombin Time | Site #1 - 94 | 0.98 | y = 0.9395x + 0.1153 |
| (Thromboplastin DS, INR) | Site #2 - 139 | 0.99 | y = 0.9955x + 0.0617 |
| Derived Fibrinogen | Site #1 - 47 | 0.98 | y = 1.0093x + 0.1442 |
| (Thromboplastin DS,<br>mg/dL) | | | |
| Activated Partial<br>Thromboplastin Time | Site #1 - 93 | 0.98 | y = 1.1449x + 1.3424 |
| (APTT-LS reagent,<br>seconds) | Site #2 - 117 | 0.96 | y = 0.9955x + 0.0617 |
| Clauss Fibrinogen | Site #1 - 50 | 0.98 | y = 0.9366x + 27.28 |
| (Thrombin reagent,<br>mg/dL) | Site #2 - 20 | 0.98 | y = 0.9843x + 3.2104 |
| Factor VIII | Site #1 - 49 | 0.97 | y = 1.04x - 0.8112 |
| (APTT-LS, % activity) | | | |
| Factor V, | Site #1 - 50 | 0.97 | y = 1.0511 - 5.6923 |
| (Thromboplastin DS, %<br>activity) | | | |
| Chromogenic ATIII | Site #1 - 58 | 0.88 | y = 0.9417x + 7.6005 |
| (ATIII Chromogenic Kit,<br>%) | | | |
| | | | Table 3. Summary of Method Comparison Studies Between the TS400C & the MLA-900C/1000C | | |
|--|--|--|---------------------------------------------------------------------------------------|--|--|
|--|--|--|---------------------------------------------------------------------------------------|--|--|
Note: Site 1 used the MLA-1000C, Site 2 the MLA-900C.
In conclusion, the similar intended use, technological characteristics and the combined performance data support the claim that the ThromboScreen® 400C is substantially equivalent to the MLA-900C and the MLA-1000C.
{4}------------------------------------------------
# 11.0 ATTACHMENTS
- ThromboScreen® 400C Operator's Manual. 1.
- The Introduction section of the MLA-1000C Operator's Manual. (With the exception of the 2. Automatic Primary Tube Sampler, the MLA-900C is the same system as the MLA-1000C. Therefore Refer to page 1-1 of the MLA-1000C the MLA-900C Operator's Manual was not included. Operator's Manual.)
- 3. TS400C Power Supply approval certificates.
- 4. Manufacturer's Certificate of Software Traceability.
- Manufacturer's Certificate of Y2K Compatibility. 5.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three overlapping shapes that resemble birds in flight. The logo is black and white.
DEPARTMENT OF HEALTH & HUMAN SERVICES
3 1999 SEP
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Laura A. Worfolk, Ph.D. Principal Research Scientist Pacific Hemostasis 11515 Vanstory Drive, Suite 125 11315 Vanstory Dr. Program 28078-8144 ·
K992130 K992130
Trade Name: Pacific Hemostasis ThromboScreen® 400C (TS400C) Re: Regulatory Class: II Product Code: KQG Dated: June 22, 1999 Received: June 23, 1999
Dear Dr. Worfolk:
We have reviewed your Section 510(k) notification of intent to market the device referenced We have reviewed your Section 510(c) notification of the minent (for the indications for use
above and we have determined the devices marked in interstate commerce above and we have determined the devices marketed products (top interstate commerce
stated in the enclosure) to legally marketed previce Americal in interstate commence stated in the enclosure) to legally marketed predical Device Amendments, or to devices that
prior to May 28, 1976, the enactment date of the Federal Food, Drug, and prior to May 28, 1976, the enatment and on the Federal Food, Drug, and have been reclassified in accordance with the provisions of the Peneral controls
Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls Cosmetic Act (Act). You may, therefore, market the Act include requirements for annual
provisions of the Act. The general controls provisions of the Act include requirement provisions of the Act. The general controls provisions of the Host. For I
registration, listing of devices, good manufacturing practice, labeling, and prohibitions against
a misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III
n treass If your device is classified (see above) into elder class in (opertials major regulations)
(Premarket Approval), it may be subject to such additions. Title 21, Parts 800 to 8 (Premarket Approval), it may be subject to such auch as mile 21, marts 800 to 895.
affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 affecting your device can be found in the counds compliance with the Current Good
A substantially equivalent determination assumes compliance with the Current Regulation (C A substantially equivalent determination assumes compilation (QS) for
Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) Manufacturing Practice requirements, as ser form in the Quad that, through periodic QS
Medical Devices: General regulation (21 CFR Part 820) and that, through periodic of Medical Devices: General regulation (ZI CFK Fart 620) will verify such assumptions. Failure to
inspections, the Food and Drug Administration (FDA) will verify such assumpti inspections, the Food and Drug Administration (r bring basilion, FDA may publish
comply with the GMP regulation may result in regulatory action. Please note: this comply with the GMP regulation in the Federal Register. Please note: this further announcements concerning your device in the reasing and affect any obligation you might
response to your premarket notification submission does not affectronic Produc response to your premarket notification submission actives under the Electronic Product
have under sections 531 through 542 - The Act lows or reaulations have under sections 551 through of ID CFE Federal laws or regulations.
{6}------------------------------------------------
### Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may Onder the Chillour Baberatory Inspiration. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.
This letter will allow you to begin marketing your device as described in your 510(k) premarket I his letter will and w Jour of substantial equivalence of your device to a legally marketed notification. The PDF Intellig of Gestion for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at additionally, 0071100111, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Steven Sutman
Steven I. Gutman, M.D, M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
Page__of_
| i10(k) Number (if known): | K992130 |
|---------------------------|---------|
|---------------------------|---------|
Device Name: _________________________________________________________________________________________________________________________________________________________________
Indications For Use:
# STATEMENT OF INDICATIONS FOR USE
The Pacific Hemostasis ThromboScreen® 400C is a photo-optical instrument used for the performance of in-vitro diagnostic coagulation testing of citrated plasma samples in the clinical laboratory. The ThromboScreen® 400C has both clot and chromogenic testing capabilities. Assays performed on the instrument include routine clotting tests such as Prothrombin Time (PT), Activated Partial Thromboplastin Time (APTT), Fibrinogen (Clauss and Derived methods), and PT and APTT-based factor assays. Chromogenic tests include assays such as Antithrombin III, Protein C and Heparin Xa.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
| | Concurrence of CDRH, Office of Device Evaluation (ODE) |
|---------------|----------------------------------------------------------------------------------------------------------|
| | <img alt="Signature" src="signature.png"/> (Division Sign-Off) Division of Clinical Laboratory Devices |
| 510(k) Number | K992130 |
Prescription Use
OR
Over-The-Counter Use
Prescri (Per 21 CFR 801.109)
OR
ver-The-Counter osc.
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.