AVANTI POLAR LIPIDS VDRL ANTIGEN SLIDE TEST KIT WITH 10 X 0.5 ML AMPULES, AVANTI POLAR LIPIDS BULK BDRL ANTIGEN, AVANTI
Applicant
Avanti Polar Lipids, Inc.
Product Code
GMQ · Microbiology
Decision Date
Sep 20, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.3820
Device Class
Class 2
Indications for Use
Avanti's VDRL antigen and Buffered Saline are intended for use in the Venereal Disease Research Laboratory (VDRL) slide test for syphilis. The VDRL slide test is a non-treponemal, micro-flocculation test, that provides both qualitative and semi-quantitative results. The test is for the detection of IgM and IgG anticardiolipin antibodies in serum specimens.
Device Story
The VDRL Antigen Slide Test Kit is an in vitro diagnostic reagent set used to detect syphilis. The device consists of an antigen suspension (cardiolipin, phosphatidylcholine, and cholesterol in ethanol) and a buffered saline solution. When mixed with patient serum containing IgM and IgG anti-cardiolipin antibodies, the antigen flocculates, forming a three-dimensional lattice structure. This reaction is observed by a clinician using a light microscope at 10x magnification. The test provides qualitative and semi-quantitative results. It is used in clinical laboratory settings to identify antibodies produced in response to Treponema pallidum infection, assisting in the diagnosis of syphilis.
Clinical Evidence
Performance evaluated using a sample population of 100 documented syphilis cases. Specificity was 100%. Sensitivity was 86.5%. Reproducibility was assessed via a blind, coded panel tested at two independent sites; inter-day and intra-day testing showed a maximum difference of one doubling dilution for the same clinician, and all reactive specimens were within two doubling dilutions of the true end point when tested by different clinicians.
Technological Characteristics
Reagent kit containing cardiolipin (0.1 g/L), cholesterol (9.0 g/L), and lecithin (1-1.9 g/L) in ethanol, plus a buffered saline solution (formaldehyde, sodium chloride, disodium phosphate, monopotassium phosphate). Principle: non-treponemal micro-flocculation. Detection: visual observation of lattice formation via light microscopy (10x ocular/objective).
Indications for Use
Indicated for the detection of IgM and IgG anti-cardiolipin antibodies in serum specimens to aid in the diagnosis of syphilis caused by Treponema pallidum.
Regulatory Classification
Identification
Treponema pallidum nontreponemal test reagents are devices that consist of antigens derived from nontreponemal sources (sources not directly associated with treponemal organisms) and control sera (standardized sera with which test results are compared) used in serological tests to identify reagin, an antibody-like agent, which is produced from the reaction of treponema microorganisms with body tissues. The identification aids in the diagnosis of syphilis caused by microorganisms belonging to the genus Treponema and provides epidemiological information on syphilis.
Predicate Devices
VDRL Antigen For Syphilis Serology (Becton Dickinson Microbiology Systems)
Cenogenics ADRL/STS Test (Cenogenics Corporation)
VDRL Antigen (Lee Laboratories, Inc.)
VDRL Antigen (Centers for Disease Control and Prevention)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for Avanti Polar Lipids, Inc. The logo includes the date SEP 20 1999 at the top left. To the left of the text is a graphic of a circle with smaller circles around it.
K992124
## Safety and Effectiveness Summary 1.
This 510(k) summary of safety and effectiveness data is being submitted in accordance with the requirements of the Safe Medical Devices Act of 1990 and the Medical Device Amendments of 1992.
- Submitter: a.
Avanti Polar Lipids, Inc. 700 Industrial Park Drive Alabaster, AL 35007 Tel: (205) 663-2494 (205) 663-0756 Fax:
- Contact: b.
Rowena Shaw, Vice President
- Date Submitted C.
June 22, 1999
- d. Device Name:
Trade Name: Avanti Polar Lipids VDRL Antigen Slide Test Kit
Common Name: VDRL Antigen Slide Test Kit
Classification Name: ANTIGENS, NON-TREPONEMAL, ALL
- Device Description e.
The Venereal Disease Research Laboratory slide test is a test for the detection of syphilis. The test employs an antigen containing cardiolipin, phosphatidylcholine (lecithin), and cholesterol dissolved in ethanol. The antigen is suspended in a buffered saline solution, which flocculates when combined with serum containing IgM and IgG anti-cardiolipin antibodies. The IgM and IgG antibodies are produced in response to infection by Treponema pallidum the causitive agent of syphilis. The flocculation forms a three-dimensional lattice structure that can be seen at low magnification (10x oculars and 10x objective) with a light
{1}------------------------------------------------
microscope.
- f. Intended Use:
Avanti's VDRL antigen and Buffered Saline are intended for use in the Venereal Disease Research Laboratory (VDRL) slide test for syphilis. The VDRL slide test is a non-treponemal, micro-flocculation test, that provides both qualitative and semi-quantitative results. The test is for the detection of IgM and IgG anticardiolipin antibodies in serum specimens.
## Substantial Equivalence Claim: g. Avanti Polar Lipids, Inc. VDRL Antigen Slide Test Kit is substantially equivalent to the following currently marketed devices:
- VDRL Antigen For Syphilis Serology. Becton Dickinson Microbiology i. Systems, Becton Dickinson and Company, Sparks, MD 21152 USA.
- Cenogenics ADRL/STS Test. Cenogenics Corporation, Morganville, NJ. ii. 07751 USA.
- iii. VDRL Antigen. Lee Laboratories, Inc. 1475 Athens Highway, S.W., Grayson, Georgia 30221 USA.
- VDRL Antigen. Centers for Disease Control and Prevention. Atlanta, GA iv. USA
- Technological Characteristics: h.
Lipid components present in VDRL Antigen
| Component | Concentration<br>Avanti Polar Lipids VDRL<br>Antigen | Concentration<br>Becton Dickinson's VDRL<br>Antigen |
|---------------------------------------------|--------------------------------------------------------------------|----------------------------------------------------------------------|
| Cardiolipin derived from ox<br>heart tissue | 0.1 grams / liter | 0.3 grams / liter |
| Cholesterol derived from<br>lanolin | 9.0 grams / liter | 9.0 grams / liter |
| Lecithin derived from egg<br>yolk | sufficient for standard<br>reactivity (1 - 1.9 grams per<br>liter) | sufficient for standard<br>reactivity (1.8 - 2.0 grams per<br>liter) |
The components listed in the above table are dissolved in ethanol.
{2}------------------------------------------------
## Components present in buffered saline solution
| Component | Concentration<br>Avanti Polar Lipids<br>Buffered Saline | Concentration Becton<br>Dickinson VDRL Buffered<br>Saline |
|------------------------------------------------|---------------------------------------------------------|-----------------------------------------------------------|
| Formaldehyde, 37% aqueous,<br>(A.C.S.) | 0.5 ml per liter | 0.5 ml per liter |
| Sodium Chloride (A.C.S.) | 10.0 grams per liter | 10.0 grams per liter |
| Disodium phosphate, anhydrous<br>(A.C.S.) | 0.037 grams per liter | 0.037 grams per liter |
| Monopotassium phosphate,<br>anhydrous (A.C.S.) | 0.170 grams per liter | 0.170 grams per liter |
The components listed in the above table are dissolved in deionized water.
- i. Performance Data
- i. Specificity Data
The specificity of Avanti Polar Lipids, Inc. VDRL Antigen Slide Test Kit was 100% for a sample population of 100 documented cases of syphilis.
- ii. Sensitivity Data
The sensitivity of Avanti Polar Lipids, Inc. VDRL Antigen Slide Test Kit was 86.5% for a sample population of 100 documented cases of syphilis.
- iii. Reproducibility Data
Specimens from a blind, coded panel were measured at two independent test sites. Reproducibility was demonstrated with inter-day and intra-day testing. Weakly reactive, reactive and non-reactive specimens were included in the panel. A maximum inter-day and intra-day difference of one doubling dilution was observed when specimens were tested by the same clinician. All reactive specimens gave an end point within two doubling dilutions of the true end point when tested by different clinicians.
- j. Conclusion
The safety and effectiveness of Avanti Polar Lipids VDRL Antigen Slide Test Kit is substantially equivalent to legally marketed VDRL Antigen Slide Test Kits, as demonstrated in the comparison of the performance of Avanti Polar Lipids VDRL Antigen Slide Test Kit to these VDRL Antigen Slide Test Kits.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized emblem that resembles an abstract caduceus, a symbol often associated with healthcare. The emblem features three parallel lines that curve and converge, creating a sense of movement and unity.
Public Health Service
SEP 20 1999
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Ms. Rowena Shaw Vice President Avanti Polar Lipids, Inc. 700 Industrial Park Drive Alabaster, Alabama 35007
Re: K992124 Trade Name: Avanti Polar Lipids VDRL antigen Slide Test Kit Regulatory Class: II Product Code: GMQ Dated: June 22, 1999 Received: June 23, 1999
Dear Ms. Shaw:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}------------------------------------------------
Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html"
Sincerely yours.
Steven Butman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
510(k) Number (if known):____K992124
Avanti Polar Lipids VDRL Antigen Slide Test Kit Device Name:
Indications For Use:
Avanti's VDRL antigen and Buffered Saline are intended for use in the Venereal Disease Research Laboratory (VDRL) slide test for syphilis. The VDRL slide test is a non-treponemal, microflocculation test, that provides both qualitative and semiquantitative results. The test detects the presence of IgM and IgG anticardiolipin antibodies in serum. The IgM and IgG antibodies are produced in response to infection by Treponema pallidum the causitive agent of syphilis.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Sally 7. Selzpak for i Dubois
Division St. Off.
Division of Clinical Laboratory Devices 510(k) Number
Prescription Use (Per 21 CFR 801.109)
OR
Over-The-Counter Use
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.