K992121 · Cook, Inc. · KRD · Dec 20, 1999 · Cardiovascular
Device Facts
Record ID
K992121
Device Name
EMBOLIZATION COIL SYSTEM
Applicant
Cook, Inc.
Product Code
KRD · Cardiovascular
Decision Date
Dec 20, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.3300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Arterial and venous embolization in the peripheral vasculature
Device Story
Sterile, single-use embolization coil system; includes introducer system with premounted platinum coil and delivery wire. Coil deployed via mechanical unscrewing of interlocking threads between coil and delivery wire (counterclockwise rotation). Used in peripheral vasculature; operated by clinicians via percutaneous entry using catheter or microcatheter. Detach Locking Device (pin vise) ensures delivery wire stability during detachment. System provides mechanical occlusion of vessels; benefits include controlled coil placement and detachment for embolization procedures.
Clinical Evidence
Bench testing only. Testing included tensile strength evaluations for Detach 11 and Detach 18 coils and performance testing within a Microferret catheter mounted in a phantom to verify reliable design and performance.
Indicated for arterial and venous embolization in the peripheral vasculature.
Regulatory Classification
Identification
A vascular embolization device is an intravascular implant intended to control hemorrhaging due to aneurysms, certain types of tumors (e.g., nephroma, hepatoma, uterine fibroids), and arteriovenous malformations. This does not include cyanoacrylates and other embolic agents, which act by polymerization or precipitation. Embolization devices used in neurovascular applications are also not included in this classification, see § 882.5950 of this chapter.
Special Controls
*Classification.* Class II (special controls.) The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Vascular and Neurovascular Embolization Devices.” For availability of this guidance document, see § 870.1(e).
{0}------------------------------------------------
DEC 2 0 1999
131
K992121
510(k) Premarket Notification Embolization Coil System COOK INCORPORATED
#### Safety and Effectiveness Information
Submitted By: Lisa Webb, RAC Regulatory Affairs Coordinator COOK INCORPORATED 925 South Curry Pike P.O. Box 489 Bloomington, IN 47402 (812) 339-2235 June 21, 1999
| Device: | Trade Name: | Detach 11® and Detach |
|---------|-------------|-----------------------|
|---------|-------------|-----------------------|
Proposed Classification Name:
188
Arterial Embolization Device
### Predicate Devices:
Detach 11@/18® is similar in terms of intended use, materials of construction and technological characteristics to the predicate devices reviewed, the Embolization Coil Positioner Set, Hilal Embolization Microcoils™, the Guglielmi Detachable Coil® and the Fibered Platinum Coil.
### Device Description
This embolization coil system is supplied sterile and is intended for one time use. The device is comprised of an introducer system with a premounted detachable embolization coil. The embolization coil is deployed when the interlocking threads between the coil and the delivery wire are "unscrewed" by turning the delivery wire counterclockwise approximately 25 times.
The introducer system consists of a delivery wire and a delivery wire inserter. The delivery wire inserter consists of a plastic delivery wire holder with the delivery wire and a cannula inserter containing the coil. A Detach Locking Device is also needed to use the set. This part is sold separately because several coils can be delivered through one Detach Locking Device. These components are further described below.
- Coil: The coil is constructed of platinum and is available in the following shapes: curled, straight, Tornado™, J-Coil Shape and Multiple J-Coil Shape. The coil is also available in three degrees of softness: Detach 18® standard, Detach 18® soft, and Detach 110.
- Delivery Wire: The delivery wire is comprised of four components: a handle, the delivery wire, a marker band and a detach tip. These are constructed of platinum and PTFE coated stainless steel. The delivery wires for Detach 11® and Detach 18® are available in a
{1}------------------------------------------------
length of 200 cm. The diameter of the Detach 11@ delivery wire is 0.011 inches. The diameter of the Detach 18® delivery wire is 0.018 inches.
- 트 Delivery Wire Inserter: The delivery wire inserter is also comprised of four components. These include a stainless steel tip, a polycarbonate tip fitting, polyethylene tubing and a nylon fitting.
- 트 Detach Locking Device: The Detach Locking Device is basically a pin vise which ensures that the delivery wire does not move forward during coil detachment. This device does not contact the skin or the blood.
### Substantial Equivalence
Four devices are currently marketed in the U.S. which are believed to be substantially equivalent to Detach 11@/18@. These devices include an Embolization Coil Positioner Set (COOK INC), Hilal Embolization Microcoils™ (COOK INC), the Guglielmi Detachable Coil® (Target Therapeutics®) and a Fibered Platinum Coil (Target Therapeutics®). All devices are introduced via the percutaneous method of entry using a catheter or microcatheter introducer.
The Embolization Coil Positioner Set was reviewed as substantially equivalent under K942189 and is indicated for arterial and venous embolization. The device is constructed of stainless steel and synthetic fiber with a coil wire diameter of 0.018 to 0.038 inches. The coils are available in straight or curled shapes with an emboli size range of 2 to 20 mm. A push-button release mechanism is the method of deployment.
Hilal Embolization Microcoils™ were reviewed as substantially equivalent under K901337 and are indicated for the embolization of arteriovenous malformations and other vascular lesions of the brain, spinal cord and spine. The device is constructed of platinum coil and synthetic fiber with a coil wire diameter of 0.018 inches. The coils are available in straight and curled shapes with an emboli size range of 3 to 10 mm. Deployment is achieved by a wire guide which pushes the coil out of the catheter.
The Guglielmi Detachable Coil® was reviewed as substantially equivalent under K951256 and K960705 and is indicated for embolization of intracranial aneurysms, arteriovenous malformations, arteriovenous fistulae and arterial venous embolizations in the peripheral vasculature. The device is constructed of platinum with a coil wire diameter of 0.010 to 0.018 inches. The coils are available in a helical shape with an emboli size range of 2 to 20 mm. The coil is deployed by electrolytic detachment from the wire guide.
The Fibered Platinum Coil was reviewed as substantially equivalent under K955293 and is indicated for arterial and venous embolization in the peripheral vasculature. The device is
{2}------------------------------------------------
#### 510(k) Premarket Notification Embolization Coil System COOK INCORPORATED
constructed of platinum and synthetic fiber with a coil wire diameter of 0.010 to 0.035 inches. The coils are available in the following shapes: straight, C-shaped, helical and complex helical. The emboli size range is 2 to 30 mm. Deployment is achieved by a wire guide which pushes the coil out of the catheter.
The Detach 11@/18® will be indicated for arterial and venous embolization. This device will be constructed of platinum with a coil wire diameter of 0.011, 0.014 and 0.018 inches. The coils will be available in curled, straight, Tornado™, J-coil and multiple J-coil shapes. The emboli size range will be 2 to 12 mm. The coil is deployed when interlocking threads between the coil and the delivery wire are "unscrewed."
The similar indications for use and technological characteristics of the Detach 11® and Detach 18® Embolization Coil Systems as compared to the predicate devices support a determination of substantial equivalency.
## Test Data
Detach 11@/18® was subjected to the following tests to assure reliable design and performance under the specified testing parameters. These tests were comprised of:
- � Tensile tests: Detach 188
- � Tensile tests: Detach 11®
- � Tensile tests: Detachable Coils
- � Performance Test in a Microferret™ Catheter Mounted in a Phantom
The results of these tests provide reasonable assurance that the device has been designed and tested to assure conformance to the requirements for its use for arterial and venous embolization.
{3}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Image /page/3/Picture/2 description: The image is a black and white circular logo for the U.S. Department of Health & Human Services. The logo features the department's emblem, which is a stylized caduceus with a human figure in the center. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the upper portion of the circle.
DEC 20 1999
Ms. Lisa Webb Regulatory Affairs Coordinator Cook Incorporated 925 South Cury Pike P.O. Box 489 Bloomington, IN 47402
K992121 Re: Embolization Coil System Requlatory Class: III (Three) Product Code: KRD September 23, 1999 Dated: Received: September 24, 1999
Dear Ms. Webb:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to A substantially equivalent determination assumes compliance with 895. the Current Good Manufacturing Practice requirements, as set forth in the Quality System Requlation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Please note: this response to your premarket Federal Register. notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
{4}------------------------------------------------
Page 2 - Ms. Lisa Webb
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
S/Kute L. Campbell
Celia M. Witten, Ph.D., M.D. Acting Director Division of Cardiovascular, Respiratory and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Addendum to K992121 510(k) Premarket Notification Embolization Coil System COOK INCORPORATED
510(k) Number (if known): K992121
Device Name:
Embolization Coil System
Indications for Use:
Arterial and venous embolization in the peripheral vasculature
Bess L. Leasure
(Division Sign-Off) Division of Cardiovascular, Respiratory, and Neurological Devices 510(k) Number __ Kggaj 2
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
い Prescription Use (Per 21 CFR 801.109)
______________________________________________________________________________________________________________________________________________________________________________
OR
Over-the-Counter Use
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.