K992116 · Siemens Medical Solutions USA, Inc. · CCK · Jan 12, 2000 · Anesthesiology
Device Facts
Record ID
K992116
Device Name
SIEMEMNS INFINITY ETC02 POD
Applicant
Siemens Medical Solutions USA, Inc.
Product Code
CCK · Anesthesiology
Decision Date
Jan 12, 2000
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.1400
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The intended use of Siemens INFINITY etCO2 pod is to measure end tidal carbon dioxide, inspired carbon dioxide, and respiration rate in both mainstream and side-stream measurement modes (side-stream is not intended for use in the neonatal population). In conjunction with an INFINITY modular bedside monitor, visual and audible alarms will be produced if any of these parameters vary beyond preset limits, and timed or alarm recordings will be produced.
Device Story
INFINITY etCO2 Pod is a modular bedside monitoring accessory for Siemens SC7000/SC8000/SC9000XL series; measures end-tidal CO2, inspired CO2, and respiration rate via mainstream or side-stream sampling. Uses dual-wavelength, non-dispersive infrared (NDIR) technology to analyze gas samples; connects to modular monitors for data processing. Provides real-time CO2 values and respiration rates; triggers visual/audible alarms when parameters exceed preset limits; generates timed or alarm-based recordings. Operated by healthcare professionals in clinical environments. Output allows clinicians to monitor patient respiratory status and ventilation adequacy; aids in clinical decision-making regarding patient respiratory support.
Clinical Evidence
No clinical data provided; substantial equivalence supported by bench testing and comparison of technological characteristics to predicate devices.
Technological Characteristics
Dual-wavelength, non-dispersive infrared (NDIR) gas analyzer. Measures CO2 partial pressure (0-100 mmHg). Supports mainstream and side-stream sampling. Connectivity via Siemens INFINITY modular bedside monitor interface. Standards: EN 864:1996, IEC 60601-1. Form factor: modular pod.
Indications for Use
Indicated for adult, pediatric, and neonatal populations (side-stream mode contraindicated for neonates) in healthcare settings by physicians, nurses, or technicians to measure end-tidal CO2, inspired CO2, and respiration rate. Not for use in MRI magnetic fields.
Regulatory Classification
Identification
A carbon dioxide gas analyzer is a device intended to measure the concentration of carbon dioxide in a gas mixture to aid in determining the patient's ventilatory, circulatory, and metabolic status. The device may use techniques such as chemical titration, absorption of infrared radiation, gas chromatography, or mass spectrometry.
{0}------------------------------------------------
### 510(k) SUMMARY
as required per 807.92(c)
## 1. Submitters Name, Address:
Siemens Medical Systems, Inc. Electromedical Systems Group, PCS Danvers, MA 01923 Tel: (978) 907-7500 Fax: (978) 750-6879 Official Correspondent: David Simard, Director Quality Assurance & Regulatory Affairs Contact person for this submission: Penelope H. Greco Date submission was prepared: June 11, 1999
#### Trade Name, Common Name and Classification Name: 2.
- A. Trade Name: Siemens INFINITY etCO2 Pod
#### Common Name, Classification Name, Class and Regulation Number: B.
| Common Name | Classification<br>Number | Class | Regulation Number |
|----------------------------------------------|--------------------------|-------|-------------------|
| Analyzer, Gas, Carbon-Dioxide, Gaseous-Phase | 73CCK | II | 868.1400 |
#### Predicate Device Identification: 3.
Siemens SC9000 etCO2 Module 510(k) K954632
Novametrix CO2SMO Plus 510(k) K963380
#### 4. Device Description:
The INFINITY etCO2 Pod is an addition to Siemens SC8000/SC7000/SC9000XL INFINITY modular bedside monitoring series. When connected to an INFINITY modular bedside monitor the etCO2 pod will measure end tidal carbon dioxide, inspired carbon dioxide, and respiration rate.
#### ડ. Intended Use:
The intended use of Siemens INFINITY etCO2 pod is to measure end tidal carbon dioxide, inspired carbon dioxide, and respiration rate in both mainstream and side-stream measurement modes (side-stream is not intended for use in the neonatal population). In conjunction with an INFINITY modular bedside monitor, visual and audible alarms will be produced if any of these parameters vary beyond preset limits, and timed or alarm recordings will be produced.
### -
### COMPANY CONFIDENTIAL
### Siemens Medical Systems, Inc.
Electromedical Systems Group, PCS
16 Electronics Avenue Danvers, MA 01923 USA
Tel: (978) 907-7500 Fax: (978) 750-6879 Telex: 511958 (Siemensm SD)
{1}------------------------------------------------
## 6. Table of Device Similarities and differences to predicate device
| Predicate Device | Predicate Device | Applicant | Explanation of<br>Differences | |
|---------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
| Manufacturer | Siemens Medical Systems<br>SC9000 etCO2 Module | Novametrix Medical Systems<br>CO2SMO Plus | Siemens Medical Systems<br>INFINITY etCO2 Pod | |
| 510(k) Number | K954632 | K963380 | To be assigned | |
| Intended Use | To measure end tidal carbon dioxide,<br>inspired carbon dioxide, and<br>respiration rate in both mainstream<br>and side-stream measurement modes<br>(side-stream is not intended for use in<br>the neonatal population). In<br>conjunction with the SC9000, visual<br>and audible alarms will be produced<br>if any of these parameters vary<br>beyond preset limits, and timed or<br>alarm recordings will be produced. | To provide spirometric, carbon<br>dioxide and pulse oximetry<br>monitoring | To measure end tidal carbon dioxide,<br>inspired carbon dioxide, and respiration rate<br>in both mainstream and side-stream<br>measurement modes (side-stream is not<br>intended for use in the neonatal<br>population). In conjunction with the SC<br>7000 / SC 8000 / SC 9000XL monitors,<br>visual and audible alarms will be produced<br>if any of these parameters vary beyond<br>preset limits, and timed or alarm recordings<br>will be produced. | Siemens etCO2 pod does<br>not measure or display<br>spirometry or pulse<br>oximetry |
| Intended Population | Adult/Pediatric | Adult/Pediatric/Neonatal | Adult/Pediatric/Neonatal | |
| Intended<br>Environment | A healthcare environment by<br>healthcare professionals | Same | Same | |
| Measuring Capabilities | | | | |
| Displayed parameters | EtCO2, iCO2, Respiration rate (RRc) | Spirometry, Pulse oximetry,<br>EtCO2, iCO2, Respiration rate<br>(RRc) | EtCO2, iCO2, Respiration rate (RRc) | Siemens etCO2 pod does<br>not measure or display<br>spirometry or pulse<br>oximetry |
| Display Scales | 0-40, 0-80 mmHg | 0-50, 0-75, 0-100 mmHg | 0-40, 0-60, 0-80, 0-100 mmHg | More display flexibility |
| Measuring method | Dual-wavelength, non-dispersive<br>infrared | Same | Same | |
| Measuring<br>capabilities | Mainstream and Sidestream | Same | Same | |
| Measuring range | 0 to 99 mmHg CO2 partial pressure | 0 to 100 mmHg partial pressure | 0 to 100 mmHg CO2 partial pressure | *A CO2 ≥ 100 is<br>considered out of range |
| Averaging | Breath, 10s, 20s | Breath, 10s, 20s, Instantaneous | Breath, 10s, 20s, Instantaneous | Improvement |
2
## COMPANY CONFIDENTIAL
{2}------------------------------------------------
7. Assessment of non-clinical performance data for equivalence: Section T
- 8. Assessment of clinical performance data for equivalence: Section V
- 9. Biocompatability: Not applicable
- 10. Sterilization: Not applicable
- 11. Standards and Guidances: EN 864: 1996 IEC 60601-1
3
## COMPANY CONFIDENTIAL
# Siemens Medical Systems, Inc.
Electromedical Systems Group, PCS
16 Electronics Avenue Danvers, MA 01923 USA
Tel: (978) 907-7500 Fax: (978) 750-6879 Telex: 511958 (Siemensm SD)
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus or a representation of human figures.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JAN 1 2 2000
Mr. David Simard Siemens Medical Systems, Inc. 16 Electronics Avenue Danvers, MA 01923
Re: K992116 Siemens INFINITY etCO2 Pod Regulatory Class: II (two) Product Code: 73 CCK October 15, 1999 Dated: Received: October 15, 1999
Dear Mr. Simard:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions aqainst misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 A substantially equivalent determination assumes compliance to 895. with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Requlation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in requlatory action. In addition, FDA may publish further announcements concerning your device in the Federal Reqister. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}------------------------------------------------
Page 2 - Mr. David Simard
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
feamothw. Wutoduca fro,
elia M. Witten, Ph.D., M.D. Acting Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Page 1 of 1
510(k) Number (if known): _
Device Name: Siemens INFINITY etCO2 Pod
Indications for Use:
Siemens INFINITY etCO2 is indicated for use in the adult, pediatric and neonatal populations, in an environment where patient care is provided by healthcare professionals, i.e. Physicians, Nurses, Technicians, when the professional determines that the device is required to measure end tidal carbon dioxide, inspired carbon dioxide, and respiration rate in either mainstream or side-stream measurement mode is not intended for use in the neonatal patient population). In conjunction with the SC 7000 / SC 9000XL monitors, visual and audible alarms will be produced if any of these parameters vary beyond preset limits, and timed or alarm recordings will be produced.
MRI Compatibility Statement: The Siemens INFINITY etCO2 Pod is not intended for use in a MRI magnetic field.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
OR
Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________
(Optional Format 1-2-96)
to AW. Watscher
Division Sign-Off)
Division of Cardiovascular, Respiratory,
and Neurological Devices
510(k) Number K9921116
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.